IEEE 2026 NSS MIC RTSD

Please note! All times in the online program are given in Madrid - Europe (CET) times!

Madrid - Europe ()
Oct 23, 2026, 6:53:52 PM
Your time ()
n/a
Madrid - Europe ()
Oct 23, 2026, 6:53:52 PM
Our Sponsors:

To search for a specific ID please enter the hash sign followed by the ID number (e.g. #123).

Signal and data processing

Session chair: Lee, Jae Sung, Professor (Seoul National University, Department of Nuclear Medicine, Seoul, South Korea); Hashimoto, Fumio, Dr. (University of Florida, Gainesville, USA)
 
Shortcut: MIC-19
Date: Saturday, 14 November, 2026, 9:00 - 11:00
Location: MANUEL DE FALLA
Session type: MIC Session

Contents

Click on an contribution to preview the abstract content.

The underlined authors are the presenters. The authors in bold are the main authors.

9:00 MIC-19-01

Noise-Aware Brownian Bridge Diffusion Model for Denoising of Low-Dose PET Images (#2951)

W. De Mits1, E. Taragola1, B. Vervenne1, M.E. Daube-Witherspoon2, C. Vanhove1, S. Matej2, S. Vandenberghe1, J.S. Karp2, F.M. Muller1, 2

1 Ghent University, Faculty of Engineering and Architecture, Ghent, Belgium
2 University of Pennsylvania, Department of Radiology, Philadelphia, Pennsylvania, United States of America

Abstract

This study proposes a Noise-Aware Brownian Bridge Diffusion Model (NA-BBDM) for low-dose positron emission tomography (PET) denoising. The proposed NA-BBDM models a conditional diffusion path between paired low- and full-dose PET images and incorporates the estimated noise level from the input image into the network. The noise level is estimated from the variance of the residual between the low-dose image and a Gaussian-smoothed version. The diffusion path uses a parabolic variance schedule with zero variance at both endpoints (i.e., low- and full-dose images), parameterized by peak variance and number of diffusion steps. Both are treated as noise-dependent hyperparameters and optimized through grid search. Training data consisted of 25 18F-FDG datasets acquired on the PennPET Explorer and subsampled to emulate dose levels from 1/5 to 1/300 of total counts. Performance was compared with a U-Net and evaluated on independent 18F-FDG and unseen 11C-carfentanil (CFN) datasets. The NA-BBDM produced images with reduced noise speckles (still present with the U-Net) and better-preserved high-uptake regions at ultra-low-dose levels. Quantitatively, the NA-BBDM also achieved lower SUV variability and mean squared error for 11C-CFN, especially below 1/60th dose. Our results suggest that the NA-BBDM provides a robust framework for ultra-low-dose and cross-tracer PET denoising.

Acknowledgment

This work was supported by the Research Foundation Flanders (Vervenne: 1195125N, Muller: 11P0E24N) and the National Institutes of Health, R01 CA-113941 (Karp, PI). Data Courtesy (UPenn): J Dubroff, C Wiers, P Bravo. 

References
[1] B. Yu, et al., “Robust whole-body PET image denoising using 3D diffusion models,” Eur J Nucl Med Mol Imaging, 2025.
[2] B. Li, et al., “BBDM: Image-to-image translation with Brownian bridge diffusion models,” in Proc. IEEE on Comp Vision and Pattern Recogn, 2023.
[3] B. Dai, et al., “Performance evaluation of the PennPET Explorer with expanded axial coverage,” Phys Med Biol, 2023.
[4] H. Xie, et al., “Unified noise-aware network for low-count PET denoising,” IEEE Trans Radiat Plasma Med Sci, 2024.
[5] F. Muller, et al., “Impact of deep learning denoising on kinetic modelling for low-dose dynamic PET,” Eur J Nucl Med Mol Imaging, 2025. 
Keywords: Denoising, Noise-Aware, PET Imaging, Brownian Bridge Diffusion, Deep Learning
9:15 MIC-19-02

N2N-DiffPET: Self-supervised Diffusion for Low-dose PET Image Denoising (#2281)

K. Fan2, 1, Y. Wu1, Q. Li1, H. Liu2, N. Guo1

1 Massachusetts General Hospital and Harvard Medical School, Department of Radiology, Boston, Massachusetts, United States of America
2 Zhejiang University, Department of Optical Engineering, Hangzhou, China

Abstract

Low-dose positron emission tomography (PET) imaging can reduce radiation exposure and shorten acquisition time, but reduced photon counts introduce strong count-dependent noise and degrade quantitative reliability. Supervised deep learning methods usually require paired low-dose and normal-dose PET images, which are difficult to acquire in clinical practice. We propose N2N-DiffPET, a self-supervised diffusion framework for low-dose PET denoising without normal-dose supervision. Inspired by recent self-supervised diffusion MRI denoising, N2N-DiffPET adapts iterative diffusion refinement to PET using paired noisy PET observations generated from independent low-dose measurements by event splitting or Poisson resampling. To better model PET noise, N2N-DiffPET constructs a fusion trajectory between the paired noisy PET images and perturbs it using pair-derived PET noise rather than synthetic Gaussian noise. Co-registered MRI is incorporated as an anatomical condition to improve structural preservation. During inference, bidirectional deterministic denoising is performed, and the two directional outputs are averaged to reduce directional bias. Normal-dose PET images are not used for training and are only used for evaluation. Experiments on BrainWeb-based PET/MR simulation data show that N2N-DiffPET suppresses noise while preserving anatomical structures and regional uptake, outperforming conventional denoising methods and self-supervised baselines in visual quality and quantitative metrics.

References
[1] J. Lehtinen, J. Munkberg, J. Hasselgren, S. Laine, T. Karras, M. Aittala, and T. Aila, “Noise2Noise: Learning image restoration without clean data,” arXiv preprint arXiv:1803.04189, 2018.
[2] C. Wu, Q. Kong, Z. Jiang, and S. K. Zhou, “Self-supervised diffusion MRI denoising via iterative and stable refinement,” arXiv preprint arXiv:2501.13514, 2025.
[3] C. A. Cocosco, V. Kollokian, R. K.-S. Kwan, and A. C. Evans, “BrainWeb: Online interface to a 3D MRI simulated brain database,” Proceedings of the 3rd International Conference on Functional Mapping of the Human Brain, 1997.
Keywords: Low-dose PET denoising, Self-supervised learning, Diffusion model, Noise2Noise
9:30 MIC-19-03

Multi-Dose PET Image Denoising using Consistency Flow Matching (#1864)

J.H. Lee1, 2, S. Bae1, 2, J.S. Lee3, 4

1 Seoul National University, Interdisciplinary Program in Bioengineering, Seoul, Republic of Korea
2 Seoul National University, Integrated Major in Innovative Medical Science, Seoul, Republic of Korea
3 Seoul National University, Department of Nuclear Medicine, Seoul, Republic of Korea
4 Brightonix Imaging Inc., Seoul, Republic of Korea

Abstract

Low-dose PET reduces patient radiation burden by decreasing injected radiotracer dose or scan duration, but the resulting reduction in photon counts introduces statistical noise that degrades diagnostic image quality. Most deep learning denoising methods are trained at a single fixed dose reduction factor and do not generalize to unseen dose levels. We propose Consistency Flow Matching (CFM), a multi-dose denoising framework that augments flow matching with a cross-dose consistency regularization. During training, pairs of low-dose images at two different dose reduction factors are jointly processed, and a consistency loss enforces agreement between their predicted normal-dose outputs, promoting a dose-invariant flow field. Dose conditioning is achieved through Spatial Feature-wise Linear Modulation with
a log-sinusoidal dose embedding. Experiments on the UltraLow-Dose PET Challenge dataset show that CFM improves perceptual image quality and out-of-distribution generalization over standard flow matching, confirming the benefit of cross-dose consistency regularization.

AcknowledgmentThis research was supported by the AI Computing Infrastructure Enhancement (GPU Rental Support) User Support Program funded by the Ministry of Science and ICT (MSIT), Republic of Korea. (No. RQT-25-090157)
References
[1] Ronneberger, O, Fischer, P, Brox, T 2015, 'U-Net: Convolutional networks for biomedical image segmentation', Proc. MICCAI
[2] Lipman, Y et al. 2023, 'Flow matching for generative modeling', Proc. ICLR
[3] Song, Y, Dhariwal, P, Chen, M, Sutskever, I 2023, 'Consistency models', Proc. ICML
[4] Park, T, Liu, MY, Wang, TC, Zhu, JY 2019, 'Semantic image synthesis with spatially-adaptive normalization', Proc. CVPR
[5] Xue, S et al. 2025, 'UDPET: Ultra-low dose PET imaging challenge dataset', Proc. MICCAI, vol. 15972, Springer
Keywords: Ultra-Low-Dose PET Denoising, Flow Matching, Cross-Dose Consistency Regularization, Multi-Dose Generalization
9:45 MIC-19-04

High-Temporal-Resolution Data-Driven Respiratory Motion Estimation in Cardiac PET Imaging (#1890)

H. Wei1, C. Sun1, Z. Feng1, Y. Li1, D. Zhang1, Q. Ye1, Y. Lu1

1 United Imaging Healthcare, Shanghai, China

Abstract

Cardiac PET is widely used for myocardial perfusion imaging and viability assessment, both of which require high image quality and quantitative accuracy. In free-breathing acquisitions, respiratory motion can induce cardiac displacement, particularly along the superior-inferior (SI, z-axis) direction, leading to image blurring, attenuation-emission mismatch, and quantitative bias. Reliable estimation of this translation is therefore essential for effective motion correction. In this work, we propose COD-S (Centroid of Distribution-Scaled), a data-driven method to estimate respiratory-induced SI cardiac translation in cardiac PET. COD-S integrates the high sensitivity and high temporal resolution of COD analysis with the physically scaled displacement provided by image registration. Specifically, the respiratory motion trajectory is  decomposed into waveform and amplitude components, where the waveform is derived from COD and the amplitude are calibrated using registration-based estimates. This produces a high-temporal-resolution and physically interpretable cardiac translation trajectory without external respiratory hardware, providing a practical input for motion correction in cardiac PET.

References
[1] D. Zhang, Y. Li, C. Sun, H. Zeng, G. Yang, F. Kang, J. Wang, and Y.Lu, “A novel data-driven respiratory gating method in PET,” in Proc. IEEE NSS/MIC/RTSD), Vancouver, BC, Canada, 2023, pp. 1.
[2] C. Sun, E. M. Revilla, J. Zhang, K. Fontaine, T. Toyonaga, J.-D. Gallezot, T. Mulnix, J. A. Onofrey, R. E. Carson, and Y. Lu, “An objective evaluation method for head motion estimation in PET—Motion corrected centroid-of-distribution,” NeuroImage, vol. 264, pp. 119678, Dec. 2022.
[3] J. Zhang, C. Sun, E. M. Revilla, D. Zhang, S. P. Mok, C. Liu, K. Fontaine, R. Carson, and Y. Lu, “An objective evaluation method for respiratory motion correction in TOF PET,” in Proc. IEEE NSS/MIC/RTSD), Boston, MA, USA, 2020, pp. 1.
Keywords: Cardiac PET, Respiratory motion, Centroid-of-distribution, High temporal resolution
10:00 MIC-19-05

Spatial-variant fast non-linear energy calibration for a high resolution SSLE pixelated PET detector (#3118)

L. Jiménez Algarra1, V. Sánchez Tembleque1, L. Batres Mongil1, D. Pérez Benito1, J. J. Vaquero López1, 2

1 Departamento de Bioingeniería, Universidad Carlos III de Madrid, Madrid, Spain, Departamento de Bioingeniería, Leganés, Spain
2 Instituto de Investigación Sanitaria Gregorio Marañón, Madrid, Spain, Madrid, Spain

Abstract

High-resolution PET detectors require position dependent energy calibration due to the spatial variability of their optical response. This work addresses that challenge for a novel hexagonal LYSO detector structured by subsurface laser engraving (SSLE) with a double honeycomb displaced crystal array, coupled to a custom 61-channel hexagonal SiPM. This design provides two depth-of-interaction (DOI) layers as well as a high-granularity for event’s positioning. We first present a DOI separation method based on a dispersion metric computed
from the per-channel light distribution, enabling crystal pixel specific calibration for each layer independently. After precise
vertex identification, energy calibration is then performed per
crystal pixel by fitting four reference points: the 202 and 307
keV intrinsic 176Lu peaks and the 511 and 1275 keV peaks
from a 22Na source to a non-linear model accounting for SiPM
saturation. These points are fitted to a non-linear response
model that accounts for SiPM saturation and electronic nonlinearity.
The proposed calibration produces stable photopeak
positions across the detector and achieves an energy resolution
of 15.2% at 511 keV, while preserving the DOI-resolved flood
maps required for high-granularity event’s positioning by PET
imaging techniques.

Keywords: Terms—Depth of Interaction, energy calibration, light sharing, subsurface laser engraving, LYSO
10:15 MIC-19-06

Random Coincidence Correction for Inter-Crystal Scatter Events in PET (#2588)

H. Shim1, 2, G.B. Ko2

1 Korea Advanced Institute of Science and Technology, Department of Nuclear and Quantum Engineering, Daejeon, Republic of Korea
2 Korea Advanced Institute of Science and Technology, Mechanical Engineering Research Institute, Daejeon, Republic of Korea

Abstract

Inter-crystal scatter (ICS) events in PET, in which a 511 keV photon deposits energy across multiple crystals, offer a route to higher effective sensitivity but are conventionally discarded by photopeak windowing. Prior work has focused on positioning the first-interaction crystal, yet positioning alone does not translate ICS recovery into a system-level sensitivity gain: accepting ICS events requires lowering the lower-level discriminator, which sharply increases the singles rate and raises three issues that have not been jointly addressed, how to group raw singles into candidate ICS events, how to reject uncorrelated groupings of independent-photon singles, and whether the resulting triple coincidences are compatible with the dual-coincidence random-correction framework. This study presents a unified pipeline that addresses all three. Candidate ICS events are formed from list-mode singles and screened for type-1 randoms (false ICS) by a joint timing–distance gate, suppressing the singles-level ICS-to-PE ratio from 15.5–18% to 7–7.5% with negligible activity dependence. Validated PE and ICS events are then paired into PE–PE and PE–ICS prompt coincidences, and the singles-prompt (SP) framework is extended to ICS events by symmetric deflation of the measured ICS rate, yielding a type-resolved SP estimator for type-2 random correction. Validated through GATE simulations of the Pharos PET scanner (Brightonix Imaging; Seoul, Korea), the type-resolved SP estimator achieves R̂/Rtrue = 0.97–1.01 for PE–ICS coincidences across six activity conditions (200–667 μCi), outperforming the unified strategy and standard delayed-window and singles-rate methods. The complete pipeline delivers approximately 13% additional net true coincidences over a PE-only baseline, with essentially activity-invariant gain across the operating range examined.

AcknowledgmentThis work was supported by the National Research Foundation of Korea(NRF) grant funded by the Korea government(MSIT) (RS-2026-25490510).
References
[1] M. S. Comanor, P. R. G. Virador, and W. W. Moses, "Algorithms to identify detector Compton scatter in PET modules," IEEE Trans. Nucl. Sci., vol. 43, no. 4, pp. 2213–2218, Aug. 1996.
[2] G. Pratx and C. S. Levin, "Bayesian reconstruction of photon interaction sequences for high-resolution PET detectors," Phys. Med. Biol., vol. 54, no. 17, pp. 5073–5094, Sep. 2009.
[3] E. Lage, V. Parot, S. C. Moore, A. Sitek, J. M. Udías, S. R. Dave, M. Park, J. J. Vaquero, and J. L. Herraiz, "Recovery and normalization of triple coincidences in PET," Med. Phys., vol. 42, no. 3, pp. 1398–1410, Mar. 2015.
[4] J. F. Oliver and M. Rafecas, "Modelling random coincidences in positron emission tomography by using singles and prompts: a comparison study," PLoS ONE, vol. 11, no. 9, e0162096, Sep. 2016.
[5] I. Torres-Espallardo, M. Rafecas, V. Spanoudaki, D. P. McElroy, and S. I. Ziegler, "Effect of inter-crystal scatter on estimation methods for random coincidences and subsequent correction," Phys. Med. Biol., vol. 53, no. 9, pp. 2391–2411, 2008.
Keywords: Gate simulation, inter-crystal scatter, randoms correction, Brain PET scanner, sensitivity
10:30 MIC-19-07

A Software-Based Coincidence Sorters for Simultaneous PET Emission and Lu-176 Intrinsic Transmission Data Extraction (#1792)

X. Zhao1, Z. Wang1, Z. Xie1

1 Peking University, Beijing, China

Abstract

Conventional PET coincidence sorters focus on emission events, disregarding intrinsic Lu-176 background radiation in LYSO (Lutetium-Yttrium Oxyorthosilicate) crystals as unwanted noise. To enable simultaneous acquisition of emission and intrinsic transmission data, we developed a software-based multi-window coincidence processor capable of extracting mixed-energy coincidence events from list-mode singles data. The processor utilizes geometric constraints and a dynamic time-of-flight (TOF)-based limit to filter candidate pairs. To suppress severe 511 keV’s backscatter contamination in the 202 keV transmission window, a dual-condition rejection strategy was integrated, reducing residual crosstalk to 15.9%. Quantitative validation against Monte Carlo simulation ground truth showed good agreement in count statistics, sinograms, and radial profiles. Tomographic reconstructions successfully recovered phantom structures, validating the fidelity of the simultaneously extracted emission and intrinsic transmission coincidences.

AcknowledgmentThis work was supported by National Natural Science Foundation of China (62394311) and National Biomedical Imaging Facility Grant.
References
[1] Omidvari, N, Cheng, L, Leung, EK, et al. 2022, ‘Lutetium background radiation in total-body PET—A simulation study on opportunities and challenges in PET attenuation correction’, Frontiers in Nuclear Medicine, 2, 963067, Lausanne: Frontiers Media SA.
[2] Leung, EK, Judenhofer, MS, Cherry, SR, et al. 2018, ‘Performance assessment of a software-based coincidence processor for the EXPLORER total-body PET scanner’, Physics in Medicine & Biology, 63(18), 18NT01, Bristol: IOP Publishing.
[3] Li, T, He, L, Liu, Z, et al. 2024, ‘CT-Less Attenuation Correction for PET using Backward Inter-Crystal Scattered Events’, 2024 IEEE Nuclear Science Symposium (NSS), Medical Imaging Conference (MIC) and Room Temperature Semiconductor Detector Conference (RTSD), 1-1, New York: IEEE.
Keywords: PET, coincidence processing, intrinsic radiation
10:45 MIC-19-08

Characterisation of Positron Emission Particle Tracking (PEPT) Performance for In-vivo Blood Flow Estimation in Preclinical PET (#1869)

L. Vass1, T. Gibson1, R. De Torres1, P. K. Marsden1

1 King's College London, School of Biomedical Engineering and Imaging Sciences, London, United Kingdom

Abstract

Positron Emission Particle Tracking (PEPT) enables tracking of individual radiolabelled particles and is widely used in industrial applications (e.g. fluid dynamics), but remains underexplored for biomedical use. We recently demonstrated whole-body in vivo tracking of single nanoparticles in mice using a preclinical PET scanner, showing transit through the vasculature from the injection site. However, murine blood flow varies from a few to several hundred mm/s. Within this range, the conditions under which PEPT can reliably resolve motion remain poorly characterised. Trajectory estimation depends on a complex interplay between activity, particle velocity, acquisition settings (e.g. energy window, temporal sampling), and algorithm parameters, which have not been explored for murine blood flow estimation.

We address this by establishing a controlled framework to evaluate PEPT performance on a 3-ring PET/CT system (Mediso nanoScan 123S). A modified FLSUN 3D printer generated programmable trajectories of Ga-68 radiolabelled beads (0.4–350 kBq) at speeds of 6–170 mms⁻¹. Ground-truth motion was validated via high-frame-rate video and frame-by-frame analysis. Trajectories were reconstructed using the Birmingham method; estimated speeds and positions were compared to ground truth to quantify accuracy and localisation error.

Our results show that tracking a single particle at speeds up to 50 mms⁻¹ with RMSE <2.5 mm requires activity >1.2 kBq. At ≥0.7 kBq, speeds had 18–34% error with large RMSE. At ≤0.5 kBq, trajectories were not reliably resolved. Localisation error increased with decreasing activity and increasing speed, exceeding 10 mm RMSE in low-activity, high-speed regimes. These findings suggest activities >1.2 kBq are required for reliable intravascular speed estimation.

References
[1] Windows-Yule, C. R. K., M. T. Herald, A. L. Nicuşan, et al. “Recent Advances in Positron Emission Particle Tracking: A Comparative Review.” Reports on Progress in Physics 85, no. 1 (2022): 016101. https://doi.org/10.1088/1361-6633/ac3c4c.
[2] Pellico, Juan, Laurence Vass, Amaia Carrascal-Miniño, et al. “In Vivo Real-Time Positron Emission Particle Tracking (PEPT) and Single Particle PET.” Nature Nanotechnology 19, no. 5 (2024): 668–76. https://doi.org/10.1038/s41565-023-01589-8.
[3] Nguyen, Hieu T. M., Neeladrisingha Das, Rohollah Nasiri, and Guillem Pratx. “In Vivo Positron Emission Particle Tracking (PEPT) of Single Cells Using an Expectation Maximization Algorithm.” IEEE Transactions on Medical Imaging 45, no. 4 (2026): 1673–85. https://doi.org/10.1109/TMI.2025.3640076.
[4] Parker, D. J., C. J. Broadbent, P. Fowles, M. R. Hawkesworth, and P. McNeil. “Positron Emission Particle Tracking - a Technique for Studying Flow within Engineering Equipment.” Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment 326, no. 3 (1993): 592–607. https://doi.org/10.1016/0168-9002(93)90864-E.
Keywords: positron emission particle tracking, preclinical PET, Total-body, data analysis