MIC Poster Session 2

Session chair: Schaart, DennisR., Dr. (Delft University of Technology, Medical Physics & Technology, Delft, Netherlands); Tashima, Hideaki, Dr. (National Institutes for Quantum Science and Technology, Department of Advanced Nuclear Medicine Sciences, Chiba, Japan)
 
Shortcut: MIC-10
Date: Thursday, 12 November, 2026, 17:30 - 19:00
Location: Poster Area
Session type: MIC Session

Contents

Click on an contribution to preview the abstract content.

The underlined authors are the presenters. The authors in bold are the main authors.

17:30 Poster panel: 2

Poster Number:
MIC-10-002

Detection efficiency of a lead-loaded plastic scintillator for positron emission tomography detectors : A Monte Carlo study (#1086)

M. Yamazaki1, M. Sato2

1 Gunma Prefectural College of Health Sciences, Department of Radiological Technology, School of Radiological Technology, Gunma, Japan
2 Shinshu University, Graduate School of Medicine, Science and Technology, Nagano, Japan

Abstract

Lead-loaded plastic scintillators are promising candidates for next-generation positron emission tomography (PET) detectors, combining the fast timing response and low cost of organic scintillators with enhanced gamma-ray stopping power due to heavy-element doping. This study evaluates the 511-keV photon detection performance of EJ-256, a commercially available plastic scintillator containing 5 wt% lead, relative to a conventional polyvinyltoluene (PVT) scintillator using Monte Carlo simulations. All simulations were performed in Geant4 (v11.1.1). Monoenergetic 511-keV gamma rays were incident on scintillator blocks with a fixed cross-section of 4 × 4 mm² and thicknesses ranging from 20 to 200 mm in 5-mm increments. For each configuration, 100,000 primary events were simulated, and the resulting energy-deposition spectra were analyzed in ROOT. Detection efficiency was defined as the fraction of incident photons depositing 505–515 keV within the scintillator volume. EJ-256 exhibited a distinct fullenergy peak at 511 keV, whereas the reference PVT scintillator showed only a broad Compton continuum. Detection efficiency for EJ-256 increased from 0.9% at 20-mm thickness to 2.9% at 200 mm, consistent with enhanced photon absorption at greater thicknesses. However, efficiency gains became marginal beyond approximately 135 mm, indicating practical saturation. The energy resolution remained approximately constant at ~10% full width at half maximum across the investigated thickness range. These results indicate that moderate lead loading enhances 511-keV photon detection in plastic scintillators without compromising spectral stability, supporting the potential of EJ-256 as a cost-effective candidate for PET detector modules.

AcknowledgmentThis study was supported by the individual research fund and overseas travel grant of Gunma Prefectural College of Health Sciences.
References
[1] S. Agostinelli et al.,“Geant4—a simulation toolkit,”Nuclear Instruments and Methods in Physics Research A, vol. 506, pp. 250–303, 2003.
[2] J. Allison et al.,“Recent developments in Geant4,”Nuclear Instruments and Methods in Physics Research A, vol. 835, pp. 186–225, 2016.
[3] W. Moskal et al.,“Time-of-flight PET detector based on plastic scintillators,”Physics in Medicine & Biology, vol. 61, pp. 2025–2047, 2016.
[4] W. Moskal et al.,“Feasibility study of the J-PET detector for whole-body PET imaging,”Physics in Medicine & Biology, vol. 59, pp. 5395–5410, 2014.
[5] S. Surti and J. S. Karp,“Advances in time-of-flight PET,”Physics in Medicine & Biology, vol. 60, pp. R112–R137, 2015.
[6] P. Lecoq,“Pushing the limits in time-of-flight PET imaging,”IEEE Transactions on Radiation and Plasma Medical Sciences, vol. 1, pp. 473–485, 2017.
[7] Eljen Technology,“EJ-256 Lead-Loaded Plastic Scintillator Data Sheet,”Eljen Technology, 2023.
[8] H. Zaidi and K. S. Koral,“Scatter modelling and correction strategies in fully 3-D PET,”Nuclear Medicine Communications, vol. 26, pp. 1181–1184, 2005.
[9] S. Jan et al.,“GATE: a simulation toolkit for PET and SPECT,”Physics in Medicine & Biology, vol. 49, pp. 4543–4561, 2004.
Keywords: PET, plastic scintillator, Monte Carlo simulation, detection efficiency, Geant4
17:30 Poster panel: 5

Poster Number:
MIC-10-005

Exploring CTR improvement through structural refinement of orthogonally stacked PET detector (#1253)

A. Sanzen1, Y. Onishi1, R. Ota1

1 Hamamatsu Photonics K.K., Central Research Laboratory, Hamamatsu, Japan

Abstract

Time of flight positron emission tomography (TOF-PET) plays an important role in the diagnosis of Alzheimer’s disease and cancer because it enables localized estimation of lesion positions. Improving coincidence time resolution (CTR) is essential for enhancing the precision of lesion position estimation. Currently, commercially available PET systems achieve a CTR of approximately 200 ps full width at half maximum(FWHM), while the next goal is to reach 100 ps FWHM at the system level. However, commercial PET systems employ thick scintillators for high detection efficiency, resulting in a trade-off between CTR and detection efficiency due to photon transport time spread. To overcome this trade-off, we proposed xDetector, a readout scheme in which detector groups are orthogonally stacked. xDetector employs shorter scintillators than conventional detectors to improve CTR, while maintaining detection efficiency through multiple layers stacked. In addition, by orthogonally stacking detector groups via air coupling, it is possible to estimate the interaction position within the scintillator, which contributes to further CTR improvement. In previous study scintillator crystals and multi pixel photon counters (MPPC)s with 4.2 mm pitch were optimized for CTR measurements together with a high-frequency readout circuit. However, it was demonstrated that the CTR of xDetector with 3.2 mm pitch was 10% faster than that with 4.2 mm pitch, which can be attributed to faster time response of 3 mm MPPC than 4 mm. Thus, in this study, we aimed to further improve CTR by leveraging geometric flexibility and adopting small crystals with the same cross-sectional pitch of MPPCs compared to previous configurations. Through this investigation, we explore the potential of xDetector towards realizing 100 ps CTR.

References
[1] Y. Onishi, A. Sanzen, and R. Ota, ”Optimizing configuration of orthogonally stacked PET detectors towards 100 ps CTR with DOI-based timing correction”, IEEE NSS/MIC Yokohama, Japan 2025
[2] Y. Onishi and R. Ota, “Alleviating the trade-off between coincidence time resolution and sensitivity using scalable TOF-DOI detectors”, Phys. Med. Biol., vol 70 2025 06500
Keywords: xDetector, coincidence time resolution, multi pixel photon counter, depth of interaction
17:30 Poster panel: 8

Poster Number:
MIC-10-008

A Cost-Effective Neural Network Approach for DOI Estimation in a State-of-the-Art PET Detector (#1688)

Q. Liu1, 2, D. Arutinov3, K.G. Ciantar1, G. Wagenknecht3, N.J. Shah1, 4, C. Lerche1

1 Forschungszentrum Jülich GmbH, Institute of Neuroscience and Medicine 4, Jülich, Germany
2 Chengdu University of Technology, College of Nuclear Technology and Automation Engineering, Chengdu, China
3 Forschungszentrum Jülich GmbH, Peter Grünberg Institute 4, Jülich, Germany
4 RWTH Aachen University, Department of Physics, Aachen, Germany

Abstract

Depth-of-interaction (DOI) estimation can reduce parallax error in PET imaging, but conventional pixelated detectors usually require additional DOI-encoding hardware or modified detector geometries. This study investigates a cost-effective neural network approach to recover DOI information from a state-of-the-art air-coupled pixelated PET detector using only compact engineered features, which could be extracted directly by front-end ASICs without digitizing all silicon photomultiplier (SiPM) channels. A Geant4 optical simulation was performed, and neural networks were trained for both 3D position regression and DOI-bin classification. The feature-only model achieved DOI performance comparable to models using the full raw SiPM counts, with an MAEz of 1.107±0.001 mm, and an F1 score of 0.790 for DOI-bin classification, with most classification errors occurring between neighboring bins. These results demonstrate a practical and cost-effective pathway to enable DOI capability without changing the detector geometry, while reducing readout complexity, system cost and power consumption.

References
[1] Zaidi, H, Salimi, Y, Amini, M, Mathoux, G, Mainta, I, Garibotto, V  2025, ‘Performance Evaluation of the Biograph Vision.X: A High Time-of-Flight Coincidence Time Resolution PET/CT Scanner’, Journal of Nuclear Medicine, 66, 251754.
[2] Teimoorisichani, M, Cho, S, Sharghi, VK, Judenhofer, M, Cabello, J, Panin, V, Aykac, M, Conti, M 2023, ‘A Depth-encoding Technique for the Biograph Vision PET/CT Scanner’, 2023 IEEE Nuclear Science Symposium, Medical Imaging Conference and International Symposium on Room-Temperature Semiconductor Detectors (NSS MIC RTSD), 1–1, Vancouver, BC, Canada: IEEE.
[3] Yoshida, E, Yamaya, T 2024, ‘PET Detectors with Depth-of-Interaction and Time-of-Flight Capabilities’, Radiological Physics and Technology, 17(3), 596–609.
[4] Lerche, CW, Herrero-Bosch, V, Ferrando-Jodar, N, Gadea-Girones, R, Sanchez-Martinez, F, Mora-Más, FJ 2010, ‘Efficient Readout Electronics for Multi-Anode Photomultiplier’, SPIE Optical Engineering + Applications, 78050V, San Diego, California, United States.
Keywords: Positron emission tomography, Depth of interaction, Pixelated detector, Neural networks, Engineered features
17:30 Poster panel: 11

Poster Number:
MIC-10-011

X-ray Imaging Performance of Mg Co-doped Y3(Al,Ga)5O12:Ce for Photon-counting CT (#1759)

H. Suezumi1, 2, K. Kamada3, 4, M. Yoshino2, 3, L. Gushchina3, K. Nakanishi5, R. Murakami2, 3, S. Ishizawa2, 3, Y. Yokota2, 4, A. Yoshikawa2, 4

1 Tohoku University, Graduate School of Engineering, Sendai, Japan
2 Tohoku University, Institute for Materials Research, Sendai, Japan
3 C&A corporation, Sendai, Japan
4 Tohoku University, New Industry Creation Hatchery Center, Sendai, Japan
5 Nagoya University, Department of Integrated Health Science, Graduate School of Medicine, Nagoya, Japan

Abstract

Photon-counting computed tomography (PCCT) has attracted attention because of its potential for high spatial resolution, low-dose imaging. Although semiconductor detectors are widely studied for PCCT, indirect-conversion detectors with fast and bright scintillators combined with silicon-photomultipliers arrays (SiPMs) have recently emerged as promising alternatives owing to their potential for the scalability and cost effectiveness. Though Gd3(Al,Ga)5O12 or (Lu,Y)2SiO5 shows an excellent scintillation properties, Kα escape peak of Gd (43 keV), and Lu (55 keV) degrade the accurate photon counting. Instead, Mg co-doped Y3(Al,Ga)5O12:Ce (YAGG) garnet scintillators with lower Kα escape peak (<15 keV) is developed for X-ray imaging applications. The effect of Mg co-doping on scintillation properties results in representing 5-7 folds increase in the light yield of 35,000 photons/MeV@662 keV and the decay time of <50 ns. In addition, synchrotron X-ray imaging performance demonstrates that the spatial resolution of Mg co-doped Y3Al3Ga2O12:Ce is comparable to that of LuAG:Ce with the 1 mm sample thickness. Although the density of YAGG:Ce,Mg (5.08–5.33 g/cm) is lower than that of LuAG:Ce (6.67 g/cm), there is a concern regarding X-ray stopping power and deterioration of spatial resolution, the linear attenuation coefficients at 12.4 keV are sufficiently high (273–350 cm−1 for YAGG:Ce,Mg and 943 cm−1 for LuAG:Ce), resulting in X-ray absorption efficiencies exceeding 99.99% for 1 mm thickness samples, indicating negligible influence on imaging performance in the lower energy range. Moreover, we also demonstrate the feasibility of YAGG:Ce,Mg to medical field from the success in clear imaging of small, dried fish. These results show that YAGG:Ce,Mg scintillators are promising candidates for next-generation X-ray imaging detectors and can be applied for indirect-conversion PCCT applications.

References
[1] M. Nikl and A. Yoshikawa, "Recent R&D Trends in Inorganic Single‐Crystal Scintillator Materials for Radiation Detection," Advanced Optical Materials 3, no. 4 (2015): 463–481.
[2] M. J. Willemink, M. Persson, A. Pourmorteza, N. J. Pelc, and D. Fleischmann, "Photon-Counting CT: Technical Principles and Clinical Prospects," Radiology 289, no. 2 (2018): 293–312.
[3] K. Kamada, T. Endo, K. Tsutumi, T. Yanagida, Y. Fujimoto, A. Fukabori, et al., "Composition Engineering in Cerium-Doped (Lu, Gd)3(Ga, Al)5O12 Single-Crystal Scintillators," Crystal Growth & Design 11, no. 10 (2011): 4484–4490.
[4] K. Shimazoe, D. Kim, M. Hamdan, Y. Kobayashi, K. Kamada, M. Yoshino, et al., "An energy-resolving photon-counting X-ray detector for computed tomography combining silicon-photomultiplier arrays and scintillation crystals." Communications Engineering 3.1 (2024): 167.
[5] H. Suezumi, K. Kamada, L. Gushchina, M. Yoshino, K. J. Kim, S. Ishizawa, et al., "Bulk Single Crystal Growth and Scintillation Properties of Ce and Mg Co-Doped Y3Ga3Al2O12 for Advanced X-ray Imaging," Scientific Reports (2026).
[6] M. Yoshino, S. Yamamoto, K. Nakanishi, K. Yogo, K. Kamada, N. Koshikawa, et al., "Development and Performance Evaluation of a Thin GAGG: Ce Scintillator Plate for High Resolution Synchrotron Radiation X-ray Imaging," Journal of Instrumentation 19, no. 10 (2024): P10030.
Keywords: Radiation imaging, Scintillators, Photon-counting computed tomography, Synchrotron radiation
17:30 Poster panel: 14

Poster Number:
MIC-10-014

Optimal Radiator Refractive Index for Cherenkov-Radiator-Integrated PET Detectors: A Simulation Study (#1942)

T. Ishikawa1, 2, R. Ota3, 4, H. Haneishi5, T. Yamaya2, 5

1 Chiba University, Graduate School of Science and Engineering, Chiba, Japan
2 National Institutes for Quantum Science and Technology, Institute for Quantum Medical Science, Chiba, Japan
3 Hamamatsu Photonics K.K., Central Research Laboratory, Hamamatsu, Japan
4 University of California, Davis, Department of Biomedical Engineering, Davis, California, United States of America
5 Chiba University, Center for Frontier Medical Engineering, Chiba, Japan

Abstract

In positron emission tomography, time-of-flight information can localize source positions along lines of response. Cherenkov-radiator-integrated microchannel-plate photomultiplier tubes (CRI-MCP-PMTs) with a 3.2-mm-thick lead glass window (n~1.5) have achieved 30-ps coincidence time resolution (CTR). However, improving coincidence detection efficiency remains a challenge, as increasing the radiator thickness degrades the CTR. In this study, we focused on the radiator refractive index. Based on the Frank-Tamm formula, we hypothesized that ultraviolet-transparent materials with a high refractive index would increase Cherenkov photon production, reducing event loss. Furthermore, increased Cherenkov photon detection would improve effective single-photon time resolution, thus enhancing CTR. We tested this hypothesis using Geant4 simulations of point-source measurements with paired CRI-MCP-PMTs incorporating optical and detector characteristics. Assuming wavelength-independent refractive indices from 1.7 to 3.0, we found coincidence detection efficiency increased from 1.1% to 3.0% whereas CTR deteriorated from 41.5 to 49.6 ps. Contrary to our expectations, the reduced speed of light in the medium increased timing jitter, overriding the CTR benefits from detecting more Cherenkov photons. A similar simulation for lead fluoride radiators (n~1.7–2.1)—expected to improve efficiency while maintaining the CTR—showed a worse CTR (46.4 ps) was obtained than for the wavelength-independent cases (~42.0 ps). Wavelength-resolved analysis showed a CTR of 53.3 ps for 240–340 nm photons, which was unexpectedly worse than that of 39.9 ps for 540–640 nm photons, despite the higher theoretical photon production in the shorter band. One possible cause is the steep refractive index dispersion in the shorter band. In summary, while higher refractive indices improve coincidence detection efficiency, they can degrade CTR. Minimizing refractive index dispersion is critical for fast timing.

References
[1] R. Ota et al., “Coincidence time resolution of 30 ps FWHM using a pair of Cherenkov-radiator-integrated MCP-PMTs,” Phys. Med. Biol., vol. 64, no. 7, p. 07LT01, Mar. 2019.
[2] R. Ota et al., “Lead-free MCP to improve coincidence time resolution and reduce MCP direct interactions,” Phys. Med. Biol., vol. 66, no. 6, p. 064006, Mar. 2021.
[3] T. Ishikawa et al., “Imaging Simulation of a Dual-Panel PET Geometry With Ultrafast TOF Detectors,” IEEE Trans. Radiat. Plasma Med. Sci., vol. 10, no. 3, pp. 418–426, Mar. 2026.
[4] T. Ishikawa et al., “Monte Carlo Simulation of Cherenkov-Radiator-Integrated MCP-PMT with Full Modeling of Optics and Waveform,” in 2025 IEEE NSS-MIC-RTSD, Nov. 2025.
[5] R. Ota et al., “Timing-performance evaluation of Cherenkov-based radiation detectors,” Nucl. Instrum. Methods Phys. Res. A, vol. 923, pp. 1–4, Apr. 2019.
[6] N. Kratochwil et al., “A roadmap for sole Cherenkov radiators with SiPMs in TOF-PET,” Phys. Med. Biol., vol. 66, no. 19, p. 195001, Sep. 2021.
[7] R. D. Shannon et al., “Refractive Index and Dispersion of Fluorides and Oxides,” J. Phys. Chem. Ref. Data, vol. 31, no. 4, pp. 931–970, Oct. 2002.
Keywords: Positron emission tomography, time-of-flight PET detector, Monte Carlo simulation, Cherenkov radiator, microchannel-plate photomultiplier tube
17:30 Poster panel: 17

Poster Number:
MIC-10-017

Fabrication of diamond detectors with graphitic contacts for tissue equivalent dosimetry (#2040)

I. Lopez Paz1, C. Guardiola1, G. Rius1

1 Institut de Microelectronica de Barcelona (IMB-CNM, CSIC), Bellaterra, Spain

Abstract

Diamond-based dosimeters are popular for medical dosimetry in view of their fast response, radiation hardness and tissue equivalence. However, some design improvements to achieve better accuracy in signal readout and approach tissue equivalence are possible by the removal of the metallic contacts, sometimes covering the active area of said devices.
In this contribution, we explore the use of a simple processing and scalable solution to obtain non-metal, low Z electrical contact. Thus, the fabrication of diamond detectors with graphitic contacts at the Microelectronics Institute of Barcelona are shown. Several approaches are explored, some of which never utilised before for radiation detector fabrication, namely pyrolytic photoresist film, nickel catalysis and ion surface graphitisation.
A prototype of such device using the pyrolytic photoresist film process for the first time has been produced and was tested under a 60 kV X-ray tube. We report the fabrication and preliminary performance results of such device. Although the substrate material quality is not ideal for radiation detection, it was possible to observe radiation response using the PPF as an interface material. The prototype features a linearity better than 3% as tested in a laboratory X-ray chamber, proving it a promising technology for detector operation. Finally, the technology challenges are finally discussed to assess the feasibility of PPF-diamond technology for dosimeter batch production.

AcknowledgmentThis work has been funded by the IMB-CNM (CSIC) TRIGGER projects PlACeD and CarDos4FLASH, supported by the ‘María de Maeztu Unit of Excellence’ accreditation (CEX2023-001397-M / AEI/10.13039/501100011033). This study was also funded RYC2024-050777-I project, financed by the MICIU / AEI / 10.13039/501100011033 and FSE+.
 
Keywords: Diamond, radiation detectors, dosimetry
17:30 Poster panel: 20

Poster Number:
MIC-10-020

Experimental Evaluation of Classification MLP for Position Estimation in Semi-Monolithic LYSO Cassettes for Ultra-High-Performance Brain PET (#2111)

E. Elmoujarkach1, F. Pagano2, F. Loignon-Houle2, A. J. Gonzalez2, D. Sanchez3, J. V. Catret3, M. Sabuncu1, 4, N. A. Karakatsanis1, S. A. Nehmeh1

1 Weill Cornell Medical College, Department of Radiology, New York, New York, United States of America
2 CSIC-Universitat Politècnica de València, Institute for Instrumentation in Molecular Imaging (I3M), Valencia, Spain
3 Oncovision, Valencia, Spain
4 Cornell University, School of Electrical and Computer Engineering, Ithaca, New York, United States of America

Abstract

Accurate localization of annihilation γ-ray interaction positions within detector block is essential for achieving high spatial resolution in positron emission tomography (PET). Conventional high-resolution PET detectors often use finely pixelated scintillator arrays resulting in increased system cost due to the larger number of crystal elements and number of readout-channels. Semi-monolithic scintillator detectors offer an alternative solution by enabling signal-based position estimation with reduced readout. In this work, we experimentally evaluated a classification-based multilayer perceptron (class-MLP) for estimating the γ-photons interaction position direction within each semi-monolithic LYSO detector slab across a cassette of 40 mini-modules (mM), each comprised of 16 LYSO slabs coupled to 8x8 SiPMs and read by 16 readout channels, developed for the Ultra-High-Performance Brain PET (UHB-PET) system. Calibration data were acquired from each of the 40 mM, and operated on a benchtop setup using a 0.5mm slit-collimated source translated in 0.5mm steps. The class-MLP used 24 multiplexed readout and classified events into 49 position bins spanning 0-24mm along the monolithic. Independent MLPs were trained for the individual mM. After correction for the finite collimator resolution, the mean positioning-error FWHM was 1.29±0.13mm (range: 0.85 to 1.41mm). The corresponding FWTM was 3.91±0.29mm (range: 3.29 to 4.73mm). Our results demonstrate that class-MLP can achieve near-millimeter detector mean spatial resolution along the monolithic dimension in 20mm-thick semi-monolithic LYSO detectors with minimal variability across all mM of a UHB-PET cassette. Future work will focus on acquiring calibration data from all cassettes in the fully assembled UHB-PET scanner to characterize its system-wide detector spatial resolution using the current calibration strategy per mM.

AcknowledgmentThis work was supported by the National Institutes of Health (NIH) under award number 1R01CA275942-01A1.
References
[1] Du, J., and Du, S., 'Performance Comparison of DOI-Encoding PET Detectors Based on 1.1-mm Pitch BGO Arrays With Different Reflectors', IEEE Trans. Radiat. Plasma Med. Sci., vol. 8, no. 3, pp. 257-262, 2024
[2] LaBella, A. et al., 'High-Resolution Depth-Encoding PET Detector Module with Prismatoid Light-Guide Array', J. Nucl. Med 61 (10) 1528-1533, 2020
[3] Karakatsanis, NA, et al, 'Monte Carlo Evaluation of the Ultra-High Physical Performance Brain PET (UHB-PET) scanner with semi-monolithic detectors', IEEE NSS/MIC/RTSD, 2024
[4] Pagano, F., et al., 'Semi-Monolithic Detectors for TOF-DOI Brain PET: Optimization of Time, Energy, and Positioning Resolutions With Varying Surface Treatments', IEEE Trans. Radiat. Plasma Med. Sci., vol. 10, no. 2, pp. 276-287, 2026
[5] Elmoujarkach, E, et al., 'Sub-Millimeter 2D Intrinsic Resolution Recovery in Semi-Monolithic PET Detector using Neural Networks - a Monte Carlo Simulation', IEEE NSS/MIC, 2025
[6] Pagano, F, et al., 'Performance evaluation of a multiplexing circuit combined with ASIC readout for cost-effective brain PET imaging', Phys. Med. Biol. 70 205001, 2025
[7] Freire, M., et al., 'Position estimation using neural networks in semi-monolithic PET detectors', Phys. Med. Biol. 67 245011, 2022
[8] Loignon-Houle, F., et al,'Multi-Input Interaction Positioning and Signal Demultiplexing With Deep Learning in Semi-Monolithic Detectors', IEEE NSS/MIC/RTSD, 2024
Keywords: UHB-PET, brain PET, semi-monolithic detectors, neural network, detector resolution
17:30 Poster panel: 23

Poster Number:
MIC-10-023

AI-based DOI determination for hexagonal monolithic LYSO scintillators with hexagonal cells engraved with laser: A simulation/experiment study (#2234)

M. Abushawish1, J. Udias1, 2, C. Golhen1, C. Galindo-Morata1, P. Galve3, 1, V. Sánchez-Tembleque4, L. J. Algarra4, D. Pérez Benito4, L. Batres4, J.J. Vaquero4

1 Universidad Complutense de Madrid, Grupo de Física Nuclear, EMFTEL & IPARCOS, Madrid, Spain
2 Hospital Clínico San Carlos, Health Research Institute of the Hospital Cínico San Carlos (IdISSC), Madrid, Spain, Madrid, Spain
3 Spanish National Research Council (CSIC), Institute for Physical and Information Technologies “Leonardo Torres Quevedo”, ITEFI,, Madrid, Spain
4 Universidad Carlos III de Madrid, Departamento de Bioingeniería, Universidad Carlos III de Madrid, Madrid, Spain

Abstract

Laser engraving of monolithic scintillators to create pixel boundaries has been around already for some time, with the promise of good pixel identification as in conventional pixelated detectors, but at a much lower cost. However, their use is hindered by the fact that the pixel boundaries are semi-transparent, making it difficult to identify the pixel of interaction with the algorithms applied to fully pixelated detectors.   The RETIMAGER project proposes laser-engraved (LE) pixelated LYSO blocks, with optimized transparency of pixel boundaries that might be exploited to obtain depth-of-interaction (DOI) based on the light spread in the SiPM array, while keeping crystal pixel identification. Hexagonal SiPM arrays, scintillators and pixel cells are introduced, improving packing fraction and reducing gaps in between detectors, but adding difficulty when applying ordinary vertex identification strategies. We present the experimental DOI performance  for several RETIMAGER detectors, engraved with different optical densities (and thus trasparencies) por hexagons of 1.25 mm side. Data is acquired in a testbench with a collimated source, and employed to train an NN algorithm, to predict DOI. For a 13 mm thick crystal, DOI assignation uncertainty ranges from 2.4 to 3.2 mm FWHM, depending on the trasparency of the boundaries, with the highest trasparency detectors exhibiting the best results. Experimental results follow the trend of the simulations, which can be used to further optimize the DOI performance and to reduce bias.

AcknowledgmentWe acknowledge support from Spanish Government and Next GenerationEU Recovery and Resilience Facility (RRF)  CPP 2021-008751 NEWMBI; European Union, the European Innovation Council's Pathfinder Open Programme: RETIMAGER, 101099096;  Spanish Government and Next Generation EU/PRTR: DI2M: Smart detectors for molecular imaging (2021/C005/00147498); P. Galve acknowledges grant JDC2023-051754-I, funded by MCIU/AEI/10.13039/501100011033 and ESF+.
Keywords: PET, Monte Carlo Simulation, Depth of interaction, DOI, SSLE
17:30 Poster panel: 26

Poster Number:
MIC-10-026

Exploring the Performance of Light-Sharing Based Single-ended BGO DOI-PET Detectors (#2339)

Y.H. Sul1, M. Yi2, 3, G.B. Ko4, J.S. Lee1, 5

1 Seoul National University, Department of Nuclear Medicine, Seoul, Republic of Korea
2 Seoul National University, Interdisciplinary Program in Bioengineering, Seoul, Republic of Korea
3 Seoul National University, Integrated Major in Innovative Medical Science, Seoul, Republic of Korea
4 Korea Advanced Institute of Science and Technology, Department of Nuclear and Quantum Engineering, Daejeon, Republic of Korea
5 Brightonix Imaging Inc., Seoul, Republic of Korea

Abstract

Bismuth germanate (BGO) is a high-density and cost-effective scintillator for PET detectors, but depth-of-interaction (DOI) encoding with single-ended readout remains challenging due to its relatively low light yield. This study investigated a light-sharing based BGO detector using paired crystals for DOI encoding. The effects of reflector geometry and lateral surface treatment were evaluated to optimize detector performance. Among the unpolished configurations, the trapezoidal reflector provided the best DOI encoding capability, achieving an average DOI resolution of 5.29 mm FWHM and an energy resolution of 20.4%. Partial surface polishing improved energy and timing performance but degraded DOI encoding capability. After applying timing correction based on DOI and rise-time correction, the timing resolution of the unpolished trapezoidal configuration improved from 427 ps to 353 ps FWHM. These results suggest the feasibility of a single-ended DOI approach for BGO PET detectors with potential time-of-flight (TOF) applications.

Acknowledgment

This work was supported by the National Research Foundation of Korea(NRF) grant funded by the Korea government(MSIT) (Grant No. RS-2024-00354123).

References
[1] M. Ito, S. J. Hong, and J. S. Lee, “PET detectors with depth-of-interaction capability,” Biomed. Eng. Lett., vol. 1, no. 2, pp. 70–81, 2011.
[2] J. Du, “Performance of dual-ended readout PET detectors based on BGO arrays and BaSO₄ reflector,” IEEE TRPMS, vol. 6, no. 5, pp. 522–528, 2021.
[3] R. S. Miyaoka, T. K. Lewellen, H. Yu, and D. L. McDaniel, “Design of a DOI PET detector module,” IEEE TNS, vol. 45, no. 3, pp. 1069–1073, 1998.
[4] J. W. Cates, S. Gundacker, E. Auffray, P. Lecoq, and C. S. Levin, “Improved SPTR for analog SiPMs with front-end readout that reduces influence of electronic noise,” PMB, vol. 63, no. 18, Art. no. 185022, 2018.
[5] M. Yi, G. B. Ko, and J. S. Lee, “Pushing the limit of BGO-based dual-ended Cherenkov PET detectors through photon transit time correction,” PMB, vol. 69, no. 2, Art. no. 025005, 2024.
[6] N. Kratochwil, S. Gundacker, P. Lecoq, and E. Auffray, “Pushing Cherenkov PET with BGO via CTR classification and correction,” PMB, vol. 65, no. 11, Art. no. 115004, 2020.
Keywords: Bismuth germanate, depth-of-interaction, single-ended readout, light sharing, timing correction
17:30 Poster panel: 29

Poster Number:
MIC-10-029

Experimental Evaluation of Coincidence Time Resolution in LYSO Scintillators with Photonic Crystal Pattern (#2465)

S. Hyeon1, S. Choi2, G.B. Ko2, M.S. Lee1

1 Daegu Gyeongbuk Institute of Science and Technology (DGIST), Electrical Engineering and Computer Science, Daegu, Republic of Korea
2 Korea Advanced Institute of Science and Technology (KAIST), Nuclear and Quantum Engineering, Daejeon, Republic of Korea

Abstract

Time-of-flight (TOF) positron emission tomography (PET) performance is governed by the coincidence time resolution (CTR), which is fundamentally constrained by the number and arrival-time distribution of scintillation photons reaching the photosensor. To enhance both quantities, we fabricated nano-scale line- and hole-type photonic crystal (PhC) patterns with a 1.2 μm period on the polished 3×3×20 mm³ surfaces of lutetium-yttrium oxyorthosilicate (LYSO) scintillators via flex-mold soft lithography using a UV-curable optical resin (n = 1.5). Six SiPM coupling conditions were evaluated: bare LYSO with and without grease, and line- and hole-type PhC patterns each with and without grease.

The intrinsic scintillation photon arrival-time distribution was reconstructed by deconvolving each measured waveform with the single-photoelectron (SPE) response of a photomultiplier tube (PMT), and the absolute light gain of each PhC geometry was independently assessed in an airgap PMT configuration by comparing the photopeak before and after patterning. PhC-patterned crystals achieved a 42.8% reduction in deconvolved rise time and light gains of 1.20 (line) and 1.18 (hole). To quantify the redistribution of photon arrivals, fractional yields were computed within three time windows — early (0–1 ns), mid (1–10 ns), and late (10–50 ns) — normalized to the total yield over 0–50 ns. The early-window yield increased from 0.7% to 0.9% (a 33.5% relative gain), the mid-window yield rose from 19.8% to 21.2%, and the late-window yield decreased from 79.5% to 77.9%, confirming a redistribution of photon arrivals toward the earlier windows.
In SiPM CTR measurements, the line-type PhC combined with grease achieved the best CTR of 180.6 ps, a 29.5% improvement over the bare crystal. These findings demonstrate that nano-scale PhC patterning improves CTR in LYSO-based TOF-PET by enhancing both the absolute light yield and the prompt photon density.

AcknowledgmentThis work was supported by National Research Foundation of Korea (Grant No. RS-2025-02316993).
References
[1] Hyeon, S, et al., 2025, 'Experimental Evaluation of Flex-Molded Nanopatterns and Coupling Interfaces for Light Extraction Improvement for Scintillator-based PET detectors', 2025 IEEE NSS/MIC Conf. Record
Keywords: LYSO, Photonic crystal, Coincidence time resolution
17:30 Poster panel: 32

Poster Number:
MIC-10-032

X-ray fluorescence imaging for breast cancer characterization within the IMMPRINT project (#2531)

K. Brzeziński1, S. Parashari1, O. Elesgaray1, K. Kumar2, C. Hoeschen2, C. Lacasta1, M. Mikuz3, A. Studen3, J. Treis4, J. Ninković4, G. Llosá1

1 Instituto de Física Corpuscular (IFIC), CSIC-UV, Valencia, Spain
2 Otto von Guericke University, Magdeburg, Germany
3 Jozef Stefan Institut, Ljubljana, Slovenia
4 MPG Halbleiterlabor, Garching, Germany

Abstract

X-ray fluorescence imaging (XFI) is an emerging molecular imaging technique that combines high sensitivity with high spatial resolution, offering significant penetration depth in tissues while minimizing ionizing radiation exposure compared to other existing methods. This makes XFI a promising tool for tumor characterization and personalized treatment. The IMMPRINT project aims to develop a proof-of-principle demonstrator that leverages X-ray fluorescence computed tomography as a novel hybrid modality for tumor profiling, with a focus on breast cancer. The pnCCD silicon sensors produced by MPG HLL, Munich, have been chosen for the proof-of-principle studies. Fluorescence photons are generated when gold nanoparticles are excited by incident X-rays and are detected by the pnCCD. A specific system has been designed to operate up to four pnCCDs, including a cooling system to stabilize sensors’ temperature between 0ºC and 20ºC, and it is under construction. Simulations are also performed which support the system design.

AcknowledgmentThis work has received funding from the European Union’s EURATOM research and innovation program, IMMPRINT Project (G.A. 101061037). S.P. acknowledges the Severo Ochoa fellowship (CEX2023-001292-S grant), funded by MCIN/AEI/10.13039/501100011033.
References
[1] Staufer, T and Grüner, F. 2023 Int. J. Mol. Sci. 24(13), 10990.
[2] Kumar, K. et al. 2024 Int. J. Mol. Sci. 25(9), 4733.
[3] Kumar, K. et al. 2024 Proc. SPIE 13924, 661–667
[4] Meidinger N. et al. 2006 Nuc. Instr. Meth. A. vol 568, 141-148
Keywords: X-ray fluorescence, pnCCD, XFI imaging, cooling system, IMMPRINT project
17:30 Poster panel: 35

Poster Number:
MIC-10-035

Characterization of DOI-enabled Three-Layer Trapezoidal Pixelated LYSO Detector (#2616)

S. Parashari1, K. M. Popov2, F. Hueso-Gonzalez1, M. Nitta2, P. G. Thirolf2, K. Parodi2, G. Llosa1

1 Instituto de Fisica Corpuscular (IFIC), CSIC-UV, Valencia, Spain
2 Ludwig-Maximilians-Universität, Department of Physics, Munich, Germany

Abstract

In the context of the “Small Animal Proton Irradiator for Research in Molecular Image-guided Radiation Oncology” (SIRMIO) project, a high-resolution depth-of-interaction (DOI) LYSO-based detector has been developed for a small-animal in-beam PET scanner. In addition to PET applications, the DOI detector may also be suitable for the detection of high-energy prompt gamma rays originating from excited nuclei such as 12C and 16O, enabling potential applications in Compton camera imaging for ion-beam therapy monitoring. However, the interaction mechanisms of high-energy gamma rays differ from those of 511 keV annihilation gamma rays,  particularly in terms of the relative contributions of photoabsorption and Compton scattering processes. In this study, we experimentally investigated the performance of the DOI detector for high-energy gamma-ray detection to gain a better understanding of its applicability to prompt gamma imaging. The detector consists of a three-layer pixelated scintillation block coupled to a 64-channel silicon photomultiplier array. The use of the PETsys TOFPET2 ASIC, employed in the Compton camera MACACOp developed by the IRIS group, is a potential alternative to the current electronics employed in the SIRMIO project, which enables individual channel readout. The study demonstrates the feasibility of operating the detector with PETsys electronics, and characterizes its response with 60Co and 22Na sources.

AcknowledgmentThis work was supported by the Spanish MICIN/AEI (MIDAS, PID2022-143246OB-I00). SP acknowledges support from the Severo Ochoa grant CEX2023-001292-S (MCIN/AEI/10.13039/501100011033). KMP acknowledges funding from the European Union’s Erasmus+ programme. LMU members acknowledge support from ERC grant 725539 and from T. Yamaya and H.G. Kang (NIRS/QST) for the LMU detector development.
References
[1] G. Lovatti, et. al., PMB, 68 (23), (2023) 235005.
[2] T. Binder, et. al., Front. Phys. 10 (2022) 954204.
[3] R.Viegas, et. al., Rad. Phys. Chem. 202 (2023) 110507.
Keywords: PET, Compton camera, DOI detector
17:30 Poster panel: 38

Poster Number:
MIC-10-038

PETRARCA: a Positron Emission Tomography Real-Time Analog Re-Configurable ANN Detector (#2827)

M. Pandocchi1, 2, M. Amadori1, 2, M. Ronchi1, 2, G. Borghi1, 2, M. Carminati1, 2, C. Fiorini1, 2

1 Politecnico di Milano, Dipartimento di Elettronica Informazione e Bioingegneria, Milano, Italy
2 Istituto Nazionale di Fisica Nucleare, Sezione di Milano, Milano, Italy

Abstract

Positron emission tomography systems are becoming increasingly complex due to the growing number of detector channels and the large amount of data generated, especially in total-body imaging configurations. In this work, we present PETRARCA, a positron emission tomography detector system integrating an analog neural network ASIC (ANNA) for real-time edge computing of gamma-ray interaction coordinates within a monolithic scintillation crystal. The system consists of a 32 mm × 32 mm × 10 mm monolithic LYSO crystal coupled to an 8 × 8 SiPM matrix read out by dedicated front-end electronics. The detector architecture includes both a conventional digital acquisition path and a parallel analog inference path based on the ANNA ASIC, enabling simultaneous acquisition of sensor data and hardware neural-network predictions. A first evaluation of the system has been carried out using simulated data and the real ASIC response to emulate a reduced-scale configuration using a 4 × 4 SiPM array coupled to a 26 mm × 26 mm × 10 mm monolithic crystal. The performances of the ASIC in predicting the interaction coordinates were tested, and proved to have a mean absolute error (MAE) of around 1 mm on both the x and y axes. The preliminary results demonstrate the feasibility of analog neural-network edge computing for spatial localization in PET detectors and support further developments toward fully integrated real-time detector systems.

References
[1] Nanni et al., “Total body positron emission tomography/computed tomography: Current status in oncology,” Semin. Nucl. Med., vol. 55, no. 1, pp. 31–40, 2025
[2] S. Di Giacomo et al., “Implementing an Integrated Neural Network for Real-Time Position Reconstruction in Emission Tomography With Monolithic Scintillators,” IEEE Trans. Radiat. Plasma Med. Sci., vol. 8, no. 5, pp. 501-510, May 2024.
[3] A. Gonzalez-Montoro et al., “Evolution of PET detectors and event positioning algorithms using monolithic scintillation crystals,” IEEE Trans. Radiat. Plasma Med. Sci., vol. 5, pp. 282–305, 2021.
[4] I. D’Adda et al., “SITH: An SiPM Readout ASIC for Prompt Gamma-Ray Detection in Hadrontherapy Applications,” IEEE Trans. Nucl. Sci., vol. 71, no. 9, pp. 2140–2150, 2024.
Keywords: PET, Neural Networks, Edge Computing, Gamma-rays
17:30 Poster panel: 41

Poster Number:
MIC-10-041

Scintillation Properties of (Lu,Y)₂O₃:La Ceramic Pixels for PET (#2881)

J. Glodo1, S. Kim1, A. Kim1, M. Muller1, C. Sosa1, J. Xie2, S. R. Cherry2, J. Du2

1 Radiation Monitoring Devices, Inc., Watertown, Massachusetts, United States of America
2 University of California, Davis, Department of Biomedical Engineering, Davis, California, United States of America

Abstract

Lutetium oxide–based scintillators are of growing interest for positron emission tomography (PET) because of their high density and high effective atomic number, which provide efficient stopping of 511 keV annihilation photons. Lanthanum-doped (Lu,Y)2O3 (LYO:La) is a transparent ceramic variant that retains the favorable stopping power of Lu2O3 while being easier to fabricate, making it an attractive candidate for next-generation PET detector arrays. In this work we report a scintillation characterization of LYO:La pixels in two geometries relevant to PET, 1×1×10 mm3 for high-resolution preclinical and dedicated-organ imaging, and 3×3×10 mm3 for whole-body clinical systems. Pixels were cut from LYO:La ceramic disks, polished on all faces, and evaluated with a range of side reflectors (Teflon, Tyvek, ESR film, and Spectralon) in both flat and upright coupling orientations. We measured X-ray-excited radioluminescence emission spectra, 22Na (511 keV) pulse height spectra, relative light yield, energy resolution, scintillation decay time profiles, and timing resolution. The results illustrate the influence of pixel geometry, coupling face, and reflector choice on light collection and energy resolution, and show that LYO:La pixels offer a promising combination of stopping power, light yield, and timing performance for PET detector modules.

AcknowledgmentThis work was supported by the US NIH NIBIB through competitively awarded contracts number 4R44MH135525-02.
Keywords: Lu2O3, Ceramic Scintillator, PET, Positron Emission Tomography, Medical Imaging
17:30 Poster panel: 44

Poster Number:
MIC-10-044

Ultrahigh Spatial Resolution Perovskite-based Direct X-ray Imagers for Low-lag Medical Applications (#3045)

B. Dunham1, S. Motakef1, K. Yang2, R. Gupta2, A. Datta1

1 CapeSym, Inc., Research and Developement, Natick, Massachusetts, United States of America
2 Massachusetts General Hospital, Department of Radiology, Boston, Massachusetts, United States of America

Abstract

Dynamic medical X-ray imaging requires detectors that combine high quantum efficiency, low residual lag, and high spatial resolution without the optical spread and afterglow of scintillator-based flat panels. We report a fully inorganic perovskite-based direct-conversion perovskite X-ray imager fabricated from phase-pure orthorhombic Cesium Lead Bromide (CPB) sensor material and integrated with active-pixel readout arrays. Phase-pure CPB was obtained by controlled crystal growth and converted into dense blade-coated films; non-phase-pure starting materials containing secondary phases produced nonuniform films and degraded images, confirming the importance of phase control for reproducible detector performance. The optimized CPB detector demonstrated high linear X-ray response, sensitivity of 55,400 μC Gyair⁻¹ cm⁻² at 30 keV and 0.5 V/μm, a limit of detection of 417 pGyair s⁻¹, a mobility-lifetime product of 8.0 × 10⁻⁴ cm² V⁻¹, and a calculated detective quantum efficiency near 85% at practical dose levels. In flat-panel imaging, CPB imagers reached 6.9 lp/mm at 20% modulation transfer function on a 100 μm-pitch amorphous-silicon array and resolved 20 lp/mm after integration with a CMOS readout. Most importantly for low-lag medical applications, improvements in phase purity and charge-transport stability reduced residual detector lag from 8.9% to 1.1%. These results show that all-inorganic perovskite direct conversion can provide a path to low-dose, low-lag, high-resolution imaging for fluoroscopy, tomosynthesis, dental cone-beam computed tomography, and other image-guided procedures.

Acknowledgment

This research was supported by the U.S. National Institutes of Health under award R44EB038123.

Keywords: direct conversion flat-panel detector, low lag, medical imaging, perovskite, X-ray imaging.
17:30 Poster panel: 47

Poster Number:
MIC-10-047

Timing and Spatial Properties of HRPPD coupled to LGSO/LYSO for TOF-PET Applications (#3103)

Y. Wang1, M. Popecki2, H. Kim1, A. Lyashenko2, M. Nemallapudi1, M. Minot2, C.-M. Kao1, C.-T. Chen1

1 University of Chicago, Radiology, Chicago, Illinois, United States of America
2 Incom Inc., Research and Development, Charlton, Massachusetts, United States of America

Abstract

Large-area microchannel plate photomultipliers (MCP-PMT) such as the Large-Area Picosecond Photon Detector (LAPPD) and High-Rate Picosecond Photon Detector (HRPPD), are attractive for time-of-flight positron emission tomography (TOF-PET) due to their superior single-photon time resolution (SPTR) and low-cost potential. However, when coupling them with popular TOF-PET scintillators, it was previously reported worse-than 250 ps coincidence time resolutions (CTR) for annihilation photons. In this work, by applying a custom high-frequency (HF) amplifier to the outputs of an HRPPD (10×10cm2 window) coupled to a 3×3×3 mm³ LGSO crystal and an SiPM-based reference detector, we measured a 144 ps CTR. In addition, we showed that individual crystals of a 4×4 array of 3×3×10 mm³ LYSOs can be resolved using this HRPPD whose output was 4×4 pads having a 25 mm pitch. We also constructed a detector model for investigating key factors affecting CTR. These factors include scintillation yield, light transport in the crystal and HRPPD window, photon detection efficiency and charge collection of the HRPPD, and electronics contributions, with parameters derived from analytical calculations, Geant4 simulations, or experimental measurements. This model predicted a CTR that was consistent with the measured value. Based on this model, a development roadmap with quantitative performance targets and near-term milestones is presented.

References
[1] Lecoq, P 2017, 'Pushing the Limits in Time-of-Flight PET Imaging', IEEE Transactions on Radiation and Plasma Medical Sciences, 1(6), pp. 473–485.
[2] Lyashenko, AV et al. 2026, 'HRPPD photosensors for RICH detectors with a high resolution timing capability', Nuclear Instruments and Methods in Physics Research A, 1082, 170964.
[3] Korpar, S et al. 2024, 'Timing Resolution of a LAPPD Prototype Measured with CERN PS Test Beams', Nuclear Instruments and Methods in Physics Research A, 1069, 169838.
[4] Minot, MJ et al. 2025, 'Advancing HRPPD Technologies for Next-Generation High-Performance TOF-PET Systems', IEEE NSS/MIC.
[5] Zobundzija, Z et al. 2025, 'Pixel and Transmission Line Readouts in Capacitive-Coupled HRPPD MCP-PMTs for High-Precision Timing', IEEE NSS/MIC.
Keywords: Coincidence time resolution (CTR), flood map, high rate picosecond photon detector (HRPPD), microchannel plate photomultiplier tube (MCP-PMT), time-of-flight positron emission tomography (TOF-PET).
17:30 Poster panel: 50

Poster Number:
MIC-10-050

Analytical modeling and inclusion of object scatter in non-TOF 3D PET for low-count data (#1275)

J. Roser1, H. P. Vo1, M. Rafecas1

1 Universität zu Lübeck, Institute of Medical Engineering, Lübeck, Schleswig-Holstein, Germany

Abstract

Compton scattering in Positron Emission Tomography (PET) is commonly seen as an important degrading factor for the image quality. In particular, object scattering is known to reduce  image contrast and distort quantification. Conventional scatter correction techniques do not model scatter directly in the system matrix, despite its potential to better handle low-count data scenarios. In this work, we propose an analytical model for single Compton scattering events and subsequently derive expressions of the system matrix and sensitivity. We validate the obtained expressions with Monte Carlo simulations. Ongoing work includes studying the impact of such approach by using low-count simulated data, and a comparison of the proposed method with conventional scatter correction techniques.

AcknowledgmentThis work was supported by the European Union’s Horizon 2024 research and innovation programme under the Marie Skłodowska-Curie grant agreement no. 101207067, and partly by the German Research Foundation (DFG) under grant agreement no. 508064995.
References
[1] Zaidi, H, Koral, KF 2004, ‘Scatter modelling and compensation in emission tomography’,
Eur J Nucl Med Mol Imaging 31, 761–782. DOI: 10.1007/s00259-004-1495-z
[2] Watson, CC, et al. 1996, ‘A Single Scatter Simulation Technique for Scatter Correction in 3D PET’, In: Grangeat, P., Amans, JL. (eds) Three-Dimensional Image Reconstruction in Radiology and Nuclear Medicine. Computational Imaging and Vision, vol 4. Springer, Dordrecht. DOI: 10.1007/978-94-015-8749-5
[3] Tamal, M, et al. 2006, ‘Noise Properties of Four Strategies for Incorporation of Scatter and Attenuation Information in PET Reconstruction Using the EM-ML Algorithm’, IEEE Transactions on Nuclear Science, vol. 53, no. 5, pp. 2778-2786. DOI: 10.1109/TNS.2006.880973
[4] Markiewicz, PJ, et al. 2007, ‘High accuracy multiple scatter modelling for 3D whole body PET’, Phys. Med. Biol. 52 829. DOI: 10.1088/0031-9155/52/3/021
[5] Schaar, M, Rafecas, M 2019, ‘Modelling of Scatter in the System Matrix for 3D PET Image Reconstruction: a Comparative Study’, 2019 IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), Manchester, UK. DOI:  10.1109/NSS/MIC42101.2019.9059797
[6] Roser, J, et al. 2026, ‘Inclusion of Inter-crystal Scattering in PET: Analytical Models and Dedicated Reconstruction’, preprint. DOI: 10.48550/arXiv.2601.05717
Keywords: PET, object scatter, analytical model, sensitivity, system matrix
17:30 Poster panel: 53

Poster Number:
MIC-10-053

Noise Modeling of Three Material Dual Energy Computed Tomography with a Pathlength Constraint (#1443)

A. E. Ferretti1, E. Sidky1, J.P. Phillips1, X. Pan1

1 University of Chicago, Department of Radiology, Chicago, Illinois, United States of America

Abstract

Dual energy computed tomography (DECT) utilizes transmission measurements through a sample with two distinct X-ray spectra to produce a two material decomposition. By introducing a path-length constraint to DECT measurements, it is possible to obtain a three material decomposition. In both cases, the stability of the resulting decompositions and noise properties of the virtual monochromatic images (VMIs) are sensitive to acquisition choices like beam energy. We aim to extend a theoretical noise-propagation framework for choosing optimal beam energies and VMI energy for two material decompositions to the three material case. We then demonstrate the utility of this framework for a proposed use-case of contrast enhanced breast tomography. In particular, by assuming linear propagation of noise through material sinogram estimation and filtered backprojection, a compound Poisson noise process in X-ray measurements, and Gaussian-distributed distance constraints, we obtain analytic expressions for the Jacobian of the material sinogram-to-data transformation, the variance of VMI pixels, and the VMI contrast to noise ratio (CNR). We then use the VICTRE software to generate a realistic computational breast phantom, modeled as consisting of glandular tissue, adipose, and iodine contrast agent. For this phantom, we find that tube potential pairings of 30-40 kVp and 55-80 kVp give a high value of the material sinogram-to-data transformation Jacobian determinant, indicating good stability and minimal noise propagation. For this phantom and spectra, we compare analytically predicted and empirically realized VMI CNR versus monoenergy, and find good agreement between them with a maximum CNR of 9.6 for a VMI monoenergy of 65.5 keV in both cases.

Acknowledgment

This work is supported in part by NIH grant No. R01-CA287302. The contents of this article are solely the responsibility of the authors and do not necessarily represent the official views of the National Institutes of Health.

References
[1] Kis, B.J., Sarnyai Z., Kákonyi, R., Erdélyi, M., Szabó, G., 2012, ‘Single-energy material decomposition using X-ray path length estimation’, Journal of Computer Assisted Tomography, vol. 36, no. 6, pp. 768–777
[2] Terzioglu, F., Sidky, E., Phillips, J.P., Reiser, I., Bal, G., Xiaochuan, P., 2024, ‘Optimizing dual‐energy CT technique for iodine‐based contrast‐to‐noise ratio, a theoretical study’, Medical Physics, vol. 51, no. 4, pp. 2871–2881
[3] Bal, G., Gong, R., Terzioglu, F., 2022, ‘An inversion algorithm for P-functions with applications to multi-energy CT’, Inverse Problems, vol. 38, no. 3, p. 035011
[4] Whiting, B. R., ‘Signal statistics in x-ray computed tomography’, 2002, Medical Imaging 2002: Physics of Medical Imaging, vol. 4682, pp. 53-60, International Society for Optics and Photonics
[5] U.S. Food and Drug Administration, ‘VICTRE: In Silico Breast Imaging Pipeline’ (RST24MX01.01). https://cdrh-rst.fda.gov/victre-silico-breast-imaging-pipeline
[6] Hubbell, J.H., Seltzer, S.M., 1995 ‘Tables of X-ray mass attenuation coefficients and mass energy-absorption coefficients 1 keV to 20 MeV for elements Z = 1 to 92 and 48 additional substances of dosimetric interest’, Technical report, National Inst. of Standards and Technology
[7] Salvat, F., Fernandez-Varea, J. M., Sempauand, J., 2008, ‘PENELOPE2006, A Code System for Monte-Carlo Simulation of Electron and Photon Transport’, Organisation for Economic Co-Operation and Development, Nuclear Energy Agency - OECD/NEA, Le Seine Saint-Germain, 12 boulevard des Iles, F-92130 Issy-les-Moulineaux (France),
[8] Kim, B., et al., 2021, ‘Adaptive image rescaling for weakly contrast-enhanced lesions in dedicated breast CT: A phantom study’, Journal of the Korean Society of Radiology, vol. 82, no. 6, p. 1477-1492
[9] Poludniowski, G., Omar, A. , Bujila, R., Andreo, P., 2021, ‘Technical note: Spekpy v2.0—a software toolkit for modeling x‐ray Tube Spectra’, Medical Physics, vol. 48, no. 7, pp. 3630–3637
Keywords: Dual Energy Computed Tomography, Noise Propagation, X-ray Imaging
17:30 Poster panel: 56

Poster Number:
MIC-10-056

Optimization of Pinhole Diameter in Fully Stationary Multi-Pinhole Brain SPECT: Prioritizing Spatial Resolution via CNN-Based Correction (#1530)

T. Akatsuka1, K. Ogawa2

1 Hosei University, Applied Informatics, Graduate School of Science and Engineering, Koganei, Japan
2 Hosei University, Applied Informatics, Faculty of Science and Engineering, Koganei, Japan

Abstract

Multi-pinhole SPECT enables static data acquisition without rotating the detectors, reducing patient burden. In brain SPECT, particularly for diagnosing neurodegenerative diseases like Alzheimer's, high spatial resolution is critical for visualizing minute regional functions. However, increasing the number of pinholes to secure sufficient sensitivity causes projection overlaps on the detector. Furthermore, physical factors such as gamma-ray attenuation and scattering within the object degrade the quantitative accuracy of reconstructed images. In addition to these issues, a significant trade-off exists regarding the pinhole aperture size. While a larger diameter (e.g., 5 mm) increases photon counts and reduces statistical noise, it exacerbates projection blur and degrades spatial resolution. Conversely, a smaller diameter (e.g., 2 mm) preserves resolution but suffers from increased noise. Previously, we proposed a convolutional neural network (CNN) approach using U-Net++ for a three-detector multi-pinhole SPECT that simultaneously performs overlap separation and various physical corrections. In this study, we applied this framework to brain SPECT to determine the optimal aperture size by comparing 2-mm and 5-mm pinhole diameters. Our results demonstrated that the 2-mm aperture, which maintains spatial resolution, outperformed the 5-mm aperture across quantitative image quality metrics, including NMSE, PSNR, and SSIM, confirming its superiority for detailed brain imaging.

References
[1] Ogawa, K, Ichimura, Y, 2014, 'Simulation study on a stationary data acquisition SPECT system with multi-pinhole collimators attached to a triple head gamma camera system', Ann. Nucl. Med., vol. 28, pp. 716–724. 
[2] Mok, GSP, Tsui, BMW, Beekman, FJ, 'The effects of object activity distribution on multiplexing multi- pinhole SPECT', 2011, Phys. Med. Biol., vol. 56, pp. 2635–2650. 
[3] N. Zeraatkar, Kalluri, KS, Auer, B, et al., 2021, 'Investigation of Axial and Angular Sampling in Multi-Detector Pinhole-SPECT Brain Imaging' Phys. Med. Biol., vol. 66, p. 065004.
[4] Moore, SC, Cervo, M, Metzler, S.D, et al., 2015, 'An Iterative Method for Eliminating Artifacts from Multiplexed Data in Pinhole SPECT', in Proc. 13th Fully 3D, pp. 515–517S.
[5] Akatsuka, T, Ogawa, K, 2025, 'Removal of Projection Data Overlaps Using Deep Learning in SPECT with a Large Number of Pinholes', IEEE NSS/MIC, https://doi.org/10.1109/NSS/MIC/ RTSD 57106.2025.11286491.
[6] Motomura, N, Ogawa, K, Shidahara, M, et al., 2007, 'Development of Digital Phantoms for Evaluation of Fundamental Performance on SPECT', IEEE NSS MIC, CD-ROM. 
[7] Allen Inst. Brain Sci., 'Allen Human Brain Atlas', https://human.brain-map.org/mri_viewers/data.
[8] Caron, M, 2022, 'MRI brain dataset', Kaggle, https://www.kaggle.com/datasets/medericcar/mribrain.
[9] Yagi, S, Usui, K, Ogawa, K, 2025, 'Scatter and beam hardening effect corrections in pelvic region cone beam CT images using a convolutional neural network', 2025, Radiol. Phys. Technol., vol. 18, pp. 457–468.
Keywords: SPECT, multi-pinhole collimator, convolutional neural network
17:30 Poster panel: 59

Poster Number:
MIC-10-059

Validated Optical Simulation for High-Energy Gamma Response in a 3-Layer LYSO DOI Detector (#1668)

S. Jiang1, G. Lovatti1, F. Evangelista1, P. Thirolf1, K. Parodi1, M. Nitta1

1 Ludwig-Maximilians-Universität München, München, Bavaria, Germany

Abstract

We have developed a 3-layer staggered LYSO de tector with DOI capability within the project SIRMIO (Small animal proton irradiator for research in molecular image-guided radiation-oncology) for proton range verification via PET imag ing. To evaluate the feasibility of detecting prompt gamma rays for quasi-instantaneous monitoring of proton beams, we propose a workflow that includes the validation of optical simulation settings and the implementation of the saturation effect of the SiPM photonsensor in the Geant4 simulation. Measurement data with 152Eu, 22Na and 137Cs radioactive sources are utilized for modeling the saturation effect of SiPM, and also for validating optical boundary settings in Geant4. The measured spectra of 22Na and 137Cs point sources are well reproduced with the proposed simulation workflow. The responses to prompt gamma rays are then simu lated. According to the floodmaps and profiles, LYSO pixel can be identified at 2.2, 4.4 and 6.2 MeV, but the contrast decreases strongly when energy reaches 6.2 MeV. The peak-to-valley ratios drop by up to 53% at 4.4 MeV and 62% at 6.2 MeV (relative to 2.2 MeV), primarily due to the energy deposition occurring more frequently in multiple LYSO pixels. These predictions suggest the detector’s potential for detecting prompt gamma rays, while accurately recovering or rejecting inter-crystal scatter events is essential to maintain detector performance at higher gamma energy.

References
[1] K. Parodi et al., Acta Oncol., vol. 58, no. 10, p. 1470, 2019.
[2] T. M. Fitzpatrick, Ph.D. dissertation, lmu, 2022.
[3] J. Allison et al., Nucl. Instrum. Meth. A, vol. 835, p. 186, 2016.
[4] K. Kotera et al., arXiv:1510.01102, 2015.
[5] M. Janecek et al., IEEE Trans. Nucl. Sci., vol. 57, no. 3, p. 964, 2010.
Keywords: DOI-PET, LYSO, Saturation, Optical simulation
17:30 Poster panel: 62

Poster Number:
MIC-10-062

A Mesh-Type Reference Computational Phantom Platform for Multi-Modality Monte Carlo Simulation in OpenGATE (#1743)

J. Jegal1, K.B. Kim1, Y.S. Yeom3, J. Yang4, C.H. Choi2, J.-I. Kim2, Y. Seo5, Y. Huh2, 5

1 Korea Institute of Radiological and Medical Sciences, Department of Radiation Ocology, Seoul, Republic of Korea
2 Seoul National University Hospital, Department of Radiation Ocology, Seoul, Republic of Korea
3 Yonsei University, Departement of Radiation Convergence Engineering, Wonju, Republic of Korea
4 UT Southwestern medical Center, Department of Radiology, Texas, United States of America
5 University of California, Department of Biomedical Imaging and Radiology, San Francisco, United States of America

Abstract

Mesh-type Reference Computational (MRCP) Phantoms provide anatomically realistic tetrahedral organ structures, but their practical use in OpenGATE-based imaging and dosimetry simulations remains limited. In our previous work, a simplified tetrahedral phantom was imported using a tessellated-solid-based workflow and evaluated with heart and body geometries. In this study, we extended the approach to a whole-body tetrahedral phantom framework using parameterised tetrahedral geometry construction and region-aware dose scoring. The updated workflow reads mesh, material, organ-region, and visualization information, constructs the complete phantom in OpenGATE, and enables selected-organ three-dimensional dose scoring while preserving full-phantom particle transport. A copy-number-based filtering method was introduced to associate tetrahedral elements with anatomical regions, avoiding material-name-based selection. The complete phantom, including all anatomical organs, was successfully loaded with a memory usage of 23.84 GB. Three-dimensional deposited energy and dose distributions were generated for selected organs. Multi-threaded performance was evaluated using 8 million x-ray beam particles, reducing computation time from 2453.13 s with 1 CPU thread to 429.43 s with 8 CPU threads, corresponding to a 5.7-fold speed improvement. In the 8-thread simulation, dose scoring for four selected organs, including blood, heart wall, and both lungs, was completed in approximately 5 minutes. These results demonstrate the feasibility of multi-threaded, organ-specific MRCP dosimetry and provide a basis for extending mesh-type phantom simulation to CT, PET, SPECT, and linac workflows in OpenGATE.

AcknowledgmentNone.
Keywords: MRCP, Tetrahedron, Monte Carlo Simulation, GATE, CT
17:30 Poster panel: 65

Poster Number:
MIC-10-065

Inter-Crystal Scattering in LYSO-SiPM PET: Integrated Analysis via Geant4/GATE Simulation, CNN, and Multi-Pitch Experiments (#1859)

T. Deng1, Q. Zhang1, H. Zhang1, X. Cai1, C. Zhang1, Y. Yin1

1 Lanzhou University, School of Nuclear Science and Technology, Lanzhou, China

Abstract

This study integrates GEANT4/GATE Monte Carlo simulations with hardware experiments to investigate and correct inter-crystal scattering (ICS) in Lutetium Yttrium Oxyorthosilicate (LYSO) Positron Emission Tomography (PET) detectors. In the simulation phase, 107 valid events from a 511 keV gamma-ray point source were tracked across a 16 x 16 LYSO array (2.1 mm pitch) to analyze 3D energy spread and depth of interaction (DOI) characteristics. To correct ICS-induced parallax errors, we designed a physics-guided Convolutional Neural Network (CNN) with an energy masking constraint to recover the initial interaction coordinates. Furthermore, we established a physical experimental platform using arrays of various pitches (2.1 mm, 3.2 mm, and 4.1 mm). Deploying the pretrained CNN on physical hardware and conducting preliminary iterative training with empirical data demonstrated the predictive potential of machine learning for measured signals. During system image reconstruction, perfect CNN prediction of ICS events could theoretically improve spatial resolution from 1.45 mm to 1.21 mm. By synergistically applying algorithmic ICS correction and DOI information, the full width at half maximum (FWHM) can be further compressed to 0.99 mm. This study provides a conceptual solution and initial attempt for mitigating inter-crystal scattering in future high-resolution PET detectors.

Keywords: Positron Emission Tomography, LYSO, Spatial resolution, GEANT4/GATE, Convolutional neural networks
17:30 Poster panel: 68

Poster Number:
MIC-10-068

Design of a brain SPECT scanner based on high intrinsic resolution CZT detectors and multi-pinhole collimators (#1985)

K. Erlandsson1, F. Fraioli1, A. Cherlin2, B. F. Hutton1, K. Thielemans1

1 University College London, Institute of Nuclear Medicine, London, United Kingdom
2 Kromek Ltd, Department of imaging, Sedgefield, United Kingdom

Abstract

SPECT is useful for investigating a range of neurological and psychiatric conditions, such as Parkinson’s disease, epilepsy and schizophrenia. Most clinical brain SPECT studies are performed with rotating gamma-cameras with parallel-hole collimators, which results in sub-optimal image quality. Multi-pinhole (MPH) collimators can provide an improved sensitivity vs. resolution trade-off, while Cadmium-Zinc-Telluride (CZT) detectors offer better intrinsic spatial resolution as well as superior energy resolution compared to scintillation detectors. In collaboration with Kromek Ltd, we have developed a molecular breast imaging system based on CZT detectors and MPH collimators. Here we report initial results on designing a dedicated brain SPECT scanner using similar technology. The system will consist of a polygon of planar detectors and will be completely stationary, allowing for dynamic imaging. We performed computer simulations for various design configurations, based on a 20-cm diameter spherical field-of-view (FOV), assuming an intrinsic detector resolution of 0.4x0.4 mm and a depth-of-interaction (DOI) resolution of 1 mm. The simulations were based on an analytical system model, and images were reconstructed using an ML-EM algorithm. Our initial evaluations involved two different digital phantoms: A grid of 7x7 point-sources with a non-zero background; and the Hoffman brain phantom. With the target resolution set to 9 mm at the center and a degree of multiplexing of ~30%, the sensitivity was ~6×10-4. An average reconstructed resolution of ~5 mm FWHM was obtained, going from 9 mm at the center down to 2-3 mm towards the edge of the FOV. In conclusion, our initial results show that high image quality can be obtained with this type of SPECT system.

AcknowledgmentThis work  was supported by funding from the UKRI EPSRC Impact Acceleration award to UCL (EP/X525649/1). INM staff are partly funded by the NIHR University College London Hospitals Biomedical Research Centre.
References
[1] K. Erlandsson et al. 2023, IEEE Trans. Radiat. Plas. Med. Sci., 7(3):225-232.
Keywords: Brain SPECT, CZT detectors, multi-pinhole collimator.
17:30 Poster panel: 71

Poster Number:
MIC-10-071

Modeling analog SiPMs with temporal crosstalk integration for fast timing in medical imaging (#2021)

N. Kratochwil1, E. Roncali1, G. Ariño-Estrada1, 2

1 University of California, Davis, Department of Biomedical Engineering, Davis, California, United States of America
2 Barcelona Institute of Science and Technology in Barcelona, Institut de Fisica d’Altes Energies, Barcelona, Spain

Abstract

Silicon photomultipliers (SiPMs) coupled to scintillators are widely used in medical imaging, particularly in fast-timing applications such as time-of-flight positron emission tomography (TOF-PET) . While significant progress has been made in modeling detector response and correlated noise, recent advances in SiPM technology and fast readout electronics require a reassessment of the full analog signal formation process. In particular, the transition from idealized digital timestamps to realistic analog signals introduces information loss due to bandwidth limitations, electronic noise, and correlated effects such as optical crosstalk.
We present a comprehensive simulation framework that bridges this gap by modeling both analog waveform formation and generation-dependent optical crosstalk with explicit temporal structure. The framework enables systematic separation of different degradation mechanisms affecting timing performance. We investigate the impact of bandwidth limitation, electronic noise, and correlated noise on coincidence time resolution, as well as the information loss transitioning form digital to analog signals.
Our results show a clear degradation of timing resolution when moving from ideal timestamps to realistic analog readout conditions, in agreement with experimental measurements. Furthermore, we observe that optical crosstalk can, in certain scintillator configurations, improve timing performance by effectively enhancing early signal amplitude, while simultaneously degrading higher-level temporal information relevant for event characterization. The presented framework provides a flexible tool for disentangling these effects and guiding SiPM and readout optimization for future fast-timing detector systems.

AcknowledgmentThis work was supported by NIH grant R01EB034062.
References
[1] Kratochwil et al 2026, https://doi.org/10.1109/TRPMS.2026.3665464.
[2] Vinogradov 2012, https://doi.org/10.1016/j.nima.2011.11.086.
[3] Herweg et al 2025, https://doi.org/10.3389/fphy.2025.1603556.
[4] Kratochwil et al 2020, https://doi.org/10.1088/1361-6560/ab87f9.
[5] Gundacker et al 2015, https://doi.org/10.1016/j.nima.2014.10.020.
[6] Gundacker et al 2020, https://doi.org/10.1088/1361-6560/ab63b4.
[7] Kratochwil et al 2025, https://doi.org/10.1016/j.sna.2025.116782.
Keywords: Silicon photomuliplier (SiPM), Optical crosstalk, analog signal formation, fast timing detectors
17:30 Poster panel: 74

Poster Number:
MIC-10-074

Study on the axial extension of the small-fish PET scanner MERMAID (#2065)

M. Kolodziej1, J. Roser1, S. Seeger1, 2, H. P. Vo1, R. Kantorek1, 2, M. Rafecas1

1 University of Luebeck, Institute of Medical Engineering, Luebeck, Schleswig-Holstein, Germany
2 University of Luebeck, Isotope Laboratory of the Natural Science Section, Luebeck, Schleswig-Holstein, Germany

Abstract

MERMAID is the first PET scanner prototype dedicated to study small aquatic animals. Particularly, it aims to support the use of adult zebrafish as a model of human diseases. The current prototype, which consists of four detection modules composed of LYSO:Ce crystals, has demonstrated the capability of providing images with submillimeter resolution. An extension of MERMAID by another four modules is planned, aimed  to increase the system sensitivity and allow for shorter scan times, making it more suitable for in-vivo imaging. In the presented work, various configurations for the extended MERMAID system are characterized by means of simulations and tailored reconstruction framework; modifications of imaging protocol for the system are  investigated.

AcknowledgmentThe project is partially supported by the German Research Foundation (DFG) under grant agreement no. 496099829 and the DFG Cluster of Excellence PMI under grant agreement no. 390884018.
References
[1] S. Seeger et al., 2023 IEEE NSS MIC RTSD, 2023. doi: 10.1109/NSSMICRTSD49126.2023.10338697.
[2] J. Roser et al., IEEE TRPMS, 2026. doi: 10.1109/TRPMS.2026.3691192.
Keywords: Positron emission tomography, PET, small-fish PET, zebrafish, simulation
17:30 Poster panel: 77

Poster Number:
MIC-10-077

Design and Modeling of BetaBox 2.0: a System for Quantitative Measurement of Beta Emitters in Single Cells (#2271)

K. Klein1, J.A. Leal Doblado1, D. Prout1, P. Clark1, M. van Dam1, A. Chatziioannou1

1 University of California Los Angeles (UCLA), Pharmacology, Los Angeles, California, United States of America

Abstract

Single cell radiation measurements are an integral part of interrogating the biological heterogeneity of cancerous tumors. Identifying cell types present in bulk tumors via FDG uptake can elucidate cell types resistant to treatments, that would potentially be obscured via macroscopic FDG PET imaging. The BetaBox 1.0 was previously constructed to perform these single cell radiation measurements, but suffers from a lack of scale and inability to isolate cells after radiation measurement for further study. To address these challenges, we are constructing the BetaBox 2.0, which will be a high-sensitivity, low-noise device capable of performing many simultaneous single cell radiation measurements via the detection of beta particles in a pixelated array of organic scintillators coupled to SiPMs. The design of BetaBox 2.0 is considered and modeled in GATE simulations, which are used to determine an optimal organic scintillator thickness for the device. These simulations are then expanded on via experimental measurements taken with a SiPM pixel read out by PETsys readout electronics. These datasets jointly suggest the BetaBox 2.0 will be capable of isolating the signal from single cells, and quantitatively assessing their radiation uptake.

AcknowledgmentThis work is supported in part by the National Institute of Health under Grants R01CA289908, and T32EB002101.
Keywords: Radiation Measurement, Beta/Alpha Dosimetry, Quantitative Imaging
17:30 Poster panel: 80

Poster Number:
MIC-10-080

Hardware Ray-Tracing-Accelerated Optical Photon Transport in TOPAS on Apple Silicon (#2319)

Y. Goh1, K. M. Back1, U.-S. Lee1, S.-W. Kim1, S. Y. Song1

1 Asan Medical Center, Department of Radiation Oncology, Seoul, Republic of Korea

Abstract

Optical photon transport in scintillator-based imaging detectors dominates TOPAS Monte Carlo wall time because each primary spawns tens of thousands of secondary photons, and TOPAS itself has no released GPU backend. We present MetalOpticalPhoton, a TOPAS extension that ports the optical physics chain to Apple Metal hardware ray tracing while reusing existing TOPAS material/surface/geometry tables. On an Apple M3 Ultra (80 GPU cores), execution wall time on a 100K-event scintillator-plus-lens benchmark is 25–28× shorter than a 30-thread M3 CPU baseline and about 10× shorter than a 90-thread AMD 7995WX baseline, sustaining roughly 33×106 photons/s GPU throughput. Two corrections to vanilla OpenTOPAS—a parallel-world envelope coincident-boundary loop suppressing 75–80% of the sensor-integrated CPU optical fluence, and a primary polarization initialization typo—were required for a self-consistent CPU reference and are relevant to TOPAS optical Monte Carlo independently of GPU acceleration. After patching, CPU–GPU agreement reaches the sub-1% level reported for production GPU optical Monte Carlo (sensor-integrated fluence sum ratio 0.993 ± 0.002 over five seeds).

AcknowledgmentThis work was supported by the National Research Foundation of Korea (NRF), funded by the Ministry of Science and Information and Communication Technology of Korea (grant number 2026OK9045).
References
[1] Allison, J et al. 2016, ‘Recent developments in Geant4’, Nucl. Instrum. Methods Phys. Res. A, vol. 835, pp. 186–225, Amsterdam: Elsevier.
[2] Perl, J, Shin, J, Schümann, J, Faddegon, B, Paganetti, H 2012, ‘TOPAS: an innovative proton Monte Carlo platform for research and clinical applications’, Med. Phys., vol. 39, no. 11, pp. 6818–6837, Hoboken: Wiley.
[3] Blyth, SC 2019, ‘Opticks: GPU optical photon simulation via NVIDIA OptiX’, EPJ Web Conf., vol. 214, art. no. 02027, Les Ulis: EDP Sciences.
[4] Apple Inc. 2026, ‘Metal Shading Language Specification: Ray Tracing’, Cupertino, CA: Apple Inc.
Keywords: Apple Metal, GPU acceleration, Monte Carlo simulation, optical photon transport, TOPAS
17:30 Poster panel: 83

Poster Number:
MIC-10-083

Identification of ICS Pixels in a high-resolution 3-Layer DOI-PET Detector Using a CNN (#2619)

S. Jiang1, G. Lovatti1, F. Evangelist1, P. Thirolf1, K. Parodi1, M. Nitta1

1 Ludwig-Maximilians-Universität München, München, Bavaria, Germany

Abstract

We have developed a 3-layer staggered LYSO DOI detector for the SIRMIO (Small animal proton irradiator for research in molecular image-guided radiation-oncology) project, with a crystal pitch of 1.0 mm. Due to the small crystal size, inter-crystal scatter (ICS) events degrade spatial resolution, while discarding them causes significant sensitivity loss. In our detector, ICS events account for approximately 50% of all interactions at 511 keV and result in a median lateral positioning error of approximately 2 mm. To solve this problem, we propose a dual-head convolutional neural network (CNN) for separating the 8×8 MPPC total response of two-pixel ICS events into two individual response matrices, from which the pixel positions and energy fractions can be obtained—a precondition for recovering the ICS events. The CNN correctly identifies both scattering pixels in 69.3% of test events, and in 26.4% of cases one pixel is correctly identified. Among the events with one pixel mismatched, the lateral error of the misidentified pixel falls within 1.0 mm in 91.0% of cases. Meanwhile, the median energy fractions prediction error is 0.016 per pixel. The results demonstrate that CNN-based MPPC response decomposition can serve as the enabling foundation for complete ICS recovery in 3-layer staggered DOI detectors, addressing a limitation explicitly acknowledged in recent physics-based ICS studies.

References
[1] F. Evangelista et al., in 2024 IEEE NSS MIC RTSD, 2024.
[2] K. Parodi et al., Acta Oncol., vol. 58, no. 10, p. 1470, 2019.
[3] M. S. Lee et al., Biomedical Engineering Letters, vol. 14, no. 6, 2024.
[4] Q. Liu et al., Physics in Medicine and Biology, 2026.
[5] S. Jiang et al., submitted to IEEE NSS MIC RTSD 2026 (#1668).
Keywords: Inter-crystal scatter, CNN, DOI-PET, LYSO
17:30 Poster panel: 86

Poster Number:
MIC-10-086

Impact of PET energy window extent on trues normalization and reconstructions (#2712)

S. Noë1, S.A. Zaman Pour1, A. Reeves2, A. Rezaei1, J. Nuyts1, G. Schramm1, K. Thielemans2

1 KU Leuven, Department of Imaging and Pathology, Nuclear Medicine & Molecular imaging, Leuven, Belgium
2 University College London, Institute of Nuclear Medicine and the Hawkes Institute, London, United Kingdom

Abstract

PET systems require (among others) normalization corrections for the geometric sensitivity of the scanner, the result of the spatial arrangement of the detector crystals and energy readout area. It is estimated for a given extent of the photon energy acceptance window, and must be re-estimated if the lower energy threshold (LET) is changed, e.g. when optimizing the number of noise-equivalent counts for a given object, or to improve joint estimation of activity and attenuation. To investigate the effect of changing the LET on the geometric sensitivity, we performed a simulation study in SimSET and compared sensitivity sinograms, reconstructions and properties of accepted photons. We found that the effects are dependent on detector location and photon incidence angle, and likely due to the finite extent of the energy readout area.

AcknowledgmentThis work was supported by the i4Health Centre for Doctoral Training, UCL, CCP SyneRBI funded by STFC through the DRI programme.
References
[1] Theodorakis, L, Loudos, G, Prassopoulos, V, Kappas, C, Tsougos, I, Georgoulias, P 2013, 'A review of PET normalization: Striving for count rate uniformity', Nucl. Med. Commun., vol. 34, no. 11, pp. 1033–1045
[2] Badawi, RD, Marsden, PK, Cronin, F, Sutcliffe, JL 1996, 'Optimization of noise-equivalent count rates in 3D PET', Phys. Med. Biol., vol. 41, pp. 1755–1776
[3] Reeves, A, Arridge, S, Thielemans, K 2025, 'Spatial Optimisation of Lower Energy Discriminator for PET Data Using Noise Equivalent Counts', in Proc. IEEE NSS/MIC/RTSD, IEEE, pp. 1–1
[4] Brusaferri, L, Emond, EC, Bousse, A, Twyman, R, Whitehead, AC, Atkinson, D, Ourselin, S, Hutton, BF, Arridge, S, Thielemans, K 2022, 'Detection Efficiency Modeling and Joint Activity and Attenuation Reconstruction in Non-TOF 3-D PET from Multiple-Energy Window Data', IEEE Trans. Radiat. Plasma Med. Sci., vol. 6, no. 1, pp. 87–97
[5] Wagadarikar, AA, Ivan, A, Dolinsky, S, McDaniel, DL 2014, 'Sensitivity improvement of time-of-flight (ToF) PET detector through recovery of Compton scattered annihilation photons', IEEE Trans. Nucl. Sci., vol. 61, no. 1, pp. 121–125
Keywords: Positron Emission Tomography, Normalization, Monte Carlo Simulation
17:30 Poster panel: 89

Poster Number:
MIC-10-089

MIMOSA: Medical IMaging MOnte Carlo Simulation Algorithms for fast proton therapy range monitoring (#2751)

D. D. Garvey1, G. Llosá1, F. Hueso González1

1 IFIC, CSIC-UV, Valencia, Spain

Abstract

Monte Carlo simulations play a vital role in many aspects of proton therapy treatment, including proton depth of penetration monitoring. A key limiting factor of full simulation engines is computation time: simulations can take several days, far longer than the time between treatment fractions. This project aims to develop a Monte Carlo simulation software, MIMOSA, for prompt gamma-ray based proton range verification, that is fast enough to be deployed before the next treatment fraction. Clinically acceptable gamma pass rates above 99% for 2%/2 mm and 97% for 1%/1 mm with respect to TOPAS were achieved on both homogenous and heterogenous phantoms. To benchmark computational speed, state-of-the-art GPU codes MOQUI and FRED were tested on the same computer. Comparatively, MIMOSA achieved a notable speed increase, able to simulate 1e9 primary protons with a 200 MeV energy in approximately 3 minutes. Along with maintaining accuracy and increasing computational speed, the project uses the hardware agnostic SYCL standard, and will be available with an open-source license and extensive documentation. Ongoing work is devoted to simulating realistic radiotherapy treatment plans, and including prompt gamma-ray transport and detector response.

AcknowledgmentThis work has been funded by the Conselleria de Educación, Investigación, Cultura y Deporte (Generalitat Valenciana), Spain, under grant number CDEIGENT/2019/011, by CSIC's JAEPRE23-19 grant, and by the MIDAS project PID2022-143246OB-I00 (MICIU/AEI). IFIC is supported by the Severo Ochoa Center of Excellence grant CEX2023-001292-S funded by MCIU/AEI
References
[1] F. Hueso-Gonzalez and T. Bortfeld, “Compact method for proton range verification based on coaxial prompt gamma-ray monitoring: A theoretical study,” IEEE Transactions on Radiation and Plasma Medical Sciences, vol. 4, no. 2, p. 170–183, Mar. 2020. [Online]. Available: http://dx.doi.org/10.1109/TRPMS.2019.2930362
[2] J. Perl, J. Shin, J. Schumann, B. Faddegon, and H. Paganetti, “TOPAS: An innovative proton Monte Carlo platform for research and clinical applications,” Medical Physics, vol. 39, no. 11, pp. 6818–6837, November 2012
[3] H. Lee, J. Shin, J. M. Verburg et al., “MOQUI: an open-source GPU-based Monte Carlo code for proton dose calculation with efficient data structure,” Physics in Medicine & Biology, vol. 67, no. 17, p. 175001, 2022, published August 30, 2022. [Online]. Available: https://doi.org/10.1088/1361-6560/ac8716
[4] M. D. Simoni, M. Fischetti, E. Gioscio, M. Marafini, R. Mirabelli, V. Patera, A. Sarti, A. Schiavi, A. Sciubba, and G. Traini, “FRED: A fast Monte Carlo code on GPU for quality control in particle therapy,” Journal of Physics: Conference Series, vol. 1548, no. 1, p. 012020, may 2020. [Online]. Available: https://doi.org/10.1088/1742-6596/1548/1/012020
[5] A. Alpay and V. Heuveline, “Adaptivity in AdaptiveCpp: Optimizing performance by leveraging runtime information during JIT-compilation,” in Proceedings of the 13th International Workshop on OpenCL and SYCL, ser. IWOCL ’25. New York, NY, USA: Association for Computing Machinery, 2025. [Online]. Available: https://doi.org/10.1145/3731125.3731127
Keywords: MIMOSA, open source, simulation, proton therapy, range verification
17:30 Poster panel: 92

Poster Number:
MIC-10-092

Quantitative Impact of Inter-Crystal Scatter on Image Quality and Sensitivity across LabPET-II Scanner Architectures: A Comparative GATE Simulation Study (#2983)

A. Bernier1, 2, M. Gaudreault5, J.-F. Beaudoin3, R. Lecomte4, 5, J.-B. Michaud1, 2

1 University of Sherbrooke, Department of Electrical and Computer Engineering, Sherbrooke, Québec, Canada
2 University of Sherbrooke, Interdisciplinary Institute for Technological Innovation, Sherbrooke, Québec, Canada
3 CRCHUS, Sherbrooke Molecular Imaging Center, Sherbrooke, Sherbrooke, Québec, Canada
4 Imaging Research & Technology, Sherbrooke, Québec, Canada
5 University of Sherbrooke, Department of Medical Imaging and Radiation Sciences, Sherbrooke, Québec, Canada

17:30 Poster panel: 95

Poster Number:
MIC-10-095

Effects of Beam Filtration on Time-of-Flight-Based Scatter Rejection in Computed Tomography (#3162)

J. Lee1, 2, S. Kim1, 2, G. Ko4, W. Sung1, 3

1 The Catholic University of Korea, Department of Biomedical Engineering, College of Medicine, Seoul, Republic of Korea
2 The Graduate School of The Catholic University of Korea, Department of Medical Sciences, Seoul, Republic of Korea
3 The Catholic Medical Center of The Catholic University of Korea, CMC Institute for Basic Medical Science, Seoul, Republic of Korea
4 The Korea Advanced Institute of Science and Technology (KAIST), Department of Nuclear and Quantum Engineering, Daejeon, Republic of Korea

Abstract

Computed tomography image quality is degraded by scattered photons, which reduce image contrast and quantitative accuracy. Time-of-flight-based scatter rejection has been proposed as a method to distinguish primary and scattered photons using their arrival-time differences, but the influence of beam filtration under polychromatic x-ray spectrum has not been fully evaluated. This study investigated the effects of tube voltage and beam filtration on time-of-flight-based scatter rejection using Monte Carlo simulation. X-ray spectrum were generated at 40 and 120 kVp, and four filtration conditions were compared: no filter, 1.06mm aluminum, 0.15mm copper, and 0.6mm tin. Polymethyl methacrylate phantoms with thicknesses of 50 and 200 mm were simulated. Scatter rejection performance was evaluated using the scatter rejection rate as a function of timing jitter from 0 to 1000 ps, and the area under the curve was used as a summary metric. At 120 kVp, tin filtration increased the area under the curve by 11.1% in the 200mm phantom and by 30.6% in the 50mm phantom compared with the no-filter condition. In contrast, at 40 kVp, the filtration effect was small, with a maximum area under the curve difference of 0.010 in the 50mm phantom. These results indicate that tin filtration improves timeof-flight scatter rejection at 120 kVp, whereas filter optimization has a limited effect at 40 kVp.

AcknowledgmentThis work was supported by KOREA HYDRO & NUCLEAR POWER CO., LTD (No. 2025-Tech-17).
This work was supported by the National Research Foundation of Korea(NRF) grant funded by the Korea government(MSIT)(RS-2026-25471484).
This research was supported by Basic Medical Science Facilitation Program through the Catholic Medical Center of the Catholic University of Korea funded by the Catholic Education Foundation.
References
[1] G. Lemaire, R. Espagnet, J. Rossignol, L.-D. Gaulin, E. Villemure, A. Corbeil Therrien, Y. Bérubé-Lauzière, M.-A. Tétrault, and R. Fontaine, “Time-of-flight x-ray measurements for computed tomography,” arXiv preprint arXiv:2311.18087, Nov. 2023.
[2] J. Rossignol, R. Martinez Turtos, S. Gundacker, D. Gaudreault, E. Auffray, P. Lecoq, Y. Bérubé-Lauzière, and R. Fontaine, “Time-of-flight computed tomography—proof of principle,” Phys. Med. Biol., vol. 65, no. 8, Art. no. 085013, Apr. 2020, doi: 10.1088/1361-6560/ab78bf.
[3] G. Bélanger, J. Rossignol, Y. Bérubé-Lauzière, and R. Fontaine, “Study of dose reduction in computed tomography using a time-of-flight scatter rejection mechanism in a 3D voxelized phantom,” Phys. Med. Biol., vol. 70, Art. no. 235030, Dec. 2025, doi: 10.1088/1361-6560/ae222f.
Keywords: Time of Flight, Monte Carlo simulation, Computed tomography
17:30 Poster panel: 98

Poster Number:
MIC-10-098

Hybrid PGI-PET Imaging for Ion-Range Monitoring in Hadron Therapy (#2044)

J. Balibrea Correa1, V. Babiano-Suárez1, J. Lerendegui-Marco1, P. Torres-Sánchez1, I. Ladarescu1, C. Domingo-Pardo1, A. Tarifeño-Saldivia1, M. Pallás2, B. Brusasco2, J. Wulff3, C. Baümer3, S. Rossomme4, A. Warnier4

1 Instituto de Física Corpuscular (CSIC-University of Valencia), Física Nuclear Experimental, Valencia, Spain
2 Universitat Politècnica de Catalunya, Física Nuclear, Barcelona, Spain
3 West German Proton Therapy Centre, Essen, Germany
4 Ion Beam Applications, Louvain, Belgium

Abstract

Hadron therapy is a highly conformal cancer treatment technique that enables superior dose delivery to the tumor while sparing surrounding healthy tissues. However, its full clinical benefit is still limited by uncertainties in proton range within tissue, mainly due to the indirect estimation of ion stopping power from X-ray CT images. These uncertainties lead to safety margins that reduce treatment precision. Real-time, high-accuracy ion-range verification is essential to improve dose delivery and further enhance the therapeutic potential of proton therapy. Among the available approaches, PGI and PET are regarded as the most promising, as they provide complementary information. PGI offers direct correlation with the Bragg peak during irradiation, whereas PET yields high-resolution activity distributions from beta+ emitters produced along the proton path.

In this work, we report recent developments and experimental results obtained with a hybrid PGI-PET system for in situ proton range verification. The system is based on four high-efficiency two-plane Compton imagers arranged in a cross configuration, specifically optimized to enhance PGI sensitivity at the expense of a reduced PET field of view. Experimental campaigns were carried out at commercial IBA Proteus Plus and Proteus One proton therapy facilities using 100 and 125 MeV pencil beams delivered at minimum clinical currents. Polyethylene phantoms were irradiated at controlled positions and data were acquired during both in-spill and off-spill periods.

The first reconstructed 2D images demonstrate that PGI can directly identify the Bragg peak position through the maximum of the PGI distribution over a broad field of view, while PET provides a narrower and higher-resolution distribution associated with the induced beta+ activity. These results confirm the feasibility and complementarity of the hybrid PGI-PET approach and support its potential for accurate, clinically relevant proton range verification.

Keywords: Ion range verification, Positron Emission Tomography, Prompt Gamma Imaging, Proton therapy
17:30 Poster panel: 101

Poster Number:
MIC-10-101

Development of a PET/MRI Compatible Fluorescence Microscope for Simultaneous Cell Tracking in a Small Animal PET/MRI Scanner (#2535)

P. Renn1, J. Schäfers1, N. Jurisa1, C. Pilger2, T. Huser2, K. Schäfers1

1 University of Münster, European Institute for Molecular Imaging, 48149, North Rhine-Westphalia, Germany
2 Bielefeld University, Biomolecular Photonics Research Group, Department of Physics, 33615, North Rhine-Westphalia, Germany

Abstract

Cell tracking in vivo is a key challenge in translational medicine. Positron emission tomography (PET) and magnetic resonance imaging (MRI) offer complementary strengths for this task, yet the validation of tracking algorithms against a reliable ground truth remains unresolved. Fluorescence microscopy can fulfill this role, but its integration into the highly constrained PET/MRI environment poses significant technical challenges. Here, we present a fluorescence microscope (FM) designed for simultaneous operation inside a small-animal PET/MRI scanner, with all custom structural and housing components 3D printed in polylactic acid (PLA). The assembled instrument has a total length of 95 cm and fits within the 35 mm PET insert bore, accommodating sample diameters of up to 21 mm. The imaging probe is rotated via a non-ferromagnetic, friction-driven mechanism, enabling multi-angle acquisition without repositioning the sample, and reliable rotation over the full 360° range was demonstrated under manual control. The system follows an epi-fluorescence configuration with 10× magnification, an excitation wavelength of 488 nm, and a 500 nm longpass filter in the detection path. Characterization with a calibration slide and a Convallaria majalis test specimen yielded a pixel pitch of 0.144 μm/pixel (6.96 pixels/μm), a spatial resolution of 2.32 μm (FWHM), a signal-to-noise ratio (SNR) of 7.21, and a field of view (FoV) of 0.55 × 0.31 mm2 (3816 × 2160 pixels), sufficient to resolve individual cells. These results establish the feasibility of integrating fluorescence microscopy into a PET/MRI scanner and provide the basis for future multimodal cell tracking studies with a reliable optical ground truth.

AcknowledgmentThis work was funded by the Deutsche Forschungsgemeinschaft (DFG) – CRC 1450 – 431460824.
References
[1] Wapler, MC, Testud, F, Hucker, P, Leupold, J, v. Elverfeldt, D, Zaitsev, M & Wallrabe, U 2021, ‘MR-compatible optical microscope for in-situ dual-mode MR-optical microscopy’, PLOS ONE, vol. 16, no. 5, p. e0250903: PLOS.
[2] Cui, M, Zhou, Y, Wei, B, Zhu, X-H, Zhu, W, Sanders, MA, Ugurbil, K & Chen, W 2017, ‘A proof-of-concept study for developing integrated two-photon microscopic and magnetic resonance imaging modality at ultra-high field of 16.4 tesla’, Scientific Reports, vol. 7, no. 1, p. 2733: Nature Publishing Group.
[3] Zhu, X, Li, R, Li, H, Gu, L, Chen, L, Xu, J, Chen, J, Li, Y, Rudin, M, Thompson, G, Zhou, N & Ren, W 2026, ‘MR-Scope: a multi-scale and simultaneous MRI-miniaturized microscopy system for elucidating neurovascular coupling’: bioRxiv.
[4] Li, C, Yang, Y, Mitchell, GS & Cherry, SR 2011, ‘Simultaneous PET and multispectral 3-dimensional fluorescence optical tomography imaging system’, Journal of Nuclear Medicine: Official Publication, Society of Nuclear Medicine, vol. 52, no. 8, pp. 1268–1275: Society of Nuclear Medicine.
Keywords: Cell Tracking, Fluorescence Microscopy, MRI, PET, Multimodal Imaging
17:30 Poster panel: 104

Poster Number:
MIC-10-104

Toward Simultaneous PET/SPECT/MR System for Theranostic Imaging (#2904)

Y. Jin1, J. Du2, S. R. Cherry2, L.-J. Meng1, 3

1 University of Illinois at Urbana and Champaign, Department of Nuclear, Plasma, and Radiological Engineering, Urbana, Illinois, United States of America
2 University of California at Davis, Department of Biomedical Engineering, Davis, California, United States of America
3 University of Illinois at Urbana and Champaign, Beckman Institute for Advanced Science and Technology, Urbana, Illinois, United States of America

Abstract

We report a design study for a combined preclinical scanner that integrates PET, SPECT, and MRI for simultaneous acquisition of PET, SPECT, and MR images for broad-band multi-isotope imaging capability critical for theranostics applications. The PET subsystem provides a very high-sensitivity and benefits from a wide variety of existing tracers. The SPECT subsystem allows for broad-band multi-isotope imaging capability critical for imaging therapeutic radioisotopes and their daughters. MRI offers unmatched soft-tissue contrast and high-resolution anatomical reference. The PET/SPECT insert consists of 12 detector rings. The central four PET rings are integrated with four rings of micro camera elements (MCEs) that are “looking at” the object through the small gaps between the PET detector blocks. This compact design provides approximately 16% PET sensitivity at the center and 1-mm PET imaging resolution across the 8-cm transverse by 28-cm axial FOV. The 64 CZT SPECT cameras provide approximately 1-mm spatial resolution over a 6-cm transverse and 9-cm axial FOV. To support simultaneous PET/SPECT operation, the design incorporates our previously developed anti-coincidence rejection strategy to suppress PET-induced contamination in SPECT data. Prior F-18/Tc-99m PET/SPECT simulations demonstrated up to 83.7% rejection of scatter contamination in projection-domain analysis, supporting the feasibility of acquisition-level cross-modality rejection. These design elements resulted in a compact and high performance tri-modality imaging platform for simultaneous PET/SPECT/MR imaging for preclinical theragnostic applications.

AcknowledgmentThis work was supported by the National Institutes of Health under grant numbers R01HL145786 (National Heart, Lung, and Blood Institute) and U01EB031798 (National Institute of Biomedical Imaging and Bioengineering).
References
[1] M. L. James and S. S. Gambhir, “A molecular imaging primer: modalities, imaging agents, and applications,” Physiol. Rev., vol. 92, no. 2, pp. 897–965, Apr. 2012, doi: 10.1152/physrev.00049.2010.
[2] S. R. Cherry, “Multimodality in vivo imaging systems: twice the power or double the trouble?,” Annu. Rev. Biomed. Eng., vol. 8, pp. 35–62, 2006, doi: 10.1146/annurev.bioeng.8.061505.095728.
[3] A. Delker et al., “Biodistribution and dosimetry for combined [177Lu]Lu-PSMA-I&T/[225Ac]Ac-PSMA-I&T therapy using multi-isotope quantitative SPECT imaging,” Eur. J. Nucl. Med. Mol. Imaging, vol. 50, no. 5, pp. 1280–1290, Apr. 2023, doi: 10.1007/s00259-022-06092-1.
[4] M. S. Judenhofer et al., “Simultaneous PET-MRI: a new approach for functional and morphological imaging,” Nat. Med., vol. 14, no. 4, pp. 459–465, Apr. 2008, doi: 10.1038/nm1700.
[5] G. Delso et al., “Performance measurements of the Siemens mMR integrated whole-body PET/MR scanner,” J. Nucl. Med. Off. Publ. Soc. Nucl. Med., vol. 52, no. 12, pp. 1914–1922, Dec. 2011, doi: 10.2967/jnumed.111.092726.
[6] S. Vandenberghe and P. K. Marsden, “PET-MRI: a review of challenges and solutions in the development of integrated multimodality imaging,” Phys. Med. Biol., vol. 60, no. 4, pp. R115-154, Feb. 2015, doi: 10.1088/0031-9155/60/4/R115.
[7] E. M. Zannoni, “Development of the MRC-SPECT-II system: a MR-compatible small-animal SPECT system based on semiconductor imaging spectrometers and a Synthetic Compound-Eye gamma camera design,” Thesis, University of Illinois at Urbana-Champaign, 2021. Accessed: May 06, 2026. [Online]. Available: https://hdl.handle.net/2142/113249
[8] Y. Jin, J. Du, S. R. Cherry, and L.-J. Meng, “Exploring Anti-Coincidence Technique for Scattering Contamination Rejection in a Simultaneous PET/SPECT Imaging System,” IEEE Trans. Radiat. Plasma Med. Sci., pp. 1–1, 2025, doi: 10.1109/TRPMS.2025.3649623.
[9]  L. Cai, X. Lai, Z. Shen, C.-T. Chen, and L.-J. Meng, “MRC-SPECT: A sub-500 μm resolution MR-compatible SPECT system for simultaneous dual-modality study of small animals,” Nucl. Instrum. Methods Phys. Res. Sect. Accel. Spectrometers Detect. Assoc. Equip., vol. 734, no. B, pp. 147–151, Jan. 2014, doi: 10.1016/j.nima.2013.08.080.
Keywords: PET/SPECT/MR, multimodality imaging, CZT detector, anti-coincidence rejection, theragnostic imaging
17:30 Poster panel: 107

Poster Number:
MIC-10-107

Development of a Preclinical Simultaneous Ultrasensitive SPECT and PET Imager (#3109)

R. Zhu1, C. Ling1, F. Ahmed3, A. Jha3, 1, J. O’Sullivan1, Y.-C. Tai1, 2

1 Washington University in St. Louis, Imaging Science, saint louis, Missouri, United States of America
2 Washington University in St. Louis, Mallinckrodt Institute of Radiology, saint louis, Missouri, United States of America
3 Washington University in St. Louis, Biomedical Engineering, saint louis, Missouri, United States of America

Abstract

We previously proposed a Broadband Ultrasensitive Gamma Imager (BUGI) that uses a sensing collimator imager to encode and detect gamma rays without a metal collimator. When combined with fast scintillators, this approach enables simultaneous high-sensitivity SPECT and time-of-flight-capable PET imaging, with potential applications in monitoring alpha-particle radiopharmaceutical therapy (RPT). Unlike conventional simultaneous PET/SPECT systems, which use high-energy collimators that reduce PET sensitivity to SPECT levels, BUGI raises SPECT sensitivity toward that of PET.In this study, we evaluated two preclinical BUGI detector designs using GATE Monte Carlo simulations: a pinhole-inspired geometry and a parallel-hole-inspired geometry. The scanner consists of an octagonal arrangement of 8 detector blocks, each containing 1 × 4 arrays of self-collimating mini-modules built from 3×3×3 mm³ GAGG crystals. A dual-isotope cylindrical phantom was simulated with a ⁹⁹ᵐTc background, a ⁹⁹ᵐTc spherical tumor, and an ¹⁸F spherical tumor, with a tumor-to-background ratio of 4:1. SPECT events and PET coincidence events were reconstructed independently using a GPU-based MLEM framework.Both detector geometries produced usable simultaneous SPECT and PET images with comparable tumor contrast . The parallel-hole design showed better background uniformity, improved preservation of the cylindrical phantom shape, and sharper tumor edges.The pinhole design showed an octagonal artifact pattern that may be reduced with increased angular sampling. PET images were sharp and visually similar between the two designs. Future work will focus on developing and validating multi-energy-window SPECT and PET reconstruction methods for the BUGI system.

AcknowledgmentThis work was supported in part by the Mallinckrodt Institute of Radiology at Washington University at St. Louis through an internal fund allotment. We also acknowledge the support from NIH (R01-EB031962, R01-CA240711) and NSF CAREER Award (2239707).
Keywords: dual modality PET/SPECT
17:30 Poster panel: 110

Poster Number:
MIC-10-110

A Dual-Ring Super-Resolution PET System (#1259)

Y. Lang1, 2, Q. Xie1, 3, C.-M. Kao4

1 Huazhong University of Science and Technology, Biomedical Engineering Department, Wuhan, China
2 Huazhong University of Science and Technology, Institute of Artificial Intelligence, Wuhan, China
3 University of Science and Technology of China, Department of Electronic Engineering and Information Science, Hefei, China
4 University of Chicago, Department of Radiology, Chicago, Illinois, United States of America

Abstract

Building upon previous works that explored the feasibility of super-resolution (SR) positron emission tomography (PET) by applying the principle of structured illumination microscopy (SIM), in this work we propose the concept of a dual-ring SR PET system design. Such a PET system contains a rotatable inner detector ring that serves the dual purpose of creating the needed measurement modulation for performing SIM-based SR as well as increasing event detection to eliminate a major drawback of the previous approach in which the modulator could considerably diminish the imaging sensitivity. We simulated a 2d PET system comprising 576 4.2 mm (transaxial) x 4.2 mm (axial) x 20 mm (radial) LYSO crystals on a 770 mm diameter ring. On the other hand, the inner ring consisted of repeating 2.8 mm-width attenuating material or detector and 5.6 mm-width air. The attenuating material was either 5 mm thick tungsten (W) or 10 mm thick LYSO. The outer detector was 20 mm thick LYSO. Axially, the inner ring was uniform and had the same 4.2 mm length as the outer detector ring. An SR dataset was made of three measurements acquired at three rotational positions of the inner ring. We also considered point-spread function (PSF) reconstruction when not using the inner ring, and also the situation when detector wobble motion was introduced to increase sampling. Our simulation results show that the dual-ring image can resolve 1.2 mm sources in the case of noise-free data and the image quality is visually superior to the PSF and Wobble+PSF images. For a noisy data set, it yields an overall better image quality than the other two SIM-based SR images using W or LYSO modulators for it has higher coincidence detection sensitivity. Quantitative performance evaluation based on the contrast-recovery-coefficient versus the background noise also shows superior results with the dual-ring system.

AcknowledgmentThis work was supported in part by the Key Program of National Natural Science of China under Grant 62131009.
References
[1] Y. Lang et al., IEEE NSS/MIC, pp. 1-1, 2023.
[2] Y. Lang et al., https://doi.org/10.48550/arXiv:2409.16085.
[3] R. A. Brooks et al., IEEE Transactions on Nuclear Science26:2760–2763, 1979.
Keywords: positron emission tomography, super-resolution
17:30 Poster panel: 113

Poster Number:
MIC-10-113

Performance Evaluation of a DOI-Enabled Dedicated Brain PETCT System (#1479)

D. Liang1, Y. Lin1, H. Liu1, X. Yu1, H. Zou1, Y. Xu1, H. Zhao1, J. Tao1, W. Zhu1, Y. Lin1, W. Ma1, W. Li1, Y. Zhang1, D. Shi1, J. Zeng1, Q. Ren1, Z. Xie1, W. Liu1, Q. Peng1

1 Shenzhen Bay Laboratory, Institute of Biomedical Engineering, Shenzhen, China

Abstract

Dedicated brain positron emission tomography (PET) systems can achieve higher sensitivity and spatial resolution than whole-body PET scanners due to their compact detector geometry. However, the reduced detector aperture introduces parallax error, making depth-of-interaction (DOI) capability important for maintaining spatial resolution uniformity across the field of view. In this work, we present a comprehensive performance evaluation of a DOI-enabled dedicated brain PET system based on light-sharing window (LSW) detector technology. The system consists of ten detector panels with an axial field of view of 280.6 mm and employs 2.0 × 2.0 × 20.0 mm³ LYSO crystal arrays with DOI encoding capability.

Performance measurements were conducted following the NEMA NU 2-2018 standard, including spatial resolution, sensitivity, count rate performance, scatter fraction, and time-of-flight (TOF) resolution. The measured sensitivity reached 14.31 cps/kBq at the center of the field of view and 15.08 cps/kBq at a radial offset of 10 cm. The peak noise-equivalent count rate (NECR) was 39.54 kcps at 5.11 kBq/mL. The TOF resolution was 298 ps at peak NECR and improved to 281 ps at low activity concentrations. Spatial resolution measurements demonstrated that DOI encoding significantly mitigated parallax error at large radial offsets. Preliminary clinical imaging studies further demonstrated the capability of the proposed system for high-resolution brain PET imaging
AcknowledgmentThis work was supported by the National Natural Science Foundation of China (Grant no. 82327809), the National Key Research and Development Program of China (Grant No. 2024YFF0507700 ), National High-Level Hospital Clinical Research Funding (2025-PUMCH-D-003), Beijing Natural Science Foundation(L252047).
Keywords: PET, DOI, dedicated brain PET, TOF
17:30 Poster panel: 116

Poster Number:
MIC-10-116

Design and Preliminary Validation of a Heart-Dedicated TOF-PET Scanner with Sub-250ps Timing Resolution (#1614)

Y. Lin1, H. Zou1, D. Liang1, Z. Zhao1, H. Liu1, X. Yu1, J. Tao1, H. Zhao1, X. Zhang1, W. Zhu1, W. Li1, D. Shi1, Y. Zhang1, Q. Peng1

1 Shenzhen Bay Laboratory, Institute of Biomedical Engineering, Shenzhen, China

Abstract

General-purpose positron emission tomography scanners often face challenges with the high-frequency motion and small target volumes inherent in cardiac imaging. This study introduces a heart-dedicated positron emission tomography prototype featuring twenty detector panels in a cylindrical geometry with a 493.28 millimeter aperture and a 236.4 millimeter axial field of view. The detectors utilize lutetium-yttrium oxyorthosilicate crystals coupled to multi-pixel photon counters with depth-of-interaction capability. Performance was validated through Geant4 Application for Tomographic Emission simulations, detector bench tests, and in vivo experiments. Results demonstrate a coincidence resolving time of 234 picoseconds full width at half maximum and a spatial resolution of 1.0 millimeter under idealized conditions. Furthermore, electrocardiogram-gated imaging in rat and rabbit models successfully captured eight cardiac phases from end-systole to end-diastole, providing distinct myocardial structures without significant motion blur. These results validate the system’s high sensitivity and timing precision, establishing a foundation for extracting functional parameters such as ejection fraction. Future work will implement depth-of-interaction correction across the full field of view to maintain high peripheral resolution for precise cardiovascular research and clinical diagnostics.

AcknowledgmentThis work was supported by the National Key Research and Development Program of China (Grant No. 2024YFF0507700 ), the National Natural Science Foundation of China (Grant no. 82327809),  National High-Level Hospital Clinical Research Funding (2025-PUMCH-D-003), Beijing Natural Science Foundation(L252047).
Keywords: Cardiac imaging, Heart-dedicated PET, Time-of-flight, Depth-of-interaction, Electrocardiogram-gated reconstruction
17:30 Poster panel: 119

Poster Number:
MIC-10-119

Design and Quantitative Characterization of a High-Performance Brain-Dedicated PET System via Virtual Pinhole PET Technique (#1919)

Y. Chen1, Z. Yuan1, Y. Chen1, C. Li1, K. Wang1, Y. Hou1, Y.-C. Tai2, J. Jiang1

1 Beijing Normal University, School of Physics and Astronomy, Beijing, China
2 Washington University in St. Louis, Mallinckrodt Institute of Radiology, St. Louis, Missouri, United States of America

Abstract

Positron emission tomography (PET) is an indispensable tool in neuroscience, yet its efficacy in early lesion detection and advanced brain research is often constrained by the fundamental trade-offs between spatial resolution and sensitivity. To surpass these limitations, this study introduces the NeuroEXPLORER with Insert (NXI), a high-performance dedicated brain PET system that integrates Virtual Pinhole (VP-PET) technology into the NeuroEXPLORER (NX) platform. We developed a comprehensive quantitative reconstruction framework incorporating a full suite of physical correction techniques tailored for the NXI’s asymmetric geometry. The system's performance was rigorously validated through Monte Carlo simulations using GATE. Point-source evaluations demonstrated a enhancement in overall sensitivity, driven by an increased proportion of high-resolution insert-scanner (IS) coincidence events in the vicinity of the insert. An ablation study performed on uniform cylinder data, demonstrated the effectiveness of the integrated correction suite, including normalization, attenuation, random, and scatter corrections. In Derenzo phantom experiments, the NXI system successfully resolved 0.6 mm hot rods, a benchmark that remained unresolvable by the standard NX configuration. Furthermore, high-resolution digital brain phantom studies confirmed the system's clinical potential, showing a 5.74% increase in data acquisition volume and a 19.31% improvement in the contrast-to-noise ratio (CNR) of gray and white matter compared to the NX system. By synergizing optimized hardware design with specialized reconstruction algorithms, this work verifies that the NXI architecture substantially elevates both sensitivity and spatial resolution, providing a robust technical pathway for the development of next-generation dedicated brain PET instrumentation.

Acknowledgment

This work was supported in part by NSFC (12475336), FRFCU (310400209521) and BNSF (QY25290).

References
[1] Li, H et al. 2024, 'Performance Characteristics of the NeuroEXPLORER, a Next-Generation Human Brain PET/CT Imager', Journal of Nuclear Medicine, 65(8), 1320-1326, Reston, VA: Society of Nuclear Medicine and Molecular Imaging
[2] Jiang, J et al. 2023, 'A virtual-pinhole PET device for improving contrast recovery and enhancing lesion detectability of a one-meter-long PET scanner: a simulation study', Phys Med Biol, 68(14), Bristol: IOP Publishing
[3] Jan, S et al. 2004, 'GATE: a simulation toolkit for PET and SPECT', Phys Med Biol, 49(19), 4543-4561, Bristol: IOP Publishing
[4] Yuan, Z et al. 2026, 'GPU-based list-mode TOF PET image reconstruction with complete correction techniques', Med Phys, 53, e70216, Hoboken, NJ: John Wiley & Sons
Keywords: Brain-dedicated PET, Virtual Pinhole PET, Quantitative Image Reconstruction
17:30 Poster panel: 122

Poster Number:
MIC-10-122

Design and Performance Characterization of a Sparsely Populated DOI capable Breast PET/MRI Insert (#2070)

J. Hohnholz1, L. U. Rauscher1, M. Judenhofer2, E. Olyaei1, 3, H. P. Vo4, M. Rafecas4, B. Pichler1, 5, F. Schmidt1, 3

1 Werner Siemens Imaging Center, Department of Preclinical Imaging and Radiopharmacy, University of Tuebingen, Tuebingen, Baden-Württemberg, Germany
2 Siemens Medical Solutions USA, Inc., Knoxville, Tennessee, United States of America
3 University Hospital Tuebingen, Nuclear Medicine and Clinical Molecular Imaging, Tuebingen, Baden-Württemberg, Germany
4 Institute of Medical Engineering, University of Luebeck, Luebeck, Schleswig-Holstein, Germany
5 Cluster of Excellence iFIT (EXC2180) “Image Guided and Functionally Instructed Tumor Therapies”, Tuebingen, Baden-Württemberg, Germany

Abstract

To improve diagnostic accuracy in hybrid breast imaging, we develop a dedicated PET/MR breast insert for integration into a clinical whole-body PET/MR system. The final insert consists of two cylindrical detector rings (⌀ = 178 mm, height = 95 mm), each comprising five rings with 32 detector blocks per ring. A sparsely populated (18/32 blocks) single-ring prototype was realized, using dual-layer, depth-of-interaction capable 10×10/9×9 LSO scintillator arrays (1.6×1.6×9.8 / 5.2 mm³) and SiPM readout connected via 30 cm flexible cables to MRI-compatible front-end electronics. Detector calibration and bias voltage optimization demonstrated changes in CRT of up to 100 ps between 42-47V. A bias voltage of 45 V was selected as a stable operating point, maintaining consistent energy resolution and coincidence count rates. Reliable crystal identification was achieved for both layers, with a peak-to-valley ratio of 3.7. At this operation point, the prototype achieved energy resolutions of 13.1 ± 1.8% FWHM (top layer) and 11.8 ± 2.4% (bottom layer). CRT depended on the applied energy window, yielding 383 ± 34 ps FWHM for 430–610 keV and improving to 349 ± 32 ps for a 511±50 keV window, while degrading to 483 ± 175 ps for 250–760 keV. Compared to previous system configurations (>500 ps), the updated readout enabled a substantial improvement in timing performance. These results demonstrate robust energy and timing performance of a sparsely populated, DOI-capable PET/MR insert under realistic integration constraints, including extended signal routing and MRI-compatible system design. The findings support the scalability towards a fully populated double-ring system currently under construction, enabling systematic evaluation of PET/MR performance under stress-test sequences and realistic breast imaging conditions within a clinical PET/MR scanner.

AcknowledgmentFunding DFG 508064995
References
[1] L. U. Rauscher et al., 2025 IEEE NSS MIC RTSD Conference, Nov 2025.
Keywords: dedicated breast PET, hybrid imaging, PET/MR, MR compatibility, PET performance
17:30 Poster panel: 125

Poster Number:
MIC-10-125

Enabling DOI capability in Low-Cost BGO PET System for Small-Animal Imaging (#2378)

H. Zou1, S. Han2, X. Yu2, W. Wang1, L. Zhao1, L. Zhao1, D. Liang1, H. Zhao1, J. Tao1, H. Liu1, Y. Zhang1, Q. Peng1

1 Shenzhen Bay Laboratory, Institute of Biomedical Engineering, Shenzhen, China
2 Huazhong University of Science and Technology, School of Mechanical Science and Engineering, Wuhan, China

Abstract

This work presents a low‑cost small‑animal PET system based on BGO crystals with depth‑of‑interaction (DOI) capability. A 10×8 BGO array (1.5×1.5×20 mm³) is coupled to an MPPC array in a 2×2 configuration, and a 6‑mm light‑sharing window (LSW) encodes DOI‑dependent light sharing. Using a rapid calibration method (vertical line source + histogram matching), the detector achieves an average DOI resolution of 4.9 mm FWHM and an energy resolution of 17.4% at 511 keV. System‑level evaluation shows a spatial resolution of 1.43 mm FWHM at the center and resolves 1.5‑mm hot rods. The design eliminates labor‑intensive mechanical collimation, offering an affordable, maintainable solution for high‑sensitivity preclinical imaging.

AcknowledgmentThis work was supported by the National Natural Science Foundation of China (Grant no. 82327809), the National Key Research and Development Program of China (Grant No. 2024YFF0507700 and 2024YFF0727400),  National High-Level Hospital Clinical Research Funding (2025-PUMCH-D-003), Beijing Natural Science Foundation(L252047). 
References
[1] Gong, K., Cherry, S. R. & Qi, J. On the assessment of spatial resolution of PET systems with iterative image reconstruction. Phys. Med. Biol. 61, N193–N202 (2016)
Keywords: depth-of-interaction, small animal PET, low‑cost PET, light-sharing window, BGO scintillator
17:30 Poster panel: 128

Poster Number:
MIC-10-128

Nested plastic scintillator PET detector geometries have distinct spatial resolution advantages (#2558)

M. Rädler1, 2, P. Moskal1, 2

1 Jagiellonian University, Centre for Theranostics, Kraków, Poland
2 Jagiellonian University, Faculty of Physics, Astronomy and Applied Computer Science, Kraków, Poland

On behalf of the J-PET collaboration

Abstract

The J-PET (Jagiellonian Positron Emission Tomography) collaboration is developing a large Field-Of-View (FOV) Total-Body J-PET (TB-J-PET) scanner [1,2,3,4] enabling, among others, low dose imaging, short scan times, and dynamic imaging. The key distinctive feature, i.e., the utilization of cost-effective plastic scintillators, could facilitate a wider adoption of large FOV scanners. In this work, we study the benefit of combining the TB-J-PET with a Brain PET Insert (BI) via Monte Carlo simulations. We show that the proposed detector configuration has unique advantages enhancing the spatial resolution. For this purpose, we first developed a rigorous and accurate scheme to estimate spatial resolution from Maximum Likelihood Expectation Maximization (MLEM) reconstructions. We thereby address known challenges of MLEM-based spatial resolution assessment [5], which is discouraged by NEMA standards that continue to rely on the suboptimal Filtered Back-Projection (FBP). We then investigate the impact of the detector geometry on the resolution. The BI uses narrower scintillators than the TB-J-PET (4 vs. 6 mm). We show that this is a resource‑efficient approach to achieve high resolution inside the BI, since the coincidences between TB-J-PET and BI (~50% of sensitivity) have particularly low uncertainty. We also show that this detector configuration is geometrically optimal and suppresses the non-collinearity effect. We back up our findings with theoretical analyses. We also investigate other factors influencing spatial resolution, including event selection, detector scattering, phantom scattering, and positron range. We provide intuitive insights, offering compelling arguments for the experimental realization of the detector geometry under investigation.

AcknowledgmentNational Science Centre of Poland (2021/42/A/ST2/ 00423, 2021/43/B/ST2/02150, 2022/47/I/NZ7/03112), SciMat and qLife under the Excellence Initiative at Jagiellonian University, and PLGrid (ACK Cyfronet AGH, PLG/2024/017688).
References
[1] Moskal, P. et al. 2020, ‘Prospects and clinical perspectives of total-body PET imaging using plastic scintillators’, PET clinics, 15.4, 439-452
[2] Moskal, P. et al. 2020, ‘Simulating NEMA characteristics of the modular total-body J-PET scanner - an economic total-body PET from plastic scintillators’, Physics in Medicine & Biology, 66.17, 175015
[3] Baran, J. et al. 2025, ‘Realistic total-body J-PET geometry optimization: Monte Carlo study’, Medical Physics, 52.4, 2961-2975
[4] Tayefi Ardebili, K et al. 2025, ‘Development of a Cost-Effective Total Body J-PET From Plastic Scintillators: Definitive Design’, IEEE NSS MIC RTSD conference, 1-2
[5] Gong, K. et al. 2016, ‘On the assessment of spatial resolution of PET systems with iterative image reconstruction’, Physics in Medicine & Biology, 61.5, N193–N202
Keywords: PET, brain PET, plastic scintillators, Monte Carlo simulations, GATE simulations
17:30 Poster panel: 131

Poster Number:
MIC-10-131

Adapt-PET: A Hardware-Algorithm Co-evolution Framework for Versatile, Ultra-High-Resolution Targeted Imaging (#2661)

R. Liao1, Z. Yuan1, K. Wang1, Y. Chen1, Y. Chen1, Y. Tai2, J. Jiang1

1 Beijing Normal University, School of Physics and Astronomy, Beijing, China
2 Washington University in St. Louis, Mallinckrodt Institute of Radiology, St. Louis, China

Abstract

PET remains a cornerstone of molecular imaging,yet the trade-off between spatial resolution,sensitivity,and system versatility persists.While integrating high-resolution inserts into existing systems can enhance performance within a targeted field-of-view,conventional designs lack geometric symmetry in insert-scanner(IS)and insert-insert(II)LORs.This asymmetry necessitates exhaustive,configuration-specific normalization data,severely limiting the mobility and clinical adaptability of the insert.To overcome these limitations,we propose the Adapt-PET(Adaptive-geometry PET)design strategy,which utilizes a "Hardware-Algorithm Co-evolution" approach featuring a movable,DOI-capable insert.The insert geometry is derived by proportionally downscaling the base scanner’s transaxial dimensions while maintaining axial consistency,followed by systematic crystal subdivision.This configuration enables a unified geometric mapping strategy that projects IS and II LORs onto a standard scanner-scanner(SS)space.By exploiting intrinsic symmetries,our approach achieves geometric compression ratios comparable to baseline systems,effectively decoupling system normalization from hardware geometry and eliminating the need for re-normalization upon insert relocation.The proposed strategy was validated through GATE simulations of two prototypes integrated with a Siemens Biograph Vision scanner.Positioned at an inner radius of 200-mm,each insert comprised 24 modules of 30×40 subdivided crystal arrays(1.047×0.749×15 mm³)with a 6-bin,2.5-mm DOI resolution.Evaluations using a uniform cylinder show that the system achieves near-quantitative reconstruction accuracy.Derenzo phantom imaging results reveal that the peak-to-valley ratio can be improved by up to 20%,while torso phantom results demonstrate a substantial enhancement in the detectability of 4-mm lesions.This study establishes a practical framework for flexible,organ-targeted PET imaging through hardware-algorithm synergy.

AcknowledgmentNSFC (No.12475336)
References
[1] Tai, Yuan-Chuan, et al. 2008, ‘Virtual-pinhole PET’, Journal of Nuclear Medicine, 49.3, 471-479, Reston, VA: Society of Nuclear Medicine and Molecular Imaging
[2] Zhou, Jian, Qi, Jinyi 2009, ‘Theoretical analysis and simulation study of a high-resolution zoom-in PET system’, Physics in Medicine & Biology, 54.17, 5193-5208, Bristol, UK: IOP Publishing
[3] Yuan, Ziquan, et al. 2026, ‘GPU‐based list‐mode TOF PET image reconstruction with complete correction techniques’, Medical Physics, 53.1, e70216, Hoboken, NJ: John Wiley & Sons
[4] Qi, Jinyi, et al. 1998, ‘High-resolution 3D Bayesian image reconstruction using the microPET small-animal scanner’, Physics in medicine & biology, 43.4, 1001-1013, Bristol, UK: IOP Publishing
Keywords: Virtual Pinhole PET, Normalization correction, Geometric symmetry, Tumor detectability, Depth-of-interaction
17:30 Poster panel: 134

Poster Number:
MIC-10-134

17:30 Poster panel: 137

Poster Number:
MIC-10-137

Performance evaluation of the third generation of TOF-DOI Prism-PET detector modules (#3032)

S. Shabani Sichani1, 2, Z. Wang1, W. Tan1, D. A. Muller3, J. Qi4, G. C.-Y. Chiang1, A. H. Goldan1

1 Weill Cornell Medicine, Radiology, New York, New York, United States of America
2 Cornell University, Physics, Ithaca, New York, United States of America
3 Cornell University, Applied Physics, Ithaca, New York, United States of America
4 University of California, Davis, Biomedical Engineering, Davis, California, United States of America

Abstract

Poor spatial resolution and partial volume effect (PVE) limit PET’s ability to accurately quantify radiotracer uptake in small brainstem nuclei such as the locus coeruleus, one of the earliest sites to undergo neuropathological changes in the early stages of tau accumulation associated with Alzheimer’s disease. Prism-PET is designed to improve spatial resolution and mitigate the PVE by combining ultrahigh-resolution detector modules with high scanner sensitivity in a practical and cost-effective system.

Latest generation of Prism-PET detector module was evaluated using both qualitative and quantitative performance assessments. The module’s flood map was examined to assess crystal separation, pattern stability, and the reliability of crystal identification across the detector module. The tested module produced a clear and stable flood map with well-separated crystal patterns, indicating robust event localization. Quantitatively, we measured the detector’s energy resolution, depth-of-interaction (DOI) resolution, and coincidence timing resolution. The tested module achieved 10.34% energy resolution, 3.23 mm FWHM DOI resolution, and 337 ps coincidence timing resolution. In the future work, we will report on the performance of the third generation Prism-PET scanner.

References
[1] Tan, W., Wang, Z., Zeng, X., Boccia, A., Wang, X., Li, Y., Li, Y., Fung, E.K., Qi, J., Zeng, T., Gupta, A. and Goldan, A.H. 2025, ‘High-resolution motion compensation for brain PET imaging using real-time electromagnetic motion tracking’, Medical Physics, vol. 52, no. 1, pp. 201–218, United States: Wiley/AAPM.
[2] Zeng, X., Wang, Z., Tan, W., Petersen, E., Cao, X., LaBella, A., Boccia, A., Franceschi, D., de Leon, M., Chiang, G.C.-Y., Qi, J., Biegon, A., Zhao, W. and Goldan, A.H. 2023, ‘A conformal TOF–DOI Prism-PET prototype scanner for high-resolution quantitative neuroimaging’, Medical Physics, vol. 50, no. 6, pp. 3401–3417, United States: Wiley/AAPM.
[3] Li, Y., LaBella, A., Zeng, X., Wang, Z., Petersen, E., Cao, X., Zhao, W. and Goldan, A.H. 2023, ‘Interleaved signal multiplexing readout in depth encoding Prism-PET detectors’, Medical Physics, vol. 50, no. 7, pp. 4234–4243, United States: Wiley/AAPM.
[4] Wang, Z., Cao, X., LaBella, A., Zeng, X., Biegon, A., Franceschi, D., Petersen, E., Clayton, N., Ulaner, G.A., Zhao, W. and Goldan, A.H. 2022, ‘High-resolution and high-sensitivity PET for quantitative molecular imaging of the monoaminergic nuclei: A GATE simulation study’, Medical Physics, vol. 49, no. 7, pp. 4430–4444, United States: Wiley/AAPM.
[5] Merlin, T., Stute, S., Benoit, D., Bert, J., Carlier, T., Comtat, C., Filipovic, M., Lamare, F. and Visvikis, D. 2018, ‘CASToR: A generic data organization and processing code framework for multi-modal and multi-dimensional tomographic reconstruction’, Physics in Medicine & Biology, vol. 63, no. 18, Art. no. 185005, Bristol, UK: IOP Publishing.
Keywords: Alzheimer’s disease, depth-of-interaction, neuroimaging, Prism-PET, ultrahigh-resolution PET
17:30 Poster panel: 140

Poster Number:
MIC-10-140

Clinical Validation of MoDeCo: A Motion Detection and Correction Method for a Dedicated Brain PET Scanner (#3187)

E. Muñoz1, D. Sánchez1, P. Bascuñana3, M.N. Cabrera-Martín3, J.V. Catret2

1 Oncovision, Image Department, Valencia, Spain
2 Oncovision, CTO, Valencia, Spain
3 Hospital Clínico San Carlos, Nuclear Medicine, Madrid, Spain

Abstract

Patient motion during PET acquisition is a well-known source of image degradation. In dedicated brain PET systems such as CareMiBrain (Oncovision), patients occasionally slide downward during acquisition despite head immobilization, compromising diagnostic quality in a clinically relevant subset of studies. We present MoDeCo (Motion Detection and Correction), a modified iterative reconstruction algorithm that detects and compensates for intra-acquisition rigid head motion without requiring external tracking hardware. The listmode data are split into temporal fragments; fast preliminary reconstructions are co-registered to the first fragment to estimate rigid transformations, and a modified MLEM/OSEM algorithm applies them within each subiteration to combine the partial back-projections. Head motion was quantified on 256 [18F]FDG patient studies acquired at Hospital Clínico San Carlos (Madrid), and a qualitative clinical comparison was performed on 15 selected studies. Maximum head displacement during acquisition exceeded 5 mm in 14% of patients. In the high-motion subset (n = 10), MoDeCo was rated significantly superior to standard OSEM in 60% of cases and moderately superior in 20%; mild-motion studies (n = 5) were rated as equivalent. No case showed standard reconstruction outperforming MoDeCo. MoDeCo improves image quality in patients with significant motion while preserving equivalence in those without appreciable motion.

Acknowledgment

The authors thank the Nuclear Medicine Department of Hospital Clínico San Carlos (Madrid) for support in patient recruitment, image acquisition, and clinical evaluation.

References
[1] P. M. Bloomfield et al., "A motion correction scheme for neurological PET," Phys. Med. Biol., vol. 48, no. 8, pp. 959-978, 2003.
[2] Y. Iwao, H. Tashima, E. Yoshida, F. Nishikido, T. Ida, and T. Yamaya, "Seated versus supine: consideration of the optimum measurement posture for brain-dedicated PET," Phys. Med. Biol., vol. 64, no. 12, p. 125003, Jun. 2019, doi: 10.1088/1361-6560/ab221d.
[3] Oncovision, "CareMiBrain dedicated brain PET scanner," Technical documentation, Valencia, Spain, 2023.
[4] R. R. Fulton et al., "Correction for head movements in PET using an optical motion-tracking system," IEEE Trans. Nucl. Sci., vol. 49, no. 1, pp. 116-123, Feb. 2002.
Keywords: CareMiBrain, clinical validation, dedicated Brain PET, motion correction
17:30 Poster panel: 143

Poster Number:
MIC-10-143

Impact of beam hardening and detector response function corrections when using photon-counting imaging detectors (#1206)

H. Hayashi1, R. Nishigami2, D. Kobayashi2, T. Kasue2, N. Kimoto3, T. Asahara4

1 Kanazawa University, College of Transdisciplinary Sciences for Innovation, Kanazawa, Japan
2 Kanazawa University, Graduate School of Medical Sciences, Kanazawa, Japan
3 Junshin Gakuen University, Department of Radiological Science, Faculty of Health Sciences, Fukuoka, Japan
4 Okayama University, Faculty of Health Sciences, Okayama, Japan

Abstract

Photon-counting detectors (PCDs) can perform quantitative imaging by utilizing the differences in the individual energies of X-ray photons, but this analysis requires various corrections, such as charge sharing, to determine the correct attenuation amount. The purpose of this study is to quantitatively clarify the importance of beam hardening and detector response corrections, which can be applied by software processing, when performing imaging with polychromatic X-rays using a line-sensor-type PCD having an anti-coincidence mode. Our PCD was a CdTe detector manufactured by Varex Imaging (pixel size 0.1 mm, thickness 0.75 mm). The response characteristics were calculated using the Monte Carlo simulation code EGS5, and the charge summing effect related to the "anti-coincidence mode" in the four clustered pixels was reproduced in a simple way. The evaluation method involved examining the reproducibility of quantitative values in the effective atomic number (Zeff) images. Simulations using various virtual phantoms and experiments using a non-destructive testing system with the PCD-type line sensor were conducted. As a result, we demonstrated that by including our proposed correction procedure with the anti-coincidence mode, it is possible to generate highly quantitative Zeff images. In conclusion, this study verified that when using PCDs, it is important to implement software-based beam hardening and detector response function corrections in addition to using the anti-coincidence modes.

Acknowledgment

The authors declare the following financial interests/personal relationships which may be considered as potential competing interests: This work was supported by JSPS KAKENHI Grant Number [KAKENHI: 24K03306 (Hayashi), 25K21586 (Asahara), 25K21600 (Kimoto), 25KJ1355 (Kobayashi)], and the Sasakawa Scientific Research Grant.

References
[1] Hou B., Chen Q., Yi L., Sellin P., Sun H.T., Wong L. J., Liu X., 2024, ‘Materials innovation and electrical engineering in X-ray detection’, Nature Reviews Electrical Engineering 1, 639-655.
[2] Kimoto N., Hayashi H., Asakawa T., Lee C., Asahara T., Maeda T., Goto S., Kanazawa Y., Katsumata A., Yamamoto S., Okada M., 2021, ‘Effective atomic number image determination with an energy-resolving photon-counting detector using polychromatic X-ray attenuation by correcting for the beam hardening effect and detector response’, Applied Radiation and Isotopes 170, 109617.
[3] Kimoto N., Hayashi H., Lee C., Maeda T., Ando M., Kanazawa Y., Katsumata A., Yamamoto S., Okada M., 2021, ‘A novel algorithm for extracting soft-tissue and bone images measured using a photon-counting type X-ray imaging detector with the help of effective atomic number analysis’, Applied Radiation and Isotopes 176, 109822.
[4] Asahara T., Okada S., Hayashi H., Maeda T., Nishigami R., Kobayashi D., Kurose C., Kimoto N., Goto S., Hisatomi M., Yanagi Y., Iguchi T., 2025, ‘Helpfulness of effective atomic number image in forensic dental identification: photon-counting computed tomography is suitable’, Computers in Biology and Medicine 184, 109333.
[5] Asahara T., Mitani M., Kimoto N., Nishigami R., Takegami K., Morimitsu Y., Akagi N., Miyoshi T., Kanazawa Y., Iguchi T., Hayashi H., 2025, ‘Applicability of effective atomic number (Zeff) image analysis of coronary plaques measured with photon counting computed tomography’, Investigative Radiology, in press.
[6] Nishigami R., Kobayashi D., Kimoto N., Asahara T., Maeda T., Haba T., Kanazawa Y., Yamamoto S., Hayashi H., 2025, ‘Optimization of energy windows to calculate quantitative X-ray images using an energy-resolving photon-counting detector: A simulation study’, Radiation Physics and Chemistry 229, 112460.
[7] Nishigami R., Kimoto N., Asahara T., Maeda T., Kobayashi D., Goto S., Haba T., Kanazawa Y., Yamamoto S., Hayashi H., 2026, ‘Bone-enhanced high contrast X-ray images derived from attenuation estimation related to ultra-low energy X-rays – An application of an energy-resolving photon-counting detector (ERPCD) –’, Radiation Physics and Chemistry 238, 113243.
[8] Kimoto N., Nishigami R., Asahara T., Kanazawa Y., Katsumata A., Yamamoto S., Hayashi H., 2026, ‘Counting loss correction algorithm for an energy-resolving photon-counting detector for X-ray imaging of biological objects with effective atomic numbers between 3 and 20’, Radiation Physics and Chemistry 243, 113723.
Keywords: Photon-counting detector, Charge sharing correction, Effective atomic number image
17:30 Poster panel: 146

Poster Number:
MIC-10-146

Evaluation of Lead-Free Composite for X-ray Shielding using LD-V1 Film (#1319)

D. T. Tran1, 2, M. Kim1, S. Park1, 2

1 Korea Institute of Radiological and Medical Sciences, Seoul, Republic of Korea
2 University of Science and Technology, Radiological and Medico-Oncological Sciences, Seoul, Republic of Korea

Abstract

This study evaluated a lead-free BaSO4–silicone composite for diagnostic X-ray shielding (60–100 kVp) using LD-V1 film, combining MCNP6 simulation with experimental validation. Evaluation of a 2 mm thick composite demonstrated shielding efficiencies ranging from 68.79% to 83.50%. Simulated results showed good agreement with experimental measurements, with a maximum deviation of 10.28%. Furthermore, the two-dimensional dose distribution map revealed an average radiation reduction of approximately 60.34% across the material. This result confirms effective attenuation of X-ray radiation across the irradiated field. The dose reduction was also spatially uniform over the entire protected area. In addition, the consistent dose reduction pattern indicates a homogeneous dispersion of BaSO4 particles within the silicone matrix. These findings establish the BaSO4–Si composite as an effective lead-free alternative and demonstrate the utility of LD-V1 film for both quantitative dosimetry and spatial validation in clinical X-ray systems.

AcknowledgmentThis work was supported by the Commercialization Promotion Agency for R&D Outcomes(COMPA) grant funded by the Korea government(Ministry of Science and ICT) (RS-2025-02309416).
Keywords: Diagnostic X-ray, radiochromic LD-V1 film, radiation shielding, Lead-free composite, Monte Carlo simulation
17:30 Poster panel: 149

Poster Number:
MIC-10-149

Evaluation of image resolution for Novel Dual-energy X-ray Absorptiometry System (#1441)

H. Leem1, Y. Ahn1, S. Min1, M. Oh1, C. Kwon1

1 Osteosys Corp., RnD center, Seoul, Republic of Korea

Abstract

DXA has evolved beyond standard BMD measurement to include sarcopenia diagnosis and bone quality assessment via the TBS. As TBS is now a recognized modifier in the WHO FRAX Tool , image resolution has become the paramount technical factor determining the performance of DXA systems. This study evaluates the image resolution of iNSiGHT C510 (Osteosys, Korea), the novel DXA model designed for high-resolution whole-body imaging within 3 minutes.

The image resolution of iNSiGHT C510 was quantitatively assessed using  a  line phantom. The system achieved a vertical resolution of 0.63 lp/mm and a horizontal resolution of 0.9 lp/mm, demonstrating superior resolution capabilities compared to the industry-leading GE Prodigy. Anatomical discriminability was further verified using the spine phantom  with increasing BMD levels. Analysis of peak-to-valley ratios, which represent the contrast clarity between bone and intervertebral discs, confirmed that iNSiGHT C510 maintains excellent discriminability despite its longitudinal scan architecture. These quantitative results suggest that the system minimizes ROI setting errors, ensuring high clinical reproducibility for TBS calculations that are highly sensitive to subtle signal variations.

This study confirms that iNSiGHT C510 provides image resolution equivalent to global standards. Hardware architecture establishes a technical foundation for reducing identification errors in spinal measurements and improving ROI precision. Consequently, iNSiGHT C510 satisfies the performance requirements demanded by the high-precision sarcopenia diagnostic market. Future work will involve integrating ToF technology to suppress scatter noise inherent in wide fan-beam architectures.
Acknowledgment

This research was financially supported by the Ministry of Trade, Industry and Resources (MOTIR), Korea, under the “Global Industrial Technology Cooperation Center(GITCC) program” supervised by the Korea Institute for Advancement of Technology (KIAT).(Task No. P0030253)

References
[1] Osterhoff, G, McCloskey, EV 2016, 'Bone mechanical properties and changes with osteoporosis', Injury, vol. 47, Suppl. 2, pp. S11–S20.
[2] McCloskey, EV et al. 2016, 'A meta-analysis of trabecular bone score in fracture risk prediction and its relationship to FRAX', Journal of Bone and Mineral Research, vol. 31, no. 5, pp. 940–948.
[3] Martineau, P, Leslie, WD 2015, 'Spine trabecular bone score precision, a comparison between GE Lunar standard and high-resolution densitometers', Journal of Clinical Densitometry, vol. 18, no. 1, pp. 92–97.
[4] Binkley, N et al. 2015, 'Dual-energy X-ray absorptiometry image resolution and TBS precision', Journal of Clinical Densitometry, vol. 18, no. 4, pp. 434–435.
[5] Ha, YC, Kim, S, Yoo, JI 2025, 'Correlation of Lean Mass Measurements Between a Novel Whole-body X-ray Bone Densitometer (iNSiGHT C510) and Magnetic Resonance Imaging: A Single-center Comparative study', Journal of Clinical Densitometry, vol. 28, p. 101585.
[6] Takata, W et al. 2016, 'Simulated increased soft tissue thickness artefactually decreases trabecular bone score: a phantom study', BMC Musculoskeletal Disorders, vol. 17, no. 1, p. 13.
[7] Boudousq, V et al. 2005, 'Image resolution and magnification using a cone beam densitometer: optimizing data acquisition for hip morphometric analysis', Osteoporosis International, vol. 16, pp. 813–822.
[8] Hind, K, Cooper, W, Oldroyd, B, Davies, A, Rhodes, L 2015, 'A cross-calibration study of the GE-Lunar iDXA and Prodigy for the assessment of lumbar spine and total hip bone parameters via three statistical methods', Journal of Clinical Densitometry, vol. 18, no. 1, pp. 86–92.
[9] Rossignol, J, Martinez Turtos, R, Gundacker, S, Gaudreault, D, Auffray, E, Lecoq, P, Bérubé-Lauzière, Y, Fontaine, R 2020, 'Time-of-flight computed tomography - proof of principle', Physics in Medicine and Biology, vol. 65, no. 8, p. 085013.
Keywords: Dual-energy X-ray Absorptiometry, Bone Mineral Density, Fracture Risk Assessment, Trabecular Bone Score, Time-Of-Flight
17:30 Poster panel: 152

Poster Number:
MIC-10-152

Physics-Guided Retinexformer for Beam Hardening Artifact Correction in Rock CBCT (#1537)

Y. Ma1, L. Li1

1 Tsinghua University, Department of Engineering Physics, Beijing, China

Abstract

Beam hardening in cone-beam computed tomography (CBCT) of high-density rock causes cupping artifacts and nonlinear CT value deviations, hindering quantitative mineral analysis. Conventional deep learning methods often over-smooth microstructures due to a lack of physical constraints. We propose a physics-informed Retinexformer framework that reformulates the polychromatic X-ray degradation as a multiplicative Retinex model. Under this paradigm, the non-uniform hardening field is treated as a degraded illumination map, while the pristine rock structure represents the ideal reflectance. A pre-trained Denoising Diffusion Implicit Model (DDIM), combined with mask-guided moment matching, is utilized to explicitly generate a statistically aligned physical hardening field prior. Furthermore, a multi-scale feature encoder extracts high-frequency structural details from the residuals between the input and the prior to guide the denoising process. Experiments on a 10-cm diameter underground rock dataset show that our method significantly corrects CT value deviations while preserving the fine structure. Quantitatively, it achieves a PSNR of 37.13 dB and an SSIM of 0.940, reducing RMSE to 0.0149—significantly outperforming SOTA baselines (NAFNet, Retinexformer, MambaIRv2). Notably, our model achieves this with only 2.00M parameters and 19.42G FLOPs, offering a highly efficient and interpretable solution for CT artifacts correction in dense media.

AcknowledgmentThe authors would like to acknowledge Yuanhao Cai and the co-authors of Retinexformer for making their source code publicly available, which significantly facilitated the implementation and development of our proposed architecture.
References
[1] L. Chen et al., “Simple baselines for image restoration,” European Conference on Computer Vision (ECCV), Springer, pp. 17-33, 2022, doi: https://doi.org/10.1007/978-3-031-20071-7_2.
[2] H. Guo et al., “Mambairv2: Attentive state space restoration,” Proceedings of the Computer Vision and Pattern Recognition Conference, 2025.
[3] Y. Cai et al., “Retinexformer: One-stage Retinex-based Transformer for Low-light Image Enhancement,” Proceedings of the IEEE/CVF International Conference on Computer Vision (ICCV), pp. 12504-12513, 2023.
[4] J. Song, C. Meng, and S. Ermon, “Denoising diffusion implicit models,” in International Conference on Learning Representations (ICLR), 2021, doi: https://doi.org/10.48550/arXiv.2010.02502.
Keywords: Beam hardening artifacts, CBCT, diffusion model, deep learning
17:30 Poster panel: 155

Poster Number:
MIC-10-155

17:30 Poster panel: 158

Poster Number:
MIC-10-158

Mitigating Scatter-Induced BMD Errors in Wide Fan-Beam DXA Using Anti-Scatter Grids (#2260)

M. Oh1, Y. Ahn1, S. Min1, H. Leem1, C. Kwon1

1 Osteosys Corp., R&D Center, seoul, Republic of Korea

Abstract

Dual-energy X-ray absorptiometry (DXA) is the clinical standard for bone mineral density (BMD) assessment. Modern wide fan-beam DXA architectures (fan angle ≥ 20°) shorten whole-body acquisition to under three minutes but enlarge the irradiated volume, raising the scatter-to-primary ratio with field size. Conventional DXA omits anti-scatter grids, relying on the long object-to-image-detector distance (OID > 580 mm) for scatter rejection via the air-gap effect. Whether this assumption holds for wide fan-beam DXA has not been directly tested. We investigated residual scatter in an experimental wide fan-beam DXA system using three complementary experiments on step-wedge and anthropomorphic phantoms. First, we implemented an outer-channel scatter probe and detected measurable scatter at outer channels even at OID > 580 mm. Second, with an 8:1 anti-scatter grid we quantified the dual-energy mass attenuation difference Δ(μ/ρ); at the maximum phantom configuration (25 cm polycarbonate + 4.9 cm aluminum), scatter caused an inversion of Δ(μ/ρ) in the no-grid configuration (−0.018 cm²/g); the 8:1 grid restored Δ(μ/ρ) to 0.076 cm²/g — a partial recovery only. Third, a 14:1 grid successfully suppressed the spurious BMD elevation. Compared to the no-grid baseline which showed a severe thickness-dependent BMD bias (coefficient of variation of approximately 19%), the 14:1 grid reduced the variability to 3% across 0–140 mm of added polycarbonate layers. These findings establish that explicit scatter compensation — whether through anti-scatter grids or software-based correction — is essential for thick-patient measurements at wide fan-beam DXA.

AcknowledgmentThis research was financially supported by the Ministry of Trade, Industry and Resource (MOTIR), Korea, under the “Global Industrial Technology Cooperation Center (GITCC) program” supervised by the Korea Institute for Advancement of Technology (KIAT) (Task No. P0030253). 
References
[1] Blake, GM, Fogelman, I 2007, ‘Role of dual-energy X-ray absorptiometry in the diagnosis and treatment of osteoporosis’, J. Clin. Densitom., vol. 10, no. 1, pp. 102–110.
[2] Messina, C et al. 2020, ‘Body composition with dual-energy X-ray absorptiometry: from basics to new tools’, Quant. Imag. Med. Surg., vol. 10, no. 8, pp. 1687–1698
[3] Blake, GM, Fogelman, I 2010, 'An update on dual-energy X-ray absorptiometry', Semin. Nucl. Med., vol. 40, no. 1, pp. 62–73.
[4] Sorenson, JA, Floch, J 1985, ‘Scatter rejection by air gaps: an empirical model’, Med. Phys., vol. 12, no. 3, pp. 308–316.
[5] Persliden, J, Carlsson, GA 1997, ‘Scatter rejection by air gaps in diagnostic radiology: calculations using a Monte Carlo collision density method’, Phys. Med. Biol., vol. 42, no. 1, pp. 155–175.
[6] Neitzel, U 1992, ‘Grids or air gaps for scatter reduction in digital radiography: a model calculation’, Med. Phys., vol. 19, no. 2, pp. 475–481.
[7] Libouban, H et al. 2002, ‘Comparison of pencil-, fan-, and cone-beam dual X-ray absorptiometers for evaluation of bone mineral content in excised rat bone’, J. Clin. Densitom., vol. 5, no. 4, pp. 355–361.
[8] Boone, JM, Cooper III, VN 2000, ‘Scatter/primary in mammography: Monte Carlo validation’, Med. Phys., vol. 27, no. 8, pp. 1818–1831.
[9] Njeh, CF et al. 1999, ‘Radiation exposure in bone mineral density assessment’, Appl. Radiat. Isot., vol. 50, no. 1, pp. 215–236.
Keywords: Anti-scatter grid, bone mineral density, dual-energy X-ray absorptiometry, wide fan-beam, scatter
17:30 Poster panel: 161

Poster Number:
MIC-10-161

Empirical Decomposition for Quantification of Multi-Contrast-Agent (#2472)

P. Chen1, 2, Z. Wang1, C. Wang1, 2, T. Jia1, S. Xie1, 2, C. Wei1

1 Institute of High Energy Physics, Beijing Engineering Research Center of Radiographic Techniques and Equipment, Beijing, China
2 University of Chinese Academy of Sciences, School of Nuclear Science and Technology, Beijing, China

Abstract

Contrast agents play an important role in spectral CT because they enhance the visibility of vessels, organs, tumors, and lesions, thereby improving image interpretation and lesion identification. However, accurately quantitative contrast-agent decomposition remains challenging. Conventional material decomposition algorithms still suffer from limited quantitative accuracy even in contrast-agent imaging. This problem becomes more severe when multiple contrast agents coexist. To address this limitation, we propose an empirical multi-agent decomposition method. The proposed method first applies conventional empirical decomposition to obtain initial decomposition results and estimate the corresponding decomposition errors. Then, bivariate error surfaces are constructed to model the relationship between iodine and gadolinium concentrations and their decomposition errors. These fitted error surfaces are subsequently used to compensate for the errors produced by conventional empirical decomposition. Simulated experiments demonstrate that the proposed method can simultaneously decompose iodine and gadolinium with improved quantitative accuracy, and the global MAPE and NRMSE are effectively reduced compared with the image-domain decomposition method for the test phantoms.

References
[1] S. Hamid, M. U. Nasir, A. So, G. Andrews, S. Nicolaou, and S. R. Qamar,“Clinical applications of dual-energy ct,” Korean Journal of Radiology, vol. 22, no. 6, pp. 970–982, jun 2021 .
[2] C. Wang, Z. Wang, M. Li, Y. Liu, X. Zhang, C. Wei, and L. Wei,“Empirical correction decomposition method (ECDM): enhancing accuracy of quantitative measurement in spectral CT,” Physics in Medicine & Biology, vol. 70, no. 1, p. 015021, Jan. 2025.
[3] M. Weiss and T. Meisen,“Reviewing material-sensitive computed tomography: From handcrafted algorithms to modern deep learning,” NDT, vol. 2, no. 3, pp. 286–310, jul 2024 .
Keywords: Photo-counting CT, spectral CT, material decomposition, contrast agent
17:30 Poster panel: 164

Poster Number:
MIC-10-164

Empirical Correction of Detector Response for Quantitative X-Ray Transmission Modelling and its Application to Computed Tomography (#2574)

Q. Chen1, B. James1, Y. Liu1

1 The Commonwealth Scientific and Industrial Research Organisation, Mineral Resources, Lucas Heights, Australia

Abstract

X-ray computed tomography (CT) relies fundamentally on accurate transmission measurements to reconstruct meaningful attenuation maps. However, detector response non-linearity systematically biases transmission measurements away from the values predicted by the polychromatic forward model, propagating these errors into reconstructed data. To address this, a combined spectral modelling and empirical calibration framework was developed for a high-resolution flat-panel detector, enabling prediction of measured transmission values across a broad range of sample thicknesses and compositions. A polychromatic spectral model was constructed by combining tube-spectrum estimates with Beer–Lambert attenuation using NIST/XCOM coefficients, convolved with an effective CsI scintillator response function. Predicted transmission values were compared against experimental measurements obtained using high purity elemental foils spanning a wide range of thicknesses. Consistent and reproducible discrepancies between modelled and measured values revealed a structured detector response non-linearity not captured by the forward model. A first-order empirical correction was derived from the model–experiment residuals, yielding a fitted detector-response exponent of α = 1.47 applied to model predictions to reproduce experimental detector behaviour. The calibrated model can therefore generate realistic transmission values consistent with measured detector output across a range of sample configurations. These findings highlight the broader potential of this framework, particularly for dual-energy CT, where material decomposition relies critically on the relative contrast between acquisitions at different tube voltages. A calibrated forward model of this kind offers a pathway toward improved relative CT threshold accuracy, supporting more reliable separation of materials with overlapping attenuation properties and more robust quantitative interpretation of reconstructed data.

References
[1] Mesina, M. B., de Jong, T. P. R. and Dalmijn, W. L. 2007, “Automatic sorting of scrap metals with a combined electromagnetic and dual energy X-ray transmission sensor,” International Journal of Mineral Processing, vol. 82, no. 4, pp. 222–232.
[2] Zhang, Y., Yoon, N. and Holuszko, M. E. 2021, “Assessment of sortability using a dual-energy X-ray transmission system for studied sulphide ore,” Minerals, vol. 11, no. 5, Art. no. 490.
[3] Moshkbar-Bakhshayesh, K., Afarideh, H. and Azimirad, R. 2023, “Inspection of cargo using dual-energy X-ray radiography: A review,” Radiation Physics and Chemistry, vol. 212, Art. no. 111180.
[4] Taguchi, K. and Iwanczyk, J. S. 2013, “Vision 20/20: Single photon counting X-ray detectors in medical imaging,” Medical Physics, vol. 40, no. 10, Art. no. 100901.
[5] Huang, S.-Y., Yang, K., Abbey, C. K. and Boone, J. M. 2012, “A semiempirical linear model of indirect, flat-panel X-ray detectors,” Medical Physics, vol. 39, no. 4, pp. 2108–2118.
[6] Berger, M. J., Hubbell, J. H., Seltzer, S. M., Chang, J., Coursey, J. S., Sukumar, R., Zucker, D. S. and Olsen, K. 2010, “XCOM: Photon Cross Sections Database,” NIST Standard Reference Database 8 (XGAM), National Institute of Standards and Technology.
Keywords: Cabinet CT, detector response, flat-panel detector, material analysis, X-ray spectrum modelling.
17:30 Poster panel: 167

Poster Number:
MIC-10-167

Synthetic generation of the dental anatomy for cone-beam CT simulation from patient data (#2593)

I. Tatachak1, 3, L. Kabongo3, N. Papadakis2, X. Ripoche3, S. Rit1

1 Creatis, INSA-Lyon, Universite Claude Bernard Lyon 1, CNRS, Inserm, CREATIS UMR 5220, U1294, F-69373, Lyon, France
2 Univ. Bordeaux, CNRS, Inria, Bordeaux INP, IMB, UMR 5251, Bordeaux, France
3 ACTEON Group, Mérignac, France

Abstract

This paper develops a pipeline for generating synthetic cone-beam computed tomography (CBCT) projections from real dental CBCT acquisitions, providing a reference for supervised learning. To this end, we segment and synthesize fine structures from real CBCT acquisitions to obtain a material map of the head volume. Using the characteristics of the CBCT scanner, notably the effective spectrum, we simulate the acquisitions by modeling a polychromatic X-ray source and adding compound Poisson noise. The results demonstrate the realism of the generated synthetic CBCT acquisitions, as well as the plausibility of the simulated artifacts.

AcknowledgmentThis work was supported by the French National Association for Technical Research (ANRT) and the France 2030 BPI project Cone Beam AI  and was performed within the framework of the SIRIC LYriCAN+ (INCa-DGOS-INSERM-ITMO cancer 18003) and the LABEX PRIMES (ANR-11-LABX-0063) of Université de Lyon, within the program ‘Investissements d’Avenir’ (ANR- 11-IDEX-0007) operated by the ANR.
References
[1] W. P. Segars, G. Sturgeon, S. Mendonca, J. Grimes, and B. M. W. Tsui, “4D XCAT phantom for multimodality imaging research,” Medical Physics, vol. 37, no. 9, pp. 4902–4915, 2010.
[2] M. ZANKL, “Adult male and female reference computational phantoms (icrp publication 110),” Japanese Journal of Health Physics, vol. 45, no. 4, pp. 357–369, 2010.
[3] K. Merken, J. Monnens, N. Marshall, N. Johan, D. M. Brasil, G. M. Santaella, C. Politis, R. Jacobs, and H. Bosmans, “Development and validation of a 3d anthropomorphic phantom for dental cbct imaging research,” Medical Physics, vol. 50, no. 11, pp. 6714–6736, 2023.
[4] A. Takeya, K. Watanabe, and A. Haga, “Fine structural human phantom in dentistry and instance tooth segmentation,” Sci Rep, vol. 14, no. 1, p. 12630, 2024.
[5] N. van Nistelrooij, L. Kr¨amer, S. Kempers, M. Beyer, F. Bolelli, T. Xi, S. Berg´e, M. Heiland, K. H. Maier-Hein, S. Vinayahalingam, and F. Isensee, “Toothseg: Robust tooth instance segmentation and numbering in cbct using deep learning and self-correction,” IEEE Journal of Biomedical and Health Informatics, 2025.
[6] F. Isensee, P. F. Jaeger, S. A. A. Kohl, J. Petersen, and K. H. Maier-Hein, “nnU-Net: a self-configuring method for deep learning-based biomedical image segmentation,” Nature Methods, vol. 18, no. 2, pp. 203–211, 2021.
[7] L. A. Feldkamp, L. C. Davis, and J. W. Kress, “Practical cone-beam algorithm,” J. Opt. Soc. Am. A, vol. 1, no. 6, pp. 612–619, 1984.
[8] S. Rit, S. Brousmiche, J. Finet, G. C. Sharp, and P. Steininger, “Reconstruction Toolkit (RTK) v2, an Insight Toolkit (ITK) module for tomographic reconstruction,” in International Conference on the use of Computers in Radiation therapy (ICCR), Lyon, France, 2024.
[9] M. Genisa, S. Shuib, Z. Rajion, E. Arief, and M. Hermana, “Density estimation based on the hounsfield unit value of cone beam computed tomography imaging of the jawbone system,”  proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine, vol. 232, p. 954411918806333, 10 2018.
Keywords: Computed Tomography, simulation, synthetic measurements
17:30 Poster panel: 170

Poster Number:
MIC-10-170

Image Quality Evaluation of Cone Beam Computed Tomography (CBCT) in an O-arm Fluoroscopy Machine Using the SedentexCT IQ Phantom (#2974)

J.P.R. S. De Guzman1, 2, A. Peralta2

1 Philippine General Hospital, Department of Radiology, Manila, Philippines
2 University of Santo Tomas, The Graduate School, Manila, Philippines

Abstract

The O-arm fluoroscopy machine, an application of cone beam computed tomography (CBCT) for interventional procedures, lacks a dedicated image quality (IQ) phantom for system image performance evaluation. This study aimed to investigate the viability and limitations of using the SedentexCT IQ phantom, a phantom originally designed for dental CBCT, to assess the image quality of resultant CBCT images of an interventional O-arm unit. Six CBCT image quality tests were performed (i.e. Uniformity, Image Noise, Geometric Accuracy, Voxel Intensity Values, Contrast Resolution, and Spatial Resolution), and corresponding parameters were assessed as baseline data in the low dose, standard dose, high dose, and enhanced cranial modes of the O-arm machine. The procedures were structured to the EFOMP-ESTRO-IAEA CBCT quality control protocol. An observed limitation of the phantom was the evaluation of the spatial resolution (lp/mm), since the lowest resolution of the SedentexCT IQ phantom is 1.0 lp/mm. Overall, since the phantom was applicable for the required CBCT image quality tests, the SedentexCT IQ phantom can be considered as a viable tool for image quality testing of the O-arm machine.

AcknowledgmentThe accomplishment of this research was attained because of the wonderful people who have contributed, showed support, and served as inspiration to this research and the researcher. I would like to express my sincere gratitude to my advisers, Ms. Agnette P. Peralta and Ms. Melanie Marquez. Ma’am Agnette’s admirable dedication to medical physics was an inspiration to contribute and make advancements in knowledge for the medical physics community in the Philippines. I am thankful for the endless support of Ma’am Melanie; she gave me the opportunity to progress with this research. Her uplifting and joyful personality always gave me hope and perseverance in finishing this research.
Keywords: Image Quality, Cone Beam Computed Tomography, Phantom
17:30 Poster panel: 173

Poster Number:
MIC-10-173

Validation of conventional mammography material decomposition using VICTRE simulated breast data (#3018)

J. Phillips1, E. Sidky1, I. Reiser1, Z. Zhang1, K. Kulkarni1, X. Pan1

1 The University of Chicago, Department of Radiology, Chicago, Illinois, United States of America

Abstract

The purpose of this work is to validate basis material decomposition in breast cancer screening mammography using a conventional single-kV scan. We employ mammography system physics modeling and a pathlength constraint to enable estimation of the basis material images. Using 3D breast phantoms and low noise Monte Carlo projection data generated with the FDA VICTRE software, we decompose transmission data into pathlength basis images of adipose and glandular tissue. The decomposition recovered adipose and glandular maps with mean absolute errors of 3.47 mm and 1.83 mm respectively. Accuracy along the skinline and nipple is degraded due to material mismatch. The results indicate material decomposition in mammography using a single energy acquisition is feasible to enable the quantitative recovery of adipose and glandular material maps.

AcknowledgmentThis research was supported in part by the National Institut of Biomedical Imaging and Bioengineering of the National Institutes of Health under award numbers R01CA287302.
References
[1] D'Orsi, CJ 1991, 'Early detection of breast cancer: mammography', Breast Cancer Research and Treatment, vol. 18, suppl. 1, pp. S107--S109, Dordrecht: Springer.
[2] Duncan, KA, Needham, G, Gilbert, FJ, Deans, HE 1998, 'Incident round cancers: what lessons can we learn?', Clinical Radiology, vol. 53, no. 1, pp. 29--32, Amsterdam: Elsevier.
[3] Smith, BJ, Dey, J, Medlock, L, Solis, D, Kirby, K 2025, 'Maximum-likelihood estimation of glandular fraction for mammography and its effect on microcalcification detection', Physical and Engineering Sciences in Medicine, vol. 48, no. 2, pp. 743--758, Cham: Springer Nature.
[4] Nissan, N, Comstock, CE, Sevilimedu, V, Gluskin, J, Mango, VL, Hughes, M, Ochoa-Albiztegui, RE, Sung, JS, Jochelson, MS 2024, 'Diagnostic accuracy of screening contrast-enhanced mammography for women with extremely dense breasts at increased risk of breast cancer', Radiology, vol. 313, no. 1, p. e232580, Oak Brook, IL: Radiological Society of North America.
[5] Leong, LT, Malkov, S, Drukker, K, Niell, BL, Sadowski, P, Wolfgruber, T, Greenwood, HI, Joe, BN, Kerlikowske, K, Giger, ML, Shepherd, JA 2021, 'Dual-energy three-compartment breast imaging for compositional biomarkers to improve detection of malignant lesions', Communications Medicine, vol. 1, no. 1, p. 29, London: Nature Publishing Group.
[6] Kis, BJ, Sarnyai, Z, K\'akonyi, R, Erd\'elyi, M, Szab\'o, G 2012, 'Single-energy material decomposition using x-ray path length estimation', Journal of Computer Assisted Tomography, vol. 36, no. 6, pp. 768--777, Philadelphia, PA: Lippincott Williams \& Wilkins.
[7] Poludniowski, G, Omar, A, Bujila, R, Andreo, P 2021, 'Technical Note: SpekPy v2.0---a software toolkit for modeling x-ray tube spectra', Medical Physics, vol. 48, no. 7, pp. 3630--3637, Hoboken, NJ: Wiley.
[8] Badano, A, Graff, CG, Badal, A, Sharma, D, Zeng, R, Samuelson, FW, Glick, SJ, Myers, KJ 2018, 'Evaluation of digital breast tomosynthesis as replacement of full-field digital mammography using an in silico imaging trial', JAMA Network Open, vol. 1, no. 7, p. e185474, Chicago, IL: American Medical Association.
[9] Huber, PJ 1964, 'Robust estimation of a location parameter', Annals of Mathematical Statistics, vol. 35, no. 1, pp. 73--101, Hayward, CA: Institute of Mathematical Statistics.
Keywords: Mammography, material decomposition, quantitative
17:30 Poster panel: 176

Poster Number:
MIC-10-176

17:30 Poster panel: 179

Poster Number:
MIC-10-179

Development of a 64-Channel Energy and Timing Acquisition System for TOF-DXA (#3146)

S.-H. Chun1, J. Jung1, J. Nam1, S. Kwon1, J.H. Jung1, Y. Choi1

1 Sogang University, Department of Electronic Engineering, Intelligent Signal Processing System (ISPS) Laboratory, Seoul, Republic of Korea

Abstract

Time-of-flight dual-energy X-ray absorptiometry (TOF-DXA) is proposed as a method for enhancing material decomposition by incorporating timing information into conventional dual-energy attenuation measurements. In this study, a 64-channel energy and timing acquisition system was developed for TOF-DXA.

The detector consisted of a 64-channel LYSO array with a pixel size of 3 × 3 × 3 mm³, coupled to four 16-channel Broadcom NUV-MT SiPM arrays. Detector signals were split into independent energy and timing paths. For the energy path, signals were processed through differential amplifiers and digitized using 10-bit multichannel analog-to-digital converters. Timing signals were processed using front-end comparators and a Kintex-7 FPGA-based tapped delay line (TDL) time-to-digital converter. The TDL was implemented using a single-chain architecture based on CARRY4 primitives.

This dual-path configuration enabled synchronized acquisition of energy and timing information on an event-by-event basis. The FPGA logic measured energy values through peak detection and synchronized them with timing data stored in a timing FIFO to ensure event-by-event alignment. The energy resolution was evaluated using a Co-57 source at 122 keV, while the timing resolution was evaluated in coincidence with a reference detector using a Na-22 source. X-ray spectra were acquired using an X-ray generator operated at 83 kVp and 0.2 mA with a Ce filter (K-edge: 40.4 keV) to evaluate system performance in the DEXA energy range.

The energy resolution was 16.3% at 122 keV, and the coincidence timing resolution was 104 ps at 511 keV. The low- and high-energy peaks were clearly resolved in the acquired X-ray spectrum.

These results demonstrated that the developed 64-channel acquisition system could simultaneously measure energy and timing information for TOF-DXA. Optimization of the acquisition board is currently in progress, and comprehensive system performance results will be presented at the conference.

References
[1] S. B. Heymsfield, et al., “Dual-energy X-ray absorptiometry: review of physical concepts,” American Journal of Physiology-Endocrinology and Metabolism, vol. 271, no. 6, pp. E941–E951, 1996.
[2] G. M. Blake and I. Fogelman, “An Update on Dual-Energy X-Ray Absorptiometry,” Seminars in Nuclear Medicine, vol. 40, no. 1, pp. 62–73, 2010.
[3] J. Jung, et al., “Development of SiPM-based DEXA detector for large detection area,” Nuclear Instruments and Methods in Physics Research Section A, vol. 1084, p. 171218, 2026.
[4] D. Portaluppi, et al., “Multi-Channel FPGA Time-to-Digital Converter With 10 ps Bin and 40 ps FWHM,” IEEE Transactions on Instrumentation and Measurement, vol. 71, p. 2002109, 2022.
Keywords: Time-of-flight, Dual-energy X-ray absorptiometry, Silicon Photomultiplier, Front-end electronics, Field programmable gate array
17:30 Poster panel: 182

Poster Number:
MIC-10-182

GARM: Geometry-Aware Reconstruction Method for Static CT Generalizable across Diverse Scanning Strategies (#1076)

P. Song1, 2, C. Chen1, 2, Z. Chen1, 2, Y. Xing1, 2

1 Tsinghua University, Department of Engineering Physics, Beijing, China
2 Tsinghua University, Key Laboratory of Particle & Radiation Imaging, Ministry of Education, Beijing, China

Abstract

Static computed tomography (SCT) enables fast, task-specific imaging through ultrafast and flexible firing among distributed focal spots. However, it introduces irregular and incomplete data acquisition, which poses great challenges to reconstruction. In this work, we propose Geometry-Aware Reconstruction Method (GARM) for SCT across diverse scanning strategies. The Geometry-Guided Dimension Reduction (G2DR) Network first transforms irregular 3D projections into slice-wise 2D fan-beam projections using geometric information derived from ray positions. A 2D dual-domain network then suppresses sparse-view artifacts guided by geometric and Radon-space information. Simulated experiments demonstrate that GARM achieves high-quality reconstruction from irregular and incomplete projections with strong generalization capability across diverse scan strategies.

AcknowledgmentThis work was supported in part by the Natural Science Foundation of Beijing (No.L242001), the National Natural Science Foundation of China (No. 12327809) and China Postdoctoral Science Foundation (No. 2025M780861)
 
References
[1] Zheng, A, Gao, H, Zhang, L, et al. 2020, ‘A dual-domain deep learning-based reconstruction method for fully 3D sparse data helical CT’, Physics in Medicine & Biology, vol. 65, no. 24, p. 245030, Bristol: IOP Publishing.
[2] Zhang, T, et al. 2020, ‘Stationary computed tomography with source and detector in linear symmetric geometry: Direct filtered backprojection reconstruction’, Medical Physics, vol. 47, no. 5, pp. 2222–2236, Hoboken: Wiley.
[3] Chen, C, et al. 2025, ‘Static CT With Sources and Detectors Distributed in a Multi-Segment Manner: System Analysis and Analytical Reconstruction’, IEEE Transactions on Computational Imaging, vol. 11, pp. 251–264, Piscataway: IEEE.
Keywords: Static Computed Tomography, sparse-view reconstruction, irregular trajectory, deep learning.
17:30 Poster panel: 185

Poster Number:
MIC-10-185

Impact of Small-Angle Event Selection in Compton Cone Imaging (#1168)

E. Medina1, M. Kolstein1, J. Caravaca2, M. Chmeissani1, B. Mehadji4, G. Ariño-Estrada1, 3

1 Institut de Física d'Altes Energies - Barcelona Institute of Science and Technology (IFAE-BIST), Medical Physics Group, Bellaterra, Barcelona, Spain
2 Lawrence Berkeley National Laboratory, Berkeley, CA, United States of America
3 University of California Davis, Department of Biomedical Engineering, Davis, CA, United States of America
4 University of California Davis, Department of Radiology in the School of Medicine, Davis, CA, United States of America

Abstract

Compton cameras are imaging systems that detect Compton-scattered gamma rays using two position-sensitive detectors, typically called the scatterer and the absorber. 
Each interaction pair defines a Compton cone enabling three-dimensional image reconstruction.
In this work, we investigate how the selection of low scattering angle events influences image quality. 
A GATE-based simulation of a Si/CdTe Compton camera system was developed to generate Compton cone data. 
A 511 keV rod phantom with source cylinders of radii from 1.5 mm to 0.75 mm was implemented, and 100 M primary photons were simulated. 
Images were reconstructed from subsets obtained by applying a maximum threshold on Compton electron energy in the scatterer. 
Energy selections of 5, 10, and 15 keV (corresponding to scattering angles of 8°, 11.4°, and 14°) were used, along with a reference dataset without any energy cut for standard CC reconstruction comparison.
We show that for two rods of 1.5 mm and 1.25 mm radius, the peak-to-valley ratio (PVR) without any energy selection is approximately 3.5. By selecting energy deposits below 15 keV (14°) the PVR nearly triples, with further improvements observed for lower energy thresholds, reaching a value of 13.8 for a 5 keV energy cut
For smaller rod radii, the PVR remains below 3, although low-energy event selection still yields enhancements of approximately 60%, 30%, and 17% for radii of 1 mm, 0.75 mm, and 0.51 mm, respectively.
These findings support event-selection strategies for CC systems capable of triggering at low energies in the scatterer.
Restricting to the most informative events can reduce computational load and improve reconstruction performance, with the strongest benefits observed for rods with radii above 1 mm.

AcknowledgmentThis work was supported by the European Research Council Starting Grant #101041972 and by the ATRAE grant #ATR2023-143812 from the Spanish Research Agency.
References
[1] Kim, SM, Lee, JS 2024, ‘A comprehensive review on Compton camera image reconstruction: from principles to AI innovations’, Biomedical Engineering Letters, 14.6, 1175–1193, Seoul: Springer Nature.
 
[2] Delogu, P, et al. 2024, ‘Validation of Timepix4 energy calibration procedures with synchrotron X-ray beams’, Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1068, 169716, Amsterdam: Elsevier.
[3] Kolstein, M, De Lorenzo, G, Chmeissani, M 2014, ‘Evaluation of list-mode ordered subset expectation maximization image reconstruction for pixelated solid-state Compton gamma camera with large number of channels’, Journal of Instrumentation, 9.04, C04034–C04034, London: IOP Publishing.
Keywords: Compton camera, gamma-ray imaging, GATE simulation, small-angle scattering
17:30 Poster panel: 188

Poster Number:
MIC-10-188

Deep Learning-based 3D SPECT Reconstruction from Biplanar Projections and CT Scout Views (#1262)

J. Huang1, 2, Q. Yao1, 2, H. Han2, H. Xie2, G. Cao1, Y. Lv2

1 ShanghaiTech University, School of Biomedical Engineering,, Shanghai, China
2 Shanghai United Imaging Healthcare, Shanghai, China

Abstract

Dual-view SPECT planar imaging is a widely used method in clinical bone scan screening due to its efficiency and low radiation dose compared to whole-body SPECT/CT imaging. However, its diagnostic efficacy is often limited by anatomical overlap and the absence of depth information. Achieving high-quality 3D image from only two projections remains a challenge for traditional methods. Furthermore, existing deep learning-based approaches for ultra-sparse-view SPECT reconstruction are mainly designed for cardiac imaging and cannot be directly adapted to bone scintigraphy. To address these limitations, we propose a deep learning framework for 3D SPECT reconstruction from orthogonal biplanar projections and CT Scout views. The framework comprises a 3D condition map generation module and a structure-guided 2D conditional diffusion module, facilitating a two-stage reconstruction process including an initial coarse reconstruction followed by iterative refinement via a diffusion model. Quantitative results demonstrate that the proposed method achieves better recovery coefficients for bone tissues and metastatic lesions, as well as improved LPIPS scores for perceptual quality. Qualitative assessments further reveal that the reconstructed images exhibit highly realistic textures and shows strong agreement with standard full-view scans. This method provides a fast, low-radiation 3D reconstruction strategy with considerable potential to optimize clinical bone scan workflows, enhance diagnostic accuracy and improve patient comfort.

Keywords: SPECT, Dual-view 3D Reconstruction, CT scout view, Diffusion Model
17:30 Poster panel: 191

Poster Number:
MIC-10-191

The Inverse Problem of Spatiotemporal Emission Reconstruction from Prompt Gamma Timing Data (#1478)

J. Werner1, M. Schmidt1, J. Roser1, J. Kasprzak1, Y. Feng1, V. Ferrero2, F. Pennazio3, M. Rafecas1

1 University of Luebeck, Institute of Medical Engineering, Lübeck, Schleswig-Holstein, Germany
2 Universitá degli Studi di Torino, Department of Physics, Torino, Italy
3 Istituto Nazionale di Fisica Nucleare, Sezione di Torino, Torino, Italy

Abstract

Reconstructing the emission distribution of prompt gamma radiation in time and space offers a unique advantage: it reveals the primary particle motion inside the irradiation target during particle therapy. Similar to typical tomographic emission reconstruction scenarios, the measurement process based on Prompt Gamma Timing can be described by a linear system model and Poisson-distributed measurement data. However, the structure of the emission domain (emission time and position of the prompt gammas) and the detection domain (detector index and event time difference relative to a primary particle signal) are unique. In this work, we inspect some features of this inverse problem and analyze related fundamental limitations based on reconstructions with ideal data and the singular value decomposition of the system matrix. The results show a need for regularization and help identify possible improvements to the approach. Additionally, understanding this inverse problem in-depth also supports the development of other Prompt-Gamma-Timing-based techniques.

AcknowledgmentThis work was supported in part by the German Research Foundation (DFG) by Grant No. 51 658 7313 (PROSIT).
References
[1] Pennazio, F et al. 2022, ‘Proton therapy monitoring: spatiotemporal emission reconstruction with prompt gamma timing and implementation with PET detectors’, Phys. Med. Biol., 67, 065005
[2] Werner, J et al. 2024, ‘Stopping power and range estimations in proton therapy based on prompt gamma timing: motion models and automated parameter optimization’, Phys. Med. Biol.,69, 14NT02
[3] Werner, J et al. 2025, ‘Stopping power monitoring during proton therapy by means of prompt gamma timing: first experimental results with a homogeneous phantom’, arXiv, 2511.21344
[4] Ferrero, V et al. 2022, ‘Estimating the stopping power distribution during proton therapy: A proof of concept’, Front. Phys., 10:971767
[5] Werner, J et al. 2026, ‘Analytical model of prompt gamma timing for spatiotemporal emission reconstruction in particle therapy’, Phys. Med. Biol., 71, 075005
[6] Tong, S et al. 2011, ‘Properties and Mitigation of Edge Artifacts in PSF-Based PET Reconstruction’, IEEE Transactions on Nuclear Science, 85-5, 2264-2275
Keywords: Prompt Gamma Timing, Proton Therapy, Inverse Problems, Reconstruction, Singular Value Decomposition
17:30 Poster panel: 194

Poster Number:
MIC-10-194

Recent updates to STIR, a community effort of a quarter century (#1658)

N. Efthimiou1, D. Deidda2, V. Dao1, Z. Li1, C. Tsoumpas1, K. Thielemans3

1 UMCG, Department of Nuclear Medicine, Groningen, Netherlands
2 National Physical Laboratory, Teddighton, United Kingdom
3 UCL, Institute of Nuclear Medicine, London, United Kingdom

Abstract

The Software for Tomographic Image Reconstruction (STIR) is a long-standing open-source platform for PET and SPECT reconstruction that has supported tomographic imaging research for more than two decades. Public STIR documentation dates to the early 2000s, and the project remains actively developed, with a current public release in November 2025 and a GitHub repository reporting 9,409 commits, 146 stars, and 108 forks.   Recently there have been major contributions, that have expanded substantially the framework in scanner support, motion correction, and raw-data interoperability. Advances include support for the Siemens Biograph Vision Quadra long axial field-of-view PET/CT system, a data-driven rigid head motion-correction workflow for brain PET, and PETSIRD as a vendor-independent raw-data interface. Here, we demonstrate the impact of motion on clinically relevant quantitative measures. Together, these developments reinforce STIR’s position as a mature, community-driven platform for modern PET reconstruction research.

Acknowledgment

C. Tsoumpas, N. Efthimiou and Z. Li are supported by the NWO AES Vici Talent program (doi: 10.61686/REEAR02777). Z. Li was also supported by a Joint Scholarship from the China Scholarship Council and the University of Groningen (202210100011) and his placement at UCL by EPSRC CCP in Synergistic Reconstruction for Biomedical Imaging (EP/T026693/1). Viet Dao was supported by an E0SRC CASE PhD scholarship in partnership with Positrigo. This work used the Dutch national e-infrastructure with the support of the SURF Cooperative using grant no. EINF-18121.
STIR is partially maintained by CCP SyneRBI which is funded by the EPSRC grant EP/T026693/1 and by STFC through the UKRI Digital Research Infrastructure programme

References
[1] Thielemans, K, Tsoumpas, C, Mustafovic, S, Beisel, T, Aguiar, P, Dikaios, N & Jacobson, MW 2012, ‘STIR: software for tomographic image reconstruction release 2’, Physics in Medicine and Biology, vol. 57, no. 4, pp. 867–883.
[2] Brusaferri, L et al. 2020, ‘Joint Activity and Attenuation Reconstruction From Multiple Energy Window Data With Photopeak Scatter Re-Estimation in Non-TOF 3-D PET’, IEEE Transactions on Radiation and Plasma Medical Sciences, vol. 4, no. 4, pp. 410–421.
[3] Biguri, A et al. 2021, ‘Recent Progress in STIR 5.0’, 2021 IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), pp. 1–3
[4] Dao, V, Mikhaylova, E, Ahnen, ML, Fischer, J, Thielemans, K & Tsoumpas, C 2022, ‘Evaluation of STIR Library Adapted for PET Scanners with Non-Cylindrical Geometry’, Journal of Imaging, vol. 8, no. 6, article 172.
[5] Efthimiou, N, Emond, E, Wadhwa, P, Cawthorne, C, Tsoumpas, C & Thielemans, K 2019, ‘Implementation and validation of time-of-flight PET image reconstruction module for listmode and sinogram projection data in the STIR library’, Physics in Medicine and Biology, vol. 64, no. 3, article 035004.
[6] Prenosil, GA, Sari, H, Fürstner, M, Afshar-Oromieh, A, Shi, K, Rominger, A & Hentschel, M 2022, ‘Performance Characteristics of the Biograph Vision Quadra PET/CT System with a Long Axial Field of View Using the NEMA NU 2-2018 Standard’, Journal of Nuclear Medicine, vol. 63, no. 3, pp. 476–484.
[7] Efthimiou, N, Whitehead, AC, Stockhoff, M, Thyssen, C, Archibald, SJ & Vandenberghe, S 2019, ‘Preliminary investigation of the impact of Axial Ring Splitting on Image Quality for the Cost Reduction of Total-Body PET’, 2019 IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), Manchester, UK: IEEE. https://doi.org/10.1109/NSS/MIC42101.2019.9059650
[8] Dao, VA, Mikhaylova, E, Lesnic, E, Aykroyd, RG, Tsoumpas,C, ‘Data Driven Motion Detection, Estimation and Correction for Brain Pet Scanners' pp. 1–1, Dec. 2025, doi: 10.1109/NSS/MIC/RTSD57106.2025.11286795.
[9] Efthimiou, N 2020, ‘New Challenges for PET Image Reconstruction for Total-Body Imaging’, PET Clinics, vol. 15, no. 4, pp. 453–461.
[10] Thielemans, K., E. Asma, M. Cook, et al. 2025. ‘Final Draft of PETSIRD v1.0: A Proposed Standard for PET Raw Data’. 2025 IEEE Nuclear Science Symposium (NSS), Medical Imaging Conference (MIC) and Room Temperature Semiconductor Detector Conference (RTSD), November, 1–2. https://doi.org/10.1109/NSS/MIC/RTSD57106.2025.11287540
Keywords: long axial field-of-view PET, motion correction, PETSIRD, medical image reconstruction, open soure
17:30 Poster panel: 197

Poster Number:
MIC-10-197

Developments in fabrication and design flexibility of amorphous silicon microchannel plates (#2584)

G. Konstantinou1, L. Barraud1, C. Baillif1, N. Wyrsch1

1 EPFL, Neuchatel, Switzerland

Abstract

The technology of amorphous silicon-based micro-channel plates (AMCPs) uses a versatile approach to stack hydrogenated amorphous silicon in thicknesses up to 100 μm, using plasma-enhanced chemical vapor deposition  (PECVD) and etch microchannels of diameter below 2 μm every 5 μm, on a hexagonal pattern, by deep reactive ion etching. With gains up to 1000 for reverse bias less than 300 V, the design is possible to build on arrays of sensitive transimpedance amplifiers, integrated in read-out circuits with high channel density, such as for instance 15 by 15 μm. This ensures single photon detection with quantum efficiency up to 30% and noise rates below 50 Hz/cm², if combined with state of the art photocathodes such as the hi-QE from Photonis. Current fabrication runs focus on direct integration of the stack on ASICs, to be integrated with the photocathode over the next months. Measurements show timing response with jitter below 10 ps, making this high-precision photodetector ideal for radiation detectors combining ultra-fast photon emission and low light yield, such as Cherenkov.

AcknowledgmentThe project is funded by the SNF SPARK grant no CRSK-2_237272
Keywords: Photodetection, Fabrication, Amorphous Silicon, Microchannel plates
17:30 Poster panel: 200

Poster Number:
MIC-10-200

A Universal Geometric Calibration Method for CBCT Systems with Arbitrary Trajectories (#1877)

T. Jia1, C. Wang1, 2, B. Liu1, 2, Y. Wang1, 2, L. Wei1, 2, Z. Wang1, 2, C. Wei1, 2

1 Institute of High Energy Physics, Chinese Academy of Sciences, Beijing Engineering Research Center of Radiographic Techniques and Equipment, Beijing, China
2 University of Chinese Academy of Sciences, Nuclear Science and Technology, Beijing, China

Abstract

This paper presents a universal calibration method based on spatial vector decomposition to address the challenge of dynamically changing and difficult-to-calibrate geometric parameters in cone-beam computed tomography (CT) systems with complex scanning trajectories. The proposed method enables high-precision and robust calibration for a broad range of CT imaging systems. First, the projection matrix at each scanning angle is computed using a calibration phantom. This matrix is then decomposed into spatial vectors representing the system's geometric state, including the position of the X-ray source, the detector reference point, and orientation vectors. These vectors can be directly used for image reconstruction or converted into device-specific parameters. A vector-based detector mapping algorithm is also introduced to facilitate efficient projection data correction. Experiments on Micro-CT, CL and robotic-arm CT systems confirm that the method significantly improves image clarity, suppresses artifacts and distortions, and accurately reconstructs the actual motion trajectories of the X-ray source and detector. The method demonstrates high accuracy and versatility, providing a reliable calibration solution for high-degree-of-freedom CT systems.

References
[1] N. S. O’Brien, R. P. Boardman, I. Sinclair and T. Blumensath. Recent advances in X-ray cone-beam computed laminography[J]. Journal of X-ray Science and Technology, 2016, 24: 691-707.

[2] Grunz J P, Kunz A S, Gietzen C H, et al. 3D cone-beam CT of the ankle using a novel twin robotic X-ray system: assessment of image quality and radiation dose[J]. European Journal of Radiology, 2019, 119: 108659.

[3] F. Noo, R. Clackdoyle, C. Mennessier, T. A. White and T. J. Roney. Analytic method based on identification of ellipse parameters for scanner calibration in cone-beam tomography[J]. Physics in Medicine & Biology, 2000, 45: 3489.

[4] Y. Sun, Y. Hou, F. Zhao and J. Hu. A calibration method for misaligned scanner geometry in cone-beam computed tomography[J]. Ndt & E International, 2006, 39: 499-513.

Keywords: Computed tomography, Robotic arm, Computed laminography, Geometric calibration
17:30 Poster panel: 203

Poster Number:
MIC-10-203

CT-less Attenuation Maps as Priors for Robust Joint Activity–Attenuation Estimation in TOF and Non-TOF PET (#2048)

E. Emond1, Z. Chalampalakis2, A. Mehranian3, K.-H. Su5, S. Ahn6, F. Kotasidis4

1 GE HealthCare GmbH, Düsseldorf, North Rhine-Westphalia, Germany
2 GE HealthCare, Buc, France
3 GE HealthCare, Oxford, United Kingdom
4 GE HealthCare, Zurich, Switzerland
5 GE HealthCare, Waukesha, United States of America
6 GE HealthCare Technology and Innovation Center, Niskayuna, United States of America

Abstract

CT‑less attenuation correction in PET has gained increasing interest as a means to reduce patient radiation exposure, with recent AI‑based methods providing anatomically plausible attenuation estimates from emission data and complementary inputs. However, residual inaccuracies in such attenuation maps can still impact PET quantification.

Joint estimation of activity and attenuation provides a physics‑based framework to mitigate these effects, but remains less well constrained in the absence of time‑of‑flight (TOF) information.

This work evaluates the use of AI‑generated CT‑less attenuation maps as spatial priors for joint estimation of activity and attenuation in TOF and non‑TOF PET. Using a maximum‑a‑posteriori formulation, the CT‑less attenuation map is employed both as an initial estimate and as a regularising prior.

Experiments on paediatric PET datasets show that this approach stabilises joint reconstruction and recovers anatomically plausible attenuation maps, including in the non‑TOF case. These results support the use of CT‑less attenuation maps as effective priors for robust PET reconstruction.

Acknowledgment

The authors acknowledge the Cincinnati Children’s Hospital Medical Center, USA, for providing the data and Sangtae Ahn and Kuan-Hao Su (GE HealthCare) for their contributions to the methodological and code development of MAP-AA within the GE HealthCare PET reconstruction MATLAB package.

Keywords: Attenuation estimation, MLAA, PET reconstruction
17:30 Poster panel: 206

Poster Number:
MIC-10-206

Convergence-Informed AI-Assisted Reconstruction for Dual-Panel PET Imaging (#2078)

A. Saberi Manesh1, M. Amini1, Y. Salimi1, G. Hajianfar1, K. Doroud2, H. Arabi1, H. Zaidi1

1 University of Geneva, Department of Medical Imaging, Geneva University Hospital, Geneva, Genève, Switzerland
2 Picotech, St Genis Pouilly, France

Abstract

Dual-panel positron emission tomography (PET) scanners offer compact, cost-effective, and organ-dedicated imaging geometries, but their limited angular coverage introduces anisotropic resolution loss and elongation artifacts. This work presents a convergence-informed artificial intelligence (AI)-guided iterative reconstruction framework for the ProVision dual-panel PET system. The proposed approach exploits voxel-wise convergence information (VCI) extracted from the first 15 iterations of list-mode multi-ray maximum likelihood expectation maximization (MLEM) reconstruction. VCI was incorporated into five reconstruction strategies: analytic VCI-guided MLEM, multilayer perceptron-based VCI-guided MLEM, standalone 3D Swin-UNet reconstruction, AI-guided maximum a posteriori reconstruction (AI-MAP), and AI-initialized MLEM (AI-MLEM). Training and evaluation were performed using Monte Carlo simulations, a scaled NEMA image quality phantom, an external simulated patient test set, and an experimental anthropomorphic pelvis phantom. Compared with conventional MLEM, the proposed hybrid methods substantially improved contrast recovery, background uniformity, and geometric fidelity. AI-MLEM increased the contrast recovery coefficient from 53.16% to 82.35%, reduced background variability from 12.38% to 8.25%, and reduced the elongation ratio from 1.48 to 1.15. On the external simulated patient test set, AI-MLEM improved structural similarity from 0.446 to 0.816 and reduced normalized mean absolute error from 0.405 to 0.215. Experimental phantom results further confirmed improved lesion conspicuity and reduced vertical spreading. These results demonstrate that early-iteration convergence behavior provides useful voxel-wise guidance for mitigating limited-angle artifacts while preserving data consistency in dual-panel PET reconstruction.

AcknowledgmentThis work was supported by the Eurostars programme of the European Commission under grant E! 114021 ProVision and the Private Foundation of Geneva University Hospitals under Grant RC-06-01.
References
[1] A. Sanaat, M. Amini, H. Arabi, and H. Zaidi, "The quest for multifunctional and dedicated PET instrumentation with irregular geometries," Annals of Nuclear Medicine, vol. 38, no. 1, pp. 31–70, 2024/01/01 2024, doi: 10.1007/s12149-023-01881-6.
[2] H. Arabi, A. S. Manesh, and H. Zaidi, "Innovations in dedicated PET instrumentation: from the operating room to specimen imaging," Physics in Medicine & Biology, vol. 69, no. 11, p. 11TR03, 2024.
[3] M. Xia et al., "On Hallucinations in Artificial Intelligence-Generated Content for Nuclear Medicine Imaging (the DREAM Report)," (in eng), J Nucl Med, vol. 67, no. 2, pp. 166–174, Feb 2 2026, doi: 10.2967/jnumed.125.270653.
[4] A. S. Manesh et al., "Dedicated Image Reconstruction with Elongation Correction for a Dual-Panel Prostate PET Scanner," in 2024 IEEE Nuclear Science Symposium (NSS), Medical Imaging Conference (MIC) and Room Temperature Semiconductor Detector Conference (RTSD), 2024: IEEE, pp. 1–1.
[5] M. Amini et al., "Acquisition Protocol Optimization of a Dual-Panel PET Scanner Dedicated to Prostate Cancer Imaging," in 2025 IEEE Nuclear Science Symposium (NSS), Medical Imaging Conference (MIC) and Room Temperature Semiconductor Detector Conference (RTSD), 2025: IEEE, pp. 1–2.
[6] M. A. A. Saberi Manesh, H. Arabi, K. Doroud, C. Williams, H. Zaidi, "Conceptual design of a Dual-Panel Dedicated Prostate PET Scanner: A Monte Carlo Simulation Study," IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), 2022.
Keywords: Positron emission tomography, dual-panel, image reconstruction, deep learning, voxel-wise convergence information
17:30 Poster panel: 209

Poster Number:
MIC-10-209

Anatomically Guided Adaptive Gaussian Representation for PET Image Reconstruction (#2133)

Z. Li1, Z. Wang2, X. Zhao1, Y. Yuluo1, Y. Huang1, Z. Xie1, 2

1 Peking University, The Institute of Medical Technology, Beijing, China
2 Peking University, National Biomedical Imaging Center, Beijing, China

Abstract

We propose a continuous Gaussian representation for PET reconstruction, modeling the activity distribution with normalized Gaussian basis functions. This reformulates PET reconstruction as a compact differentiable optimization problem by coupling an analytical Gaussian forward model with Poisson likelihood. The key component is an anatomically guided adaptive density control strategy, which dynamically prunes, splits, and clones Gaussian primitives according to contribution, gradient, and MRI-derived mask constraints, thereby allocating representation capacity to informative regions while suppressing redundant components. Image-domain Huber-TV regularization is incorporated to stabilize reconstruction. Experiments on 2D simulations demonstrate competitive performance across count levels and clear advantages under low-count conditions, suggesting the potential of adaptive continuous representations for PET reconstruction.

AcknowledgmentThis work was supported by National Natural Science Foundation of China (62394311).
References
[1] Shepp, L.A. and Y. Vardi, Maximum Likelihood Reconstruction for Emission Tomography. IEEE Transactions on Medical Imaging, 1982. 1(2): p. 113-122.
[2] Kerbl, B., et al., 3d gaussian splatting for real-time radiance field rendering. ACM Trans. Graph., 2023. 42(4): p. 139:1-139:14.
[3] Zha, R., et al., R $^ 2$-Gaussian: Rectifying Radiative Gaussian Splatting for Tomographic Reconstruction. arXiv preprint arXiv:2405.20693, 2024.
[4] Kingma, D.P. and J. Ba, Adam: A Method for Stochastic Optimization. CoRR, 2014. abs/1412.6980.
Keywords: PET Reconstruction, Gaussian Representation, analytical forward model, adaptive density control
17:30 Poster panel: 212

Poster Number:
MIC-10-212

MRI guided PET reconstruction in meningioma [68Ga]DOTATOC patients, pituitary adenoma [18F]FET, and neurodegenerative [18F]FDG patients (#2361)

E. S. Christensen1, 2, A. Clemmensen1, 2, P. Lehmann3, J. Cabello4, O. M. Henriksen1, I. Law1, 5, A. Kjaer1, 2, T. L. Andersen1, 5

1 Copenhagen University Hospital, Department of Clinical Physiology and Nuclear Medicine, Copenhagen, Denmark
2 University of Copenhagen, Department of Biomedical Sciences, Cluster for Molecular Imaging, Copenhagen, Denmark
3 Siemens Healthcare A/S, Ballerup, Denmark
4 Siemens Medical Solutions USA, Knoxville, Tennessee, United States of America
5 University of Copenhagen, Department of Clinical Medicine, Copenhagen, Denmark

Abstract

The latest generation of simultaneous positron emission tomography and magnetic resonance imaging (PET/MRI) scanners is now at the forefront of hybrid imaging. As of now, the reconstruction of both modalities is independent, but research is ongoing in MRI-guided (MRIg) PET reconstruction, where information from the MRI is used as prior information in the iterative PET reconstruction.
We present three different clinical cohorts: meningioma patients (n=8) scanned with [68Ga]DOTATOC, a single pituitary adenoma scanned dynamically with [18F]FET and a neurodegenerative cohort (n=10) scanned with [18F]FDG. The cohorts have been analyzed with standard OSEM reconstruction (b=0) versus MRIg reconstruction. The meningioma cohort showed an increased volume with standardized uptake values (SUV) above 2.5 when using MRIg reconstruction, and an increased SUVmax in the tumor. The pituitary adenoma showed increase time-activity curve both for β=1 and β=2. The neurodegenerative cohort showed stable SUV ratios of the cerebral-to-cerebellar cortex down to 3 minutes.

This preliminary work shows the clinical value in utilizing the information from the MRI in PET reconstruction with applications both to increase image quality and resolution, but also to facilitate faster imaging.

References
[1] Henriksen OM, Korsholm K, Loft A, et al. First Clinical Experiences with the Ultra-Fast Time-of-Flight BIOGRAPH One Next-Generation Hybrid PET/MRI System. Diagnostics. 2026;16:398.
[2] Afshar-Oromieh A, Wolf MB, Kratochwil C, et al. Comparison of 68Ga-DOTATOC-PET/CT and PET/MRI hybrid systems in patients with cranial meningioma: Initial results. Neuro-Oncology. 2015;17:312-319.
[3] Pruis IJ, Verburg FA, Balvers RK, et al. [18F] FET PET/MRI: an accurate technique for detection of small functional pituitary tumors. Journal of Nuclear Medicine. 2024;65:688-692.
[4] Acquah IK. Comparison of Methods of µ-Map Generation: MR-Based Method in PET/MR Imaging Versus Pseudo-CT Method in Radiotherapy Dose Planning, University of Ghana; 2019.
[5] Kazantsev D, Bousse A, Pedemonte S, Arridge SR, Hutton BF, Ourselin S. Edge preserving bowsher prior with nonlocal weighting for 3D spect reconstruction. Paper presented at: 2011 IEEE International Symposium on Biomedical Imaging: From Nano to Macro, 2011.
[6] Bowsher JE, Yuan H, Hedlund LW, et al. Utilizing MRI information to estimate F18-FDG distributions in rat flank tumors. Paper presented at: IEEE Symposium Conference Record Nuclear Science 2004., 2004.
[7] Na S, Kang DW, Kim GH, et al. The usefulness of 18F-FDG PET to differentiate subtypes of dementia: The systematic review and meta-analysis. Dementia and Neurocognitive Disorders. 2024;23:54.
Keywords: Positron Emission Tomography, Meningioma, DOTATOC, FET, FDG
17:30 Poster panel: 215

Poster Number:
MIC-10-215

A Hybrid Neural Network Model for DaTscan Diagnosis of Neurodegenerative Disorders (#2498)

E. Roupas1, A. Koukopoulou1, E. Mitsou1, E. Stiliaris2

1 National and Kapodistrian University of Athens, Medical School, Athens, Greece
2 National and Kapodistrian University of Athens, Department of Physics, Athens, Greece

Abstract

This study presents the development of a hybrid Neural Network designed to simultaneously reconstruct medical images from their sinograms and to predict diagnostic parameters for neurodegenerative disorders, specifically Parkinson’s disease. The training pipeline utilizes simulated Dopamine Transporter Scan (DaTscan) brain images and their corresponding sinograms, generated via a custom Monte Carlo based framework that employs analytically driven functions to model realistic hotspot and background noise characteristics. The architecture integrates image reconstruction and disease-specific parameter estimation into a unified framework. Key diagnostic features include the bilateral asymmetry of radiotracer uptake within the caudate nuclei; deviations in spatial distribution or intensity within predefined Regions of Interest (RoIs) serve as primary indicators of neuropathology. This dual-function approach advances automated, AI-assisted functional neuroimaging, offering a robust pathway for early, non-invasive diagnosis of neurodegenerative conditions.

References
[1] Duan T, Chen W, Ruan M, Zhang X, Shen S and Gu W, 2025, 'Unsupervised deep learning-based medical image registration: a survey', Physics in Medicine & Biology 70, 02TR01
[2] Paschalis P, Giokaris ND, Karabarbounis A, Loudos GK, Maintas D, Papanicolas CN, Spanoudaki V, Tsoumpas Ch and Stiliaris E, 2004, 'Tomographic image reconstruction using Artificial Neural Networks', Nuclear Instruments & Methods A 527, 151-156
[3] Argyrou M, Maintas D, Tsoumpas Ch and Stiliaris E, 2012, 'Tomographic Image Reconstruction based on Artificial Neural Network (ANN) Techniques', IEEE MIC Record M17-12, 3324-3327
[4] Hwang D and Zeng GL, 2006, 'Convergence study of an accelerated ML-EM algorithm using bigger step size', Physics in Medicine & Biology 51, 237
[5] Wang Z, Bovik AC, Sheikh HR, Simoncelli EP, 2004, 'Image quality assessment: from error visibility to structural similarity', IEEE Transactions on Image Processing 13, 600–612
Keywords: Dopamine Transporter Scan (DaTscan), Machine Learning, CNN, Tomographic Reconstruction, DaTscan Classification
17:30 Poster panel: 218

Poster Number:
MIC-10-218

A Global-Local Latent Diffusion Model for Limited-Angle CT Reconstruction (#2642)

Y. Wang1, 2, C. Wei1, 2, Q. Xu1, 3

1 Institute of High Energy Physics, Chinese Academy of Sciences, Beijing, China
2 University of Chinese Academy of Sciences, School of Nuclear Science and Technology, Beijing, China
3 Jinan Laboratory of Applied Nuclear Science, Jinan, China

Abstract

Limited-angle computed tomography (LACT) reconstruction is highly ill-posed because missing projection views lead to severe artifacts and incomplete boundaries. This work proposes a global-local latent diffusion framework for limited-angle CT reconstruction. The proposed method performs diffusion reconstruction in the latent space of a variational autoencoder and jointly incorporates U-Net-based local modeling and transformer-based global modeling into the denoising process. The U-Net component captures multiscale spatial details and boundary continuity, whereas the transformer component models long-range dependencies and global structural relationships associated with missing-angle information. These complementary priors are adaptively integrated to guide latent-space denoising. During sampling, a physics-guided data-consistency constraint is further introduced. Specifically, the intermediate latent prediction produced by the model is decoded back to the image domain, processed with a projection-consistency operation, and then re-encoded into the latent diffusion trajectory. Preliminary experiments on a simulated data set show favorable reconstruction quality, indicating that the proposed global-local diffusion framework can improve structural fidelity and projection consistency for limited-angle CT reconstruction.

AcknowledgmentNational Natural Science Foundation of China (NSFC) (No.12475313) and Excellent Youth Fund of Shandong Natural Science Foundation (ZR2024YQ066)
References
[1] R. Rombach, A. Blattmann, D. Lorenz, P. Esser, and B. Ommer, “High-Resolution Image Synthesis with Latent Diffusion Models,” Proceedings of the IEEE Computer Society Conference on Computer Vision and Pattern Recognition, vol. 2022-June, pp. 10674–10685, Dec. 2021.
[2] W. Peebles and S. Xie, “Scalable Diffusion Models with Transformers,” Proceedings of the IEEE International Conference on Computer Vision, pp. 4172–4182, Dec. 2022.
Keywords: Computed tomography, latent diffusion, limited-angle reconstruction, projection consistency, transformer
17:30 Poster panel: 221

Poster Number:
MIC-10-221

Iterative proton CT reconstruction from list-mode data of a sandwich time-of-flight scanner (#2821)

A. Coussat1, J.M. Létang1, S. Rit1

1 INSA-Lyon, Université Claude Bernard Lyon 1, CNRS, Inserm, CREATIS, UMR 5220, U1294, F-69373, Lyon, France

Abstract

Proton CT is a promising alternative to X-ray CT for proton therapy treatment planning, allowing direct estimation of the relative stopping power map within the patient body without relying on conversion from Hounsfield units. Conventional list-mode proton CT scanners measure the energy loss of each individual proton to estimate the integral of the relative stopping power along its path, the water-equivalent path length, but do not meet the requirements for clinical use due to their low acquisition rates. An alternative proton CT scanner design was recently proposed where the time-of-flight of each proton is measured between two detectors located before and after the patient along the proton beam. The main advantages of this sandwich time-of-flight design are its compactness and high acquisition rates. However, conversion from sandwich time-of-flight data to water-equivalent path length is not possible. In this work, we propose an iterative reconstruction of the sought relative stopping power map based on the non-linear forward problem. The method is assessed using Monte Carlo simulations of the acquisition of four simple phantoms. Although computationally expensive, this method accurately recovers the relative stopping power map.

AcknowledgmentWith financial support from IT Cancer of Inserm within the framework of the 2021-2030 Cancer Control Strategy, on funds administered by Inserm, within the framework of the SIRIC LYriCAN+ (INCa-DGOS-INSERM-ITMO cancer 18003) and the LABEX PRIMES (ANR-11-LABX-0063) of Université de Lyon, within the program ”Investissements d’Avenir” (ANR-11-IDEX-0007) operated by the ANR.
References
[1] A. Coussat, N. Krah, J. M. Létang, and S. Rit. “Proton computed tomography reconstruction from sandwich time-of-flight using a lookup table approach”. In: Fully3D. Shanghai, China, May 2025, pp. 646–649.
[2] R. P. Johnson. “Meeting the detector challenges for pre-clinical proton and ion computed tomography”. In: Phys Med Biol 69.11 (May 2024), 11TR02.
[3] F. Ulrich-Pur, T. Bergauer, A. Hirtl, C. Irmler, et al. “Novel ion imaging concept based on time-of-flight measurements with low gain avalanche detectors”. In: Journal of Instrumentation 18.02 (2023), p. C02062.
Keywords: proton therapy, proton computed tomography, iterative reconstruction, automatic differentiation
17:30 Poster panel: 224

Poster Number:
MIC-10-224

Direct normalization correction for a brain PET system with multi-resolution detector design (#2848)

C. Yang1, H. Ye1, C. Huang2, Z. Wu2, H. Liu3, Q. Wei4, Z. Gu2, J. Wu1

1 Beijing Normal University, Center for Advanced Quantum Studies, School of Physics and Astronomy, Beijing, China
2 Shenzhen Bay Laboratory, Institute of Biomedical Engineering, Shenzhen, China
3 Tsinghua University, Department of Engineering Physics, Beijing, China
4 University of Science and Technology Beijing, School of Automation and Electrical Engineering, Beijing, Germany

Abstract

A high-sensitivity, high-resolution brain PET system with a multi-resolution detector design has been developed. However, image reconstruction without normalization correction produces pronounced image artifacts. In this study, we implemented list-mode reconstruction with direct normalization correction using normalization factors derived from simulation data for this multi-resolution PET system. The effect of normalization correction on reconstructed image quality was evaluated using Monte Carlo simulations and real scan data from a modified image-quality phantom. The results showed that normalization-corrected reconstruction substantially reduced image artifacts and improved image uniformity in simulated data, while residual nonuniformity remained in real scan data due to the lack of experimentally measured crystal efficiencies. Overall, this work demonstrates the feasibility of normalization correction for the dual-layer brain PET system and provides a basis for further experimental normalization using measured crystal efficiencies

AcknowledgmentThis study was supported by National Key Research and Development (R&D) Program of China (No.2023YFC2414800), National Natural Science Foundation of China (No. 12305379, No.12205166), and CIRP Open Fund of Radiation Protection Laboratories (TNMPP-2024-No.1).
References
[1] Z. Chen et al., "Initial System Performance Evaluation of N-PET, a High-Sensitivity and High-Resolution TOF-DOI PET Prototype Dedicated for Human Brain Imaging," 2025 IEEE Nuclear Science Symposium (NSS), Medical Imaging Conference (MIC) and Room Temperature Semiconductor Detector Conference (RTSD), Yokohama, Japan, 2025, pp. 1-2
Keywords: brain PET, list-mode reconstruction, normalization correction, multi-resolution detector design
17:30 Poster panel: 227

Poster Number:
MIC-10-227

Effect of Cardiac Reorientation on Deep Learning-Based Denoising for Low-Dose Cardiac SPECT (#2932)

M. Toumi1, Y. Yang1, M. King2, J. G. Brankov1

1 Illinois Institute of Technology, Electrical and Computer Engineering, Chicago, Illinois, United States of America
2 University of Massachusetts Chan Medical School, Department of Radiology, Worcester, Massachusetts, United States of America

Abstract

In low-dose (LD) myocardial perfusion imaging (MPI) with single-photon emission computed tomography (SPECT), radiation exposure is reduced at the expense of image quality, potentially impairing perfusion defect detection. Deep learning methods have shown promise in reducing noise in LD MPI images. However, clinical data exhibit varying cardiac orientations, increasing the complexity of the learning task for  denoising models. We hypothesize that standardizing cardiac orientation with respect to a common reference prior to training can reduce this complexity and improve denoising performance. We evaluated this approach with a previously proposed edge-aware convolutional autoencoder (EdgeCAE) across four levels of trade-off between edge preservation and mean-squared error (MSE) fidelity. The model was trained on 738 clinical subjects and tested on 300 noise realizations derived from two independent patient anatomies, with and without pre-reorientation. Our results demonstrated that cardiac reorientation consistently reduced the MSE compared to the full-dose reference images. This was also supported by visual assessment on an independent patient cohort.

AcknowledgmentResearch reported in this publication was supported by the National Institute of Biomedical Imaging and Bioengineering(NIBIB) and the National Heart, Lung, and Blood Institute (NHLBI) of the National Institutes of Health under Award Numbers R01EB036082 and R01HL154687. The content is solely the responsibility of the authors and does not necessarily represent the official views of the National Institutes of Health.
References
[1] M. D. Cerqueira et al., “Standardized myocardial segmentation and nomenclature for tomographic imaging of the heart,” Circulation, vol. 105, pp. 539–542, 2002.
[2] P. J. Slomka et al., “Cardiac imaging with myocardial perfusion SPECT,” J. Nucl. Med., vol. 60, pp. 1100–1108, 2019.
[3] I. Shiri et al., “Standard SPECT myocardial perfusion estimation from half-time acquisitions using deep convolutional residual neural networks,” J. Nucl. Cardiol., vol. 28, pp. 2761–2779, 202
[4] A. J. Ramon et al., “Initial investigation of low-dose SPECT-MPI via deep learning,” IEEE Trans. Radiat. Plasma Med. Sci., vol. 4, pp. 65–72, 2020.
[5] M. Toumi et al., “Edge-aware convolutional autoencoder for low-dose SPECT myocardial perfusion image enhancement,” IEEE NSS-MIC Conf. Rec., 2024.
[6] E. V. Garcia et al., “Quantification of rotational thallium-201 myocardial tomography,” J. Nucl. Med., vol. 26, pp. 17–26, 1985.
[7] H. H. Barrett and K. J. Myers, Foundations of Image Science. Hoboken, NJ: Wiley, 2004.
[8] T. Marin, J. G. Brankov, “Deformable left-ventricle mesh model for motion-compensated filtering in cardiac gated SPECT,” Med Phys. vol. 37, no. 10, pp. 5471-81, 2010/ doi: 10.1118/3483098. PMID: 21089783; PMCID: PMC2962663.
[9] B. Feng, “Modeling of the left ventricle (LV) by using mechanical models and image data,” Ph.D. thesis, The University of Utah, 2002.
Keywords: Deep Learning, myocardial perfusion imaging, Deep Learning-Based Denoising, Cardiac SPECT
17:30 Poster panel: 230

Poster Number:
MIC-10-230

Ghost Artifact Suppression in Multiplexed Self-Collimating SPECT via Total Variation Regularization (#2996)

S. Mehta1, H. Tripathi1, T. Methuku1, T. Skirsky2, L. Ying4, T. Ma3, R. Yao2

1 The State University of New York at Buffalo, Department of Computer Science and Engineering, Buffalo, New York, United States of America
2 The State University of New York at Buffalo, Department of Radiology, Buffalo, New York, United States of America
3 Tsinghua University, Department of Engineering Physics, Beijing, China
4 The State University of New York at Buffalo, Biomedical Department, Buffalo, New York, United States of America

Abstract

Multiplexing in SPECT renders the measurement operator underdetermined. In symmetric multi-source configurations, iterative ambiguity reinforcement sustains persistent ghost artifacts. Established mitigation strategies require supplemental single-beam nonmultiplexed projections, constraining collimator design and sacrificing sensitivity. Sparsity in gradients for common tracer distributions motivate total-variation (TV) regularization as a natural demultiplexing strategy: ghost artifacts produce extra gradient support that TV prior penalizes, breaking the ambiguity cycle. We demonstrate that MAP-TV (maximum a posteriori total variation) acts as an active reconstruction-stage demultiplexer, a supplemental approach to modifying the projection set in the data acquisition stage.

Acknowledgment

This work is supported by the National Institute of Biomedical Imaging and Bioengineering of the National Institutes of Health, USA, under Grant R21EB032993. Computation work has been carried out on the high-performance computing platform supported by the Center for Computational Research at the University at Buffalo.

References
[1] Chambolle, A, Pock, T 2011, 'A first-order primal-dual algorithm for convex problems with applications to imaging', Journal of Mathematical Imaging and Vision, vol. 40, no. 1, pp. 120–145, Springer: New York
[2] Panin, VY, Zeng, GL, Gullberg, GT 1999, 'Total variation regulated EM algorithm [SPECT reconstruction]', IEEE Transactions on Nuclear Science,  vol. 46, no. 6, pp. 2202–2210, IEEE: Piscataway.
[3] Mok, GSP, Tsui, BMW, Beekman, FJ 2011, 'The effects of object activity distribution on multiplexing multi-pinhole SPECT', Physics in Medicine and Biology, vol. 56, no. 8, pp. 2635–2650, IOP Publishing: Bristol.
[4] Pells, S, Shestakova, I, Khalil, M, Hugg, J, Majewski, S, Bhatt, P, Bhatt, D 2024, 'Correction of multiplexing artefacts in multi-pinhole SPECT through temporal shuttering, de-multiplexing of projections, and alternating reconstruction', Physics in Medicine and Biology, vol. 69, no. 12, art. no. 125004, IOP Publishing: Bristol
Keywords: Multiplexed SPECT, Ghost artifact suppression, Total Variation regularization, MAP reconstruction, Multi-pinhole collimator
17:30 Poster panel: 233

Poster Number:
MIC-10-233

Integrating In-Patient Scattered Events in PET Reconstruction via a Monte Carlo Projector (#3119)

M. Moulin1, 3, G. Montémont1, H. De Las Heras Gala1, J.M. Létang3

1 CEA-LETI, Université Grenoble-Alpes, Grenoble, France
2 INSA-Lyon, Université Claude Bernard Lyon 1, UJM- Saint Etienne, CNRS, Inserm, Lyon, France

Abstract

In this paper, we propose a new projection method for PET, called Monte Carlo Point Projectors (MCPP), which back-projects measured gamma interactions in the detector by sampling potential positions of the event source within the field of view (FoV). We apply this method to scattered events using a projector for scattered lines of response (ScLoR), as well as a combination of scattered lines of response and Compton cones of response (CoR). The simulated system is a cylinder with a radius of 15~cm and a length of 16~cm, composed of 12 Cadmium Zinc Telluride panels, with a 7~cm radius water sphere at the center to simulate in-patient scatter. We show that for this system, the point spread function (PSF) resolutions were 8.3~cm for CoR and 5.3--4.7~cm for the proposed ScLoR projector and the combination with the CoR, demonstrating the advantage of using scattered events with the MCPP method. We also studied how the missing energy of scattered events influences the PSFs obtained with MCPP, noting that lower missing energy yields better resolution. Its combination with a scatter classification method on experimental data is being evaluated. In conclusion, the MCPP approach facilitates the integration of different projectors to process a maximum of detected events.

References
[1] Efthimiou, N., Karp, J.S., Surti, S. 2022, 'Data-driven,energy-based method for estimation of scattered events in positron emission tomography’, Physics in Medicine & Biology, vol 67, p. 095010
[2] Moulin, M, Montémont, G, Létang, J 2025, ‘Efficiency of an Event Neural Network Classifier for Positron Emission Tomography’, 2025 IEEE Nuclear Science Symposium (NSS), Medical Imaging Conference (MIC) and Room Temperature Semiconductor Detector Conference (RTSD),(Yokohama Japan),p. 1
[3] Chinn, G, Foudray, MK, Levin, CS, 2006, ‘Accurately Positioning and Incorporating Tissue-Scattered Photons into PET Image Reconstruction’, 2006 IEEE Nuclear Science Symposium Conference Record, vol 3, pp. 1746-1751
Keywords: PET, Back-projection, Scattering, Monte-Carlo
17:30 Poster panel: 236

Poster Number:
MIC-10-236

List-Mode Subspace-Based Dynamic PET: 1-Second Temporal Resolution on the NeuroEXPLORER (#3210)

Y. Djebra1, Y. Najmaoui1, D. Szezurek1, Y. Chemli1, K. Fontaine1, J.-D. Gallezot1, T. Toyonaga1, M. Normandin1, G. El Fakhri1, C. Ma1, T. Marin1

1 Yale University, Yale Biomedical Imaging, New Haven, Connecticut, United States of America

Abstract

Dynamic PET imaging is becoming increasingly valuable in neuroimaging and cardiac imaging, thanks to its ability to track functional changes over time.
Conventional approaches to dynamic PET imaging need to find a balance between frame duration and noise: shorter temporal frames can provide important information on rapid physiological processes, but at the expense of increased noise.  Traditional filtering strategies aim at reducing noise, but inherently cause loss of spatial and/or temporal resolution.
In this work, we represent dynamic PET images using a subspace model during image reconstruction to substantially reduce noise while preserving spatial and temporal information, thereby exploiting the spatiotemporal correlations in dynamic images.  We demonstrate the power of the proposed approach on an in-vivo scan of a nonhuman primate performed on the high-sensitivity, high-resolution NeuroEXPLORER scanner.  Dynamic images reconstructed at 1-second frames exhibit high resolution and dramatically reduced noise compared to conventional reconstructions, enabling the visualization of early tracer perfusion, respiratory motion and contrast changes in the whole body, without any gating.  The proposed method shows promise for short-frame imaging of the brain and body.

AcknowledgmentThis work was supported in part by the National Institutes of Health under the grants R01EB035093, R01AG076153, and P41EB022544.
References
[1] H. Li et al., “Performance Characteristics of the NeuroEXPLORER, a Next-Generation Human Brain PET/CT Imager,” Journal of Nuclear Medicine, 2024.
[2] Y. Djebra et al., “LR-PET: A Subspace-Based Dynamic PET Imaging via Explicit Non-Negative Low-Rank Factorization,” in 2025 IEEE NSS/MIC/RTSD, 2025.
[3] J. Qi et al., “Iterative reconstruction techniques in emission computed tomography,” Physics in Medicine and Biology, 2006.
[4] K. Lange et al., “EM reconstruction algorithms for emission and transmission tomography,” Journal of Computer Assisted Tomography, 1984.
[5] H. M. Hudson et al., “Accelerated image reconstruction using ordered subsets of projection data,” IEEE Transactions on Medical Imaging, 1994.
Keywords: Dynamic PET imaging, Subspace, Non-Negative Matrix Factorization, Optimization, Respiratory-Motion Resolved
17:30 Poster panel: 239

Poster Number:
MIC-10-239

Hierarchical ROI-free Learning for GFR Category Stratification in Dynamic 99mTc-DTPA Renography (#1527)

Z. Xu1, J. Li2, C. Li1, J. Zhong2, W. Sun2, J. Chen3, J. Dong1, Z. Sun1, Y. Zeng4, H. Xiao2, J. Sun1

1 Hainan University, Department of Data Science and Big Data, Haikou, China
2 The First Affiliated Hospital of Hainan Medical University, Department of Nuclear Medicine, Haikou, China
3 Hainan General Hospital, Haikou, China
4 The Second Affiliated Hospital of Hainan Medical University, Department of Radiation Oncology, Cancer Treatment Center, Haikou, China

Abstract

Dynamic 99mTc-diethylenetriaminepentaacetic acid (DTPA) planar renography is widely used for quantitative renal function assessment, but Gates-based glomerular filtration rate (GFR) estimation depends on manually defined kidney and background regions of interest (ROIs), which is time-consuming and operator-dependent. We developed RFVGG, an ROI-free hierarchical deep learning framework that learns directly from dynamic renography image sequences using routinely available Gates-derived labels. This retrospective study included 728 consecutive patients who underwent dynamic 99mTc-DTPA planar renography. RFVGG was built on a shared 3D VGG-style backbone with hierarchical task-specific heads for three-, five-, and six-category GFR stratification. Model development followed five-fold cross-validation, and RFVGG was compared with CNN, ResNet-18, and DenseNet under identical evaluation settings. Classification performance was evaluated using accuracy, precision, recall, and specificity. RFVGG achieved accuracies of 0.842±0.020, 0.705±0.043, and 0.606±0.042 for the three-, five-, and six-category tasks, respectively. The corresponding precisions were 0.846±0.017, 0.668±0.037, and 0.614±0.034; recalls were 0.843±0.024, 0.689±0.045, and 0.642±0.039; and specificities were 0.915±0.012, 0.920±0.012, and 0.920±0.008, respectively. These results indicate that RFVGG can perform automated ROI-free GFR category stratification from dynamic 99mTc-DTPA planar renography without contour-based kidney or background ROI tracing

AcknowledgmentThis work was supported in part by the National Natural Science Foundation of China (12505396) and the Science and Technology Talents Innovation Project of Hainan Province (KJRC2023D30). Corresponding authors: Jingzhang Sun and Huan Xiao.
References
[1] A. Levin et al., "Kidney Disease: Improving Global Outcomes (KDIGO) CKD Work Group. KDIGO 2012 clinical practice guideline for the evaluation and management of chronic kidney disease," Kidney international supplements, vol. 3, no. 1, pp. 1-150, 2013.
[2] Y. Pi et al., "Deep regression using 99mTc-DTPA dynamic renal imaging for automatic calculation of the glomerular filtration rate," European Radiology, vol. 33, no. 1, pp. 34-42, 2023.
Keywords: dynamic renography, 99mTc-DTPA, glomerular filtration rate, GFR stratification, ROI-free deep learning
17:30 Poster panel: 242

Poster Number:
MIC-10-242

Design and Construction of Physical and Digital Phantoms for Image-Derived Input Function Validation Studies in Brain PET Imaging (#1866)

U. Khalid1, 2, C. Memis1, L. Tellmann1, M. Lennartz3, T. Meurer3, P. Lohmann1, 4, H. Iida5, M. Wiesmann6, H. Herzog1, J. Scheins1, N. J. Shah1, 7, C. W. Lerche1, C. R. Brambilla1

1 Forschungszentrum Jülich GmbH, Institute of Neuroscience and Medicine, INM-4, Jülich, North Rhine-Westphalia, Germany
2 RWTH Aachen University, Faculty of Medicine, Aachen, North Rhine-Westphalia, Germany
3 Forschungszentrum Jülich GmbH, Institute of Energy Technologies (IET), Jülich, North Rhine-Westphalia, Germany
4 RWTH Aachen, Department of Nuclear Medicine, Aachen, North Rhine-Westphalia, Germany
5 University of Turku, Turku PET center, Institute of Biomedicine, Faculty of Medicine, Turku, Finland
6 Uniklinik Aachen, Clinic for Diagnostic and Interventional Neuroradiology, Aachen, North Rhine-Westphalia, Germany
7 RWTH Aachen, Department of Neurology, Aachen, North Rhine-Westphalia, Germany

Abstract

Quantitative brain PET studies often require arterial input functions (AIF), which can be derived from radiotracer concentrations through the internal carotid arteries (ICAs) in dynamic PET images. Image-derived input function (IDIF) methods avoid invasiveness; however, challenges in precise quantification of dynamic flow conditions and correction of partial volume effects (PVE) persist. Recently developed dedicated PET systems for brain imaging, such as the RETIMAGER System, offer excellent spatial and temporal resolution, which is of high relevance for IDIF determination. In addition, hybrid systems offer complementary information for improved IDIF validation. This study shows the design and development of digital and physical phantoms for future validation of IDIF methods in Brain PET systems. A two-phase approach employs the design and construction of a cylindrical phantom with an ICA insert matching real structural geometries. Phase 1 uses a digitalized phantom with Monte Carlo (MC) simulations to validate and optimize acquisition and reconstruction parameters. Phase 2 is in progress for experimental versus simulation validations using a physical phantom with an ICA insert to simulate flow dynamics using an injector. Additionally, a 3D head phantom incorporating gray matter (GM), white matter (WM), scalp, and ICA structures has been designed and is in fabrication for future comprehensive validation.

AcknowledgmentThe authors would like to acknowledge their gratitude to Claire Rick for English proofreading. This work was supported by the Horizon EIC 2022 Pathfinder Open– RETIMAGER, Project ID:101099096.
References
[1] Y. F. Tai et al, J. Neurol. Neurosurg. Psych, vol. 75, no. 5, pp. 669–76, 2004.
[2] H. Zaidi et al., Med. Phys., vol. 38, no. 10, pp. 5667–89, 2011.
[3] P. Ziegler-Fregonara et al., J. Cereb. Blood Flow Metab., vol. 31, no. 10, pp. 1986–98, 2011.
[4] P. Ziegler-Fregonara et al., J. Cereb. Blood Flow Metab., vol. 29, no. 11, pp. 1825–35, 2009.
[5] C. R. Brambilla et al., Front. Neurosci., vol. 19, p. 1725728, 2025.
[6] M. L. Thorne et al., Ultrasound Med. Biol., vol. 35, no. 1, pp. 120–28, 2009.
[7] H. Iida et al., Ann. Nucl. Med., vol. 27, no. 1, pp. 25–36, 2013.
[8] A. Haddar et al., “RETIMAGER”, in Proc. IEEE, pp. 1–1, 2025.
Keywords: Brain PET, Dynamic PET, Image-Derived Input Function, PET quantification
17:30 Poster panel: 245

Poster Number:
MIC-10-245

TRACE: A New Phantom for Simultaneous Multi-Method Assessment of PET Spatial Resolution (#2135)

D. Choudhary1, 2, B. Mehadji1, R. D. Badawi1, N. Omidvari1, B. A. Spencer1

1 UC Davis Health, Radiology, Sacramento, California, United States of America
2 UC Davis, Biomedical Engineering, Davis, California, United States of America

Abstract

Spatial resolution assessment in positron emission tomography (PET) typically relies on the full-width at half-maximum (FWHM) of point-like line sources reconstructed with filtered back-projection (FBP), yet modern clinical imaging predominantly uses iterative reconstruction methods, motivating the need for phantoms that evaluate resolution under clinically relevant protocols. To address this, we present TRACE (Tapered Rod Array for Continuous-resolution Evaluation), a 3D-printed phantom enabling simultaneous spatial resolution assessment via four complementary methods – FBP FWHM, modulation transfer function (MTF), Rayleigh criterion, and visual assessment – in a single acquisition. TRACE consists of a tapered cone with continuously decreasing fillable conical rod diameters (4.8–1.2 mm over 10 cm) and three integrated capillary slots for NEMA-compatible line source measurements. The phantom was scanned on the uEXPLORER total-body PET scanner in transverse (XY) and axial (XZ) orientations. FBP FWHM yielded mean transverse resolutions of 3.0 ± 0.2 mm (radial) and 3.4 ± 0.2 mm (tangential), and 2.9 ± 0.5 mm axially. Under OSEM reconstruction, the Rayleigh criterion and visual assessment resolved rods down to 1.8 mm (XY) and 2.0 mm (XZ), while the 50% and 10% MTF were achieved at 2.8 mm and 1.6 mm (XY), and 2.6 mm and 1.7 mm (XZ), respectively. TRACE bridges standardized and clinically relevant resolution assessment, delivering all four metrics in a single acquisition and offering a comprehensive framework for PET spatial resolution characterization

References
[1] NU 2-2018. National Electrical Manufacturers Association; 2018
[2] S. E. Derenzo, T. F. Budinger, J. L. Cahoon, R. H. Huesman and H. G. Jackson, "High Resolution Computed Tomography of Positron Emitters," in IEEE Transactions on Nuclear Science, vol. 24, no. 1, pp. 544-558, Feb. 1977, doi: 10.1109/TNS.1977.4328738. 
[3] Spencer, Benjamin A., et al. "Performance evaluation of the uEXPLORER total-body PET/CT scanner based on NEMA NU 2-2018 with additional tests to characterize PET scanners with a long axial field of view." Journal of nuclear medicine 62.6 (2021): 861-870.
[4] Toussaint M, Lecomte R, Dussault JP. Improvement of Spatial Resolution with Iterative PET Reconstruction using UltraFast TOF. IEEE Trans Radiat Plasma Med Sci. 2021 Sep;5(5):729-737. doi: 10.1109/trpms.2020.3033561. Epub 2020 Oct 26. PMID: 35059544; PMCID: PMC8765719.
Keywords: Spatial Resolution, Phantoms, NEMA standard, PET Characterization, EXPLORER
17:30 Poster panel: 248

Poster Number:
MIC-10-248

A Reconstruction-Agnostic Paper Made Fan Phantom for Task-Relevant PET Spatial Resolution Assessment (#2307)

B. Mehadji1, R. Badawi1, 2, L. Nardo1, B. A. Spencer1

1 UC Davis, Radiology, Sacramento, California, United States of America
2 UC Davis, Biomedical Engineering, Davis, California, United States of America

Abstract

Conventional methods for assessing PET spatial resolution, such as the NEMA NU-2 protocol and ACR guidelines, provide limited insight into the effective resolution achieved with modern clinical reconstruction techniques. In this work, we propose a simple and low-cost fan-pattern paper phantom enabling modulation transfer function (MTF)-based analysis for continuous spatial resolution characterization. The phantom consists of laser-cut paper in a fan geometry with gap sizes decreasing from 5 mm to 1 mm which is briefly immersed in an F-18 solution.
Reproducibility and repeatability were evaluated on a uEXPLORER PET/CT system. Images were reconstructed using filtered back projection (FBP) and multiple clinical reconstruction algorithms, including OSEM and artificial intelligence (AI)-based methods. Spatial resolution full width at half maximum (FWHM) was derived from the MTF using the 50% frequency criterion. Results demonstrate an average radial resolution of 3.23 ± 0.034 mm for reproducibility and 3.22 ± 0.054 mm for repeatability, consistent with NEMA NU-2 measurements but with lower uncertainties, which can be attributed to the robustness of the MTF evaluation method. The proposed approach enables rapid (<5 min), robust, and continuous assessment of spatial resolution, including under clinical and AI-based reconstruction conditions. The proposed phantom provides a practical tool for PET system characterization and reconstruction optimization and will be made available to facilitate wider adoption.

Acknowledgment

This work was supported by the NIH NIBIB grant No. R03EB036611.

References
[1] NU 2-2018. National Electrical Manufacturers Association; 2018.
[2] Measurements-of-Positron-Emission-Tomographs, ACR; 2025
[3] G. Li et al., 2024, DOI: 10.2967/jnumed.123.265929
Keywords: PET characterization, PET spatial resolution, performance evaluations, NEMA standard, ACR
17:30 Poster panel: 251

Poster Number:
MIC-10-251

Quantitative Image Quality Evaluation of 177Lu SPECT Using a NEMA NU 4-based SPECT IQ Phantom (#2365)

Y. Zhang1, D. Boreel2, M. Boswinkel2, R. M. Ramakers3, K. Chatzipapas1, M. C. Goorden1

1 TUDelft, Radiation Science & Technology Department, Delft, Netherlands
2 Radboud University Medical Center, Department of medical Imaging, Nijmegen, Netherlands
3 MILabs B.V, Utrecht, Netherlands

Abstract

Abstract—Lutetium-177, a combined gamma and beta emitter,
is gaining significant attention as a theranostic isotope with clinical
applications in metastatic prostate cancer and neuroendocrine
tumors. Quantitative imaging of 177Lu in a preclinical setting
is important for therapy optimization. This study evaluates the
image quality performance of a small-animal SPECT scanner
equipped with an ultra-high-sensitivity collimator for 177Lu using
a SPECT IQ phantom adapted from the NEMA NU 4 PET
phantom.
Whole-body acquisitions were performed using a multi-bed
protocol, and projection data were reconstructed with an iterative
algorithm under different post-filtering conditions. Image quality
was assessed using recovery coefficients (RCs) for rods of varying
diameters, and image noise was quantified as the percentage
standard deviation (%STDunif ) within a uniform region.
The results show that RC varies with rod size, iteration
number, and post-reconstruction filtering. Effective recovery
was observed for 1.35 mm rods (RC ∼0.77), while RC values
decreased to approximately 0.38 and 0.23 for 1.15 mm and
0.95 mm rods, respectively, indicating that reliable recovery is
achieved for rod sizes of about 1.5 mm. Image noise increases
with iteration number and decreases with stronger smoothing,
reflecting the expected resolution–noise trade-off.
These initialfindings demonstrate the capability of the SPECT
IQ phantom for quantitative performance evaluation and protocol
optimization in high-resolution SPECT imaging.

AcknowledgmentThis work was supported by the research Grant UNRANU - Understanding the radiobiology of therapeutic medical radionuclides (project no.15311), which is funded by the Dutch Research Council (NWO).
Keywords: SPECT IQ phantom, Lutetium-177, recovery coefficient, image noise, theranostics
17:30 Poster panel: 254

Poster Number:
MIC-10-254

Investigation of the stability of energy-based scatter estimation in PET (#2546)

S. ZamanPour1, A. Rezaei1, J. Nuyts1, G. Schramm1

1 KU Leuven, Department of Imaging and Pathology, Leuven, Belgium

Abstract

Energy-based scatter estimation in PET estimates scatter from the energy information of detected photon pairs, without relying on an attenuation map. Recent work demonstrated promising results using a model that decomposes the 2D energy histogram into true and scattered coincidences via three basis functions. This model rests on two fundamental assumptions. First, that the shape of the scattered energy spectrum is sufficiently stable across varying activity distributions and time-of-flight (TOF) bins to be represented by fixed weights on two basis functions (c1 p1 + c2 p2). Second, that the scattered and true energy distributions remain statistically distinguishable given the finite energy resolution of the scanner: when low-angle Compton scatter dominates (as in LORs through high-contrast hot regions at central TOF bins) the scatter spectrum shifts towards 511 keV and the two distributions may become so spectrally similar that reliable decomposition is mathematically infeasible, regardless of counting statistics. In this work, we investigate the validity of both assumptions using long MCGPU-PET Monte Carlo simulations of an elliptical phantom with hot-sphere inserts of radii 1, 3, and 5 cm and contrast ratios of 1:10 and 1:100. Scattered energy spectra are extracted and inspected across TOF bins, hot-region sizes, and contrast ratios; challenging cases are identified for each assumption; and the feasibility of the nine-component 2D energy histogram decomposition from S.A. Zaman Pour et al. is evaluated.

Keywords: Energy-based scatter estimation, MCGPU-PET
17:30 Poster panel: 257

Poster Number:
MIC-10-257

A Quantitative Comparison of PET Image Reconstruction Methods for Amyloid-Beta Plaque Identification in Alzheimer's Disease (#2744)

A. Azeez1, M. Khalighi1, A. Sokolowski2, E. Mormino2, C. Young3, F. Boada1

1 Stanford School of Medicine, Radiological Sciences Laboratory, Palo Alto, California, United States of America
2 Stanford School of Medicine, Department of Neurology and Neurological Sciences, Palo Alto, California, United States of America
3 University of California Davis, Department of Radiology, Davis, Davis, California, United States of America

Abstract

The accurate quantification of amyloid-beta (Aβ) plaques with  Positron Emission Tomography (PET) is critical for the early diagnosis and management of Alzheimer's Disease (AD). However, the diagnostic accuracy of PET imaging is heavily dependent on the image reconstruction algorithm employed. This study provides a quantitative comparison of three distinct PET image reconstruction methods to determine their efficacy for Aβ plaque identification. Using a clinical AD dataset, each of the three algorithms was applied to generate enhanced PET images. The resulting Aβ burden was quantified using the standardized Centiloid scale(CL), and a rigorous statistical analysis was performed to evaluate the performance differences between the methods. Identifying the optimal reconstruction technique is paramount, as its implementation can lead to more sensitive early detection of AD, improved monitoring of disease progression, and ultimately, the enhancement of therapeutic outcomes.

Acknowledgment
This work was supported in part by PHS Grant EB029306.
Keywords: AI/ML Image Reconstruction, Alzheimer's Disease, PET/MR, Clinical application, Amyloid Quantification
17:30 Poster panel: 260

Poster Number:
MIC-10-260

Dose-Dependent Electron Density CT Images for PET Attenuation Correction (#2877)

D. Niekämper1, 2, S. Mostafapour2, L. Tellmann1, S. Faby4, H. Alrakh1, J. Scheins1, J. Braaksma3, D. Overbeek2, M. J.W. Greuter3, N.J. Shah1, 5, C. Tsoumpas2, C. Lerche1

1 Forschungszentrum Jülich GmbH, Institute of Neuroscience and Medicine 4, Jülich, North Rhine-Westphalia, Germany
2 University of Groningen, University Medical Center Groningen, Nuclear Medicine and Molecular Imaging, Groningen, Netherlands
3 University of Groningen, University Medical Center Groningen, Department of Radiology, Groningen, Netherlands
4 Siemens Healthineers AG, Department of Computed Tomography, Forchheim, Germany
5 RWTH Aachen University, Department of Neurology, Aachen, Germany

Abstract

Generating accurate attenuation maps from CT images is essential for quantitative PET/CT imaging, while minimising radiation exposure is also important. In conventional CT, the attenuation contributions from Compton effect and photoelectric absorption are combined into a single attenuation value and transformed together from the CT energy spectrum to the PET imaging energy, despite their different energy dependencies. In contrast, spectral CT enables separate electron density and effective atomic number images to be reconstructed. This allows to transform the relevant Compton attenuation contribution independently by applying a scaling factor to the electron density image. In this study, we compare attenuation maps obtained using this approach with those obtained via a bilinear transformation. Measurements of phantoms containing various tissue-equivalent inserts are performed using a PET/CT with twin-beam dual-energy capability and a photon-counting CT. In additional measurements, selected inserts are replaced by compartments containing activity in order to evaluate the effect of the different attenuation maps on reconstructed PET images. The measurements are performed across a wide dose range to investigate how far the radiation dose can be reduced while still obtaining accurate attenuation maps. Preliminary results show good agreement between attenuation values obtained at different dose levels, with the highest discrepancies between both transformation methods observed for bone-equivalent inserts. Using electron density and effective atomic number reconstructions from spectral scanners to generate attenuation correction values may improve quantitative PET imaging.

Acknowledgment

DN and CT were partially supported by the NWO AES Vici Talent program (doi: 10.61686/REEAR02777).

References
[1] Carney, JPJ et al. 2006, 'Method for transforming CT images for attenuation correction in PET/CT imaging', Medical Physics, 33 4
 
[2] Guy, MJ et al. 1998, ‘DETECT - Dual Energy Transmission Estimation CT - for improved attenuation  correction in SPECT and PET’, IEEE Trans. Nucl. Sci, 45 3
 
[3] Mostafapour, S et al. 2024, 'Ultra-low dose CT scanning for PET/CT', Medical Physics, 51 1
 
Keywords: PET/CT, attenuation correction, energy transformation
17:30 Poster panel: 263

Poster Number:
MIC-10-263

TOF-Aware Single Scatter Simulation in STIR: Implementation and Validation Against Monte Carlo (#2967)

M. Sabeur1, 2, S. ZamanPour4, S. Stute2, Y. Moussaoui1, 2, S. Moussaoui1, K. Thielemans3

1 Nantes Université, Ecole Centrale de Nantes, LS2N, Department of Medical Imaging, Nantes, France
2 CHU Nantes, Section of Nuclear Medicine, Nantes, France
3 University College London, Institute of Nuclear Medicine, London, United Kingdom
4 KU Leuven, Department of Imaging and Pathology, Leuven, Belgium

Abstract

Single scatter simulation (SSS) combined with tail-
fitting is the standard analytical method for scatter correction
in PET. Extending SSS to time-of-flight (TOF) data requires
non-trivial adaptations to the scatter kernel. The open-source
Software for Tomographic Image Reconstruction (STIR) library
has a widely used implementation of SSS for non-TOF data.
Here, we present a fully TOF-aware SSS implementation in STIR
with two contributions: (i) the TOF kernel εt in the scatter-arm
emission integral is computed via erf (error function) as opposed
to central point approximation in Watson, avoiding the need for
an extra normalisation step. (ii) re-use of cached integrals from
the non-TOF implementation. The implementation is validated
against MCGPU Monte Carlo on a 9-ring compact PET scanner
(27 TOF bins, 169 ps, uniform phantom), confirming strong
agreement with MC single scatter (r=0.985 full 4-D sinogram;
r=0.994 TOF-summed spatial). This contribution opens the door
to fully quantitative PET reconstructions for TOF-PET in STIR.

Acknowledgment

The authors thank the STIR open-source community. 
This work was supported by LS2N, Ecole Centrale de Nantes, and by CCP SyneRBI which is funded by the EPSRC grant EP/T026693/1 and by STFC through the UKRI Digital Research Infrastructure programme.

 

References
[1] C. Watson, “Extension of single scatter simulation to scatter correction
of time-of-flight PET,” IEEE Trans. Nucl. Sci., vol. 54, no. 5, pp. 1679–
1686, 2007. doi: 10.1109/TNS.2007.901227.
[2] K. Thielemans, C. Tsoumpas, S. Mustafovic et al., “STIR: software for
tomographic image reconstruction release 2,” Phys. Med. Biol., vol. 57,
no. 4, pp. 867–883, 2012. doi: 10.1088/0031-9155/57/4/867.
[3] Polycarpou, K. Thielemans, R. Manjeshwar, P. Aguiar, P. Marsden,
and C. Tsoumpas, “Comparative evaluation of scatter correction in 3D
PET using different scatter-level approximations”, Annals of Nuclear
Medicine, vol. 25, no. 9, pp. 643–649, Nov. 2011, doi: 10.1007/s12149-
011-0514-y.
[4] J. L. Herraiz, A. Lopez-Montes, and A. Badal, “MCGPU-PET: an open-
source real-time Monte Carlo PET simulator,” Comput. Phys. Commun.,
vol. 296, p. 109008, Mar. 2024. doi: 10.1016/j.cpc.2023.109008.
[5] R. José Santo, A. Salomon, H. W. A. M. de Jong et al., “openSSS: an
open-source implementation of scatter estimation for 3D TOF-PET,”
EJNMMI Phys., vol. 12, p. 17, 2025. doi: 10.1186/s40658-025-00730-x.
Keywords: PET, time-of-flight, single scatter simulation, STIR, Monte Carlo validation
17:30 Poster panel: 266

Poster Number:
MIC-10-266

Event-by-Event Energy-Based Scatter Correction via Bayesian Likelihood Estimation (#3026)

M. Teimoorisichani1

1 Siemens Medical Solutions USA Inc., Knoxville, Tennessee, United States of America

Abstract

Energy-based scatter (EBS) correction is increasingly utilized in PET image reconstruction. This work introduces a novel EBS framework that assigns scatter probabilities to individual listmode events via analytical Bayesian calculations. The method utilizes pre-defined energy distributions of scatter kernels as basis functions. By analyzing the global energy spectrum, the algorithm estimates the fractional probability for each event type and subsequently calculates the likelihood of Compton scattering for individual coincidences. The approach was validated using two phantom datasets acquired on a Biograph Vision PET/CT scanner (Siemens Healthineers). Reconstructed images were compared against standard vendor-implemented single-scatter simulation (SSS) with tail-fitting. Results demonstrate high consistency, with scatter fraction estimates differing by less than 3% from SSS. Furthermore, uniform and NEMA image quality phantom analyses showed less than 1% divergence in contrast recovery and background variability. These findings suggest that the proposed EBS method provides a robust, computationally efficient alternative for scatter estimation in clinical PET reconstruction.

References
[1] Watson, C., Extension of Single Scatter Simulation to Scatter Correction of Time-of-Flight PET. IEEE Transactions on Nuclear Science, 2007. 54(5): p. 1679-1686.
[2] Herraiz, J.L., A. Lopez-Montes, and A. Badal, MCGPU-PET: An open-source real-time Monte Carlo PET simulator. Computer Physics Communications, 2024. 296.
[3] Efthimiou, N., J.S. Karp, and S. Surti, Data-driven, energy-based method for estimation of scattered events in positron emission tomography. Phys Med Biol, 2022. 67(9).
[4] Hamill, J.J., et al., Energy-based scatter estimation in clinical PET. Med Phys, 2024. 51(1): p. 54-69.
Keywords: Scatter Correction, Energy-based Scatter (EBS), Listmode Reconstruction, Bayesian Estimation
17:30 Poster panel: 269

Poster Number:
MIC-10-269

Synthetic CT Generation for Attenuation Correction in Respiratory-Gated Rb-82 Cardiac PET Using 3D Diffusion Models (#3061)

B. Yu1, W. Qi2, J. Kolthammer2, K. Gong1

1 University of Florida, J. Crayton Pruitt Family Department of Biomedical Engineering, Gainesville, Florida, United States of America
2 Canon Medical Research USA, Inc., Vernon Hills, Illinois, United States of America

Abstract

Accurate attenuation correction (AC) using computed tomography (CT) is critical for quantitative PET imaging. However, in respiratory-gated cardiac PET, respiratory phase mismatch between gated PET and CT can compromise the accuracy of CT-based AC, while acquiring additional phase-matched CT scans is impractical due to radiation dose concerns. To address these limitations, we employed a 3D Denoising Diffusion Probabilistic Model (DDPM) to generate synthetic CT images directly from non-attenuation corrected (NAC) PET data for gated PET AC. The model was trained using paired non-gated Rb-82 myocardial perfusion PET scans and CT-derived attenuation maps, and directly applied to respiratory-gated PET without retraining. Visual evaluation showed that 3D DDPM generated synthetic CT images with superior anatomical fidelity and improved respiratory phase alignment with gated PET compared with 3D GAN, 3D UNet, and 2D DDPM. Regional quantitative analysis near the diaphragm further demonstrated that 3D DDPM achieved improved attenuation correction in regions most affected by respiratory mismatch.

References
[1] Yu, B., Ozdemir, S., Dong, Y., Shao, W., Pan, T., Shi, K., & Gong, K. (2025). Robust whole-body PET image denoising using 3D diffusion models: evaluation across various scanners, tracers, and dose levels. European Journal of Nuclear Medicine and Molecular Imaging, 52(7), 2549-2562.
[2] Pan, T., Thomas, M. A., & Luo, D. (2022). Data‐driven gated CT: an automated respiratory gating method to enable data‐driven gated PET/CT. Medical physics, 49(6), 3597-3611.
Keywords: PET-to-CT translation, Respiratory gated PET, Diffusion models, Attenuation correction
17:30 Poster panel: 272

Poster Number:
MIC-10-272

Beyond the Single-Tissue Voxel:A Spin-Echo–Based Multi-Tissue Decomposition Framework with Implications for Synthetic MRI and Machine Learning Diagnosis (#1194)

H. Li1, Q. Luo2, M. J. Pomeroy3, H. Li4, Y. Kee4, X. J. Zhou2, Z. J. Liang4, 1, L. Zeng5

1 Stony Brook University, Electrical and Computer Engineering, Stony Brook, New York, United States of America
2 University of Illinois College of Medicine at Chicago, Radiology, Chicago, Illinois, United States of America
3 Stony Brook University, Biomedical Engineering, Stony Brook, New York, United States of America
4 Stony Brook University, Radiology, Stony Brook, New York, United States of America
5 Utah Valley University, Computer Sience, Orem, United States of America

Abstract

Conventional synthetic MRI generates contrast images from a single effective set of voxel-wise quantitative parameters: proton density (PD), T1, and T2, which may be insufficient when any two or all three of gray matter (GM), white matter (WM), and cerebrospinal fluid (CSF) coexist within the same image voxel. This work presents an exploratory framework using the standard spin-echo signal generation model with a purity-weighted Expectation-Maximization algorithm for tissue prototype estimation that systematically quantifies the prevalence of such multi-tissue voxels. From three co-registered spin-echo MR brain images at distinct Time to Echo (TE), Time to Repetition (TR) settings, per-voxel PD, T1, and T2 relaxation maps are recovered via a robust nonlinear inversion pipeline and validated by re-synthesizing the original images. Each voxel is then decomposed into fractional contributions of GM, WM, and CSF using three methodologically independent models: a Signal-Domain Mixture Model, Parameter-Domain Mixture Model and Mean Parameter-Domain Mixture Model. All three consistently show that 52% – 62% of brain image voxels contain contributions from multiple tissue types, demonstrating model-independently that the single-effective-parameter representation is insufficient in most brain voxels and that conventional synMRI parameter maps can be systematically biased. A tissue-aware synthesis framework is proposed that generates physically grounded MRI contrasts across the full spin-echo (TE, TR) measurement space, providing a quantitative basis for more reliable synthetic MRI. Future work will extend this framework to multi-subject validation and evaluate the synthesized contrast library for machine learning-based lesion diagnosis.

References
[1] D. W. McRobbie, E. A. Moore, et al. “MRI from Picture to Proton,” in Cambridge Univ. Press Cambridge, 2nd ed.U.K.2006.
[2] JBM. Warntjes, et al. “Rapid magnetic resonance quantification on the brain: Optimization for clinical usage,” Magnetic Resonance in Medicine, vol. 60, no. 2, pp. 320–329, 2008.
[3] I. Blystad, et al. “Synthetic MRI of the brain in a clinical setting,” Acta Radiol., vol. 53, no. 10, pp. 1158–1163, 2012.
[4] M. A. González Ballester, A. Zisserman, and M. Brady, “Estimation of the partial volume effect in MRI,” Med Image Anal., vol. 6, no. 4, pp. 389–405, 2002.
[5] Y. Zhang, M. Brady, and S. Smith, “Segmentation of brain MR images through a hidden Markov random field model and the expectation-maximization algorithm,” IEEE Trans Med Imaging,
vol. 20, no. 1, pp. 45–57, 2001.
[6] Z. Liang and S. Wang, “An EM approach to MAP solution of segmenting tissue mixtures: a numerical analysis,” IEEE Trans Med Imaging, vol. 28, no. 2, pp. 297–310, 2009.
[7] P. Tofts, Ed., “Quantitative MRI of the Brain: Measuring Changes Caused by Disease.” Chichester, U.K.: Wiley, 2003.
[8] Z. Liang, S. Chang, Y. Gao, et al.  "Leveraging prior knowledge in machine intelligence to improve lesion diagnosis for early cancer detection,” Medical Physics, vol. 52, no. 7, pp.1–14, 2025.
[9] H. Li, Pomeroy, M. J, et al. "Enhancing classification of indeterminate pulmonary nodules at multiple scans in NLST dataset through integrated spectral CT radiomics." Medical Imaging 2026: Imaging Informatics. Vol. 13930. SPIE, 2026.
Keywords: synthetic MRI, partial volume effect, tissue mixture model, multi-tissue decomposition, quatitative MRI
17:30 Poster panel: 275

Poster Number:
MIC-10-275

Assessment of Cortical Defect Discriminability in Reduced-Dose Pediatric Tc-99m DMSA Renal SPECT (#1422)

W. Fu1, Y. Yang1, N. Kwatra2, 3, S.T. Treves2, 3, F. Fahey2, M. A. King4

1 Illinois Institute of Technology, Electrical and Computer Engineering, Chicago, Illinois, United States of America
2 Harvard Medical School, Radiology, Boston, Massachusetts, United States of America
3 Boston Children’s Hospital, Radiology, Boston, Massachusetts, United States of America
4 University of Massachusetts Medical Center, Radiology, Worcester, Massachusetts, United States of America

Abstract

In prior work, deep-learning (DL) based denoising was demonstrated to be capable of yielding images comparable to standard-dose studies when the imaging dose is reduced several-fold in pediatric Tc-99m DMSA renal SPECT, based on both quantitative image quality metrics and visual assessment by clinical experts. In this study, we further evaluate the diagnostic utility of DL-denoised images by employing a DL-based classifier to perform a task-based assessment of cortical defects, where a 3D convolutional neural network (CNN) was used to differentiate kidneys with cortical defects from those with normal uptake. We demonstrated this approach with a set of clinical data acquired from 956 subjects, with performance quantified using receiver-operating characteristic (ROC) analysis. The results showed that, without DL denoising, the quarter-dose data achieved a lower classification performance than the full-dose in terms of the area under the ROC curve (p-value < 0.001); however, with DL denoising, the quarter-dose data achieved a statistically comparable performance to that of the full-dose data (p-value = 0.452).

AcknowledgmentThis work was supported under NIH Grant R01-EB029315.
Keywords: DL denoising, pediatric renal SPECT, reduced dose, cortical defect, classification
17:30 Poster panel: 278

Poster Number:
MIC-10-278

Automated Coronary Artery Extraction up to Distal Vessels Based on Hessian Matrix and 3D Skeletonization to Compensate for Incomplete Labels (#1528)

R. Ide1, K. Ogawa2

1 Hosei University, Applied Informatics, Graduate School of Science and Engineering, Tokyo, Japan
2 Hosei University, Applied Informatics, Faculty of Science and Engineering, Tokyo, Japan

Abstract

While deep learning has become the mainstream approach for coronary artery extraction from coronary computed tomography angiography (CCTA) images, its reliance on training data makes it difficult to accurately extract fine distal vessels using existing datasets with insufficient annotations. Therefore, this study proposes a hybrid extraction method utilizing image feature analysis. This method consists of three stages. First, an initial extraction is performed based on anatomical features and CT values. Second, a chained connection algorithm combining tubular structure enhancement via the Hessian matrix and three-dimensional (3D) skeletonization is employed to connect vessel fragments through local search, thereby recovering continuity. Third, volume expansion is performed using an adaptive threshold corresponding to local intensity to restore the accurate vessel diameter. Application of this method to an open dataset demonstrated extreme robustness against CT value degradation caused by insufficient contrast agent perfusion. We successfully extracted and reconstructed fine distal vessels that were missing in the existing labels while maintaining their continuity. Future work involves collaborating with clinical specialists to reconstruct high-quality ground truth labels and conduct rigorous quantitative evaluations.

References
[1] Zeng, A, Wu, C, Lin, G, et al. 2023, ‘ImageCAS: A large-scale dataset and benchmark for coronary artery segmentation based on computed tomography angiography images,’ Computerized Medical Imaging and Graphics, Vol.109, https://doi.org/10.1016/j.compmedimag.2023.102287
[2] Xu, X 2023, ‘ImageCAS: A Large-Scale Dataset and Benchmark for Coronary Artery Segmentation based on CT,’ Kaggle
[3] Zhao, B, Peng, J, Zhang, K, et al. 2025, ‘TMN-Net: A Hybrid 2.5D Multi-Branch Transformer Network for Coronary Artery Segmentation in Cardiac Diagnosis,’ IEEE Access, pp. 115581-115603
[4] Tayebi, RM,  Wirza, R, Sulaiman, PSB, et al., 2014, ‘Cardiac Components Categorization and Coronary Artery Enhancement in CT Angiography,’ International Conference on Computer Assisted System in Health, https://doi.org/10.1109/CASH.2014.19
Keywords: coronary artery extraction, CCTA, Hessian matrix, 3D skeletonization, distal vessels
17:30 Poster panel: 281

Poster Number:
MIC-10-281

Differential Diagnosis of PD, PSP, and MSA: An XGBoost Approach to Retinal Layer Analysis (#1850)

M. Rodriguez-Rubio1, F. J. Martinez-Murcia1, 2, J. E. Arco3, I. A. Illán1, 2, F. Segovia1, 2, C. Jimenez-Mesa2, C. Vazquez-Garcia1, 2, J.M. Gorriz1, 2, J. Ramirez1, 2, S. Schönecker4, S. Kamp4, L. Klatte4, S. Katzdobler4, 6, L. Marth4, 6, O. Wagemann4, 6, C. Palleis4, 6, U. Fietzek4, E. Weidinger4, 6, A. Danek4, G. Hölinger4, 6, J. Havla7, A. Bernhardt4, A. Jäck4, J. Levin4, 6

1 University of Granada, Department of Signal Theory, Networking and Communications, Granada, Spain
2 Andalusian Research Institute in Data Science and Computational Intelligence, Granada, Spain
3 Mind, Brain and Behavior Research Center (CIMCYC), Granada, Spain
4 Ludwig-Maximilians-Universität München, Department of Neurology, LMU University Hospital, Munich, Bavaria, Germany
5 German Center for Neurodegenerative Diseases (DZNE), Munich, Bavaria, Germany
6 Ludwig-Maximilians-Universität München, Institute of Clinical Neuroimmunology, Biomedical Center and University Hospital, Munich, Bavaria, Germany

Abstract

Optical coherence tomography (OCT) provides non-invasive retinal biomarkers that may reflect the distinct neurodegenerative patterns of underlying proteinopathies. This study evaluates a machine learning framework to classify synucleinopathies (Parkinson’s disease [PD] and multiple system atrophy [MSA]) and tauopathies (progressive supranuclear palsy [PSP]) using retinal layer morphometry. Several classification tasks were assessed, ranging from healthy control (HC) discrimination to the differential classification of proteinopathic groups. XGBoost was employed as the primary classifier and evaluated using nested cross-validation. To minimize intra-subject variability, only the most recent visit per patient was included, with features aggregated as the average of both eyes to represent bilateral neurodegenerative involvement. The model demonstrated notable efficacy in distinguishing synucleinopathies (PD+MSA) from tauopathies (PSP), achieving an accuracy of 0.704 ± 0.057. This performance was comparable to the binary classification of HCs vs. all pathological cases (0.724 ± 0.079). In contrast, multi-class settings proved more challenging. These findings suggest that while fine-grained clinical subtype differentiation remains a challenge, OCT-derived biomarkers are highly effective at capturing the broad divergent impacts of alpha-synuclein and tau pathology on the retina.

AcknowledgmentThis research is part of the projects PID2022-137629OA-I00, {CNS2024-154296} and PID2022-137451OB-I00, funded by MICIU/AEI/10.13039/501100011033 and by ERDF/ EU, and TIC251-G-FEDER project, funded by ERDF/EU. The Friedrich-Baur-Stiftung and the “FöFoLe” Program of the Faculty of Medicine of the Ludwig-Maximilians-Universität München have supported the work of S.S.
References
[1] C. Palleis et al., “Biomarker-based classification of corticobasal syndrome,” Brain, 2026.
[1] Luo, X, et al. 2023, ‘Retinal Structure Abnormalities in Parkinson's Disease and Atypical Parkinsonism’, Biomolecules, 13, 218, Basel: MDPI.
[2] Mammadova, N, et al. 2026, ‘Retinal Thickness Profiles in Parkinsonian Syndromes: Discerning Parkinson's Disease, Multiple System Atrophy, and Progressive Supranuclear Palsy via Optical Coherence Tomography’, Biomedicines, 14, 249, Basel: MDPI.
[3] Visser, F, et al. 2024, ‘Optical Coherence Tomography as a Biomarker in the Differential Diagnosis between Parkinson's Disease and Atypical Parkinsonian Syndromes: A Narrative Review’, Applied Sciences, 14, 2491, Basel: MDPI.
[4] Zhang, Y, et al. 2023, ‘Applications of Machine Learning to Diagnosis of Parkinson’s Disease’, Brain Sciences, 13, 1546, Basel: MDPI.
Keywords: Optical Coherence Tomography (OCT), Parkinson’s Disease (PD), Atypical Parkinsonian Syndromes (APS), XGBoost, Differential Diagnosis
17:30 Poster panel: 284

Poster Number:
MIC-10-284

Multi-step Nonlinear Iteration Method for ECT Energy Spectrum Decomposition (#1885)

J. Mu1, Z. Lu1, J. Peng1, Z. Yu1, D. Bi1, S. An1

1 United Imaging Healthcare Co., Ltd, MI BU Detector R&D Center, Shanghai, China

Abstract

Theranostic radionuclide molecular imaging represents a core technology in precision nuclear medicine. However, its clinical application is hindered by scatter interference and complex background signals in spectral analysis. To address this bottleneck, this paper proposes a multi-step nonlinear iterative method for energy spectrum background subtraction and full-energy peak analysis, suitable for emission computed tomography detectors. The method realizes high-precision background subtraction and peak stripping via a systematic workflow consisting of peak localization, Log-Log-Square root transformation, interval iterative calculation, and Gaussian fitting. Experimental validation is conducted under both single-nuclide and multi-nuclide conditions. The results demonstrate that the proposed method can effectively separate full-energy peaks submerged by scatter in single-nuclide spectra. For the CsI detector, the energy resolution is improved from 12.9% to10.97%. For multi-nuclide and multi-peak spectra, it successfully separates overlapping peaks, suppresses Compton continuum interference, and optimizes energy resolution to around 10.6%. This method improves detector energy resolution, provides solid technical support for the clinical translation of theranostic radionuclide molecular imaging, and exhibits favorable adaptability and correction accuracy. Future work will focus on optimizing iterative efficiency and extending the method’s applicability to more complex clinical scenarios.

References
[1] N. Pandit-Taskar, S. M. Larson, and J. A. Carrasquillo, "Theranostics in nuclear medicine: current status and future directions," J. Nucl. Med., vol. 59, no. 1, pp. 15-22, Jan. 2018, doi:10.2967/jnumed.117.198945.
[2] M. Morhac and V. Matousek, “Peak clipping algorithms for background estimation in spectroscopic data,” Appl. Spectrosc., vol. 62, no. 1, pp. 91–106, Jan. 2008, 10.1366/000370208783412762.
[3] W. Zhang, Y. Li, H. Wang, Y. Zhang, and Z. Li, "System performance characterization of United Imaging Healthcare's Micro-SPECT/CT platform for preclinical imaging," in Proc. IEEE Nucl. Sci. Symp. Med. Imaging Conf. (NSS/MIC), Beijing, China, 2024, pp. 1-4, doi:10.1109/NSS/MIC48546.2024.10765432.
Keywords: ECT detector, energy peak extraction, Compton scattering, energy resolution
17:30 Poster panel: 287

Poster Number:
MIC-10-287

Residual Scatter Correction Using Grid-Wall Shadow Sampling in Cone-Beam CT with a Two-Dimensional Anti-Scatter Grid (#2293)

J. Park1, S. Cho1, H. Yang1, D. Jeon1, D.Y. Han2, H. Cho1

1 Yonsei University, Department of Radiation Convergence Engineering, Wonju, Republic of Korea
2 Yale School of Medicine, Department of Therapeutic Radiolog, New Haven, Connecticut, United States of America

Abstract

Residual X-ray scatter degrades image contrast, uniformity, and quantitative accuracy in cone-beam computed tomography (CBCT). A two-dimensional (2D) anti-scatter grid physically suppresses scattered photons, but conventional gain correction using a matched flood projection only compensates the multiplicative grid-wall footprint and does not remove additive residual scatter, which persists as low-frequency nonuniformity in reconstructed images. This work proposes a projection-domain residual scatter correction method that exploits grid-wall shadows as internal sampling positions for scatter estimation. After gain correction, primary signals equalize between wall-shadow and neighboring hole pixels by design; any remaining signal mismatch is therefore attributed to additive residual scatter. Sparse scatter estimates at wall-shadow locations are expanded into a full-field scatter map via Gaussian-weighted interpolation and subtracted from the raw projection before logarithmic conversion and filtered backprojection reconstruction. This approach requires no additional hardware or dedicated scatter measurements, as it utilizes the intrinsic grid-wall structure already present in every grid projection. We evaluate the method using an ACR 464 phantom scanned at 80 kVp and 50 mA with 720 views on a 1516 × 1516 flat-panel detector (0.28 mm pixel pitch). Compared with no-grid and grid-only reconstruction, the proposed correction reduces residual shading artifacts and improves image uniformity. The contrast-to-noise ratio increases from 1.63 (no grid) and 3.27 (grid only) to 3.69 with the proposed method, with improved contrast visibility. These results demonstrate that grid-wall shadows provide practical sampling positions for estimating and correcting additive residual scatter in grid-based CBCT systems.

Acknowledgment

This work was supported by the National Research Foundation of Korea(NRF) grant funded by the Korea government(MSIT)(RS-2024-00347270).

References
[1] Bayat, F et al 2024, '2D antiscatter grid and scatter sampling based CBCT method for online dose calculations during CBCT guided radiation therapy of pelvis', Med Phys., vol. 51, 3053-3066
Keywords: Cone-beam computed tomography (CBCT), Anti-scatter grid, Residual scatter correction, Grid-wall shadow sampling
17:30 Poster panel: 290

Poster Number:
MIC-10-290

Low-Dimensional Embeddings of Structural MRI for Multi-Disease Brain Morphometry (#2362)

C. Jimenez Mesa1, 2, M. Mamalakis3, 4, C. Vazquez-Garcia2, 5, F.J. Martinez Murcia2, 5, J. E. Arco2, 6, F. Segovia2, 5, J. Ramirez2, 5, A. Ortiz1, 2, J. Suckling3, J.M. Gorriz2, 3

1 University of Malaga, Department of Communications Engineering, Malaga, Spain
2 University of Granada, DaSCI Andalusian Institute of Data Science and Computational Intelligence, Granada, Spain
3 University of Cambridge, Department of Psychiatry, Cambridge, United Kingdom
4 University of Cambridge, Cancer Research United Kingdom Cambridge Institute, Cambridge, United Kingdom
5 University of Granada, Department of Signal Theory, Telematics and Communications, Granada, Spain
6 University of Granada, Mind, Brain, and Behavior Research Center, Granada, Spain

Abstract

The differential diagnosis of neurological disorders such as Alzheimer’s disease (AD), frontotemporal dementia (FTD), Parkinson’s disease (PD), and multiple sclerosis (MS) remains a clinical challenge due to overlapping neuroanatomical atrophy patterns in structural Magnetic Resonance Imaging (MRI). This work investigates whether low-dimensional embeddings derived from T1-weighted MRI can reveal a consistent neuroanatomical organisation across these conditions. Using a dataset of 528 subjects, we compared two feature extraction strategies: (i) high-dimensional voxel-wise representations and (ii) regional embeddings based on the Automated Anatomical Labeling (AAL) atlas (116 regions). Dimensionality reduction was performed using Principal Component Analysis (PCA) followed by Uniform Manifold Approximation and Projection (UMAP) to project subjects into a two-dimensional manifold embedding. Our results demonstrate that the neuroanatomical manifold is remarkably scale-invariant as both voxel-wise and regional embeddings yielded consistent topological distributions. Centroid distance analysis in the embedding space revealed a progressive deviation from the HC baseline across disorders. Interestingly, while the model was built on the HC-AD axis, MS and FTD cohorts exhibited the largest structural displacements (dMS=7.24, dFTD=5.59) compared to AD (dAD=3.57) and PD (dPD=4.27). These findings suggest that low-dimensional manifold representations derived from structural MRI are robust to spatial aggregation and may provide an interpretable framework for studying shared and disease-specific brain alterations across heterogeneous neurological disorders.

AcknowledgmentThis research is part of the PID2022-137451OB-I00 and PID2022-137629OA-I00 projects, funded by the MICIU/AEI /10.13039/501100011033 and by "ERDF/EU". C.J.M is supported by grant JDC2023-051807-I funded by MICIU/AEI/10.13039/501100011033 and by ESF+.
References
[1] Jimenez-Mesa, C., Ramirez, J., Suckling, J., Vöglein, J., Levin, J., Gorriz, J. M., & Alzheimer’s Disease Neuroimaging Initiative. (2023). A non-parametric statistical inference framework for deep learning in current neuroimaging. Information Fusion, 91, 598-611.
[2] Vázquez-García, C., Martínez-Murcia, F. J., Segovia, F., Forte, A., Ramírez, J., Illán, I., Hernández-Segura, A., Jiménez-Mesa, C., & Górriz, J. M. (2026). An explainable framework for the relationship between dementia and metabolism patterns. NeuroImage, 330, 121855.
[3] Prado, P., Medel, V., Gonzalez-Gomez, R., Sainz-Ballesteros, A., Vidal, V., Santamaría-García, H., ... & Ibañez, A. (2023). The BrainLat project, a multimodal neuroimaging dataset of neurodegeneration from underrepresented backgrounds. Scientific Data, 10(1), 889.
[4] Tzourio-Mazoyer, N., Landeau, B., Papathanassiou, D., Crivello, F., Etard, O., Delcroix, N., ... & Joliot, M. (2002). Automated anatomical labeling of activations in SPM using a macroscopic anatomical parcellation of the MNI MRI single-subject brain. Neuroimage, 15(1), 273-289.
Keywords: Manifold learning, Neurological disorders, Structural MRI, PCA, UMAP
17:30 Poster panel: 293

Poster Number:
MIC-10-293

Detector design in the AIDER Compton Camera Using Neural Networks: A Hybrid CNN-DNN Approach (#2484)

D. Rodriguez Almora1, J. Altube1, L. Barrientos1, A. Brunamonti2, K. Brzezinski1, C. Fiorini2, F. Ghisio2, A. Hanna3, A. Iltis3, S. Parashari1, J. Pérez Curbelo1, G. Llosá1

1 Instituto de Física Corpuscular (IFIC-CSIC/UVEG), Valencia,, Spain
2 Politecnico di Milano, Dipartimento di Elettronica, Informazione and Bioingegneria, Milano, Italy
3 DAMAVAN Imaging, Troyes, France

Abstract

Abstract: This paper presents the development of a high-performance Compton camera for targeted radionuclide therapy within the framework of the European project AIDER. The camera is composed of two detector planes, each made of four LaBr3 or CeBr3 monolithic crystals of 50 mm x 50 mm size coupled to Silicon Photomultiplier (SiPM) arrays. In order to determine the optimum pixel size, a hybrid analysis approach based on convolutional neural networks (CNN) and dense networks (DNN) has been developed to estimate the interaction position of the photons. Starting from Monte Carlo simulation data (GATE v7.0) and using a reference based on an analytical model, the proposed alternative incorporates the systematic error of this model as an additional input for the NNs. This achieves active correction of the edge effect and improves the overall spatial resolution (FWHM) from a range of 1.80–2.25 mm down to 0.63–0.70 mm. Results show the benefits of the combined NN approach and indicate the possibility of employing SiPM arrays with large pixel sizes, effectively restricting the amount of readout channels. Experimental tests are ongoing.

Acknowledgment

This work has received funding from the Spanish MICIU / AEI (project MIDAS, PID2022-143246OB-I00) and from the European Union’s EURATOM research and innovation program (AIDER project, G.A. 101061037). S.P. acknowledges the Severo Ochoa CEX2023-001292-S grant, funded by MCIN/AEI/10.13039/501100011033

References
[1] https://aider.ific.uv.es/aider/
[2] L. Barrientos et al., Radiat. Phys. Chem., vol. 208, p. 110922, 2023.
[3] J. Roser et al., Phys. Med., vol. 104956, 2025. doi: 10.1016/j.ejmp.2025.104928.
[4] Li, Z., et al. (2010). IEEE Transactions on Nuclear Science. Vol. 57, No. 5, pp. 2450-2456. 2010. 10.1109/TNS.2010.2051457
[5] G. Jaliparthi, P. F. Martone, A. V. Stolin, and R. R. Raylman, Phys. Med. Biol., vol. 66, no. 14, p. 145008, 2021.
[6] J. Jo and J. S. Lee, Nucl. Instrum. Methods Phys. Res. A, 2024.DOI: 10.1016/j.nima.2023.168925
[7] S. Cho and J. S. Lee, J. Korean Phys. Soc., 2024.DOI: 10.1007/s40042-023-00971-w
[8] R. Liu et al., arXiv:1807.03247, 2018.
[9] A. Nibali et al., arXiv:1801.07372, 2018.
Keywords: convolutional neural network, spatial resolution, radionuclide therapy, AIDER, Compton camera
17:30 Poster panel: 296

Poster Number:
MIC-10-296

Impact of Spatial Normalization on the Performance of Deep Learning-Based Systems for Neurodegenerative Disease Classification (#2537)

F. Segovia1, 3, R. López1, C. Vázquez-García1, 3, A. Hernández1, Á. Zorrilla1, C. Jiménez-Mesa2, 3, F. J. Martínez-Murcia1, 3, I. Á. Illán1, 3, J. Ramírez1, 3, T. Martín-Noguerol4, J. M. Górriz1, 3

1 Univiersity of Granada, Department of Signal Theory, Networking and Communications,, Granada, Spain
2 University of Malaga, Department of Communications Engineering, Málaga, Spain
3 Andalusian Institute of Data Science and Computational Intelligence (DaSCI ), Granada, Spain
4 HT Medica, Radiology department, Granada, Spain

Abstract

This study evaluates the suitability of spatial registration in deep learning for neuroimaging, focusing on the balance between reducing inter-subject anatomical variability and preserving data fidelity. We hypothesize that processing images in their native anatomical space avoids morphological distortions introduced by registration, enabling neural networks to detect subtle structural biomarkers that may be altered in normalized spaces. To test this, a standard densely connected neural network was used to classify two balanced cohorts: one focused on the diagnosis of Parkinson's disease (PD) and the other on the diagnosis of Alzheimer's disease (AD). The experimental design compared four input modalities: whole-brain volumes and gray matter maps, each in native space and registered to a standard reference space. For PD classification, native and registered spaces achieved similar accuracy, although registration produced a marked imbalance between sensitivity and specificity. For AD, the best performance was obtained using gray matter in native space, confirming that preserving topological integrity is critical in pathologies with pronounced atrophy. These findings suggest that spatial registration should not be considered an indispensable preprocessing step in deep learning-based neuroimaging, but rather a methodological choice that must be evaluated based on the task and dataset. Processing in native space is an effective alternative that simplifies preprocessing while preserving diagnostically relevant information, opening future research directions on model robustness to anatomical variability without prior transformation.

AcknowledgmentThis work is part of the projects PID2022-137629OA-I00 and PID2022-137451OB-I00 funded by MICIU/AEI/10.13039/501100011033 and by ERDF/EU, and the C-ING-183-UGR23 project, cofunded by the Consejería de Universidad, Investigación e Innovación and by European Union, funded by Programa FEDER Andalucía 2021-2027
References
[1] M. Klingenberg et al., “MRI Image Registration Considerably Improves CNN-Based Disease Classification”, in Machine Learning in Clinical Neuroimaging: 4th International Workshop, MLCN 2021, Held in Conjunction with MICCAI 2021, Strasbourg, France, September 27, 2021, Proceedings. Berlin, Heidelberg: Springer-Verlag, Sep. 2021, pp. 44–52.
[2] D. Fujibayashi et al., “Nonlinear registration as an effective preprocessing technique for Deep learning based classification of disease,” IEEE Engineering in Medicine and Biology Society. Annual International Conference, vol. 2021, pp. 3245–3250, Nov. 2021.
[3] H. He et al., “Landmark-guided region-based spatial normalization for functional magnetic resonance imaging,” Human Brain Mapping, vol. 43, no. 11, pp. 3524–3544, Apr. 2022.
[4] J. Kim et al., “An interactive cascaded deep learning framework with expert refinement for accurate striatal subregion segmentation,” Scientific Reports, vol. 16, no. 1, p. 6550, Jan. 2026.
[5] V. Ramineni et al., “STN-MRI: A Spatial Transformer Network-Based Pipeline for ROI-Aware Brain MRI Preprocessing and Alignment,” in 2025 29th International Computer Science and Engineering Conference (ICSEC), Nov. 2025, pp. 406–409.
Keywords: MRI data, deep learning, spatial registration, brain scans, medical imaging
17:30 Poster panel: 299

Poster Number:
MIC-10-299

Edge-Guided Deformation Field Fusion for Robust PET Motion Correction (#2650)

Q. Liu1, W. Qi1, J. Kolthammer1

1 Canon Medical Research USA, Vernon Hills, Illinois, United States of America

Abstract

Non-rigid motion correction for gated PET is challenging due to low contrast and high noise in homogeneous regions, where deformation estimates may drift, while strong smoothness regularization can oversmooth boundaries and suppress localized motion near lesions. We propose an edge-guided flow fusion method that combines two deformation fields predicted by identical registration architectures trained with different smoothness weights. For each moving--fixed PET pair, we compute an edge probability map from the gradient magnitude of the Gaussian-smoothed absolute intensity difference, followed by percentile normalization and logistic mapping. After thresholding and morphological dilation, the resulting mask defines an edge-dominant region. We fuse the two predicted flows by selecting the low-regularization flow near edges and the high-regularization flow in homogeneous background regions, then warp the moving PET volume using a spatial transformer. This approach preserves deformation flexibility at boundaries while maintaining smooth motion in low-contrast regions, without changing network architecture.

AcknowledgmentThe authors have no disclosures to report.
Keywords: PET, Motion correction, Deep Learning
17:30 Poster panel: 302

Poster Number:
MIC-10-302

Simulation Trained Convolutional Neural Network For Clinical Inter-Crystal Scattering Filtering (#2835)

H. Maa-Hacquoil1, E. De Benetti1, A. Reznik1, 2, O. Bubon1, 2

1 Lakehead University, Department of Physics, Thunder Bay, Ontario, Canada
2 Radialis Inc., Thunder Bay, Ontario, Canada

Abstract

PET detectors often use a Center of Gravity (CoG) algorithm applied to the output signal of sensor
modules to perform coordinate reconstruction. CoG assumes that the spread of optical photons emitted
by a pixelated scintillator crystal is symmetric about the point of interaction. A photoelectric
(PE) event, where energy is deposited within a single scintillator pixel, produces such a
distribution. Inter-crystal scatter (ICS) events, however, occur when a gamma photon deposits energy
into multiple scintillator pixels, producing multiple photon distributions that degrade CoG
accuracy and spatial resolution. Convolutional neural networks (CNNs) have proven effective
for identifying and recovering ICS events in simulated data and in real systems when trained on
experimentally acquired datasets, typically obtained using collimated measurements to isolate
scintillator pixels. However, acquiring collimated datasets is time-consuming and requires high
precision, motivating the use of simulated data as a primary training source. In this work, PET
data was generated using GATE and used to train a small CNN to identify and remove ICS events
based on softmax confidence values for a detector consisting of a 24×24 LYSO array coupled to an 8×8
SiPM array. A NU-4 image quality phantom and Na-22 point source acquired on a clinical system
demonstrated improvements in recovery coefficient (RC) and peak-to-valley ratio for all RC five rod sizes, along
with modest improvements in spatial resolution and spill-over ratio. The results demonstrate the feasibility of using simulated
data as a primary training source for ICS filtering and recovery. Thus, a hybrid approach incorporating
limited real data may further improve performance without requiring a large number of collimated acquisitions.

AcknowledgmentWe acknowledge the support from the Natural Sciences and Engineering Research of Canada (NSERC). We thank Lakehead
University, the Thunder Bay Regional Health Research Institute, and Radialis Inc.
References
[1] Maa-Hacquoil, H., Poladyan, H., Baldassi, B., Komarov, B., Rado, J., Bubon, O. and Reznik, A., 2025. Improved Edge Pixel Resolution in Modular PET Detectors with Partly Segmented Light Guides. Sensors, 25(22), p.7062.
[2] Zhang, C., Sang, Z., Wang, X., Zhang, X. and Yang, Y., 2019. The effects of inter-crystal scattering events on the performance of PET detectors. Physics in Medicine & Biology, 64(20), p.205004.
[3] Lee, S. and Lee, J.S., 2021. Inter-crystal scattering recovery of light-sharing PET detectors using convolutional neural networks. Physics in Medicine & Biology, 66(18), p.185004.
[4] Park, H. and Lee, J.S., 2020. SiPM signal readout for inter-crystal scatter event identification in PET detectors. Physics in Medicine & Biology, 65(20), p.205010.
[5] Lee, S. and Lee, J.S., 2023. Experimental evaluation of convolutional neural network-based inter-crystal scattering recovery for high-resolution PET detectors. Physics in Medicine & Biology, 68(9), p.095017.
[6] Stiles, J., Baldassi, B., Bubon, O., Poladyan, H., Freitas, V., Scaranelo, A., Mulligan, A.M., Waterston, M. and Reznik, A., 2022. Evaluation of a high-sensitivity organ-targeted PET camera. Sensors, 22(13), p.4678.
[7] Jan, S., Santin, G., Strul, D., Staelens, S., Assié, K., Autret, D., Avner, S., Barbier, R., Bardiès, M., Bloomfield, P.M. and Brasse, D., 2004. GATE: a simulation toolkit for PET and SPECT. Physics in Medicine & Biology, 49(19), pp.4543-4561.
Keywords: PET, convolutional neural network, inter crystal scattering, coordinate reconstruction, planar PET
17:30 Poster panel: 305

Poster Number:
MIC-10-305

Temporally and Spatially-Variant Penalty for Dynamic PET Reconstruction (#2858)

N. Taheri1, T. Carlier2, S. Stute2

1 CRCI2NA, INSERM UMR 1307, Nantes University, Nantes, France
2 Nuclear Medicine Department, Nantes University Hospital, Nantes, France

Abstract

Dynamic PET reconstruction is challenging due to strong temporal variations in count statistics, leading to non-uniform noise and spatial resolution across frames. In this work, we simply extend previous works to propose a temporally and spatially-variant regularization strategy for dynamic penalized maximum-likelihood (PML) reconstruction based on a time-and-space-dependent modulation factor derived from the Fisher information matrix. A modified estimation of the modulation factor is used to improve robustness in low-count conditions by removing dependence on noisy measurements. The method was evaluated using dynamic FDG simulations based on the 2D Zubal brain phantom and compared with OSEM and PML reconstruction without modulation.
Results showed improved preservation of anatomical structures and lower RMSE values in the cortex, putamen, and caudate regions while avoiding the excessive smoothing observed with conventional PML reconstruction. These results suggest that automatic temporally and spatially adaptive regularization can improve image quality and quantitative accuracy in dynamic PET imaging.

References
[1] J. N. et al., IEEE TNS, vol. 49, no. 1, pp. 56–60, 2002.
[2] S. A. et al., PMB, vol. 60, pp. 5733–51, 2015.
[3] Y. T. et al., IEEE TMI, vol. 39, no. 1, pp. 11–22, 2020.
[4] J. A. Fessler, IEEE TMI, vol. 13, no. 2, pp. 290–300, 1994.
[5] S. Stute et al., IEEE NSS/MIC Conf. Rec., pp. 1–3, 2015.
Keywords: Dynamic PET, Penalized Likelihood Reconstruction, Spatially-Variant Regularization
17:30 Poster panel: 308

Poster Number:
MIC-10-308

SAR-Based Brain Aging Validation Using Functional Connectivity Correlation Matrices from UK Biobank (#3186)

A. Hernández1, 3, I. A. Illán1, 3, J. Ramírez1, 3, F. Segovia1, 3, D. Salas-González1, 3, C. Jiménez-Mesa2, 3, C. Vázquez-García1, 3, F. J. Martínez-Murcia1, 3, J. Levin4, J. M. Górriz1, 3

1 University of Granada, Department of Signal Theory, Networking and Communications, Granada, Spain
2 University of Malaga, Department of Communications Engineering, Málaga, Spain
3 Andalusian Institute of Data Science and Computational Intelligence (DaSCI ), Granada, Spain
4 Ludwig-Maximilians-University, Department of Neurology, Munich, Germany

Abstract

This study applies the Statistical Agnostic Regression (SAR) methodology to validate predictive model for brain aging using gray matter functional connectivity derived from fMRI data from the UK Biobank database. The method evaluates significant linear relationships between age and functional connectivity correlation matrices, by resampling and estimating an upper bound on the expected risk through a Bayesian threshold under the worst-case scenario. The SAR power quantifies the probability of detecting a true linear relationship using the expected risk statistic, via Monte Carlo simulations under the null distribution. Predictive performance is assessed through Principal Component Analysis (PCA) applied to the global correlation matrices, which are constructed from subjects whose ages fall within the extreme percentiles (below the 15th and above the 85th percentile). This framework enables the identification of connectivity signatures associated with aging within a unified representation of functional brain connectivity dynamics. Functional connectivity matrices of gray matter were computed from preprocessed fMRI data by extracting region-wise time series and estimating pairwise statistical dependencies using Pearson correlation coefficients. The results support the development of a statistically significant model for predicting age from functional connectivity correlation matrices and demonstrate that combining rigorously validated predictive models with SAR and PCA methodologies can characterize age-related variability in functional connectivity and improve understanding of brain aging using large-scale population neuroimaging data.

AcknowledgmentThis work is part of the projects PID2022-137451OB-I00 and PID2022-137629OA-I00 funded by MICIU/AEI/10.13039/501100011033 and by ERDF/EU, and the C-ING-183-UGR23 project, cofunded by the Consejería de Universidad, Investigación e Innovación and by European Union, funded by Programa FEDER Andalucía 2021-2027.
References
[1] Varoquaux, G.,  2018, ‘Cross- Validation failure: Small sample sizes lead to large error bars’, NeuroImage, 180, 68-77.
[2] LeCun, Y., 2015, ‘Deep learning’, Nature, 521, 436–444.
[3] Leming, M., Górriz, J. M., Suckling, J., 2020, ‘Ensemble Deep Learning on Large, Mixed-Site fMRI Datasets in Autism and Other Tasks’, International Journal of Neural Systems, 30(7), 2050012.
[4] Górriz, J. M., et al.,  2026, ‘Statistical Agnostic Regression: A machine learning method to validate regression models’, Journal of Advanced Research, 80, 503-533.
[5] Pan, Y., et al., 2024, ‘Brain-wide functional connectome analysis of 40,000 individuals reveals brain networks that show aging effects in older adults’, Imaging Neuroscience, 2, imag–2–00394.
Keywords: SAR Validation, Machine Learning, Neuroimaging, Brain Aging, UK Biobank
17:30 Poster panel: 311

Poster Number:
MIC-10-311

Interpreting Prognostic [18F]FDG PET Radiomic Signatures Using Parametric Tumour Simulation (#1849)

L. M. Bonney1, 2, K. M. Bradley3, A. B. Hassan1, D. R. McGowan2, 4

1 University of Oxford, Sir William Dunn School of Pathology, Oxford, United Kingdom
2 Oxford University Hospitals NHS Foundation Trust, Department of Medical Physics and Clinical Engineering, Oxford, United Kingdom
3 Cardiff University, Wales Research and Diagnostic Positron Emission Tomography Imaging Centre (PETIC), Cardiff, United Kingdom
4 University of Oxford, Department of Oncology, Oxford, United Kingdom

Abstract

Radiomic features derived from [18F]FDG PET/CT imaging can quantify intratumoural spatial heterogeneity and have demonstrated prognostic utility across multiple oncological indications, including sarcoma. However, the biological interpretation of radiomic risk scores remains poorly understood, partly due to confounding in clinical datasets and the absence of ground-truth tumour parameters. A simulation-based approach offers a controlled framework for mapping image-derived features to biologically and physically meaningful tumour properties.

In this study, a parametric tumour model incorporating eight biologically motivated parameters—including uptake intensity, radial signal gradients, compartmental structure, multifocal uptake, and morphological deformation—was used to generate 1,024 synthetic tumour volumes across physiologically plausible parameter ranges. In parallel, a prognostic radiomic risk score was derived from [18F]FDG PET/CT images of 115 sarcoma patients enrolled in the Oxford Precision Oncology for Sarcoma (OxPOS) study. This trained model was subsequently applied to each simulated volume to infer a radiomic risk score, and correlation analysis identified simulation parameters most strongly associated with predicted risk.

Three parameters; baseline uptake intensity, radial signal gradient, and fractional high-uptake volume, exhibited moderate but statistically significant correlations with the model-derived risk score (|r|= 0.41, 0.38, and 0.52, respectively). These findings provide mechanistic insight consistent with known tumour biology; for example, elevated radial signal gradients are characteristic of tumours with necrotic cores.

This work demonstrates the utility of simulation for enhancing biological interpretability and communicability of radiomic prognostic models. Ongoing work aims to expand the simulation framework and integrate genomic data to further elucidate the biological underpinnings of radiomic signatures in sarcoma.

References
[1] P. Lambin et al., Radiomics: extracting more information from medical images using advanced feature analysis, Eur. J. Cancer, vol. 48, no. 4, pp. 441--446, 2012.
[2] A. Zwanenburg et al., The Image Biomarker Standardization Initiative: standardized quantitative radiomics for high-throughput image-based phenotyping, Radiology, vol. 295, no. 2, pp. 328--338, 2020.
[3] J. J. M. van Griethuysen et al., Computational radiomics system to decode the radiographic phenotype, Cancer Research, vol. 77, no. 21, pp. e104--e107, 2017.
[4] R. Boellaard et al., 18F-FDG PET/CT: EANM procedure guidelines for tumour imaging: version 3.0, The EANM Journal, vol. 1, 2025.
 
Keywords: PET/CT, Radiomics, Interpretability, Simulation
17:30 Poster panel: 314

Poster Number:
MIC-10-314

Parameter Estimation of Arterial Input Function Using Nonlinear Mixed-Effects Modeling With Reduced Sampling (#2685)

J. Jiang2, Z. Jiang1, J. Chen1, Y. Du1

1 Johns Hopkins University School of Medicine, Department of Radiology, Baltimore, Maryland, United States of America
2 Johns Hopkins University Whiting School of Engineering, Department of Applied Mathematics and Statistics, Baltimore, Maryland, United States of America

Abstract

Dynamic positron emission tomography (PET) requires dense arterial blood sampling to estimate arterial input functions (AIFs), creating substantial clinical burden and limiting practical implementation. To reduce number of samples, we developed a data-driven method for AIF estimation under sparse sampling conditions using nonlinear mixed-effects modeling. The study dataset included 90-min dynamic brain PET AIF data from 26 healthy subjects acquired with the 11C-DPA-713. A three-exponential nonlinear mixed-effects model was used to characterize the AIF, with fixed effects representing population-level kinetics and random effects capturing inter-individual variability. The AIF was modeled using a deterministic five-point subsampling strategy that selected representative early, intermediate, and late decay-phase observations. Model parameters were estimated using the stochastic approximation expectation maximization algorithm, and reduced-sampling predictions were compared with reference ground truth generated from the full dataset. The reduced five-point strategy produced AIF estimates that closely matched the full-data reference, yielding a root mean squared error of 0.37 and a mean relative error of ranging from 19% to 27.8%. Visual comparisons further demonstrated strong agreement between predicted and reference AIFs across subjects, indicating that plasma activity characteristics were preserved despite reduction in sampling density. These findings suggest that sparse sampling combined with nonlinear mixed-effects modeling can provide reliable AIF estimation while reducing the burden of arterial blood collection in dynamic PET studies.

References
[1] Zanotti-Fregonara P, Chen K, Liow JS, Fujita M, Innis RB, “Image-derived input function for brain PET studies: many challenges and few opportunities,” J. Cereb. Blood Flow Metab., vol. 31, no. 10, pp. 1986–1998, 2011, doi: 10.1038/jcbfm.2011.107.
[2] Xiu Z, Muzi M, Huang J, Wolsztynski E, “Patient-adaptive population-based modeling of arterial input functions,” IEEE Trans. Med. Imaging, vol. 42, no. 6, pp. 1698–1709, 2023, doi:10.1109/TMI.2023.3248454.
[3] Comets E, Lavenu A, Lavielle M, “Parameter estimation in nonlinear mixed effect models using saemix, an R implementation of the SAEM algorithm,” J. Stat. Softw., vol. 80, no. 3, pp. 1–41, 2017, doi: 10.18637/jss.v080.i03.
Keywords: arterial input function, dynamic positron emission tomography, nonlinear mixed-effects model, sparse sampling, pharmacokinetic modeling
17:30 Poster panel: 317

Poster Number:
MIC-10-317

Orbital MRI radiomics in thyroid eye disease: A multi-structure stability analysis using perturbation-based methods (#3226)

F. Semnani1, S.S. Aarabi1, A. Jafarizadeh2, M. Abbasi3, A. Doostparast4, M. Abdollahi2, R. Alizadehsani5, J. M. Gorriz6, 7

1 School of Medicine, Tehran University of Medical Sciences (TUMS), Tehran, Iran (Islamic Republic of)
2 Nikookari Eye Center, Tabriz University of Medical School, Tabriz, Iran (Islamic Republic of)
3 School of Medicine, Hormozgan University of Medical Sciences (HUMS), Bandar Abbas, Iran (Islamic Republic of)
4 Eye Research Center, Mashhad University of Medical Sciences, Mashhad, Iran (Islamic Republic of)
5 School of medicine, Queensland University of Technology, Brisbane, Australia
6 Data Science and Computational Intelligence Institute, Universidad de Granada,, Granada, Spain
7 Department of Psychiatry, University of Cambridge, Cambridge, United Kingdom

Abstract

Thyroid eye disease (TED), the most common autoimmune orbital disorder and a major cause of orbital morbidity, is still assessed largely through subjective clinical measures. Although orbital MRI radiomics offers potential quantitative biomarkers for disease activity and treatment response, the reproducibility of these features under real-world imaging variability has not been systematically evaluated, limiting clinical translation. We performed a perturbation-based stability analysis using 150 patients from the public TOM500 dataset, extracting 293 radiomic features from seven Image Biomarker Standardization Initiative (IBSI)-compliant families across nine bilateral orbital structures using original and Laplacian-of-Gaussian filtered MRI images. We used ten perturbations per patient with combined rigid transformations, Gaussian noise, and 1-voxel mask modifications, and assessed stability using ICC(A,1). Across 2,637 structure-feature combinations, mean ICC was 0.72, and 53% of features achieved Good or Excellent stability (ICC ≥ 0.75). The medial rectus (ICC 0.84), followed by superior rectus and optic nerve (both 0.81) were the most robust structures, whereas the eyeball (0.55) and lateral rectus (0.61) were least stable. Laplacian-of-Gaussian filtering improved texture feature reproducibility, increasing median texture ICC from 0.73 to 0.80 and enhancing stability in medial rectus, lacrimal gland, and orbital fat features. Stability rankings were nearly identical between active and inactive disease groups (Pearson r = 0.92), indicating that disease severity did not influence feature reliability. These findings provide the first large-scale reproducibility assessment of orbital MRI radiomics in TED and establish a foundation for future biomarker and predictive modeling studies.

AcknowledgmentAs a declaration of interests, J.M.G serves as Deputy General Chair of the 2026 IEEE NSS MIC RTSD.
References
[1] Bartalena, L, et al., 2021, 'The 2021 European Group on Graves' orbitopathy (EUGOGO) clinical practice guidelines for the medical management of Graves' orbitopathy'. European journal of endocrinology, 185(4), G43–G67. https://doi.org/10.1530/EJE-21-0479
[2] Mourits, MP, Prummel, MF, Wiersinga, WM, and Koornneef, L, 1997, 'Clinical activity score as a guide in the management of patients with Graves' ophthalmopathy'. Clinical endocrinology, 47(1), 9–14. https://doi.org/10.1046/j.1365-2265.1997.2331047.x
[3] Hu, H, et al., 2022, 'T2 -Weighted MR Imaging-Derived Radiomics for Pretreatment Determination of Therapeutic Response to Glucocorticoid in Patients With Thyroid-Associated Ophthalmopathy: Comparison With Semiquantitative Evaluation'. Journal of magnetic resonance imaging : JMRI, 56(3), 862–872. https://doi.org/10.1002/jmri.28088
[4] Zhang, H., et al., 2025, 'TOM500: A Multi-Organ Annotated Orbital MRI Dataset for Thyroid Eye Disease'. Scientific data, 12(1), 60. https://doi.org/10.1038/s41597-025-04427-9
[5] van Griethuysen, JJM, et al., 2017, 'Computational Radiomics System to Decode the Radiographic Phenotype'. Cancer research, 77(21), e104–e107. https://doi.org/10.1158/0008-5472.CAN-17-0339
[6] Zwanenburg, A., et al., 2020, 'The Image Biomarker Standardization Initiative: Standardized Quantitative Radiomics for High-Throughput Image-based Phenotyping'. Radiology, 295(2), 328–338. https://doi.org/10.1148/radiol.2020191145
[7] Koo, TK, and Li, MY, 2016, 'A Guideline of Selecting and Reporting Intraclass Correlation Coefficients for Reliability Research'. Journal of chiropractic medicine, 15(2), 155–163. https://doi.org/10.1016/j.jcm.2016.02.012
[8] Zhang, H, et al., 2024, 'Whole-orbit radiomics: machine learning-based multi- and fused- region radiomics signatures for intravenous glucocorticoid response prediction in thyroid eye disease'. Journal of translational medicine, 22(1), 56. https://doi.org/10.1186/s12967-023-04792-2
Keywords: Thyroid Eye Disease, Radiomics, Magnetic Resonance Imaging, Clinical Activity Score, Perturbation
17:30 Poster panel: 320

Poster Number:
MIC-10-320

Feasibility of portal imaging in proton therapy: concept and phantom experiments (#1070)

J. N. Esser1, 4, C. M. Bäcker1, A. Hammi2, B. Timmermann1, 3, J. Weingarten2, J. Wulff2, C. Bäumer1, 2

1 West German Proton Therapy Centre Essen, Medical Physics, Essen, Germany
2 TU Dortmund University, Department of Physics, Dortmund, Germany
3 University Medicine of Essen, Department of Particle Therapy, Essen, Germany
4 University of Duisburg Essen, Faculty of Medicine, Essen, Germany

Abstract

The aim of the study was to investigate whether proton therapy fields can be verified in vivo during treatment using film-based transit portal dosimetry. Potential clinical treatment plans were checked for downstream dose contributions. The complete proton portal imaging workflow was verified in experiments with an anthropomorphic phantom.   Head-and-neck, mediastinal, thoracic, abdominal and cranio-spinal axis irradiations (CSIs) yielded downstream dose areas, which enable portal imaging. The cranial fields of CSI plans were deemed as most suitable and tested with the phantom in a simplified setup. Radiochromic films, mounted on top of the patient couch, served as portal imaging detectors. Phantom tests indicate that the patient setup can be verified within an accuracy of ≈1 mm per direction. In conclusion, a novel, straight-forward procedure has been developed for in-vivo dosimetry in proton therapy for some patients with specific indications.

AcknowledgmentThis work was supported by the MERCUR graduate school “Präzisionsprotonentherapie — Praxisbezogene Physik und Chemie an der Schnittstelle zur Medizin” (grant-number St-2019-0007).
References
[1] Dogan, B. J. Mijnheer, K. Padgett, A. Nalichowski, C. Wu, M. J. Nyflot, A. J. Olch, N. Papanikolaou, J. Shi, S. M. Holmes, et al., Use of electronic portal imaging devices for pre-treatment and in vivo dosimetry patient-specific IMRT and VMAT: Report of AAPM task group 307, Medical physics 50 (8) (2023) e865 (2023).
Keywords: proton therapy, portal imaging
17:30 Poster panel: 323

Poster Number:
MIC-10-323

A Feasibility Study of Patient-Specific Integral Dose Verification Using a 3D-Printed Tumor Model Scintillator for Radiosurgery (#1239)

H.J. Yang1, T.H. Kim2, Y.K. Kim3, H. Kim1, H.-T. Chung4

1 National Cancer Center Korea, Proton Therapy Center, Goyang, Republic of Korea
2 Radexel Inc., Research and Development Team, Seoul, Republic of Korea
3 Hanyang University, Department of Nuclear Engineering, Seoul, Republic of Korea
4 Seoul National University College of Medicine, Department of Neurosurgery, Seoul, Republic of Korea

Abstract

We developed a patient-specific volumetric dosimetry system using a tumor-shaped plastic scintillator derived from patient MR images for integral dose verification in Gamma Knife radiosurgery. Three tumor model scintillators (TMSs) representing a vestibular schwannoma (0.39 cm³) were irradiated using a nine-shot plan prescribing 8 Gy to the 60% isodose line. The integral dose, defined as the energy absorbed over the entire tumor volume, was measured and compared with the value calculated by the Leksell GammaPlan® treatment planning system. Two independent calibration methods were employed: calibration at a secondary standard dosimetry laboratory and Monte Carlo simulation-based calibration using Geant4. The integral dose measured by the simulation-based calibration was (2.00 0.76)% higher than the planned value of 5.01 mJ, falling within the acceptable 3% tolerance for Gamma Knife absolute dosimetry. The laboratory-based calibration yielded an integral dose (5.84 3.56)% lower than the reference value of 4.30 mJ; the larger deviation is attributed to probe reassembly uncertainty and the angular dependence of the calibration coefficient. These results demonstrate that the TMS system can serve as a practical tool for verifying the three-dimensional integral dose calculated by a treatment planning system.

AcknowledgmentThis research was supported by the National Cancer Center Korea (NCC-2410960).
Keywords: 3D-printed scintillator, Integral dose, Radiosurgery dosimetry, Small dosimetry, Volumetric dosimetry
17:30 Poster panel: 326

Poster Number:
MIC-10-326

Attenuation Correction for Targeted Alpha Therapy Monitoring with Compton Camera MACACO III+ (#1402)

Y. Feng1, J. Roser1, J. Werner1, J. Kasprzak1, L. Barrientos2, S. Parashari2, K. Brzezinski2, G. Llosa2, M. Rafecas1

1 University of Luebeck, Institute of Medical Engineering, Luebeck, Schleswig-Holstein, Germany
2 University of Valencia, Institute for Corpuscular Physics, Valencia, Spain

Abstract

Compton cameras are being explored as a promising technique for treatment monitoring of targeted alpha therapy,  by imaging the gamma emissions produced during the decay of the alpha-emitting radionuclides. To improve image accuracy, attenuation correction is essential. Here we analyze the effects achieved by incorporating attenuation correction into the list-mode MLEM reconstruction of MACACO III+ synthetic data using ideal simulation. The results demonstrate that, without attenuation correction, contrast recovery coefficients of 97.2%, 84.7%, 73.5%, 47.8% are achieved for rods with diameters of 6, 5, 4 and 3 mm, respectively, using a total activity of 0.135 MBq of 440 keV gammas, a major gamma peak in the decay of 225Ac and a hot to background ratio 10:1, simulated over 2 hours. Combined energy reconstruction at 218 and 440 keV yields improved image contrast when attenuation correction is applied. This work was carried out within the frame of the Advanced Imaging DEtector for targeted Radionuclide therapy (AIDER) project.

AcknowledgmentThe authors acknowledge the Advanced Imaging DEtector for targeted Radionuclide therapy (AIDER) programme (Grant agreement ID: 101165088), funded under the Euratom Research and Training Programme (EURATOM).
References
[1] JANG, Albert, et al. Targeted alpha-particle therapy: a review of current trials. International journal of molecular sciences, 2023, 24. Jg., Nr. 14, S. 11626.
[2] BENABDALLAH, Nadia, et al. Practical considerations for quantitative clinical SPECT/CT imaging of alpha particle emitting radioisotopes. Theranostics, 2021, 11. Jg., Nr. 20, S. 9721.
[3] FENG, Yuemeng, et al. Evaluation of a Compton Camera System for Whole-body Imaging of Ac-225: A Simulation Study. IEEE Transactions on Radiation and Plasma Medical Sciences, 2026.
[4] ROSER, Jorge, et al. Radiopharmaceutical imaging of phantoms and patients with MACACO III Compton camera. Physica Medica, 2025, 132. Jg., S. 104928.
[5] BARRIENTOS, L., et al. 131 I imaging with MACACO Compton camera. In: 2024 IEEE Nuclear Science Symposium (NSS), Medical Imaging Conference (MIC) and Room Temperature Semiconductor Detector Conference (RTSD). IEEE, 2024. S. 1-1.
[6] REN, Yuan, et al. Compton-camera-based radiopharmaceutical imaging with an attenuation-corrected LM-MLEM reconstruction strategy. Radiation Measurements, 2025, 181. Jg., S. 107379.
[7] MUNOZ, Enrique, et al. Study and comparison of different sensitivity models for a two-plane Compton camera. Physics in Medicine & Biology, 2018, 63. Jg., Nr. 13, S. 135004.
[8] MUNOZ, Enrique, et al. Imaging of polychromatic sources through Compton spectral reconstruction. Physics in Medicine & Biology, 2022, 67. Jg., Nr. 19, S. 195017.
Keywords: Compton camera imaging, attenuation correction, spectral reconstuction, Monte Carlo simulation, targeted alpha therapy
17:30 Poster panel: 329

Poster Number:
MIC-10-329

Monte Carlo-Based PET Range Verification for Carbon-Ion Therapy (#1456)

R. Liu1, D. Mu1, J. Zhu1, A. Qiu1, T. Hu3, W. Li4, Y. Nitta4, J. Liu4, L. Wan2, P. Xiao1, J. Zhu4, Q. Xie1, 3

1 Huazhong University of Science and Technology, Department of Biomedical Engineering, Wuhan, China
2 Huazhong University of Science and Technology, School of Software Engineering, Wuhan, China
3 University of Science and Technology of China, Department of Electronic Engineering and Information Science, Hefei, China
4 Zhejiang Cancer Hospital, Hangzhou Institute of Medicine (HIM), Chinese Academy of Sciences, Department of Radiation Oncology, Hangzhou, China

Abstract

Positron emission tomography (PET) is an effective non-invasive modality for in vivo range verification in carbon-ion therapy. This study evaluates an all-digital PET system for in-beam monitoring of carbon-ion treatment plans. Clinical DICOM RT plans were converted into a GATE/Geant4 model of single-field pencil beam scanning (PBS) carbon-ion irradiation in a polymethylmethacrylate (PMMA) phantom. Simulated dose and positron-emitter distributions were compared with treatment planning system (TPS) dose and measured PET data. TPS and GATE showed good dosimetric agreement, with nearly identical planning target volume (PTV) D50% and a 3D gamma passing rate of 97.66% (2%/2 mm). PET validation yielded R50,sim = 124.33 mm and R50,exp = 125.00 mm, corresponding to ΔR50 = 0.67 mm. These results support millimeter-level PET-based range verification for carbon-ion therapy.

AcknowledgmentThis work was supported by the National Natural Science Foundation of China (Grant No. 625B2079).
References
[1] W. Enghardt, P. Crespo, F. Fiedler, R. Hinz, K. Parodi, J. Pawelke, and F. Pönisch, "Charged hadron tumour therapy monitoring by means of PET," Nucl. Instrum. Methods Phys. Res. A, vol. 525, no. 1, pp. 284–288, Jun. 2004. DOI: 10.1016/j.nima.2004.03.128
[2] T. Hu, J. Hu, X. Zhang, S. Zheng, T. Zhou, Q. Liu, W. Hu, A. Qiu, Y. Xia, B. Li, P. Xiao, and Q. Xie, "Development and Performance Evaluation of an Open-Ring PET Prototype for Intraoperative Metabolic Monitoring," IEEE Trans. Radiat. Plasma Med. Sci., vol. PP, pp. 1–1, Jan. 2026. DOI: 10.1109/TRPMS.2026.3676872
[3] L. Grevillot, D. J. Boersma, H. Fuchs, A. Aitkenhead, A. Elia, M. Bolsa, C. Winterhalter, M. Vidal, S. Jan, U. Pietrzyk, L. Maigne, and D. Sarrut, "Technical Note: GATE-RTion: a GATE/Geant4 release for clinical applications in scanned ion beam therapy," Med. Phys., vol. 47, no. 8, pp. 3675–3681, 2020. DOI: 10.1002/mp.14242
Keywords: All-Digital PET, GATE/Geant4, In-beam PET, Carbon-Ion Therapy
17:30 Poster panel: 332

Poster Number:
MIC-10-332

Performance of a state-of-the-art ToF-PET detector in a proton therapy in-beam radiation environment (#1485)

T. Fogg1, P. Dendooven1

1 University Medical Center Groningen, Particle Therapy Research Center & Department of Radiation Oncology, Groningen, Netherlands

Abstract

In-vivo positron emission tomography (PET) imaging of positron emitters produced by the therapeutic proton beam in the patient is a well-studied irradiation verification tool. Adaptive treatment planning benefits from fast feedback, down to the sub-second time scale, requiring an in-beam PET scanner operating during beam delivery for N-12 imaging. Use in clinical routine requires that the performance of such detectors is not significantly affected by radiation exposure over time. We have studied the performance of four state-of-the-art time-of-flight Hamamatsu C13500-4075YC-12 (ToF)-PET detectors in a proton therapy radiation environment. A 150 MeV proton beam (pulsed with intensities of 0.3, 1 and 3 nA at 10% and 1.5% duty cycles) was stopped in PMMA phantoms whilst PET data was acquired. The coincidence energy spectra obtained in-beam (both with the beam on and between beam pulses) show the N-12 511 keV peak. As the experiment progressed, the data output shows cumulative faults from an increasing percentage of corrupt events. Whilst the accumulation of these faults was partially mitigated by system resets during the experiment, comparison of data acquisitions with a combined Na-22 & Ga-68 source before and after the experiment show that the performance degradation is long-lasting, although the detectors show signs of recovery 5 months later.

References
[1] A.-C. Knopf and A. Lomax, “In vivo proton range verification: a review,” Phys. Med. Biol., vol. 58, no. 15, p. R131, Jul. 2013. [Online]. Available: https://doi.org/10.1088/0031-9155/58/15/R131
[2] B. Zapien-Campos, Z. Ahmadi Ganjeh, G. Perotti-Bernardini, J. Free, S. Both, and P. Dendooven, “Instantaneous in vivo distal edge verification in intensity-modulated proton therapy by means of PET imaging,” Med Phys, vol. 52, no. 7, p. e17850, Jul. 2025. [Online]. Available: https://pmc.ncbi.nlm.nih.gov/articles/PMC12257898/
Keywords: Positron Emission Tomography, Proton Therapy, Range verification
17:30 Poster panel: 335

Poster Number:
MIC-10-335

In-Beam All-Digital PET for Dose-Guided Particle Therapy: Vision, Roadmap, and Progress (#1518)

A. Qiu1, Q. Xie1, 2

1 Huazhong University of Science and Technology, Department of Biomedical Engineering, Wuhan, China
2 University of Science and Technology of China, Department of Electronic Engineering and Information Science, Hefei, China

Abstract

Proton and heavy-ion therapy exploits the Bragg peak to concentrate dose within the tumor while sparing surrounding normal tissues; however, range uncertainties necessitate conservative safety margins and restrict clinical indications. To address this, we are developing an in-beam All-Digital positron emission tomography (PET) platform for dose-guided particle therapy. Two prototype systems have been designed, built, and tested at multiple proton and heavy-ion therapy centers: a 12-module pre-clinical partial-ring system for phantom and small-animal studies, and a 16-module research-clinical open-ring system for human-subject trials. Both employ lutetium-yttrium oxyorthosilicate scintillators read out via the multi-voltage threshold digitization architecture. In-beam PET imaging of proton-induced positron activity has been demonstrated in polymethyl methacrylate phantoms with approximately 1 mm range agreement, and in live rats where biological washout effects on image contrast were observed. A three-phase development roadmap is presented, progressing from online range monitoring to adaptive therapy and ultimately real-time feedforward therapy, aiming to narrow safety margins, broaden treatment indications, and provide the verification infrastructure for FLASH therapy.

AcknowledgmentThis work was supported by the National Natural Science Foundation of China under Grant 625B2079.
References
[1] Parodi, K, Yamaya, T, Moskal, P 2023, 'Experience and new prospects of PET imaging for ion beam therapy monitoring', Zeitschrift für Medizinische Physik, vol. 33, no. 1, pp. 22–34, Munich: Elsevier
[2] Gao, M, Chen, H-H, Chen, F-H, Hong, J-H, Hsiao, I-T, Yen, T-C, Mao, J, Lu, JJ, Wang, W, D'Ascenzo, N, Xie, Q 2021, 'First Results From All-Digital PET Dual Heads for In-Beam Beam-On Proton Therapy Monitoring', IEEE Transactions on Radiation and Plasma Medical Sciences, vol. 5, no. 6, pp. 775–782, New York: IEEE
[3] Ma, Q, Mu, D, Zhang, R, Liu, Z, Wan, L, Liu, Y, Qiu, A, Yang, Z, Xie, Q 2025, 'Development and evaluation of an in-beam PET system for proton therapy monitoring', Physics in Medicine & Biology, vol. 70, no. 2, 025019, Bristol: IOP Publishing
[4] Fang, L, Zhang, B, Li, B, Zhang, X, Zhou, X, Yang, J, Li, A, Shi, X, Liu, Y, Kreissl, M et al. 2024, 'Development and evaluation of a new high-TOF-resolution all-digital brain PET system', Physics in Medicine & Biology, vol. 69, no. 2, 025019, Bristol: IOP Publishing
[5] Zhang, B, Li, B, Fang, L, Zhou, X, Li, A, Zhang, X, Liu, Y, Wang, Z, Kao, C-M, Liu, Y et al. 2025, 'Performance Evaluation of New PET/CT DigitMI 930', IEEE Transactions on Radiation and Plasma Medical Sciences, vol. 9, no. 5, pp. 578–585, New York: IEEE
[6] Hu, T, Hu, J, Zhang, X, Zheng, S, Zhou, T, Liu, Q, Hu, W, Qiu, A, Xia, Y, Li, B et al. 2026, 'Development and Performance Evaluation of an Open-Ring PET Prototype for Intraoperative Metabolic Monitoring', IEEE Transactions on Radiation and Plasma Medical Sciences, New York: IEEE
Keywords: Proton therapy, heavy-ion therapy, range monitoring, in-beam PET, All-Digital PET
17:30 Poster panel: 338

Poster Number:
MIC-10-338

Feasibility Study of In-Beam Range Verification Using Pixelated CZT Detectors in Carbon Ion Therapy (#1646)

D. Guo1, J. Shao1, X. Cai1, C. Zhang1, Y. Yin1

1 Lanzhou University, LanZhou, China

Abstract

To overcome the limitations of existing beam range verification methods in terms of detection efficiency, system complexity, and real-time capability, this study proposes a carbon ion beam range verification method based on a pixelated cadmium zinc telluride (CZT) detector and machine learning. The pixelated CZT detector was used to measure the secondary particle response generated during carbon ion irradiation of a polymethyl methacrylate (PMMA) target, and a machine learning regression model was then applied to predict the beam range. The experimental results demonstrate that the pixelated CZT detector can accurately measure secondary particle responses. In addition, the effects of different detection regions, defined by pixel ranges, were investigated at a monitor unit (MU) setting of 6 × 10⁶. Using data from a 16 × 16 pixel region, the proposed method achieved a mean absolute error (MAE) of 0.79 mm. These results indicate that the proposed approach provides a feasible technical strategy for online range verification in carbon ion beam therapy

References
[1] Wang S, Guo J H, Zhang Y, et al. High-resolution pixelated CdZnTe detector prototype system for solar hard X-ray imager. Nuclear Science and Techniques, 2019, 30(3): 42. https://doi.org/10.1007/s41365-019-0571-9
[2] Hui Zou, Trevor Hastie, Regularization and Variable Selection Via the Elastic Net, Journal of the Royal Statistical Society Series B: Statistical Methodology, Volume 67, Issue 2, April 2005, Pages 301–320, https://doi.org/10.1111/j.1467-9868.2005.00527.x
Keywords: Carbon ion therapy, Range verification, Pixelated CZT detector, Machine learning, Secondary particle monitoring
17:30 Poster panel: 341

Poster Number:
MIC-10-341

Experimental setup for radiobiological studies with laser-accelerated protons at high dose rates (#1921)

L. Velasco Peñarrubia1, L. Martín Fernández1, J. Juan Morales1, D. de Luis Blanco2, J.L. Henares González2, J.I. Apiñaniz Aginako2, M. Seimetz1

1 Instituto de Instrumentación para Imagen Molecular (i3M), CSIC-Universitat Politècnica de València, Valencia, Spain
2 Centro de Láseres Pulsados (CLPU, Salamanca, Spain

Abstract

Laser-driven sources of ionizing radiation provide unique possibilities for studying ground-breaking phenomena and for exploring novel applications of compact accelerators. An outstanding example is the controlled irradiation of biological samples with laser-accelerated, high-energy particles. After a pilot experiment at the Spanish Pulsed Laser Center (CLPU), the aim of the present study is an increase of the single-shot dose and the mean dose rate on monolayer cell cultures. Two possible scenarios at the same laboratory have been considered. On the one hand, the previous setup at the VEGA-3 laser line may be upgraded through a set of permanent quadrupole magnets to focalize the proton beam closely behind the laser target. On the other hand, a similar, more compact arrangement inside the VEGA-2 target vessel benefits from shorter flight paths and a 10-fold higher repetition rate. Particle tracking simulations with realistic beam control elements point towards a 5–15 times increase in proton fluence and single-shot dose.

AcknowledgmentGrant PID2023-148992OB-C21 funded by MICIU/AEI/10.13039/501100011033 and cofunded by the European Union, ERDF
References
[1] Daido, H., Nishiuchi, M., Pirozhkov, A.S. 2012, ‘Review of laser-driven ion sources and their applications’, Reports on Progress in Physics, 75, 056401.
[2] Chaudhary, P. et al. 2021, ‘Radiobiology Experiments With Ultra-high Dose Rate Laser-Driven Protons: Methodology and State-of-the-Art’, Frontiers in Physics, 9, 75.
[3] Reija Vecino, A. 2026, Laser-driven radiation sources for in vitro radiobiological studies, PhD thesis, Universidade de Santiago de Compostela.
[4] Reija, A., Esteban, D. et al. 2024, ‘Design and Implementation of an Energy Selector for Laser-Accelerated Protons. Instruments’, Instruments, 8, 36.
[5] Ehret, M. et al. 2024, ‘Stability and debris-mitigation of a solid tape target delivery system for intense laser-matter interactions towards high-repetition-rate’, Plasma Physics and Controlled Fusion, 66, 045003.
[6] Bin, J. et al. 2012, ‘A laser-driven nanosecond proton source for radiobiological studie’, Applied Physics Letters, 101, 243701.
[7] Pommarel, L. et al. 2017, ‘Spectral and spatial shaping of a laser-produced ion beam for radiation-biology experiments’, Physical Review Accelerators and Beams, 20, 032801.
[8] Salinas Morote, A.M. 2025, Sistema de focalización para protones acelerados por láser, Final degree project, Faculty of Physics, Universitat de València.
Keywords: proton acceleration, pulsed laser, radiobiology, particle beam
17:30 Poster panel: 344

Poster Number:
MIC-10-344

Design and Development of a BGO-based PET Scanner for Dose Verification after Proton Therapy (#1975)

Ó. Pietrzyk1, M. Freire1, M. Gil1, L. F. Vidal1, M. Balcerzyk2, 3, A. J. Gonzalez1

1 Instituto de Instrumentación para Imagen Molecular (I3M), Centro Mixto CSIC—Universitat Politècnica de València, 46022 Valencia, Spain, Valencia, Spain
2 Bioaraba Health Research Institute, Vitoria, Spain
3 IKERBASQUE, Basque Foundation of Science, Bilbao, Spain

Abstract

Proton therapy is an advanced radiotherapy modality whose clinical advantage arises from the highly localized dose deposition produced at the Bragg peak. However, this feature makes the technique highly sensitive to uncertainties in patient positioning, anatomical changes, and beam delivery. Verifying the dose delivered to patients after treatment session is therefore essential to ensure treatment safety and reducing planning margins. During proton irradiation, nuclear interactions generate positron‑emitting isotopes that can be detected with Positron Emission Tomography (PET). In this work, we focus on the detection of 18F, whose long half‑life enables offline PET acquisitions 60–150 min after treatment, when short‑lived isotopes have decayed. Because the expected in‑vivo 18F activity is very low (0.1–1 Bq/mL), we propose a PET system optimized for low‑count imaging, based on BGO scintillation crystals to avoid the intrinsic background of lutetium‑based detectors typically used in commercial scanners. 
The proposed PET prototype (INSPIRE) consists of 16 detector modules arranged in a single ring, each comprising a two‑layer pixelated BGO block (19×19 and 20×20 arrays) coupled to a 12×12 SiPM array, thus enabling Depth of Interaction (DOI) information. The system, which has an inner diameter of 26 cm and an axial length of 5 cm, is intended for brain studies. In this work we present the design of the INSPIRE system and the experimental characterization of the detector modules, including pixel resolvability, layer separation, and energy performance, together with preliminary proton‑irradiation simulations that illustrate the suitability of the 18F signal for offline PET imaging.

Acknowledgment

This work was supported by the Agencia Valenciana de Innovacion under Grant No INNVA1/2024/72. M. Freire was supported by the CIAPOS Grant No. CIAPOS/2023/141 and M. Gil was supported by the ACIF Grant No. GVRTE/2024/5586404.

Keywords: BGO, Dose verification, PET, Proton therapy
17:30 Poster panel: 347

Poster Number:
MIC-10-347

A Software Framework for Real-Time Prompt Gamma Range Verification (#2107)

L. Santini1, 3, K. Urban1, 3, E. S. Bellotti2, B. Foglia4, A. Forgione1, 3, M. Piroddi1, 3, C. Riboldi1, 3, A. Contangelo1, G. Borghi1, 3, M. Carminati1, 3, S. Agosteo2, 3, D. Mazzucconi2, 3, D. Bortot2, 3, K. Parodi4, A. Mereghetti5, C. Fiorini1, 3

1 Politecnico di Milano, Department of Electronics, Information and Bioengineerin, Milano, Italy
2 Politecnico di Milano, Department of Energy, Milano, Italy
3 National Institute for Nuclear Physics, Section of Milan, Milano, Italy
4 Ludwig-Maximilians-Universität Munich, Department for Medical Physics, Garching, Germany
5 Centro Nazionale di Adroterapia Oncologica, Pavia, Italy

Abstract

Real-time monitoring in hadron therapy is essential to account for anatomical shifts and setup uncertainties that may compromise dose delivery. This work evaluates a software framework designed for online range verification via monitoring of the Prompt Gamma (PG) profile, which is know to be correlated to the Bragg peak position. The architecture employs a decoupled simulation pipeline: PG emission is modeled using MC codes, while the detector response is characterized through a dedicated Python-based tool, PGIPY. This simulated new profile, computed spot by spot, is then compared with measurement inside the software. To ensure spot-by-spot clinical correspondence, the software utilizes machine log files to synchronize simulated data with actual delivery sequences.

Experimental validation with an homogenoeus phantom was performed at the CNAO facility (Pavia, Italy). A high degree of correlation was observed between simulated and experimental profiles, confirming that the framework is free from significant systematic modeling errors. These results demonstrate that the proposed pipeline provides a robust, validated baseline for clinical integration, ensuring that any detected profile deviations during treatment are attributable to physical range shifts.

References
[1] M. Pinto, Prompt-gamma imaging in particle therapy. Eur. Phys. J. Plus, vol. 139, p. 10, 2024.
[2] J. Krimmer, D. Dauvergne, J. M. Létang, and É. Testa, Prompt-gamma monitoring in hadrontherapy: A review. Nucl. Instrum. Meth. A, vol. 878, 2018, pp. 58--73.
[3] K. Urban, A New Software Tool for Modeling the Performance of Prompt Gamma 
Cameras for Range Verification in Hadron Therapy.
Presented at the IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), Yokohama, Japan, 2025. [Poster].
[4] K. Souris et al.,  MCsquare — fast multipurpose Monte Carlo simulation framework. Med. Phys., 2016.
[5] B. Foglia et al., Implementation and 3D extension of dose reconstruction strategies from prompt-gamma emissions in proton therapy. Physics in Medicine & Biology, vol. 70, 235026, 2025.
Keywords: Prompt gamma imaging, Software for real time PG monitoring, MC simulations
17:30 Poster panel: 350

Poster Number:
MIC-10-350

Construction and positioning accuracy of a prototype Portable Organ Dosimetry Device (PODD) for RPT (#2213)

W. C. Hunter1, R. S. Miyaoka1

1 University of Washington, Radiology, Seattle, Washington, United States of America

Abstract

Purpose: To enable personalization dosage of 177Lu-based Theranostic radiopharmaceutical therapy (RPT), we are developing a Portable Organ Dosimetry Device (PODD) for at-home use to monitor activity in regions of interest (ROI) such as tumors and organ at risk (OAR). To accurately estimate OAR activity, knowledge of the relative positions of probes and ROIs is required. This can be achieved either by consistency of initially measured locations, or by verifying locations for each measurement. Methods: We have constructed a prototype PODD, and we examine the variation of ROI-location estimates of an NEMA IQ phantom in the PODD reference frame by measuring PODD and ROI features from repeat x-ray CT scan with the phantom removed and replaced in the PODD between measurements. Results: The standard deviations of ROI translational location with the phantom in a supine orientation was 1.0mm axially, 0.2mm horizontally, and 0.1mm vertically. Standard deviations of rotational position were 0.7 degree in tilt and less than 0.1 degrees in roll about the PODD and CT scanner axis. Conclusion: Locating uncertainty of a rigid body phantom within our prototype PODD will not significantly affect the accuracy of time-integrated activity estimates using longitudinal measurements with the phantom removed and replaced in the PODD between measurements. We must repeat these measurements for a realistic deformable phantom or with real subjects.

AcknowledgmentThis work was supported in part by NIH NCI R01CA287205.
Keywords: Theranostic, personalized medicine, dosimetry, PRRT, RPT
17:30 Poster panel: 353

Poster Number:
MIC-10-353

224Ra Experimental and simulation studies with the MACACO III+ Compton camera prototype (#2389)

L. Barrientos1, K. Brzeziński1, D. Rodríguez Almora1, S. Parashari1, M. Strugari2, I. Torres-Espallardo2, G. Llosá1

1 Instituto de Física Corpuscular (IFIC-CSIC/UVEG), Valencia, Spain
2 La Fe Hospital (HUiP), Nuclear Medicine department, Valencia, Spain

Abstract

Compton cameras are showing promise for targeted radionuclide therapy (TRT). The IRIS group of IFIC (Valencia, Spain) has developed a scintillator-based Compton camera prototype for medical imaging applications which is being assessed in TRT. The prototype MACACO III+ consists of two detector planes, with the second having an active area four times larger than the first. Previous tests performed on phantoms and small animals using 131I demonstrated the suitability of the system for radiotracer imaging. The system is being employed to obtain initial experimental data with different radionuclides to validate the simulations carried out within the European project AIDER, which will develop a high performance Compton camera. In this work, experimental measurements with MACACO III+ were carried out at La Fe Hospital (Valencia, Spain) to image a vial filled with 224Ra. GATE simulations were also performed to compare with experimental data. In addition, the prototype MACACO IV, composed of two planes with four detectors each, is under development. Simulation studies also include the configuration of the new prototype to assess influence of detector geometry on the overall system performance.

AcknowledgmentThis work has received funding from the Spanish MICIU / AEI (project MIDAS, PID2022-143246OB-I00) and from the European Union’s EURATOM research and innovation program (AIDER project, G.A. 101061037). S.P. acknowledges the Severo Ochoa CEX2023-001292-S grant, funded by MCIN/AEI/10.13039/501100011033.
References
[1] Tashima, H and Yamaya, T., Phys. Technol., 2022, vol. 15(3). 187-205.
[2] LLosá, G and Rafecas, M., Eur. Phys. J. Plus, 2023, vol 138(3), 214.
[3] Fujieda, K et al., Nucl. Inst. and Meth. A, 2020, 958, 162802.
[4] Roser, J et al., Physica Medica., 2025, vol 132, 104928.
[5] Brzeziński, K et al., 2026, Article in preparation.
[6] Barrientos, L et al., Rad Phys and Chemi., 2023, vol. 208 110922.
[7] Agostinelli, S et al., Nucl. Inst. and Meth. A., 2003, vol. 506 (3):250-303.
[8] Muñoz, E et al., Phys. Med. Biol., 2018, vol. 63(13), 135004.
Keywords: Compton cameras, Targeted radionuclide therapy, 224Ra, Simulations
17:30 Poster panel: 356

Poster Number:
MIC-10-356

AIDER Compton Camera for Targeted Radionuclide Therapy (#2559)

G. Llosá1

1 Instituto de Física Corpuscular (IFIC), CSIC-UV., Valencia, Spain

On behalf of AIDER Collaboration

Abstract

Targeted radionulclide therapy (TRT) is a cancer treatment modality that employs radiopharmaceuticals to selectively deliver radiation to cancerous cells at specific target organs. Its use has increased drastically in the last years, and it has been extended to novel radiopharmaceuticals. The secondary photons emitted by the therapeutic radionuclides can be employed for imaging the compound distribution in the patient’s body for treatment assessment and dosimetry. A precise quantification of the radiation dose delivered to lesions and organs at risk is crucial for improving patient treatment safety and reducing related side effects. Conventional imaging modalities are not optimized for such secondary emission, in particular in the case of alpha-emitting radionuclides, which are gaining prominence due to their exceptional cytotoxicity. For targeted alpha therapy (TAT), image-based dosimetry is an unmet need as currently used gamma cameras are not effective in the case of these high energy and low activity emitters. The European project AIDER aims at developing a high-performance scintillator-based Compton camera to improve imaging in TRT and to enable image-based dosimetry in TAT. The project, coordinated by the IRIS group of IFIC, includes four academic partners, two hospitals, an SME and a patient association. Within the project, the system will be developed and tested with phantoms and patients. Initial work on system design optimization and image reconstruction, as well as detector and electronics development has been carried out.

Acknowledgment

This work has received funding from the European Union’s EURATOM research and innovation program (AIDER project, G.A. 101061037).

References
[1] Ballal et al. 225Ac-PSMA-617-targeted alpha therapy for the treatment of metastatic castration-resistant prostate cancer: A systematic review and meta-
analysis.. The prostate. Volume 81, Issue9 2021. Pages 580-591.
[2] Seo, Y. "Quantitative imaging of alpha-emitting therapeutic radiopharmaceuticals." Nuclear medicine and molecular imaging 53.3 (2019): 182-1s88.
[3] https://aider.ific.uv.es/aider/
[4] Sarrut D, … Krah N. GATE 10 Monte Carlo particle transport simulation: I. Development and new features. Phys Med Biol. 2026 Jan 14;71(1).
[5] L. Barrientos et al. Characterization and 131I tests with MACACO III+ Compton camera. Submitted to TRPMS.
Keywords: Compton camera, targeted radionuclide therapy, targeted alpha therapy, dosimetry, AIDER
17:30 Poster panel: 359

Poster Number:
MIC-10-359

Gamma Detection for Source Localisation in Targeted Alpha Therapy (#2678)

C. De Sio1, L. AbuSabah1, J. Duan1, Y. Shi1, S. Williams1, J. Velthuis1

1 University of Bristol, School of Physics, Bristol, United Kingdom

Abstract

Targeted alpha therapy delivers highly localised dose, so treatment planning and delivery verification depend on whether the radionuclide source remains at the intended deposition site or is redistributed by diffusion or biological transport. This simulation study tests whether decay-chain gamma emissions can provide a detectable source-localisation signal in an AlphaGlue-like geometry. Photon transport was modelled for a thin loaded glue layer on a water phantom surrounded by six detector panels, with each photon entry into a panel counted as one hit. Seven source configurations were analysed: a retained-source baseline, broad redistribution of 10%, 50%, or 90% of the source in water, and localised displacement of 20%, 40%, or 60% of the source to a compact water region. Each configuration used ten repetitions of 10,000 primary decays and was normalised to a 3 μCi reference source.

Total detector-hit yield changed only modestly, but baseline-subtracted spatial hit profiles showed strong source-dependent structure. The central glue-facing signal fell to 0.94, 0.72, and 0.47 of baseline for the broad-redistribution cases, and to 0.84, 0.67, and 0.50 for the localised-displacement cases. Profile-shape features recovered the broad diffusion fraction with an R2 value of 0.998, while a secondary-source metric identified all localised displaced-source cases. The reconstructed displaced-source position was centred at approximately 525–531 mm, compared with the simulated 500 mm source region. Time-window analysis showed that large redistributions reached three-standard-deviation thresholds within 30 minutes to 2 hours; the weakest broad redistribution required about 6 hours, and the 20% localised displacement produced detectable displaced-region excess within about 1 hour. These results indicate that gamma-hit patterns provide source-location information beyond total count rate and could support activity verification and treatment-delivery monitoring.

Keywords: AlphaGlue, theranostics, gamma detection, targeted alpha therapies, Geant4
17:30 Poster panel: 362

Poster Number:
MIC-10-362

Cellular-Scale Imaging of Theranostic Radiopharmaceutical Uptake with a Lensless Radiomicroscopy Sensor (#2746)

N. Belcari1, 2, L. Pesce3, G. Sportelli1, 2, C. Catarinicchia4, L. Menichetti4, D. Panetta4, L. Colagiorgio5, A. Leso6, D. Serafini6, A. Arzenton7, L. Cansolino8, 9, C. Alberghina10, 11, F. P. Cammarata10, 11, C. Bonanno12, 13, M. Ippolito13, F. Scopelliti13, G. Grosso14, A. Gandini14, A. Donzella15, 9, E. Mariotti16, 2, G. S. Valli16, A. Andrighetto6

1 University of Pisa, Department of Physics, Pisa, Italy
2 INFN, Sezione di Pisa, Pisa, Italy
3 University of Pisa, CISUP, Pisa, Italy
4 CNR, IFC, Pisa, Italy
5 Scuola Superiore Sant’Anna, Pisa, Italy
6 INFN, Laboratori Nazionali di Legnaro, Legnaro, Italy
7 University of Padova, Department of Physics and Astronomy, Padova, Italy
8 University of Pavia, Department of Clinical Surgical Sciences, Pavia, Italy
9 INFN, Sezione di Pavia, Pavia, Italy
10 CNR, IBSBC, Cefalù, Italy
11 INFN, Laboratori Nazionali del Sud, Catania, Italy
12 University of Catania, Department of Chemical Sciences, Catania, Italy
13 Cannizzaro Hospital, Catania, Italy
14 LENA, Pavia, Italy
15 University of Brescia, Department of Mechanical and Industrial Engineering, Brescia, Italy
16 University of Siena, Department of Physical Sciences, Earth and Environment, Italy

Abstract

We present a low-cost lensless radiomicroscopy system for cellular-scale imaging of theranostic radiopharmaceutical uptake using a commercial CMOS image sensor. The device enables simultaneous brightfield and β-emission imaging with sub-10 μm spatial resolution, allowing direct correlation between morphology and radionuclide distribution. The system was validated using β⁻-emitting 111Ag and β⁺-emitting 18F-FDG sources, demonstrating quantitative linearity and accurate radioactive decay measurements. Live-cell experiments on MDA-MB-231 breast cancer cells showed detectable intracellular uptake of 111Ag, while imaging of U87 glioblastoma spheroids revealed spatially localized 18F-FDG accumulation in a warm background environment. These results demonstrate the potential of lensless radiomicroscopy as a compact and affordable platform for quantitative in vitro evaluation of theranostic agents at single-cell and multicellular levels.

Acknowledgment

This work is partially supported under the National Recovery and Resilience Plan (NRRP), Mission 4 Component 2 Investment 3.1 - DD MUR 3264/21 - funded by the European Union – NextGenerationEU. Project code IR00023 - Concession Decree No. 101 of 16/06/2022 adopted by MUR, CUP B53C22001810006, Project PNRR-SEE LIFE - Strengthening the Italian Infrastructure of Euro-Bioimaging.

References
[1] Klein JS, Kim TJ, Pratx G. Development of a Lensless Radiomicroscope for Cellular-Resolution Radionuclide Imaging. J Nucl Med. 2023 Mar;64(3):479-484. doi: 10.2967/jnumed.122.264021.
Keywords: Theranostic, Lens-less radiomicroscopy, beta imaging
17:30 Poster panel: 365

Poster Number:
MIC-10-365

Intercrystal Compton Scattering Detection in Pixelated LYSO Detectors via Quantum Localization (#2993)

A. Hashemi1, X. Zhang1, H. Sabet1

1 Harvard Medical School/Massachusetts General Hospital, Radiology Department, Boston, Massachusetts, United States of America

Abstract

We introduce a novel quantum-inspired method for identifying intercrystal Compton scattering in pixelated scintillation detectors. The approach treats each SiPM readout as a spatial signal and converts it into a discrete eigenfunction representation using a Hamiltonian built from the measured intensity distribution. Event structure is then assessed through quantum localization, quantified by the participation ratio. Single-site events produce highly localized dominant modes, whereas Compton-scattered events produce multiple localized modes associated with distinct interaction sites. A dual criterion based on quantum localization and eigenvalue proximity is used to classify multi-site events without calibration-intensive response modeling or supervised training. The method was evaluated with Geant4 optical photon simulations and measurements from LYSO:Ce detector arrays with thicknesses from 2 to 25 mm. Experimental classification agreed closely with simulation, with differences below 1% for most thicknesses and below 5% in all cases. These results indicate that the proposed representation captures the topology of gamma-ray interactions directly from detector signals.

AcknowledgmentThis work was supported in part by NIH grant R01EB034785.
References
[1] M. S. Lee, H. S. Shim, and J. S. Lee, “Strategies for mitigating inter-crystal scattering effects in positron emission tomography: a comprehensive review,” Biomedical Engineering Letters, vol. 14, no. 6, pp. 1243–1258, Nov. 2024, doi: 10.1007/s13534-024-00427-7.
[2] S. Lee, K. Y. Kim, M. S. Lee, and J. S. Lee, “Recovery of inter-detector and inter-crystal scattering in brain PET based on LSO and GAGG crystals,” Phys. Med. Biol., vol. 65, no. 19, p. 195005, Oct. 2020, doi: 10.1088/1361-6560/ab9f5c.
[3] M. N. Ullah and C. S. Levin, “Application of Artificial Intelligence in PET Instrumentation,” PET Clinics, vol. 17, no. 1, pp. 175–182, Jan. 2022, doi: 10.1016/j.cpet.2021.09.011.
[4] A. Hashemi, S. Dutta, B. Georgeot, D. Kouamé, and H. Sabet, “A novel perspective on denoising using quantum localization with application to medical imaging,” AIP Advances, vol. 15, no. 8, p. 085012, Aug. 2025, doi: 10.1063/5.0267924.
[5] A. Hashemi, S. Dutta, B. Georgeot, D. KouamÉ, and H. Sabet, “Quantum-Inspired Denoising via Adaptive Planck Constant and Image Patching,” in 2024 IEEE Nuclear Science Symposium (NSS), Medical Imaging Conference (MIC) and Room Temperature Semiconductor Detector Conference (RTSD), Oct. 2024, pp. 1–1. doi: 10.1109/NSS/MIC/RTSD57108.2024.10657709.
[6] A. Hashemi, H. Babaei, and S. Lee, “Effects of medium range order on propagon thermal conductivity in amorphous silicon,” Journal of Applied Physics, vol. 127, no. 4, p. 045109, Jan. 2020, doi: 10.1063/1.5124821.
Keywords: Intercrystal scattering, detector, quantum physics, Compton, localization.
17:30 Poster panel: 368

Poster Number:
MIC-10-368

A Learned Unfolded Deconvolution Framework for Partial Volume Correction in Total-Body PET Imaging (#3021)

M. S. Azimi1, 2, V. Felfelian3, M. Cheraghi4, A. Alavi5, H. Arabi6, H. Zaidi6, 2

1 Obuda University, Doctoral School of Applied Informatics and Applied Mathematics, Budapest, Hungary
2 Obuda University, University Research and Innovation Center, Budapest, Hungary
3 Shahid Beheshti University, Department of Medical Radiation Engineering, Tehran, Iran (Islamic Republic of)
4 Shahid Beheshti University of Medical Sciences, Department of Nuclear Medicine, Tehran, Iran (Islamic Republic of)
5 University of Pennsylvania, Department of Radiology, Philadelphia, United States of America
6 Geneva University Hospital, Division of Nuclear Medicine & Molecular Imaging, Geneva, Switzerland

Abstract

Total-body PET imaging provides unprecedented sensitivity and extended axial coverage but remains affected by partial volume effects (PVE), particularly in small lesions and regions with spatially varying resolution. Conventional deconvolution approaches, such as Richardson–Lucy (RL), can improve spatial resolution but often suffer from noise amplification and inaccuracies in point spread function (PSF) modeling. In this work, we propose a learned unfolded deconvolution framework for robust PVE correction in total-body PET imaging. The method reformulates RL deconvolution as a fixed-depth network in which each iteration explicitly incorporates the system PSF while learnable parameters enable adaptive regularization and noise suppression. Standard-dose [¹⁸F]FDG PET images were used as network inputs to isolate the PVE correction problem from low-count noise effects. The model was trained in a self-supervised manner using RL-generated surrogate references, eliminating the need for voxel-wise ground-truth images. Evaluation was performed on multicenter total-body PET datasets acquired from uEXPLORER and Biograph Vision Quadra systems. Quantitative assessment on an independent test set demonstrated high structural fidelity (PSNR = 47.38 ± 5.71, SSIM = 0.98 ± 0.01) and low reconstruction error (RMSE = 0.005 ± 0.003), while maintaining acceptable SUV bias (12.2% ± 23.1%). Qualitative analysis showed improved lesion delineation and enhanced uptake contrast without visible noise amplification. Bootstrap-based analysis further demonstrated stable performance with narrow confidence intervals across metrics. The proposed framework provides an interpretable and robust solution for PVE correction, improving image quality and quantitative consistency in total-body PET imaging.

References
[1] M. S. Azimi et al., "Critical review of partial volume correction methods in PET and SPECT imaging: benefits, pitfalls, challenges, and future outlook," European Journal of Nuclear Medicine and Molecular Imaging, pp. 1–32, 2025.
[2] M.-S. Azimi et al., "Deep learning-based partial volume correction in standard and low-dose positron emission tomography-computed tomography imaging," Quantitative Imaging in Medicine and Surgery, vol. 14, no. 3, p. 2146, 2024.
[3] M. Azimi et al., "Attention-based deep neural network for partial volume correction in brain 18F-FDG PET imaging," Physica Medica, vol. 119, p. 103315, 2024.
[4] M. S. Azimi, F. MahdiMaleki, F. MahdiMaleki, M. Cheraghi, A. Alavi, and H. Zaidi, "Supervised vs. Unsupervised Radiomics Reproducibility Across Dose-Level Variations and Partial Volume Correction Methods in Total-Body PET Imaging," Biomedical Signal Processing and Control, In press 2026.
[5] H. Marquis, K. Erlandsson, and I. Buvat, "Addressing partial volume effects in clinical PET quantification: modern correction strategies and challenges," PET clinics, vol. 20, no. 4, pp. 407–422, 2025.
Keywords: Total-body PET, Partial Volume Correction (PVC), Learned Unfolding, Self-Supervised Learning, Image Reconstruction
17:30 Poster panel: 371

Poster Number:
MIC-10-371

Direct Theranostic Dose Estimation using a Multi-Pinhole Model of the Sparse Portable Organ Dosimetry Device (#3220)

G. Maliakal1, R. S. Miyaoka2, W. C. Hunter2, A. M. Alessio1

1 Michigan State University, Institute for Quantitative Health, Biomedical Engineering, CMSE, and Radiology, East Lansing, Michigan, United States of America
2 University of Washington, Radiology, Seattle, Washington, United States of America

Abstract

Many studies have proposed methods to estimate internal dosimetry, enabling patient-specific adjustment of radiopharmaceutical therapy.  We are developing a practical, low-cost multi-pinhole collimated system (PODD) with on the order of 10 sensors at fixed locations around a subject.  Estimation of internal dosimetry to organs at risk is extremely ill-posed with this sparse sensor placement.  In this work, we hypothesize that use of objective functions with an accurate forward model and multiple regularization terms can directly estimate total accumulated activity using simulated PODD acquisitions of realistic distributions of Lu-177 DOTATATE.  Methods: We developed a custom GPU-based forward projector enabling rapid multi-pinhole detection simulation. We propose an objective function for direct estimation of wash-in, wash-out, and total integrated activity (TIA) kinetic parameters from multi-time point, sparse sensor measurements.  This objective function includes regularization based on a) matching sensor measurements to initial time-point SPECT acquisitions and b) promoting kinetic parameters similar to population kinetics. Evaluation was performed by simulating sensor acquisitions from public 4-time point SPECT/CT images (Dewaraja et al, 2021, Lu-177 DOTATATE Anonymized Patient Datasets). Conclusion: Direct estimation of clinically relevant TIA can be performed with on the order of 10 to 20 pinhole-collimated sensors acquiring data at four times post treatment.

AcknowledgmentThis work was supported by NIH Grant 1R01CA287205
Keywords: dosimetry, direct reconstruction, radionuclide therapy, SPECT/CT, Lu177-DOTATATE
17:30 Poster panel: 374

Poster Number:
MIC-10-374

FastPET for low-count reconstruction scenarios (#1498)

V. Nadig1, D. Bharkhada2, M. Millardet3, 2, V. Panin2, J. Schaefferkoetter2, Y. Zheng2, M. Zhang2, N. Murthy2, S. Liu2, P. Schleyer2

1 Siemens Healthcare Limited, London, United Kingdom
2 Siemens Medical Solutions, Knoxville, Tennessee, United States of America
3 University of Pennsylvania, Radiology Department, Philadelphia, Pennsylvania, United States of America

Abstract

Highly sensitive long-axial-field-of-view (LAFOV) positron emission tomography (PET) scanners require fast and reliable reconstruction methods, where accurate quantification in short temporal frames is key to kinetic modeling applications. We extended FastPET - a direct neural network reconstruction method based on image loss - with an additional forwardprojection step and data loss penalty, modified the training database to specifically account for short temporal frames and evaluated its PET quantification accuracy in high- and lowcount regimes on the Biograph Vision Quadra. Compared to ordered subset expectation maximation (OSEM, 4i5s, all pass) reconstructions, FastPET based only on image loss resulted in smoothed images and an overshoot in the contrast recovery coefficient (CRC) by up to 20%. A combination of data loss and image loss counteracted this effect but still showed a degraded CRC for frame lengths ≤5s. Including low-count training data in the training stablized the image quality and slightly improved the CRC of small structures in short frames ≤5s.

AcknowledgmentThe authors would like to thank the Department of Nuclear Medicine at Bern University Hospital’s Inselspital for providing patient data. They also extend their gratitude to King’s College London for sharing image quality phantom data. This investigative work was supported by the National Institute of Biomedical Imaging and Bioengineering through grant number R01-EB031806. The content in this article is solely the responsibility of the authors and does not necessarily represent the official views of the National Institutes of Health.
References
[1] Overbeck et al. Near real-time reconstruction of dynamic [18F]FDG PET examinations for Total Body PET, Journal of Nuclear Medicine Jun 2025, 66 (supplement 1) 251340.
[2] Nadig et al. PET quantification accuracy using FastPET in ultra-short frames on the Biograph Vision Quadra PET/CT, EANM Innovation (2025): S1–S1013; 10(1): 63–73. doi: 10.1016/S3051-2913(25)00
[3] Whiteley et al., DirectPET: full-size neural network PET reconstruction from sinogram data, Journal of Medical Imaging, Vol. 7, Issue 3, 032503 (February 2020), doi: 10.1117/1.JMI.7.3.032503
[4] Whiteley et al., FastPET: Near Real-Time PET Reconstruction from Histo-Images Using a Neural Network, IEEE Transactions on Radiation and Plasma Medical Sciences, vol. 5, no. 1, pp. 65-77, Jan. 2021, doi: 10.1109/TRPMS.2020.3028364
[5] Millardet et al. Improved quantification in end-to-end deep learning Fast-PET reconstruction using multi-view histo-images of attenuation correction
factors, IEEE Transactions on Radiation and Plasma Medical Sciences 2026 January; 10(1): 63–73. doi: 10.1109/TRPMS.2025.3569198
[6] Bharkada, Panin and Millardet, Methods and apparatus for deep learning based image reconstruction, US 2025/0148662 A1
Keywords: reconstruction, FastPET, convolutional neural network, histoimages, LAFOV
17:30 Poster panel: 377

Poster Number:
MIC-10-377

Quantitative Reconstruction and Correction for Long-Axial FOV PET: Optimizing Component-Based Normalization and Scatter Estimation (#1897)

Y. Chen1, R. Liao1, Y. Hou1, J. Jiang1

1 Beijing Normal University, school of physics and astronomy, Beijing, China

Abstract

The extension of the axial field-of-view (AFOV) advances in positron emission tomography (PET), offering unprecedented sensitivity and enabling ultra-low-dose 4D total-body imaging. However, long-axial field-of-view (LAFOV) systems introduce significant challenges, including prohibitive computational loads and complex quantitative corrections necessitated by expanded axial acceptance angles. A primary obstacle is the extreme sparsity of acquired data relative to the vast lines-of-response (LOR) space. We present a robust quantitative correction strategy integrated into QuanTOF, a GPU-accelerated, list-mode reconstruction framework. Our primary contribution is the implementation of a modified component-based normalization (CBN) strategy tailored for axial-dependent variations, effectively mitigating noise amplification in sparse data environments. Furthermore, we introduce a simplified calibration protocol utilizing 0.5 m thin-ring and FOV-length line sources to derive geometric and plane efficiencies—an efficient alternative to conventional through-axial rotating rods. To ensure correction accuracy despite data sparsity, a moderate rebinning strategy was employed for scatter and random estimation. The framework was validated via GATE simulations of the 1.5 m uMI Panorama GS system. Reconstruction of a 1.5 m uniform cylinder phantom (8cm radius) demonstrated superior axial uniformity, with intensity fluctuations constrained to less than 2% within 70% of the FOV. Additionally, Derenzo phantom simulations confirmed the framework's high-resolution capabilities, successfully resolving 2.4 mm hot rods. These results show this framework successfully overcomes the inherent challenges of LAFOV data sparsity, securing the quantitative fidelity essential for total-body molecular imaging.

AcknowledgmentThis work was supported in part by the National Natural Science Foundation of China (No. 12475336), and the Fundamental Research Funds for the Central Universities (No. 310400209521).
References
[1] Zhang, H, et al. 2024, ‘Performance Characteristics of a New Generation 148-cm Axial Field-of-View uMI Panorama GS PET/CT System with Extended NEMA NU 2-2018 and EARL Standards’, Journal of Nuclear Medicine, jnumed.124.267963, Reston: Society of Nuclear Medicine and Molecular Imaging
[2] Qi, J, Matej, S, Wang, G, and Zhang, X 2021, ‘3D/4D Reconstruction and Quantitative Total Body Imaging’, PET Clinics, 16(1), 41–54, Philadelphia: Elsevier
[3] Zhang, X, Badawi, RD, Cherry, SR, and Qi, J 2018, ‘Theoretical study of the benefit of long axial field-of-view PET on region of interest quantification’, Physics in Medicine and Biology, 63(13), 135010, Bristol: IOP Publishing
[4] Zhang, X, et al. 2020, ‘Total-Body Dynamic Reconstruction and Parametric Imaging on the uEXPLORER’, Journal of Nuclear Medicine, 61(2), 285–291, Reston: Society of Nuclear Medicine and Molecular Imaging
[5] Cherry, SR, Jones, T, Karp, JS, Qi, J, Moses, WW, and Badawi, RD 2018, ‘Total-Body PET: Maximizing Sensitivity to Create New Opportunities for Clinical Research and Patient Care’, Journal of Nuclear Medicine, 59(1), 3–12, Reston: Society of Nuclear Medicine and Molecular Imaging
[6] Yuan, Z, et al. 2025, ‘GPU-based list-mode TOF PET image reconstruction with complete correction techniques’, Medical Physics, 53(1), e70216, Alexandria: American Association of Physicists in Medicine
Keywords: Positron emission tomography, PET reconstruction, LAFOV
17:30 Poster panel: 380

Poster Number:
MIC-10-380

Optimized Partial Field-of-View Reconstruction by Limited TOF Axial Rebinning in Long-Axial PET (#2235)

I. Hong1, Z. Burbar1

1 Siemens Healthineers, R&D, Knoxville, Tennessee, United States of America

Abstract

This work addresses efficient partial axial field-of-view (aFoV) reconstruction for long-axial PET systems, where reconstructing the full long-axial dataset and subsequently cropping to a target aFoV is computationally expensive. This need may arise when CT coverage is intentionally limited (e.g., organ-focused protocols or pediatric imaging). However, simply limiting the PET aFoV reduces the effective sensitivity of a long-aFoV scanner to that of a shorter-aFoV system. We propose a limited axial rebinning strategy that retains a subset of oblique lines of response that would otherwise be discarded when truncating data to a partial aFoV, thereby leveraging the sensitivity of the long-aFoV system. The approach selects a starting rebin segment and a bounded set of additional oblique segments (up to a maximum rebin segment), and rebins eligible events into target segments while preserving time-of-flight (TOF) geometry. In an FDG human study on a long-aFoV system, the proposed limited rebinning substantially increases retained true counts for a Vision-equivalent aFoV versus truncation without rebinning, while maintaining comparable axial resolution and avoiding the cost of reconstructing the full long-aFoV image.

References
[1] LR. D. Badawi et al., “First Human Imaging Studies with the EXPLORER Total-Body PET Scanner,” J. Nucl. Med., 2019.
[2] S. Surti, A. R. Pantel, and J. S. Karp, “Total Body PET: Why, How, What for?,” IEEE Trans. Radiat. Plasma Med. Sci., 2020.
[3] M. Defrise et al.,“Continuous and Discrete Data Rebinning in Time-of-Flight PET,” IEEE Trans. Med. Imaging, 2008.
[4] M. Defrise et al., “Exact and approximate rebinning algorithms for 3-D PET data,” IEEE Trans. Med. Imaging, 1997.
[5] S. Cho et al.,“Exact and approximate Fourier rebinning of PET data from time-of-flight to non time-of-flight,” PMB, 2009.
[6] I. Hong, and Z. Burbar, “PET image reconstruction optimization for long aFoV scanner by using TOF axial rebinnig,” IEEE NSS/MIC, 2025.
Keywords: Long Axial FOV pet scanner, PET reconstruction, Optimization
17:30 Poster panel: 383

Poster Number:
MIC-10-383

Generalizable CT-Free Low-Dose PET Restoration: Few-Shot Adaptation Across Multi-Scanners (#2478)

X. Qiao1, H. Wang1, B. Yao1, J. Yu2, Z. Zhuang3, Q. Huang4

1 Shanghai Jiao Tong University, Department of Nuclear Medicine, Ruijin Hospital, Shanghai, China
2 Shanghai Jiao Tong University, Department of Cardiology, Ruijin Hospital, Shanghai, China
3 Wageningen University and Research, Wageningen, Netherlands
4 Macao Polytechnic University, Faculty of Applied Sciences, Macao, China, Macao Special Administrative Region

Abstract

PET/CT radiation exposure arises from both the injected PET radiotracer and the accompanying CT scan used for attenuation and scatter correction. Reducing PET dose while eliminating CT-based correction is therefore an attractive strategy for lower-radiation PET/CT imaging, but this setting substantially degrades image quality. In this study, we propose a deep learning framework for CT-free low-dose [18F]FDG PET restoration based on a Noise-Level-Estimation network and systematically evaluate its generalizability across different PET scanners. The model restores attenuation- and scatter-corrected PET images from reduced-dose non-corrected inputs using a mixed-dose training strategy. It was trained on a main cohort from Ruijin Hospital, acquired with [18F]FDG on a uEXPLORER scanner, comprising 5,731 reconstructed PET volumes derived from 1,163 examinations. Scanner-domain validation was performed on 640 reconstructed PET volumes derived from 320 examinations acquired on three different scanners, including Discovery MI, Vision 450, and uMI 780. Direct pretrained inference, few-shot scanner-domain finetuning, and from-scratch training were compared. Direct application of the uEXPLORER-trained model showed performance degradation under scanner shifts, whereas few-shot finetuning consistently improved restoration fidelity. In the most challenging uMI 780 cohort, median PSNR increased from 35.24 dB to 47.58 dB after finetuning. These results demonstrate the feasibility of scanner-generalizable CT-free low-dose FDG PET restoration.

Acknowledgment

The authors would like to thank all colleagues and collaborators who contributed to this project.

References
[1] Lammertsma, A, et al. 2017, ‘Forward to the past: The case for quantitative PET imaging’, J. Nucl. Med., vol. 58, no. 7, pp. 1019–1024.
[2] Khosravi, B, et al. 2025, ‘Temporal hindsight, clinical foresight: Longitudinal lymphoma analysis at PET/CT’, Radiol. Artif. Intell., vol. 7, no. 3, e250149.
[3] Wang, H, et al. 2025, ‘Optimizing MR-based attenuation correction in hybrid PET/MR using deep learning: Validation with a flatbed insert and consistent patient positioning’, Eur. J. Nucl. Med. Mol. Imaging, vol. 52, no. 7, pp. 2577–2588.
[4] Chen, G, et al. 2023, ‘A total-body ultra-low dose PET reconstruction method via image space shuffle U-Net and body sampling’, IEEE Trans. Radiat. Plasma Med. Sci., vol. 8, no. 6, pp. 357–365.
[5] Wang, H, et al. 2025, ‘Noise-Level-Estimation Network for CT-Less Low Dose Total-Body PET Imaging’, 2025 IEEE Nuclear Science Symposium (NSS), Medical Imaging Conference (MIC) and Room Temperature Semiconductor Detector Conference (RTSD), Yokohama, Japan.
Keywords: PET imaging, low dose, CT-less, deep learning, total body
17:30 Poster panel: 386

Poster Number:
MIC-10-386

3D Photon Position Estimation Using Neural Networks in Monolithic-Based PET Systems (#2663)

P. Piera-Sapiña1, M. Freire1, F. Lopez-Berenguer1, A. Gonzalez-Montoro1, A. J. Gonzalez1

1 Instituto de Instrumentación para Imagen Molecular (I3M), Centro Mixto CSIC—Universitat Politècnica de València,, Valencia, Spain

Abstract

Neural Network (NN) models have emerged as a promising tool for determining 3D photon impact positions in Positron Emission Tomography (PET) detectors based on monolithic scintillators. Nevertheless, NNs typically require long acquisitions of reference data training, becoming a time-consuming process that requires extensive experimental data for each individual detector module. To reduce calibration time, we propose a new methodology based on NNs trained entirely using simulated data to predict the 3D impact position of gamma rays. The feasibility of the proposed approach was verified using a real preclinical PET system composed of two rings of 8 LYSO:Ce monolithic scintillators each.

The simulation of a single detector was implemented in GATE and reference data moving a pencil beam in 1 mm steps was simulated to train two Multilayer Perceptrons (MLPs), MLPX and MLPY, to predict the x- and y- impact positions, respectively. To validate the introduced methodology at reconstructed image level, experimental data was acquired using a 22Na point source at the center of the preclinical system and processed using both the presented framework and the system’s standard calibration methodology.

Results have demonstrated that similar spatial resolution was obtained with both approaches, thus demonstrating that NNs trained exclusively with Monte Carlo simulations can generate coherent reconstructed images and thus eliminate the need for calibration measurements for each individual detector module.

Acknowledgment

This work was supported by the Generalitat Valenciana under Grant No. CIPROM/2021/003 and funding from the University of Virginia Focused Ultrasound Cancer Immunotherapy Center and the FUS Foundation. M. Freire was supported by the CIAPOS Program for Researchers in Postdoctoral Phase of the Generalitat Valenciana under Grant No. CIAPOS/2023/141.

References
[1] Gonzalez-Montoro A, et al. 2021, IEEE Trans. on Rad. and Plasma Med. Sci., 5, 282-305
[2] Iborra A, et. al. 2019, Phys. Med. Biol., 64, 195010
[3] Lopez-Berenguer F, et al. 2025, IEEE Trans. on Rad. and Plasma Med. Sci., 10, 789-798
[4] Gonzalez-Montoro A, et al. 2017, IEEE Trans. on Rad. and Plasma Med. Sci., 1, 229-237
[5] Freire M, et al. 2020, IEEE Trans. on Rad. and Plasma Med. Sci., 4, 350-360
[6] Jan S, et al. 2004, Phys. Med. Biol., 49, 4543-4543
[7] Freire M, et al. 2022, Phys. Med. Biol., 67, 245011
Keywords: Monolithic, Preclinical PET, Calibration, Neural Networks
17:30 Poster panel: 389

Poster Number:
MIC-10-389

Depth Attention and Volume Attention Module for 3D Medical Image Segmentation (#1088)

X. Kan1, K. Ogawa2

1 Hosei University, Graduate School of Science and Engineering, Tokyo, Japan
2 Hosei University, Faculty of Science and Engineering, Tokyo, Japan

Abstract

3D medical image segmentation of CT and MRI volumes is essential for clinical diagnosis and treatment planning. While deep learning methods such as 3D U-Net, Attention U-Net, and nnU-Net achieve strong performance, conventional 3D CNNs treat all slices equally along the axial direction, causing insufficient learning of lesion-containing slices. Existing attention mechanisms such as Attention U-Net focus on height×width (H×W) planar spatial attention without explicitly modeling slice-wise importance along the depth (D) axis. This paper proposes two attention modules to address these limitations. First, a Depth Attention Module (DAM) automatically learns the importance of each slice and integrates it into each decoder layer via residual connection. Second, a volume Attention Module (VAM) extends DAM to independently learn importance along H, W, and D axes, combining three independent axis-wise attention maps to achieve disease-adaptive spatial importance learning. DAM achieves Dice improvements of 7.61%, 4.71%, and 3.72% over 3D U-Net on brain hemorrhage CT (HemSeg-200), kidney CT, and ischemic stroke MRI, respectively, and 8.52% over nnU-Net on lung cancer CT (MSD Task06). On brain hemorrhage CT, DAM further improves over Attention U-Net by 4.54% and over nnU-Net by 1.03%. VAM achieves improvements of 7.76% over 3D U-Net on brain hemorrhage CT and 2.83% over nnU-Net on lung cancer CT. Experimental results confirm the effectiveness of the proposed modules across multiple diseases and imaging modalities.

References
[1] Ronneberger, O, Fischer, P, Brox, T 2015, 'U-Net: convolutional networks for biomedical image segmentation,' in Proceedings of MICCAI, pp. 234–241
[2] Çiçek, Ö, Abdulkadir, A, Lienkamp, SS, Brox, T, Ronneberger, O 2016, '3D U-Net: learning dense volumetric segmentation from sparse annotation,' in Proceedings of MICCAI, pp. 424–432
[3] Oktay, O, Schlemper, J, Folgoc, LL, Lee, M, Heinrich, M, Misawa, K 2018, 'Attention U-Net: learning where to look for the pancreas,' in Proceedings of MIDL
[4] Isensee, F, Jaeger, PF, Kohl, SAA, Petersen, J, Maier-Hein, KH 2021, 'nnU-Net: a self-configuring method for deep learning-based biomedical image segmentation,' Nature Methods, vol. 18, no. 2, pp. 203–211
[5] Hu, J, Shen, L, Sun, G 2018, 'Squeeze-and-excitation networks,' in Proceedings of IEEE CVPR, pp. 7132–7141
[6] Song, C et al. 2024, 'HemSeg-200: a voxel-annotated dataset for intracerebral hemorrhage segmentation in brain CT scans,' in Proceedings of IEEE ISBI, pp. 3376–3377
[7] Simpson, AL et al. 2019, 'A large annotated medical image dataset for the development and evaluation of segmentation algorithms,' arXiv:1902.09063
[8] Heller, N et al. 2021, 'The KiTS19 challenge: data, methods, and results,' Medical Image Analysis, vol. 67, art. no. 101821
[9] Liew, SL et al. 2022, 'ATLAS: a large stroke neuroimaging dataset,' Scientific Data, vol. 9, art. no. 320
Keywords: 3D segmentation, CT, deep learning, CNN, medical imaging
17:30 Poster panel: 392

Poster Number:
MIC-10-392

Comparing the anatomical accuracy of a PET(O15) reconstructed algorithm (AGR) on Moyamoya Patients (#1252)

A. Azeez1, A. Zhang1, M. Zhao2, G. Zaharchuk1, F. Boada1

1 Stanford School of Medicine, Radiological Sciences Laboratory, Palo Alto, California, United States of America
2 Stanford School of Medicine, Department of Neurosurgery, Palo Alto, California, United States of America

Abstract

PET(O15) is the gold standard for assessing cerebral perfusion in vivo. Assessment of perfusion deficits before surgical interventions in Moyamoya patients with PET(O15) is hampered by poor spatial resolution and low signal-to-noise ratio (SNR). We investigate the use of anatomically-guided reconstruction (AGR) for improving the assessment of baseline perfusion metrics in Moyamoya patients. To evaluate PET(O15) perfusion metrics in Moyamoya patients with and without the use of anatomically guided reconstructions. We assess perfusion metrics across different anatomical brain parcellations to evaluate the differences between the conventional and AGR-enhanced PET(O15) images. Regions that overlap with clinically relevant regions are significantly different in signal intensity when comparing the use of the reconstruction method. Many of which overlap with those reported in the Moyamoya literature. Implementation of the anatomy-guided PET reconstruction(AGR) algorithm can provide clinicians with a better understanding of the disease progression, especially if data integrity is limited, and of low resolution. Particularly in cases of radioisotopes with short decay half-life, such as O15, or when studying diseases of tissue perfusion and oxygen metabolism.

Acknowledgment
This work was supported in part by PHS Grant EB029306.
References
[1] Demartini Jr., Z., Teixeira, B. CA., Koppe, G. L., Gatto, L. A. M., Roman, A., & Munhoz, R. P. (2022). Moyamoya disease and syndrome: A review. Radiologia Brasileira, 55(1), 31–37. https://doi.org/10.1590/0100-3984.2021.0010 [doi]
[2] Karsonovich, T., & Lui, F. (2025). Moyamoya disease. StatPearls.
[3] Lee, S. U., Oh, C. W., Kwon, O.-K., Bang, J. S., Ban, S. P., Byoun, H. S., & Kim, T. (2018). Surgical treatment of adult moyamoya disease. Current Treatment Options in Neurology, 20(7). https://doi.org/10.1007/s11940-018-0511-8 [doi]
[4] Scott, R. M., Smith, J. L., Robertson, R. L., Madsen, J. R., Soriano, S. G., & Rockoff, M. A. (2004). Long-term outcome in children with moyamoya syndrome after cranial revascularization by pial synangiosis. Journal of Neurosurgery: Pediatrics, 100(2), 142–149. https://doi.org/10.3171/ped.2004.100.2.0142 [doi]
[6] Suzuki, J. (1969). Cerebrovascular “moyamoya” disease. Archives of Neurology, 20(3), 288. https://doi.org/10.1001/archneur.1969.00480090076012 [doi]
[5] Shang, S., Zhou, D., Ya, J., Li, S., Yang, Q., Ding, Y., Ji, X., & Meng, R. (2018). Progress in moyamoya disease. Neurosurgical Review, 43(2), 371–382. https://doi.org/10.1007/s10143-018-0994-5 [doi]
[7] Zhao, M. Y., Fan, A. P., Chen, D. Y.-T., Ishii, Y., Khalighi, M. M., Moseley, M., Steinberg, G. K., & Zaharchuk, G. (2022). Using arterial spin labeling to measure cerebrovascular reactivity in Moyamoya disease: Insights from simultaneous PET/MRI. Journal of Cerebral Blood Flow & Metabolism, 42(8), 1493–1506. https://doi.org/10.1177/0271678x221083471 [doi]
[8] Schramm, G., Rigie, D., Vahle, T., Rezaei, A., Van Laere, K., Shepherd, T., … Boada, F. (2021). Approximating anatomically-guided PET reconstruction in image space using a convolutional neural network. NeuroImage, 224, 117399. https://doi.org/10.1016/j.neuroimage.2020.117399 [doi]
Keywords: AI/ML Image Reconstruction, PET/MR, Moyamoya disease, Cerebrovascular disease, Oxygen-15-labeled water
17:30 Poster panel: 395

Poster Number:
MIC-10-395

MRI-Informed SPECT-only Spatial Normalization for Reliable Brain Perfusion Quantification (#1410)

E. You1, 2, Y.J. Seol1, 2, H.G. Ryoo3, S.K. Kang4, J.S. Lee1, 2

1 Seoul National University, Interdisciplinary Program in Bioengineering, Seoul, Republic of Korea
2 Seoul National University, Integrated Major in Innovative Medical Science, Seoul, Republic of Korea
3 Seoul National University, Department of Nuclear Medicine, Seoul, Republic of Korea
4 Samsung Ventures, Seoul, Republic of Korea

Abstract

Spatial normalization enables standardized volume-of-interest quantification in brain perfusion single-photon emission computed tomography, but accurate alignment is difficult because of low resolution, noise, partial volume effects, and uptake-dependent variation. We propose a MRI-informed, SPECT-only framework that distills anatomical deformation fields and feature guidance from a magnetic resonance imaging registration teacher. The student disentangles anatomical and uptake-related representations to suppress functional-uptake interference while requiring only single-photon emission computed tomography at inference. In 173 paired imaging datasets, the method improved standardized uptake value ratio consistency between subject and template spaces, especially for fine-grained volumes of interest, supporting MRI-free, single-pass template-based quantification.

References
[1] Li, R., Figueredo, G., Auer, D., Wagner, C. and Chen, X. 2024, ‘MrRegNet: Multi-resolution Mask Guided Convolutional Neural Network for Medical Image Registration with Large Deformations’, 2024 IEEE International Symposium on Biomedical Imaging (ISBI), pp. 2349–2353, Piscataway, NJ: IEEE.
[2] Woo, S., Park, J., Lee, J.-Y. and Kweon, I.S. 2018, ‘CBAM: Convolutional Block Attention Module’, Proceedings of the European Conference on Computer Vision (ECCV), pp. 3–19, Cham: Springer.
[3] Kang, S.K., Kim, D., Shin, S.A., Kim, Y.K., Choi, H. and Lee, J.S. 2023, ‘Fast and Accurate Amyloid Brain PET Quantification Without MRI Using Deep Neural Networks’, Journal of Nuclear Medicine, vol. 64, no. 4, pp. 659–666, Reston, VA: Society of Nuclear Medicine and Molecular Imaging.
Keywords: Brain perfusion SPECT, feature disentanglement, knowledge distillation, spatial normalization.
17:30 Poster panel: 398

Poster Number:
MIC-10-398

When Do TAC and ROI Features Help? Residual Multimodal Learning for GFR Estimation in Dynamic 99mTc-DTPA Renography (#1697)

Y. Zhu1, J. Li2, C. Li1, J. Zhong2, W. Sun2, Z. Xu1, J. Chen3, Y. Zeng4, J. Sun1, H. Xiao2

1 Hainan University, Department of Data Science and Big Data, Haikou, China
2 The First Affiliated Hospital of Hainan Medical University, Department of Nuclear Medicine, Haikou, China
3 Hainan General Hospital, Pediatric Intensive Care Unit, Haikou, China
4 The Second Affiliated Hospital of Hainan Medical University, Department of Radiation Oncology, Cancer Treatment Center, Haikou, China

Abstract

Dynamic 99mTc-DTPA renography provides both spatiotemporal image information and clinically interpretable time-activity curve (TAC) and region-of-interest (ROI) descriptors for renal functional assessment. However, how these heterogeneous information sources should be integrated for automated Gates-derived total glomerular filtration rate (GFR) estimation remains unclear. This study developed a residual multimodal learning framework in which a 3D CNN image branch first estimated Gates-derived total GFR from multi-frame renography sequences, and selected TAC/ROI descriptors were then used to predict a residual correction term rather than being directly concatenated with image features. A total of 471 complete paired examinations were retrospectively included, with 376 cases for training and 95 cases for held-out testing. Representative image-only, TAC-only, direct-fusion, ROI/TAC tabular, and residual-correction models were compared under the same test split. The dynamic image-only model outperformed perfusion TAC-only, function TAC-only, and pure ROI/TAC tabular models, indicating that the multi-frame renography sequence was the dominant information source. Direct fusion of image and TAC inputs did not improve image-only prediction. In contrast, the proposed residual-correction model achieved the best overall performance, with the lowest mean absolute error of 8.21 and root mean square error of 10.93. These findings suggest that TAC/ROI descriptors are more useful as clinically interpretable residual correction cues than as standalone predictors or directly fused inputs. The proposed framework provides a practical strategy for approximating total GFR from routine dynamic 99mTc-DTPA renography while clarifying the complementary role of TAC/ROI descriptors.

AcknowledgmentThis work was supported in part by the National Natural Science Foundation of China (12505396) and the Science and Technology Talents Innovation Project of Hainan Province (KJRC2023D30).
References
[1] B. Feng et al., "The rotation method for correcting renal depth in the determination of glomerular filtration rate using Tc-99m diethylenetriamine pentaacetic acid (DTPA)-based renal dynamic imaging in patients with hydronephrosis," Quantitative Imaging in Medicine Surgery, vol. 15, no. 1, pp. 721-730, 2025.
[2] Y. Pi et al., "Deep regression using 99mTc-DTPA dynamic renal imaging for automatic calculation of the glomerular filtration rate," European Radiology, vol. 33, no. 1, pp. 34-42, 2023.
Keywords: 99mTc-DTPA, Glomerular filtration rate, Time-activity curve, Region of interest, Multimodal learning
17:30 Poster panel: 401

Poster Number:
MIC-10-401

Uncertainty-Guided Localized Diffusion Refinement for Lumbar Spine MRI Segmentation (#1924)

Q.V. Pham1, 2, S. Sung1, H.J. Choi3, S. Jeong3, S.W. Nam4, H. Kim5, C.H. Min1

1 Yonsei University, Department of Radiation Convergence Engineering, Wonju, Republic of Korea
2 Hanoi University of Science and Technology, Faculty of Engineering Physics, Hanoi, Viet Nam
3 Yonsei University Wonju College of Medicine, Department of Radiation Oncology, Wonju, Republic of Korea
4 Wonju Severance Christian Hospital, Wonju College of Medicine, Yonsei University, Department of Rheumatology, Wonju, Republic of Korea
5 Wonju Severance Christian Hospital, Wonju College of Medicine, Yonsei University, Department of Radiology, Wonju, Republic of Korea

Abstract

Lumbar spine MRI segmentation is an important prerequisite for quantitative spine image analysis; however, it remains challenging due to complex anatomy, low contrast, and high inter-patient variability. Recent diffusion-based approaches have shown potential for medical image segmentation; nevertheless, preserving anatomically consistent boundaries remains challenging, particularly in low-contrast regions. We propose a conditional diffusion-based framework for lumbar spine MRI segmentation with uncertainty-guided localized refinement. The model integrates a UNet-based denoising backbone with multi-scale feature encoding to preserve anatomical structures during the diffusion process. Furthermore, an entropy-driven uncertainty mechanism is introduced to identify ambiguous regions, enabling localized refinement and selective diffusion re-sampling in low-confidence regions to support boundary delineation. The framework is evaluated using 5-fold cross-validation on multi-contrast (T1- and T2-weighted) MRI data, achieving Dice scores of 0.87-0.89 with low surface errors (ASSD ≈ 1-2 mm, HD95 < 10 mm). The current results demonstrate consistent segmentation performance across spinal canal, vertebral bodies, and intervertebral discs, indicating the feasibility of automated multi-structure lumbar MRI segmentation. These results suggest that the proposed framework can serve as a technical foundation for automated quantitative spine MRI analysis. In future work, this framework will be extended to vertebral body and corner-region analysis for edema detection, assessment of corner changes, and quantitative MRI evaluation of axial spondyloarthritis.

References
[1] A. Kazerouni et al., Med. Image Anal., vol. 88, p. 102846, Aug. 2023, doi: 10.1016/j.media.2023.102846.
[2] Z. Xing et al., arXiv:2303.10326, 2023, doi: 10.48550/arXiv.2303.10326.
[3] M. Monzon et al., Proc. Mach. Learn. Res. (PMLR), vol. 156, pp. 1–18, 2025, accepted for MIDL 2025.
[4] J. W. Van Der Graaf et al., Sci. Data, vol. 11, no. 1, p. 264, Mar. 2024, doi: 10.1038/s41597-024-03090-w.
Keywords: Lumbar spine MRI, Medical image segmentation, Diffusion models, Uncertainty estimation, Boundary refinement
17:30 Poster panel: 404

Poster Number:
MIC-10-404

PhysiCal-PET: Physics-based neural networks for the rapid position Calibration of PET Scanners (#2036)

M. Abushawish1, P. Galve3, 1, I. Colmenarejo Allavena1, S. Gaitan Dominguez1, C. Galindo Morata1, V. M. Nouvilas1, J.M. Udias1, 2

1 Universidad Complutense de Madrid, Grupo de Física Nuclear, EMFTEL & IPARCOS, Madrid, Spain
2 Health Research Institute of the Hospital Cínico San Carlos (IdISSC), Madrid, Spain
3 Spanish National Research Council (CSIC), Institute for Physical and Information Technologies “Leonardo Torres Quevedo”, ITEFI, Madrid, Spain

Abstract

Total-body positron emission tomography (PET) increases system sensitivity and enables whole-body dynamic imaging, but its extended axial field of view makes accurate depth-of-interaction (DOI) calibration essential for maintaining uniform spatial resolution and reducing parallax error. Conventional DOI calibration requires mechanically scanned collimated sources or detector-specific reference measurements, which become impractical as the number of detector modules increases. PhysiCal-PET introduces a physics-based neural-network framework for autonomous in-system DOI calibration of PET scanners. The method trains a network to map raw detector signals to DOI using coincidence data acquired with an uncollimated positron source at known positions. For each event, the loss function enforces annihilation geometry by requiring the predicted line of response to be consistent with the source position, thereby providing supervision without measured DOI labels. The framework was tested on realistic PET simulations and applied to data from a real preclinical PET scanner. Using this procedure, the DOI response of all 144 detector modules was calibrated in less than 2 hours of point-source measurement time. These results demonstrate that the annihilation physics can replace labor-intensive DOI calibration hardware and enable rapid, routine calibration of high-resolution PET systems. PhysiCal-PET therefore provides a scalable path toward autonomous calibration for total-body PET and other systems that rely on accurate DOI positioning.

AcknowledgmentP. Galve acknowledges grant JDC2023-051754-I, funded by MCIU/AEI/10.13039/501100011033 and ESF+.
Keywords: PET, DOI, In-system Calibration, PINNs, PhysiCal-PET
17:30 Poster panel: 407

Poster Number:
MIC-10-407

Nonsubsampled Contourlet Transform-based Generative Adversarial Networks for Anatomical Structure-Preserving Low Dose CT Denoising (#2204)

H. Cho1, S. Cho1

1 Yonsei University, Department of Radiation Convergence Engineering, Wonju, Republic of Korea

Abstract

This paper presents a Nonsubsampled Contourlet Transform-Based Generative Adversarial Network (NSCT-GAN) for low-dose computed tomography (LDCT) denoising. LDCT reduces radiation exposure but introduces quantum Poisson and electronic Gaussian noise, which blurs anatomical structures. Conventional deep learning methods using mean-squared error loss produce over-smoothed images, while existing image-domain generative adversarial networks (GANs) struggle with high-frequency CT noise. To address this, the proposed NSCT-GAN extracts multi-scale, multi-directional, and shift-invariant high-frequency features using the NSCT. The network trains a discriminator on these high-frequency images to distinguish noise from anatomical edges. The generator utilizes a U-Net architecture to maintain spatial consistency. The objective function combines a Wasserstein GAN with gradient penalty adversarial loss for high-frequency reconstruction, an L1 loss for accurate CT number recovery, and a multi-scale structural similarity (MS-SSIM) loss for global and local structural consistency. The method was evaluated using the 2016 NIH-AAPM-Mayo Clinic Low-Dose CT Grand Challenge dataset and compared against five deep learning-based denoising methods: WavResNet, SMGAN, CPCE, DU-GAN, and WDLP. Quantitative results demonstrate that NSCT-GAN achieves a peak signal-to-noise ratio (PSNR) of 27.1850 dB, a structural similarity index (SSIM) of 0.8215, and a feature similarity index (FSIM) of 0.9698, yielding the highest SSIM and FSIM values among the tested models. Qualitative evaluations show the method restores continuous vascular boundaries in the lungs and maintains heterogeneous parenchymal patterns in the liver without introducing over-smoothing or artifacts.

Acknowledgment

This work was supported by the National Research Foundation of Korea(NRF) grant funded by the Korea government(MSIT)(RS-2025-16065119).

Keywords: Low-dose CT, Noise reduction, Nonsubsampled contourlet transform
17:30 Poster panel: 410

Poster Number:
MIC-10-410

Preliminary Results of Generative Model-Based Positron Range Correction in PET Imaging (#2359)

Y. Hou1, J. Ni1, Y. Chen1, F. Zhan2, 3, J. Jiang1

1 Beijing Normal University, School of Physics and Astronomy, Beijing, China
2 The First Affiliated Hospital of USTC, Department of Nuclear Medicine,, Hefei, China
3 Jiangnan University, Wuxi School of Medicine, Wuxi, China

Abstract

Positron range is an inherent physical factor that degrades the spatial resolution of positron emission tomography (PET), particularly for radionuclides with high positron energies. This study investigated the feasibility of deep learning-based positron range correction for PET brain imaging. A total of 304 brain datasets acquired using a Siemens Biograph Vision PET/CT scanner at the First Affiliated Hospital of the University of Science and Technology of China were voxelized to generate brain phantoms. These phantoms were used as activity maps for GATE simulations with three emission settings: 18F, 68Ga, and back-to-back sources. The 18F and 68Ga simulations represented mild and severe positron range effects, respectively, whereas the back-to-back simulation served as the range-free reference. Two deep learning-based correction methods were evaluated, including a conditional denoising diffusion probabilistic model (C-DDPM) and a residual U-Net (Res-UNet). Visual results showed that C-DDPM better restored fine sulcal and gyral structures and achieved closer agreement with the back-to-back reference, especially for the more severe 68Ga correction task. Although Res-UNet achieved superior PSNR, SSIM, and NRMSE values, its restored images exhibited noticeable oversmoothing and partial loss of subtle anatomical details. These findings suggest that diffusion-based generative restoration is a promising strategy for positron range correction, with the potential to reduce positron range induced blurring while preserving fine anatomical structures.

AcknowledgmentThis work was supported in part by the National Natural Science Foundation of China (No. 12475336), and the Fundamental Research Funds for the Central Universities (No. 310400209521).
References
[1] Berg, E & Cherry, SR 2018, ‘Innovations in Instrumentation for Positron Emission Tomography’, Seminars in Nuclear Medicine, vol. 48, no. 4, pp. 311–331, Elsevier.
[2] Hou, Y, Zhan, F, Liu, J, Cheng, X, Li, C, Yuan, Z, et al. 2026, ‘Cycle-constrained adversarial denoising convolutional network for PET image denoising: Multi-dimensional validation on large datasets with reader study and real low-dose data’, Medical Image Analysis, vol. 107, 103826, Elsevier.
[3] Yuan, Z, Zhan, F, Lu, H, Hou, Y, Liao, R, Li, C, et al. 2026, ‘GPU-based list-mode TOF PET image reconstruction with complete correction techniques’, Medical Physics, vol. 53, no. 1, e70216, Wiley.
[4] Ho, J, Jain, A & Abbeel, P 2020, ‘Denoising diffusion probabilistic models’, Advances in Neural Information Processing Systems, vol. 33, pp. 6840–6851.
Keywords: Positron emission tomography, Positron range correction, Image restoration, Diffusion model
17:30 Poster panel: 413

Poster Number:
MIC-10-413

Wavelet-Conditioned Rectified Flow for Efficient Whole-Body Low-Dose PET Denoising (#2431)

L. Xie1, W. Duan1, J. Yan1, H. Zhang1, S. Wang1, Z. Wu1, H. Liu1, Y. Liu1

1 Tsinghua University, Beijing, China

Abstract

Low-dose positron emission tomography (PET) reduces injected activity and acquisition time, but low-count statistics introduce strong noise, impaired lesion conspicuity, and unstable standardized uptake value (SUV) quantification in whole-body imaging. Recent 3D diffusion models improve PET denoising, and wavelet-conditioned diffusion further shows that the Haar low-frequency LLL subband can serve as a robust structural prior. However, diffusion sampling is computationally expensive for large 3D PET volumes. We propose Wavelet-Conditioned Rectified Flow (WCRF), a deterministic whole-body PET restoration framework that integrates wavelet-domain structural guidance with rectified-flow matching.

To establish robust structural guidance, WCRF isolates the lowest-frequency (LLL) subband via a single-level 3D Haar decomposition. This frequency-derived information is then combined with the original noisy input to form a resilient conditioning signal. A lightweight ControlNet-style branch injects this condition into a time-conditioned rectified-flow network, which learns the velocity field from low-dose to full-dose PET along a straight transport path. During inference, WCRF restores full 3D volumes using sliding-window Euler integration with only a small number of function evaluations. Experiments against low-dose input, 3D U-Net, 3D GAN, and 3D DDPM baselines at the 1/20-dose setting show that WCRF achieves competitive overall restoration performance among the evaluated methods, while substantially reducing inference cost relative to DDPM.
AcknowledgmentThe authors gratefully acknowledge Xue Song, Kuangyu Shi, and Axel Rominger from the University of Bern, as well as Hanzhong Wang, Rui Guo, and Biao Li from Ruijin Hospital, Shanghai Jiao Tong University School of Medicine, for providing the PET data used in this study.
References
[1] Wang, Y., Yu, B., Wang, L., Zu, C., Lalush, D. S., Lin, W., ... & Shen, D. (2018). 3D conditional generative adversarial networks for high-quality PET image estimation at low dose. NeuroImage, 174, 550-562.
[2] Yu, B., Ozdemir, S., Dong, Y., Shao, W., Pan, T., Shi, K., & Gong, K. (2025). Robust whole-body PET image denoising using 3D diffusion models: evaluation across various scanners, tracers, and dose levels. European Journal of Nuclear Medicine and Molecular Imaging (EJNMMI), 52(7), 2549-2562.
[3] Lyu, Q., Chen, T., Wang, Y., Guo, E., & Zhou, L. (2025). WID-PET: PET image reconstruction from low-dose data using a wavelet-informed diffusion model with fast inference. In International Conference on Medical Image Computing and Computer-Assisted Intervention (MICCAI) (pp. 684-693). Springer.
[4] Xue S, Guo R, Bohn KP, Matzke J, Viscione M, Alberts I, Meng H, Sun C, ZhangM, Zhang M, Sznitman R, EI Fakhri G, Rominger A, Li B and Shi K, A cross-scannerandcross-tracer deep learning method for the recovery of standard-dose imaging qualityfrom low-dose PET. European journal of nuclear medicine and molecular imaging, 2021. pp.1-14.
Keywords: Low-dose PET, deep learning, DDPM, rectified flow, wavelet structural prior
17:30 Poster panel: 416

Poster Number:
MIC-10-416

Adaptive Spatio-Temporal Denoising for Low-Dose Fluoroscopy via Motion-Gradient Weighting (#2582)

N. León1, 2, X. Ye1, J. L. Longas1, J.M. Arco1, J.M. Ortega1, M. Pérez-Liva2, J. L. Herraiz2, A. Sisniega1

1 Sedecal Molecular Imaging, Madrid, Spain
2 Universidad Complutense de Madrid, Nuclear Physics Group and IPARCOS, Universidad Complutense de Madrid, Madrid, Spain

Abstract

A spatio-temporal denoising training strategy for low-dose fluoroscopy is proposed, based on a spatially-adaptive temporal consistency metric. This approach acts on a small set of consecutive frames coupling a conventional denoising operator with an adaptive motion-gradient attention gate to selectively modulate the temporal consistency terms within the loss function, ensuring effective noise suppression and flickering reduction while maintaining high anatomical fidelity in dynamic regions.
The proposed method was used to train a denoising operator in which a three-frame temporal window is used as input. The model was trained across multiple dose reduction factors (qF), and demonstrated superior structural recovery and enhanced temporal stability in the processed image sequences, compared to single-frame models. Quantitative and qualitative evaluation of image quality, along with stability maps—defined as the Normalized Cross-Correlation (NCC) of temporal Coefficient of Variation (CV) maps—confirms high fidelity relative to high-dose reference acquisitions, effectively mitigating stochastic fluctuations without introducing motion artifacts.

AcknowledgmentThe authors would like to express their sincere gratitude to the Image Processing and AI Group at Sedecal Molecular Imaging for their invaluable technical support and expertise. We also thank the Nuclear Physics Group and IPARCOS at the Universidad Complutense de Madrid for their collaboration and the resources provided for this research. 

We acknowledge support from the Comunidad de Madrid regional government through the LUNABRAIN project (TEC-2024/TEC-43) and the Industrial PhD program IND2023/BMD-28275.

 

References
[1] Dierckx, P. et al. 2007, “Spatio-temporal adaptive filter for low-dose fluoroscopy”, IEEE Nuclear Science Symposium Conference Record, pp. 3450–3453.
 
[2] Wang, X. et al. 2019, “EDVR: Video restoration with enhanced deformable convolutional networks”, IEEE/CVF Conference on Computer Vision and Pattern Recognition Workshops (CVPRW).
 
[3] Chen, H. et al. 2017, “Low-dose CT with a residual encoder-decoder convolutional neural network”, IEEE Transactions on Medical Imaging, vol. 36, no. 12, pp. 2524–2535.
 
[4] Lehtinen, J. et al. 2018, “Noise2Noise: Learning image restoration without clean data”, Proceedings of the International Conference on Machine Learning (ICML).
 
[5] Zhao, W., Rowlands, J. A. 1995, “X-ray imaging using amorphous selenium: Feasibility of a flat panel self-scanned detector for digital radiology”, Medical Physics, vol. 22, no. 10, pp. 1595–1604.
 
[6] Cunningham, I. A. 2000, “Applied linear-systems theory”, in Handbook of Medical Imaging, vol. 1, Sonka, M., Fitzpatrick, J. M. (eds.), pp. 79–160, Bellingham, WA: SPIE.
Keywords: Deep learning, fluoroscopy, image denoising, motion-adaptive filtering, spatio-temporal analysis
17:30 Poster panel: 419

Poster Number:
MIC-10-419

An Ensemble of Deep-Learning Classifiers for 3D Anatomical Change Reconstruction in Carbon-Ion Radiotherapy (CIRT) (#2644)

P. Schlegel1, 3, R. Kirchgässner1, 2, P. Ochoa-Parra1, 2, E. Bartelme1, 2, R. Mikut4, O. Jäkel1, 2, M. Martisikova1, 2

1 DKFZ, Division of Medical Physics in Radiation Oncology, E040, Heidelberg, Baden-Württemberg, Germany
2 Heidelberg Institute for Radiation Oncology (HIRO), National Center for Research in Radiation Oncology (NCRO), Heidelberg, Baden-Württemberg, Germany
3 University, Heidelberg, Medical Faculty, Heidelberg, Baden-Württemberg, Germany
4 Karlsruhe Institute of Technology (KIT), Institute for Automation and Applied Informatics (IAI), Karlsruhe, Germany

Abstract

Carbon-ion radiotherapy (CIRT) combines high biological effectiveness with outstanding treatment precision. Since treatment plans are based on the most recent Computed Tomography (CT), anatomical changes during multi-week treatments may compromise dose delivery accuracy. To address the increasing uncertainty about the patient’s anatomical state with time elapsed since the planning CT, our group developed an in vivo monitoring method using secondary charged particles emitted during therapy. This non-invasive approach requires no additional dose and reduces uncertainty about patient anatomy and thus, loss of treatment accuracy. It uses a custom detection system to detect nuclear fragments emmited from the patient and allows approximating their origin. To reconstruct these changes in 3D, we utilize our recently published method for detecting the depth (z) of changes along the beam axis and apply it to a sliding window (x,y) along the transverse plane. However, splitting reconstructed nuclear fragment origins into smaller subsets reduces the statistical stability of each subset distribution. In this study, we present an approach to robustly distinguish whether a windowed subset belongs to a region with or without an anatomical change based on correlation matrices generated by our depth detection method. These images contain location-wise correlations of joined frequency band variations between projections of nuclear fragment measurements from different treatment days. To classify regions of change and approximate change depth, we train an ensemble of deep learning-based image classifiers with different architectures (i.e. ResNet, RegNet and ConvNeXt) and combine their outputs into an ensemble decision. The approach was validated using dedicated irradiation measurements with coin-sized cavities inserted into PMMA head phantoms. Despite limited training data, we achieved robust improvements in change localization using an ensemble of different deep learning architectures.

References
[1] Schlegel, P., Kirchgässner, R., Ochoa-Parra, P., Kelleter, L., Gertz, M., Mikut, R., Jäkel, O. and Martišíková, M. 2025, ‘A data analysis framework for in vivo monitoring in carbon-ion radiotherapy (CIRT): towards 3D reconstruction of interfractional anatomical changes’, Physics in Medicine & Biology, 70, 245009, Bristol: IOP Publishing.
[2] Kirchgässner, R., Martišíková, M., Schlegel, P., Ochoa-Parra, P., Harrabi, S., Jäkel, O., Debus, J. and Kelleter, L. 2026, ‘Out-of-beam artifact suppression in charged nuclear fragment-based carbon-ion radiotherapy monitoring’, Medical Physics, 53, e70351, Hoboken: Wiley.
[3] Ghesquière-Diérickx, L., Félix-Bautista, R., Schlechter, A., Kelleter, L., Reimold, M., Echner, G., Soukup, P., Jäkel, O., Gehrke, T. and Martišíková, M. 2022, ‘Detecting perturbations of a radiation field inside a head-sized phantom exposed to therapeutic carbon-ion beams through charged-fragment tracking’, Medical Physics, 49, Hoboken: Wiley.
[4] Kelleter, L., Schmidt, S., Subramanian, M., Marek, L., Granja, C., Jakubek, J., Jäkel, O., Debus, J. and Martišíková, M. 2024, ‘Characterisation of a customised 4-chip Timepix3 module for charged-particle tracking’, Radiation Measurements, 173, Amsterdam: Elsevier.
Keywords: Carbon Ion Radiotherapy, In Vivo Therapy Monitoring, Deep Learning, 3D Reconstruction
17:30 Poster panel: 422

Poster Number:
MIC-10-422

Reduction of Breathing and Metallic Implant Artifacts in PET with Synthetic CT Attenuation (#2684)

J. Edwards1, 2, Z. Wilson1, 2, S. Wollenweber3, R. Johnsen3, H. Ratner3, A. Mehranian3, K. M. Bradley4, D. R. McGowan1, 2

1 University of Oxford, Oncology Department, Oxford, United Kingdom
2 Oxford University Hospitals NHS Foundation Trust, Department of Medical Physics and Clinical Engineering, Oxford, United Kingdom
3 GE HealthCare, Chicago, Illinois, United States of America
4 University Hospital of Wales, Wales Research and Diagnostic Positron Emission Tomography Imaging Centre, Cardiff, United Kingdom

Abstract

Despite its widespread use for compensating attenuation in positron emission tomography (PET), computed tomography (CT)-based attenuation correction has limitations which can create imaging artifacts associated with breathing motion and metallic implants. Using deep learning networks to create synthetic attenuation correction maps represents a data-driven, non-ionizing alternative with the potential to reduce the frequency and/or intensity of such errors. In this study, the authors aimed to demonstrate that using synthetic maps would reduce breathing motion and metallic implant artifacts. To achieve this, ten anonymized whole-body F18-Fluorodeoxyglucose exams were assessed. Scans were divided into two Cohorts; Cohort 1 comprised exams with breathing motion artifacts, whilst Cohort 2 contained scans of patients with hip replacements. Each scan was corrected separately with CT-based attenuation correction and synthetic maps prior to identical reconstruction. Quantitative assessments of standardized uptake value (SUV) parameters were performed, and differences were tested for statistical significance. Additionally, each image’s lesion detectability, diagnostic confidence and overall image quality were scored on a Likert scale by an experienced radiologist. After assessment, no statistically significant differences in SUV quantification were detected (p > 0.05). Images corrected with synthetic maps were preferred by the radiologist in Cohort 1, but no preference was expressed in Cohort 2. Moreover, a single Cohort 1 case was identified in which breathing motion artifacts impacted clinical diagnosis only for the image treated with CT-based attenuation correction. Thus, it was concluded that synthetic maps reduce breathing motion artifacts to an extent which may support clinical diagnoses, although their effects on metallic implant artifacts require further investigation.

Keywords: Attenuation correction, Computed tomography, Positron emission tomography.
17:30 Poster panel: 425

Poster Number:
MIC-10-425

Frozen Stable Diffusion VAE as a Slice-Wise Encoder for Scalable 3-D Medical Image Generation (#2828)

Y. Mellak1, A. De Paepe1, A. Bousse1, D. Visvikis1

1 Université de Bretagne Occidentale, LaTIM, INSERM, UMR 1101, Brest, France

Abstract

Full-volume 3-D medical image generation at clinical resolutions remains a significant challenge due to GPU memory constraints. Existing voxel-space generative methods are difficult to scale to large clinical volumes, as memory and compute costs grow rapidly with spatial resolution. Latent diffusion approaches offer a more scalable alternative [1], but they typically require training a dedicated medical autoencoder for each modality or dataset before generative modeling can begin, making the compression stage itself a major bottleneck [2], [3], [4]. A lighter alternative is to encode axial slices independently with a 2-D autoencoder and stack the resulting latent maps into a 3-D tensor [5]; however, this still requires training a modality-specific encoder from scratch, limiting generality across modalities and resolutions.
We propose a scalable latent flow framework that eliminates this bottleneck by using a frozen pretrained Stable Diffusion VAE as a ready-to-use slice-wise encoder for medical volumes, requiring no modality-specific retraining or adaptation. Each axial slice is resampled to a common resolution, intensity-normalized, and encoded independently by the frozen encoder, yielding a high in-plane compression factor compared to dedicated medical autoencoders. The resulting latent maps are stacked into a compact 3-D tensor, on which a full-volume 3-D U-Net flow model is trained directly. The objective predicts clean latents and supervises in velocity space [6], ensuring stable optimization and sharp generations in high-dimensional latent spaces, avoiding the blurriness and instability observed with standard diffusion objectives.
We validate our framework on CT, PET, and MRI volumes at clinical resolutions, demonstrating strong encoding fidelity and coherent full-volume generation across axial, coronal, and sagittal planes on a single consumer-grade GPU, without any modality-specific autoencoder training.

References
[1] Rombach, R., Blattmann, A., Lorenz, D., Esser, P. and Ommer, B., 2022. High-resolution image synthesis with latent diffusion models. In Proceedings of the IEEE/CVF conference on computer vision and pattern recognition (pp. 10684-10695).
[2] Guo, P., Zhao, C., Yang, D., Xu, Z., Nath, V., Tang, Y., Simon, B., Belue, M., Harmon, S., Turkbey, B. and Xu, D., 2025, February. Maisi: Medical ai for synthetic imaging. In 2025 IEEE/CVF Winter Conference on Applications of Computer Vision (WACV) (pp. 4430-4441). IEEE.
[3] Khader, F., Müller-Franzes, G., Tayebi Arasteh, S., Han, T., Haarburger, C., Schulze-Hagen, M., Schad, P., Engelhardt, S., Baeßler, B., Foersch, S. and Stegmaier, J., 2023. Denoising diffusion probabilistic models for 3D medical image generation. Scientific reports, 13(1), p.7303.
[4] Pinaya, W.H., Tudosiu, P.D., Dafflon, J., Da Costa, P.F., Fernandez, V., Nachev, P., Ourselin, S. and Cardoso, M.J., 2022, September. Brain imaging generation with latent diffusion models. In MICCAI workshop on deep generative models (pp. 117-126). Cham: Springer Nature Switzerland.
[5] Kebaili, A., Lapuyade-Lahorgue, J., Vera, P. and Ruan, S., 2024, May. 3d mri synthesis with slice-based latent diffusion models: Improving tumor segmentation tasks in data-scarce regimes. In 2024 IEEE International Symposium on Biomedical Imaging (ISBI) (pp. 1-5). IEEE.
[6] Li, T. and He, K., 2025. Back to basics: Let denoising generative models denoise. arXiv preprint arXiv:2511.13720.
Keywords: 3D medical image generation, flow matching, frozen VAE encoding, latent diffusion, memory-efficient generation.
17:30 Poster panel: 428

Poster Number:
MIC-10-428

No CT, No Problem: Synthetic PET Enables Attenuation Correction in Dedicated Brain Scanners (#2919)

A. Merino-Balaguer1, D. Perez-Benito1, C. J. Brooks1, J. J. Vaquero1, 2, J. E. Iglesias3, 4, D. Izquierdo-Garcia1

1 Universidad Carlos III de Madrid, Dpt. Bioengineering, Madrid, Spain
2 Instituto de Investigación Sanitaria Gregorio Marañón, Madrid, Spain
3 Massachusetts General Hospital, Harvard Medical School, Dpt. Radiology, Boston, Massachusetts, United States of America
4 Massachusetts Institute of Technology, Cambridge, Massachusetts, United States of America

Abstract

Dedicated standalone brain positron emission tomography (PET) scanners have emerged in recent years as a promising technology for neuroimaging, offering high spatial resolution, increased sensitivity, reduced injected dose and shorter acquisition times. However, unlike hybrid PET/computed tomography (PET/CT) or PET/magnetic resonance imaging (PET/MRI) systems, these scanners do not provide an anatomical image from which attenuation information can be derived. We present a SynthSR-based framework to predict CT-derived attenuation maps from non-attenuation-corrected (NAC) PET images, aiming to enable attenuation correction in standalone brain PET. Building on SynthSR, a large and widely variable set of synthetic PET volumes is generated to sample a space larger than the potential clinically viable PET images, and thus increasing the robustness and clinical applicability of the model. The synthetic PET volumes are generated on-the-fly from anatomical labels using label-conditioned Gaussian-mixture intensity sampling, spatial/intensity augmentation and PET-specific priors estimated from real normalized PET data. A PET appearance model introduces count-based noise, correlated texture, smoothing, and controlled intensity variability. Finally, spatial transformations are used to increase sample diversity while preserving voxel-wise alignment. A three-dimensional U-Net was trained to predict normalized attenuation maps from synthetic PET, while validation and testing were performed on never-seen real PET/CT data. The model was optimized with 24,000 on-the-fly synthetic samples and evaluated on held-out real PET/CT data. The best validation checkpoint achieved MAE=0.03 (3%), PSNR=25.74 dB, and Pearson correlation r=0.859. On the independent test subject, the model achieved MAE=0.027 (2.7%), PSNR=26.35 dB, and r=0.854. These results support the feasibility of CT/MRI naive pseudo-CT generation for PET attenuation correction in novel standalone dedicated brain PET scanners.

References
[1] J. J. Vaquero and P. E. Kinahan, “Positron emission tomography: Current
challenges and opportunities for technological advances in clinical and
preclinical imaging systems,” Annu. Rev. Biomed. Eng., 2015
[2] Juan E. Iglesias et al., “SynthSR: A public AI tool to turn heterogeneous
clinical brain scans into high-resolution T1-weighted images for 3D
morphometry”, Sci. Adv.9,eadd3607, 2023
[3] D. Izquierdo-Garcia and C. Catana, “MR imaging-guided attenuation
correction of PET data in PET/MR imaging,” PET Clinics, 2016
[4] B. Billot et al., “SynthSeg: Segmentation of brain MRI scans of any
contrast and resolution without retraining,” Medical Image Analysis,
2023.
Keywords: attenuation correction, pseudo CT, standalone PET, SynthSR, deep learning
17:30 Poster panel: 431

Poster Number:
MIC-10-431

Alpha-CenterNet: A Heatmap-Based Detector for Alpha-Particle Localization in Scintillator-Camera Imagery (#3015)

D. Kang1, S. Kim2, 3, W. Sung2, 3

1 Agency for Defense Development, Chem-Bio Technology Directorate, Daejeon, Republic of Korea
2 The Catholic University of Korea, Department of Biomedical Engineering, College of Medicine, Seoul, Republic of Korea
3 The Catholic University of Korea, Department of Medical Sciences, Graduate School, Seoul, Republic of Korea

Abstract

Targeted Alpha Therapy with actinium-225 induces double-strand DNA breaks through individual alpha decays, motivating quantitative imaging of single-event distributions for dose assessment. We address point detection in a cerium-doped gadolinium aluminum gallium garnet scintillator coupled to an electron-multiplying camera, in which each alpha produces a point-spread function of approximately 2-micrometer diameter at 10 to 150 events per 100-millisecond frame. We compare three method groups on Monte Carlo simulated images: 20 classical preprocessing-and-clustering pipelines, two deep learning detectors based on the U-Net and YOLOv8-nano architectures, and Alpha-CenterNet, a heatmap-based detector that adapts CenterNet with a feature pyramid network and a modified focal loss. The highest-F1 classical pipeline attains 0.641 with 447-nanometer mean localization error. U-Net reaches F1 = 0.978 with 302-nanometer error; YOLOv8-nano reaches 185-nanometer error with F1 = 0.938; the two metrics produce divergent rankings. Alpha-CenterNet (3.78 million parameters) achieves F1 = 0.965 and 244-nanometer error, reducing localization error by 19 percent relative to U-Net at 26 percent of its parameter count, and reaching the validation F1 plateau of 0.98 in 29 versus 86 epochs—a 2.97-fold convergence speedup within the same ablation. The modified focal loss decouples F1 from localization error: F1 changes by 0.0004 while the mean error decreases by 76 nanometers, and this effect persists across all tested densities. The method generalizes to clustered and edge source distributions at unseen densities, supporting point detection for Targeted Alpha Therapy monitoring.

References
[1] S. Yamamoto et al., "Real-time trajectory imaging of alpha particles emitted from actinium-225 and its daughter radionuclides," Sci. Rep., vol. 15, no. 1, p. 2534, Jan. 2025, doi: 10.1038/s41598-025-87014-7.
[2] X. Zhou, D. Wang, and P. Krähenbühl, "Objects as points," arXiv:1904.07850, Apr. 2019.
[3] H. Law and J. Deng, "CornerNet: Detecting objects as paired keypoints," in Proc. ECCV, Munich, Germany, Sep. 2018, pp. 734–750.
Keywords: Targeted alpha therapy, Alpha-particle detection, Heatmap-based detection
17:30 Poster panel: 434

Poster Number:
MIC-10-434

AI-Assisted Circuitry Design for a Timing-Sensitive SiPM Front-End Readout (#3235)

Z. Lv1, Z. Hao1, X. Lai1, P. Xie2

1 ShanghaiTech University, School of Biomedical Engineering, Shanghai, China
2 Shanghai United Imaging Healthcare Advanced Technology Research Institute Co., Ltd., Shanghai, China

Abstract

Artificial-intelligence-assisted PCB layout is an emerging capability that promises to reduce design iteration time and enable rapid exploration of placement and routing alternatives, which is particularly valuable for complex, multi-channel analog front-end boards where manual layout is labor-intensive and experience-dependent. In this work, we explore the application of an AI-assisted layout workflow to a timing-sensitive silicon photomultiplier (SiPM) front-end readout board for time-of-flight positron emission tomography (TOF-PET) detectors. A four-channel RF front-end readout board, extending a validated single-channel timing topology, underwent automated placement and routing via AI layout tools. To establish a performance baseline, a single-channel prototype of the same RF topology was fabricated and characterized using a coincidence measurement setup with LYSO scintillators, achieving a coincidence time resolution of 100.78 ps and confirming the near-100-ps timing capability of the front-end design. Preliminary evaluation of the AI-generated four-channel layout shows that the tool achieved high routing completion and produced a viable candidate layout. Based on these encouraging results, we plan to fabricate the AI-generated board and systematically compare its timing performance, energy response, and channel consistency against our manually designed baseline, providing a direct, measurement-level assessment of AI-assisted layout quality for timing-sensitive detector electronics.

AcknowledgmentThis work was supported by the Science and Tech-nology Commission of Shanghai Municipality under grant No. 24DIPA00700. We acknowledge the ShanghaiTech Material and Device Lab and HPC Platform of ShanghaiTech University.
References
[1] N. Zhang, A. Agnesina, N. Shbat, Y. Leader, Z. Zhang, and H. Ren, “Cypress: VLSI-inspired PCB placement with GPU acceleration,” in Proceedings of the 2025 International Symposium on Physical Design (ISPD), 2025, pp. 31–41, doi: 10.1145/3698364.3705346.
[2] L. Vassallo and J. Bajada, “Learning circuit placement techniques through reinforcement learning with adaptive rewards,” in Proc. Design, Automation & Test in Europe Conference & Exhibition (DATE), 2024, pp. 1–6.
[3] L. Chen, R. Chen, S. Hu, X. Yao, Z. Tang, S. Kai, S. Xu, M. Yuan, J. Hao, B. Yu, and J. Xu, “PCBAgent: An agent-based framework for high-density printed circuit board placement,” in Proceedings of the 30th Asia and South Pacific Design Automation Conference, 2025, pp. 781–787, doi: 10.1145/3658617.3697736.
[4] P. Lecoq, “Pushing the limits in time-of-flight PET imaging,” IEEE Transactions on Radiation and Plasma Medical Sciences, vol. 1, no. 6, pp. 473–485, Nov. 2017, doi: 10.1109/TRPMS.2017.2756674.
Keywords: AI-assisted PCB layout, silicon photomultiplier, front-end electronics, timing-sensitive layout
17:30 Poster panel: 437

Poster Number:
MIC-10-437

Simultaneous Dual Tracer PET Separation in reconstruction: Validation on LAFOV Data (#1161)

Z. Li1, N. Efthimiou1, S. L.M. Viol2, G. Luurtsema1, T. van Meerten3, K. Thielemans4, 5, A. H. Brouwers1, A. A. Lammertsma1, C. Tsoumpas1

1 University Medical Center Groningen, Department of Nuclear Medicine and Molecular Imaging, Groningen, Netherlands
2 VU University Medical Center, Department of Radiology and Nuclear Medicine, Amsterdam UMC, Amsterdam, Netherlands
3 University Medical Center Groningen, Department of Hematology, Groningen, Netherlands
4 University College London, Institute of Nuclear Medicine, London, United Kingdom
5 University College London, UCL Hawkes Institute, London, United Kingdom

Abstract

Simultaneous dual tracer PET provides complementary physiological information compared with single tracer imaging. For example, combining [⁸⁹Zr]labeled monoclonal antibodies (mAbs) with [¹⁸F]fluorodeoxyglucose (FDG) enables the assessment of both targeted uptake and glucose metabolism, potentially improving patient selection [1]. Existing dual tracer separation methods exploiting differences in tracer kinetics are typically applied post-reconstruction in the image domain. However, these approaches may be sensitive to noise propagation. In this work, we evaluate a sinogram-domain iterative separation method that is incorporated directly into the reconstruction process, enabling direct generation of single tracer images.

AcknowledgmentZekai Li was also supported by a Joint Scholarship from the China Scholarship Council and the University of Groningen (202210100011) and his placement at UCL by EPSRC CCP in Synergistic Reconstruction for Biomedical Imaging (EP/T026693/1).Charalampos Tsoumpas, Nikos Efthimiou and Zekail Li are supported by the NWO AES Vici Talent program (doi: 10.61686/REEAR02777).
References
[1] Li Z, de Boer JW, van Meerten T, et al. Separation of simultaneously acquired [89Zr]atezolizumab and [18F]FDG PET scans. Eur J Nucl Med Mol Imaging. 2026;53(2):705-707.
[2] Li Z, Mohr P, Brouwers AH, Lammertsma AA, Tsoumpas C. Reconstruction-Based Image Separation for Simultaneous Dual-Tracer PET Scanning. IEEE Nuclear Science Medical Imaging Conference (MIC). 2025. 1-1.
[3] Venet L, Pati S, Feldman MD, et al. Accurate and robust alignment of differently stained histologic images based on Greedy diffeomorphic registration. Appl Sci (Basel). 2021;11:1892.
[4] Thielemans K, Tsoumpas C, Mustafovic S, et al. STIR: software for tomographic image reconstruction release 2. Phys Med Biol. 2012;57(4):867-883.
Keywords: dual tracer, image reconstruction, PET
17:30 Poster panel: 440

Poster Number:
MIC-10-440

High-Energy Proton Scattering Radiography of a Geant4 Hand Phantom under Constant-Energy Loss (#1435)

R. S. Ughade1, S. Chatzidakis1

1 Purdue University, West Lafayette, Indiana, United States of America

Abstract

Proton radiography and tomography often use residual energy or water equivalent path length (WEPL) to estimate relative stopping power (RSP). Since the maximum likelihood path (MLP) formalism of Schulte et al. (2008), proton trajectories in tissue have been modeled as curved paths constrained by entry and exit tracking. In full proton CT, this path estimation is coupled to energy dependent multiple Coulomb scattering and longitudinal energy loss, which becomes computationally and physically complex at low residual energies. Most MLP based reconstructions use the estimated path mainly to assign WEPL or RSP contributions to voxels. Here, we examine a thin object regime in which scattering itself provides imaging contrast. For high energy protons traversing a human hand, the integrated energy loss is small, allowing the scattering relevant βcp factor to be approximated as constant along the path. Under this constant energy approximation, we reconstruct a volumetric scattering density map using an M-value calculated from each proton scattering angle and kinetic energy, together with the μTRec Bayesian voxel attribution algorithm developed for muon scattering tomography. The method is demonstrated in Geant4 using an anthropomorphic hand phantom containing bones, muscles, tendons, and skin. A single orientation 200 MeV parallel proton beam with 10^7 protons is simulated and reconstructed on a 0.3 mm voxel grid. The resulting proton radiograph resolves carpals, metacarpals, and phalanges, and detects a submillimetre osteochondral gap at the thumb interphalangeal joint. These results suggest that constant energy, MLP inspired scattering radiography can provide a computationally lightweight alternative or complement to energy loss based proton imaging for thin anatomical sites.

AcknowledgmentThe present research is being conducted with the support of funding provided by the Purdue University School of Nuclear Engineering and the Purdue Research Foundation.
Keywords: proton radiography, most likely path, geant4, hand phantom, muTRec
17:30 Poster panel: 443

Poster Number:
MIC-10-443

Multiple Emission Tomography: A High-Sensitivity Single-Layer Simultaneous PET Compton-Camera (#1838)

M. Pizzichemi1, 2, G. Terragni2, F. Pennazio3, F. Sgarbi1, L. Cassano1, L. Arnone1, M. N. Colombo1, E. Fiorina3, F. Cossio3, M. Fadavi Mazinani3, M. Paganoni1, E. Auffray2, L. Presotto1

1 University of Milano-Bicocca, Milano, Italy
2 CERN, Geneva, Genève, Switzerland
3 INFN Torino, Torino, Italy

Abstract

The growing importance of theranostics and research with long‑lived isotopes demands imaging systems far more sensitive than standard PET or SPECT. Many emerging isotopes lie outside current detector capabilities, resulting in low sensitivity, high background, and poor quantification, as seen with 90Y verification and 124I or 89Sr imaging. The Multiple Emission Tomography (MET) project introduces a novel imaging device based on inorganic scintillators, capable of operating in dual modes: PET-style coincidence detection of 511 keV photon pairs and single-photon Compton-scattering detection in a single-layer configuration. The technical configuration of the MET prototype utilizes a DOI-capable light-sharing scheme. Each module consists of a 16x16 matrix of 1.53x1.53x20 mm3 depolished LYSO crystals coupled to an 8x8 array of 3x3 mm2 Hamamatsu Silicon Photomultipliers (SiPMs). The detector's readout electronics are based on the ALCOR ASIC, a low-power, mixed-signal device featuring single-photon sensitivity, a high bandwidth of up to 2 Mevents/s per channel, and a fine 25 ps time binning. Supported by a combination of analytic algorithms and a machine learning pipeline, the MET detector can successfully solve the temporal sequence of inter-crystal scatters, and determine the points of interaction and the energies deposited. To fully leverage these recovered events, a dedicated Compton image reconstruction algorithm has been developed, employing an event-by-event statistical model with ad-hoc projector functions. This technology is expected to significantly impact theranostics, collimator-less SPECT, and in-vivo range verification for CPT.

AcknowledgmentThis work is supported by the European Union-Next Generation EU, Mission 4 Component 1 CUP H53D23007430001, and carried out in association with the CERN Crystal Clear Collaboration.
Keywords: PET, SPECT, DOI, Compton, Inter-crystal scatter
17:30 Poster panel: 446

Poster Number:
MIC-10-446

A Scalable NaI(Tl)-SiPM Photon-Counting Scintillator Detector for Gold and Gadolinium XRF Imaging (#2151)

A. L.M. Silva1, B. Casanelli1, D. Batista1, T. Ferreira1, P. Correia1, R. Figueroa2, J. Veloso1

1 I3N -University of Aveiro, Physics Dpt., Aveiro, Portugal
2 Center of Excellence in Physics and Engineering in Health (CFIS), University of La Frontera (UFRO),, Temuco, Chile

Abstract

Gold and gadolinium-based nanoparticles have been investigated as theranostic agents whose characteristic X-ray fluorescence (XRF) emissions enable non-invasive, element-specific biodistribution imaging. This work presents the design, assembly, and characterization of a scalable NaI(Tl)-SiPM scintillation detector module developed as a building block for a 16-element detector array for the Confocal Orthovoltage X-Ray--Induced Radiation System (COXIRIS) Project. Each module comprises a 25.4x25.4x4 mm3 monolithic NaI(Tl) crystal optically coupled to a custom SiPM array and read out by the TOFPET2 ASIC.
The architecture targets the K-shell fluorescence lines of gold (Kα = 68 keV) and gadolinium (Kα = 43 keV), while providing a large active area at low cost.
Four modules were characterized with a 2.0 μCi 241Am source, whose 59.5 keV emission lies within the 40--80 keV target interval. The peak was clearly identified in all modules, with energy resolutions ranging from 47% to 65%. Per-pixel gain calibration produced an additional relative improvement of approximately 2% for each module. These results support the feasibility of the proposed assembly approach as a detector building block for XRF imaging applications.

AcknowledgmentThis work was supported by Project TA24I10046 from the National Research and Development Agency of Chile (ANID) and 
by Portuguese national funds through FCT – Fundação para a Ciência e a Tecnologia, I.P., under the project UID/50025/2025 (doi.org/10.54499/UID/50025/2025), UID/PRR/50025/2025 (doi.org/10.54499/UID/PRR/50025/2025), UID/PRR2/50025/2025 (doi.org/10.54499/UID/PRR2/50025/2025) and the Associate Laboratory I3N – LA/P/0037/2020 (doi.org/10.54499/LA/P/0037/2020)
References
[1] M. A. Kouri, K. Polychronidou, G. Loukas, A. Megapanou, I.-A.Vagena, A. M. Gerardos, E. Spyratou, and E. P. Eftsathopoulos, “Consolidation of gold and gadolinium nanoparticles: An extrastep towards improving cancer imaging and therapy,” Journal of Nanotheranostics, vol. 4, no. 2, pp. 127–149, Apr. 2023. [Online]. Available: https://www.mdpi.com/2624-845X/4/2/7
[2] R. G. Figueroa, M. Valente, J. Guarda, J. Leiva, E. Quilaguilque, B. Casanelli, and F. Leyton, “OXIRIS project: Development of a new XRF device for the simultaneous detection and treatment of cancer,” X-Ray Spectrometry, vol. 51, no. 3, pp. 251–261, May 2022. [Online]. Available: https://analyticalsciencejournals.onlinelibrary.wiley.com/doi/
10.1002/xrs.3272
[3] PETsys Electronics, “TOFPET2 ASIC,” https://www.petsyselectronics. com/, accessed: 2026-05-10. [Online]. Available: https://www.petsyselectronics.com/
Keywords: photon counting detector, X-ray fluorescence, theranostics, gold nanoparticles, gadolinium nanoparticles
17:30 Poster panel: 449

Poster Number:
MIC-10-449

A Preclinical PET System for Simultaneous Acquisition and Separation of Multiple Tracers (#2209)

J. W. Foster1, W. He2, M. N. Ullah2, Y. B. Shah2, G. Chinn2, C. S. Levin2, 3

1 Stanford University, Electrical Engineering, Stanford, California, United States of America
2 Stanford University, Radiology, Stanford, California, United States of America
3 Stanford University, By courtesy Electrical Engineering, Bioengineering, Physics, Stanford, California, United States of America

Abstract

Positron emission tomography (PET) is traditionally limited to scanning one tracer per session. Positron and prompt gamma emitters may be used to form triple coincidences that allow for signal separation based on energy. We present a proof-of-concept, preclinical PET system designed for detecting and distinguishing prompt gamma and annihilation photons in the same detector. We describe the data processing pipeline with advances for estimating randoms, scatter, and sensitivity of the prompt gamma photon. System performance is characterized with NEMA and Derenzo phantom studies. Next, we present multi-isotope studies using 52Mn and 18F in a custom image quality phantom. We compare two reconstruction methods and find alternating updates yield faster convergence and higher image quality. After correcting for injected activity and decay, the mean voxel intensity ratio between 18F and 52Mn across background ROIs is 0.95 ± 0.06, where 1 represents perfect quantitative agreement. The uncertainty reflects variability across ROIs and dose calibrator uncertainty (±2%), suggesting unmixing accuracy is within dose calibrator tolerance. Further, we scan the phantom after 18F decay to determine the relative image quality between single (52Mn only) and dual (52Mn with 18F) acquisition. We find the 52Mn image has similar contrast recovery (0.65/0.62) and crosstalk ratio (0.24/0.24) between the single and dual acquisition. Due to low triples sensitivity, simultaneous acquisition of both radionuclides reduces the contrast to noise from 10.2 to 4.7. Despite limitations on axial field of view, data rate, and the high background from 52Mn’s 3 gamma cascade, we achieve high quantitative accuracy. Future studies will scale up the system, use more favorable tracers, add regularization to mitigate noise propagation and include preclinical animal scans.

Acknowledgment

This work is supported by NIH F31CA265205, the Falk Medical Research Trust (SPO 268891), Stanford Bio-X IIP and an SNMMI Mars Shot Award.

References
[1] A. Andreyev, A. Sitek, and A. Celler, “EM reconstruction of dual isotope PET using staggered injections and prompt gamma positron emitters,” Medical Physics, vol. 41, no. 2, p. 022501, Feb. 2014, doi: 10.1118/1.4861714.
[2] T. Fukuchi, T. Okauchi, M. Shigeta, S. Yamamoto, Y. Watanabe, and S. Enomoto, “Positron emission tomography with additional γ-ray detectors for multiple-tracer imaging,” Medical Physics, vol. 44, no. 6, pp. 2257–2266, 2017, doi: 10.1002/mp.12149.
[3] E. Pfaehler, D. Niekämper, J. Scheins, N. J. Shah, and C. W. Lerche, “ML-EM based dual tracer PET image reconstruction with inclusion of prompt gamma attenuation,” Phys Med Biol, vol. 70, no. 1, Dec. 2024, doi: 10.1088/1361-6560/ad9660.
[4] López-Montes A, Cabello J, Conti M, Badal A, Herraiz JL, “GPU-accelerated Monte Carlo-Based Scatter and Prompt-Gamma Corrections in PET,” IEEE Nuclear Science Symposium and Medical Imaging Conference 2021, M-07-01, Nov. 2021
[5] J. Foster, G. Chinn, and C. Levin, “MLEM Method for Estimating Spatial Sensitivity Coefficients for Simultaneous Multi-Tracer Positron Emission Tomography (mPET) Reconstruction,” in 2024 IEEE Nuclear Science Symposium and Molecular Imaging Conference, Nov. 2024.
[6] A. Sitek, A. Andreyev, and A. Celler, “Reconstruction of dual isotope PET using expectation maximization (EM) algorithm,” in 2011 IEEE Nuclear Science Symposium Conference Record, Oct. 2011, pp. 4323–4326. doi: 10.1109/NSSMIC.2011.6153832.
[7] W. He, J. W. Foster, and C. S. Levin, “Alternating Update Method for Joint EM Estimation of Multiple Tracer Distributions in Multiplexed PET (mPET),” IEEE Transactions on Radiation and Plasma Medical Sciences, pp. 1–1, 2026, doi: 10.1109/TRPMS.2026.3688900.
Keywords: multiplexed PET, prompt gamma emitters, small animal PET, multi-tracer PET, data corrections
17:30 Poster panel: 452

Poster Number:
MIC-10-452

On the Pseudo Dark-Field Signal in X-Ray Scattering Tensor Tomography (#2316)

R. Deng1, R. Liu1, P. Song1, J. Lu1, L. Zhang1, Z. Wang1

1 Tsinghua University, Department of Engineering Physics, Beijing, China

Abstract

X-ray scattering tensor tomography retrieves directional dark-field projections from the focused, structured illumination produced by 2D diffractive optics, and is increasingly attractive as a route to anisotropic micro-structural imaging on laboratory sources. Quantitative tensor reconstruction requires the focused reference field to cancel out exactly in the Fourier-domain retrieval, a condition met only under monochromatic illumination. We show that beam hardening through an absorbing sample reshapes this focused reference field non-uniformly across its bright core and the surrounding plateau, so the retrieval assigns a non-zero scattering tensor to a purely absorbing object—a pseudo dark-field signal. We derive a closed-form expression for the pseudo dark-field signal, prove its exact isotropy arising from the circular symmetry of the focal spot, and identify the retrieved observables that are biased by beam hardening versus those that are not. The standard angular-consistency metric is blind to this bias, so a dedicated calibration is required. We describe a step-wedge calibration protocol that compresses the family of pseudo dark-field signals into a single thickness-dependent function suitable for per-pixel correction. Removing this artefact is a prerequisite for quantitative X-ray scattering tensor tomography on laboratory sources.

AcknowledgmentThis work was supported in part by National Natural Science Foundation of China (No. 62227804), Beijing Municipal Natural Science Foundation (No. L232115) and National Key Research and Development Program of China (No. 2023YFC2605802).
References
[1] M. Kagias, Z. Wang, P. Villanueva-Perez, K. Jefimovs, and M. Stampanoni, "2d-omnidirectional hard-x-ray scattering sensitivity in a single shot," Physical Review Letters, vol. 116, no. 9, p. 093902, 2016.
[2] M. Kagias, Z. Wang, M. E. Birkbak, E. Lauridsen, M. Abis, G. Lovric, K. Jefimovs, and M. Stampanoni, "Diffractive small angle x-ray scattering imaging for anisotropic structures," Nature Communications, vol. 10, p. 5130, 2019.
[3] F. Pfeiffer, M. Bech, O. Bunk, C. Bronnimann, C. Grünzweig, and C. David, "Hard-x-ray dark-field imaging using a grating interferometer," Nature Materials, vol. 7, no. 2, pp. 134–137, 2008.
[4] G. Giakoumakis, J. Primot, A. Jarnac, L. Guitard, and A. Stolidi, "Artifacts reduction in high-acutance phase images for x-ray grating interferometry," Optics Express, vol. 30, no. 23, pp. 41147–41156, 2022.
[5] G. Lautizi, V. Di Trapani, A. Studer, M.-C. Zdora, F. De Marco, J. Kim, F. Marone, M. Stampanoni, and P. Thibault, "Robust dark-field signal extraction for modulation-based x-ray tensor tomography," Applied Physics Letters, vol. 125, no. 26, p. 264103, 2024.
[6] M. Kagias, Z. Wang, G. Lovric, K. Jefimovs, and M. Stampanoni, "Simultaneous reciprocal and real space x-ray imaging of time-evolving systems," Physical Review Applied, vol. 15, no. 4, p. 044038, 2021.
Keywords: X-ray scattering tensor tomography, dark-field signal, beam hardening, polychromatic illumination, focusing gratings.
17:30 Poster panel: 455

Poster Number:
MIC-10-455

A Proof-of-Concept Tri-Mode PET and SPECT Imager with a Shared Scintillator-as-Collimator Detector (#2390)

X. Zeng1, H. Yang1, M. Niu1, W. Huang1, X. Jiang1, C. Huang1, Y. Zhao1, J. Tian1, X. Zhang1, Q. Wei3, J. Wu5, Y. Liang7, Z. Gu1

1 shenzhen bay laboratory, Shenzhen, China
2 University of Science and Technology Beijing, School of Automation and Electrical Engineering, Beijing, China
3 Beijing Normal University, School of Physics and Astronomy, Beijing, China
4 Cancer Hospital Chinese Academy of Medical Sciences, Shenzhen, China

Abstract

Simultaneous dual-radionuclide imaging offers complementary information for diagnosis, stratification, treatment planning, and therapy evaluation. However, the fundamental differences between PET and SPECT present significant challenges for integrated imaging. We propose a scintillator-as-collimator detector enabling both PET and SPECT. It consists of three scintillator layers separated by two light guides. Parallel holes in the upper portion function as a SPECT collimator, referred to as the scintillator-as-collimator volume (SCV). The lower portion serves as the SPECT detector (SD). In PET mode, the entire detector acts as a multi-resolution TOF-DOI PET detector. Thus, it serves as a self-collimated SPECT or TOF-DOI PET detector. A dual-panel prototype was developed with rotational acquisitions for PET, SPECT, and simultaneous imaging. SPECT events are selected using flood image mask, maximum SiPM channel energy (Pmax), and DOI-integrated charge (QDC). SPECT resolution reaches ~5 mm. PET performance is preserved, with best SD–SD CTR of 352 ps and two-layer DOI. The introduction of holes has minimal impact on PET detection, enabling high-quality PET images. In simultaneous imaging, a unified timing framework rejects 511 keV down-scatter, and residual artefacts were further mitigated using a 200–240 keV energy window for scatter correction. Phantom experiments validate tri-mode imaging. This detector eliminates external collimators, supports multi-tracer imaging, and offers a cost-effective single-detector-set PET&SPECT system.

Acknowledgment

This work is supported by the National Key R&D Program of China (No. 2023YFC2414800), the Major Program of Shenzhen Bay Laboratory (No. C1012530005), the National Natural Science Foundation of China (Nos. 12275182 and 62371314), Guangdong Basic and Applied Basic Research Foundation (Nos. 2024A1515012154), the Guangdong Provincial Project (Grant No. 2023TQ07A233) and the Shenzhen Bay Laboratory Proof-of Concept Grants (No. C1022430005).

References
[1] Huang, S. et al. (2026). Optimization of TOF and DOI Performance in a Multi-Resolution Detector for Brain-Dedicated PET. IEEE Transactions on Radiation and Plasma Medical Sciences, 10(3).
[2] Zeng X et al. (2025). Tri-Mode PET, SPECT, and PET&SPECT Imager Using a Novel Scintillator-as-Collimator Detector. 2025 IEEE NSS, MIC and RTSD, Yokohama, Japan.
Keywords: Scintillator-as-Collimator, Simultaneous imaging, PET&SPECT detector, Depth-of-interaction, Time-of-flight
17:30 Poster panel: 458

Poster Number:
MIC-10-458

Zero-Shot PET Image Super-Resolution via Cross-Modal Mutual Information Guidance (#2526)

Y. He1, Y. Lv1

1 Shanghai United Imaging Medical Technology Co., Ltd, Molecular Imaging Business Unit, Shanghai, China

Abstract

Diffusion models have shown strong potential for improving PET image quality. However, existing methods based on unconditional pretrained diffusion priors often struggle to balance realism and fidelity, particularly in preserving high-resolution details. To address this issue, we propose MI-ResShift, a zero-shot mutual information-guided diffusion model with residual shifting for PET image super-resolution. MI-ResShift introduces a differentiable local-wise mutual information layer to guide the denoising process using structural information from CT or MRI, enabling robust cross-modal feature alignment in the statistical domain. By combining mutual information guidance with residual shifting, the proposed method enhances PET detail fidelity and robustness while reducing diffusion-based inference to only four steps. Experiments on PET/MR brain datasets demonstrate that MI-ResShift improves image quality over classical interpolation methods and provides an efficient solution for cross-modal PET super-resolution.

AcknowledgmentThis work was supported in part by Shanghai United Imaging Medical Technology Co., Ltd., through its internal research program. The authors gratefully acknowledge the Molecular Imaging team for their technical support and helpful discussions.
References
[1] Z. Wang et al., "Mutual Information Guided Diffusion for Zero-Shot Cross-Modality Medical Image Translation," in IEEE Transactions on Medical Imaging, vol. 43, no. 8, pp. 2825-2838, Aug. 2024, doi: 10.1109/TMI.2024.3382043.
[2] Z. Yue, J. Wang and C. C. Loy, "Efficient Diffusion Model for Image Restoration by Residual Shifting," in IEEE Transactions on Pattern Analysis and Machine Intelligence, vol. 47, no. 1, pp. 116-130, Jan. 2025, doi: 10.1109/TPAMI.2024.3461721.
Keywords: Positron emission tomography, super-resolution, cross-modality, diffusion model, zero-shot learning
17:30 Poster panel: 461

Poster Number:
MIC-10-461

Extracting total and Compton attenuation coefficients via ghost imaging-like photon counting (#2895)

G. R. Romanchek1, 2, P. G. Bharathi2, S. Abbaszadeh2

1 University of Illinois Urbana-Champaign, Department of Nuclear, Plasma, and Radiological Engineering, Urbana, Illinois, United States of America
2 University of California Santa Cruz, Department of Electrical and Computer Engineering, Santa Cruz, California, United States of America

Abstract

Coincidence-based imaging with the back-to-back 511 keV photon pairs emitted during positron annihilation enables transmission imaging with reduced quantum noise. Conventional transmission imaging recovers only a single attenuation-related quantity per voxel or pixel. In this work, we extend coincidence-based imaging to simultaneously recover both the total and the Compton-scatter linear attenuation coefficients from a single acquisition, providing a second axis for material identification. Forward-scattered photons collected in coincidence within an analytically defined polar window, obtained by integrating the Klein-Nishina differential cross section over a 20-to-30-degree band, enable direct estimation of the Compton macroscopic cross section, while the idler channel provides the incident count for the total coefficient. We evaluate the approach using Monte Carlo simulations of opposing cadmium zinc telluride panels viewing a water phantom of thickness 2 to 10 millimeters. Idler-normalized transmission was recovered to within 0.1 percent across all thicknesses, while the extracted total and Compton coefficients exhibited 1 to 4 percent absolute error, with the Compton coefficient transitioning from over- to underestimation near 6 millimeters. These deviations arise from logarithmic amplification of small counting errors and from competing air-scatter contamination at thin phantoms and self-attenuation of scattered photons at thicker ones. The results establish that scatter data from coincidence-based annihilation imaging can recover a second material descriptor, motivating further development toward dose-efficient material identification.

AcknowledgmentThis work was supported by the National Institutes of Health under Grant R01-EB028091.
References
[1] Romanchek, G, Yu, Z, Abbaszadeh, S 2026, 'Monte Carlo validation of signal-to-noise improvements in transmission imaging using correlated annihilation photon pairs', Scientific Reports, London: Nature Portfolio.
[2] Padgett, MJ, Boyd, RW 2017, 'An introduction to ghost imaging: quantum and classical', Philosophical Transactions of the Royal Society A: Mathematical, Physical and Engineering Sciences, vol. 375, no. 2099, 20160233, London: The Royal Society.
[3] Sarrut, D, Bała, M, Bardiès, M, Bert, J, Chauvin, M, Chatzipapas, K, Dupont, M, Etxebeste, A, Fanchon, LM, Jan, S, Kayal, G, Kirov, AS, Kowalski, P, Krzemień, W, Labour, J, Lenz, M, Loudos, G, Mehadji, B, Ménard, L, Morel, C, Papadimitroulas, P, Rafecas, M, Salvadori, J, Seiter, D, Stockhoff, M, Testa, E, Trigila, C, Pietrzyk, U, Vandenberghe, S, Verdier, MA, Visvikis, D, Ziemons, K, Zvolský, M, Roncali, E 2021, 'Advanced Monte Carlo simulations of emission tomography imaging systems with GATE', Physics in Medicine & Biology, vol. 66, no. 10, 10TR03, Bristol: IOP Publishing.
Keywords: Quantum imaging, gamma-ray imaging, positron annihilation, Compton scattering, attenuation coefficient estimation
17:30 Poster panel: 464

Poster Number:
MIC-10-464

Enhancement of Charged-Particle Shielding via Lightweight Nanocomposites (#2978)

G. Grey1, N. Speck2, D. Vecchio3, K. Cotterill4, 5, A. Taub1, M. Hammig2

1 University of Michigan, Materials Science & Engineering Department, Ann Arbor, Michigan, United States of America
2 University of Michigan, Nuclear Engineering & Radiological Sciences, Ann Arbor, Michigan, United States of America
3 Amphionic, LLC., Plymouth, Michigan, United States of America
4 University of Michigan, Department of Math, Ann Arbor, Michigan, United States of America
5 University of Michigan, Department of Physics, Ann Arbor, Michigan, United States of America

Abstract

The development of lightweight radiation shielding materials is of high interest for a broad range of defense, space, and medical applications. In particular, mass-savings are desired in vehicles (ground, naval, and/or aerospace) and as wearable fabrics in high dosage environments such as those encountered in the medical setting. Here we investigate the potential of enhanced shielding via radiation attenuation of nanostructured materials which take advantage of interfacial and nanocrystalline loss processes increase the multiplicity of energy loss mechanisms within the shielding solid. In this work, colloidal metallic nanoparticles grown in an aqueous environment (such as palladium (Pd) and silver (Ag)) with sizes in the 5 - 50 nm range are imbedded in a polymeric matrix material (epoxy or aramid nanofiber (ANF)) to demonstrate photon and electron shielding while maintaining mechanical properties such as strength, durability, and form-factor flexibility. The mechanisms of energy loss at the interfaces of the nanomaterials and overall in the solid are quantified through transport modeling as well as the unorthodox usage of the Transmission Electron Microscope (TEM), in which a focused electron beam and state-of-the-art detector system are utilized to measure the effects of particle interactions within our material system to a near single particle resolution. Therefore, this work encompasses both an investigation of the potential enhancement of radiation attenuation by nanostructured materials while also introducing an innovative nanocomposite material system with manufacturing versatility enabling lightweight, flexible shielding.

Acknowledgment

The authors would like to acknowledge the technical support of the University of Michigan’s Michigan Center for Materials Characterization and Michigan Medicine Microscopy Core. G. J. Grey & N. P. Speck give special thanks to Dr. Tao Ma for guidance on TEM operation and Megan Bennett for support on ultramicrotome sample preparation.

References
[1] M. Wang et al., “Review of Passive Shielding Materials for High-Energy Charged Particles in Earth’s Orbit,” Materials (Basel), vol. 18, no. 11, p. 2558, May 2025, doi: 10.3390/ma18112558.W.-K. Chen, Linear Networks and Systems. Belmont, CA, USA: Wadsworth, 1993, pp. 123–135.
[2] N. Ipe, “Transmission of Shielding Materials for Particle Therapy Facilities,” Nuclear Technology, no. 168(2), pp. 559–563, Apr. 2017, doi: https://doi.org/10.13182/NT09-A9243.
[3] R. L. Maughan, M. J. Hardy, M. J. Taylor, J. Reay, and R. Amos, “Radiation shielding and safety for particle therapy facilities,” in Design and Shielding of Radiotherapy Treatment Facilities: IPEM Report 75, 2nd Edition, IOP Publishing, 2017. doi: 10.1088/978-0-7503-1440-4ch11.
[4] Q. Chang, S. Guo, and X. Zhang, “Radiation shielding polymer composites: Ray-interaction mechanism, structural design, manufacture and biomedical applications,” Materials & Design, vol. 233, p. 112253, Sep. 2023, doi: 10.1016/j.matdes.2023.112253.
[5] D. A. Vecchio, M. D. Hammig, and N. A. Kotov, “High-Resolution Radiation Sensors from Flexible Network Nanocomposites of Nanoparticles and Aramid Nanofibers,” ACS Nano, vol. 19, no. 12, pp. 11924–11935, Apr. 2025, doi: 10.1021/acsnano.4c15939.
[6] K. A. Cotterill et al., “Synthesis of Palladium Nanoparticles for Multifunctional Applications in Space and Hydrolysis Environments,” unpublished.
[7] J. Lyu et al., “Stretchable conductors by kirigami patterning of aramid-silver nanocomposites with zero conductance gradient,” Appl. Phys. Lett., vol. 111, no. 16, p. 161901, Oct. 2017, doi: 10.1063/1.5001094.
Keywords: Nanomaterial, Nanocomposite, Shielding, Particle Interaction, Lightweight
17:30 Poster panel: 467

Poster Number:
MIC-10-467

Detector Geometry Optimization of a Sensing Collimation Imager (SCI) (#3067)

L. Cai1, H. Xia2, S. Aggarwal1, F. Ahmed2, J. A. O’Sullivan1, 3, A. K. Jha2, 4, Y.-C. Tai3, 4

1 Washington University in St. Louis, Imaging Science, St. Louis, Missouri, United States of America
2 Washington University in St. Louis, Department of Biomedical Engineering, St. Louis, Missouri, United States of America
3 Washington University in St. Louis, Department of Electrical and Systems Engineering, St. Louis, Missouri, United States of America
4 Washington University in St. Louis, Mallinckrodt Institute of Radiology, St. Louis, Missouri, United States of America

Abstract

The Sensing Collimation Imager (SCI) is an ultrasensitive gamma detector that integrates multi-pinhole, coded aperture, and self-collimation imaging into a single compact device, achieving substantially higher sensitivity than conventional gamma cameras. Maximizing the spatial encoding capability of the SCI module requires optimizing the crystal arrangement over a high-dimensional combinatorial design space, which is too large for exhaustive search.

We propose an information-theoretic detector geometry optimization framework that uses the Fisher Information Matrix (FIM) spatial correlation map as a computationally efficient performance metric. The metric requires only computing the FWHM of the FIM column vector (O(n)), avoiding the O(n³) cost of inverting the FIM for the Cramér–Rao bound, enabling broad geometry searches. A genetic algorithm (GA) guided by this metric searches over fixed-cardinality crystal arrangements to identify high-performing designs.

Applied to the SCI module, the GA converged within 50 generations and identified an optimized arrangement producing narrower FIM correlation peaks throughout the field of view than a heuristic design. GATE Monte Carlo simulations of a Derenzo phantom confirmed improved recovery of the 8 mm sphere group with clearer feature separation, validating the optimization framework.

AcknowledgmentThis work was supported in part by the Mallinckrodt Institute of Radiology at Washington University at St. Louis through an internal fund allotment. We also acknowledge the support from NIH (R01-EB031962, R01-CA240711) and NSF CAREER Award (2239707). 
References
[1] Cai et al., "An ultra-high sensitivity sensing collimation imager: design, modelling, and assessment," Proc. SPIE 13405, Medical Imaging 2025: Physics of Medical Imaging
[2] H H, Gifford H. Cone-beam tomography with discrete data sets[J]. Physics in Medicine & Biology, 1994, 39(3): 451.
[3] Qi and R. H. Huesman, "Wavelet crosstalk matrix and its application to assessment of shift-variant imaging systems," 2002 IEEE Nuclear Science Symposium Conference Record
Keywords: Sensing Collimation Imager, detector optimization, Fisher Information Matrix, genetic algorithm
17:30 Poster panel: 470

Poster Number:
MIC-10-470

Toward Whole-Body Whole Gamma Imaging: Monte Carlo Simulation of 89Zr PET-Compton Hybrid Imaging (#3156)

M. Miyahara1, 2, H. Tashima2, M. Suga3, S. Takyu2, F. Nishikido2, G. Akamatsu2, T. Yamaya2

1 Chiba University, Graduate School of Science and Engineering, Chiba, Japan
2 National Institutes for Quantum Science and Technology, Institute for Quantum Medical Science, Chiba, Japan
3 Chiba University, Center for Frontier Medical Engineering, Chiba, Japan

Abstract

Whole gamma imaging (WGI) is a novel imaging technique that combines PET and Compton imaging by inserting a scatterer detector ring inside a PET detector ring. One of the suitable nuclides for WGI is 89Zr. 89Zr emits a 909 keV single gamma ray after undergoing positron emission or electron capture to form 89mY. Because these 909 keV gamma rays are emitted approximately four times more frequently than positrons, hybrid imaging of WGI combining both modalities is expected to improve image quality compared to conventional 89Zr-PET which uses only annihilation gamma rays. In this study, we aimed at finalizing the design of a clinical whole-body WGI system. The hybrid imaging method was applied to Geant4-simulated data of a whole-body WGI system. We modeled a system employing HR-GAGG crystals as the scatterer and LGSO crystals as the absorber. Measurement of an 89Zr-filled NEMA IEC body phantom placed at the center of the field of view was simulated. Images were reconstructed using the hybrid reconstruction method combining PET and Compton events and compared with PET images reconstructed by the list-mode OSEM algorithm using only PET events. We calculated the percent contrast (QH,j) and percent background variability (Nj) for each sphere size j (10-37 mm). The results showed that, for large spheres with diameters of 22 mm or more, hybrid imaging reduced Nj, but the maximum achievable QH,j was  lower than that of PET. In contrast, for small spheres with diameters of 17 mm or less, hybrid imaging demonstrated a specific advantage by suppressing noise amplification while accelerating contrast recovery in the early iterations. In conclusion, although the overall image quality improvement was limited in certain aspects, the potential to enable novel applications such as multi-isotope imaging and positronium lifetime imaging using appropriate position emitters was suggested.

Keywords: Whole gamma imaging, Positron emission tomography, Compton camera, Human whole-body imaging, 89Zr