High resolution imaging systems (organ-dedicated, small animal systems, X-ray imaging systems (CT, spectral CT, photo-counting CT), Quantitative imaging, image assessment and standardization

Session chair: Ma, Tianyu, (Tsinghua University, Department of Engineering Physics, Beijing, China); Zannoni, Elena Maria, Dr. (The University of Texas at Austin, Austin, USA)
 
Shortcut: M-LT-02
Date: Wednesday, 11 November, 2026, 12:45 - 13:05
Location: MANUEL DE FALLA
Session type: MIC Session

Contents

Click on an contribution to preview the abstract content.

The underlined authors are the presenters. The authors in bold are the main authors.

12:45 M-LT-02-01

Experimental Results from Novel Monolithic PET Detector and Application to Axillary Lymph Node Imaging (#1769)

T. Cotroneo1, D. Pare1, L. Saleh1, P. Vaska1, 2

1 Stony Brook University, Biomedical Engineering, Stony Brook, New York, United States of America
2 Renaissance School of Medicine at Stony Brook University, Radiology, Stony Brook, New York, United States of America

Abstract

Monolithic scintillators are a promising alternative to standard pixelated crystals in positron emission tomography, as they can be more cost-effective, reduce dead space, and capture an unobstructed light distribution, which can be used for depth of interaction estimation. We focus on the improvements seen by using an array of integrated corner reflectors (ICR) machined directly into the entrance face of the monolith, which we previously showed via simulation to improve light collection and narrow the light distribution. In this study, we built a physical prototype of the device using a CNC machining process. We compared this prototype with other monolithic block designs. We show that the ICR monolith with the CNN positioning algorithm has potential for high-resolution imaging with ~1 mm resolution at 12 mm thickness. Using GATE simulations, we constructed a full scanner with the performance of the monolithic blocks and tested the capabilities compared to existing systems. We show the improved spatial resolution and feasibility for a non-cylindrical design for axillary lymph node scanning.

AcknowledgmentWe thank the Carol M. Baldwin Foundation for Breast Cancer Research, Stony Brook University Graduate School, and the SBU OVPR seed funding program for their support.
References
[1] A. Gonzalez-Montoro et al., "Evolution of PET Detectors and Event Positioning Algorithms Using Monolithic Scintillation Crystals," IEEE T Radiat Plasma, vol. 5, no. 3, pp. 282-305, 2021, doi: 10.1109/TRPMS.2021.3059181.
[2] Y. Wang, W. Zhu, X. Cheng, and D. Li, "3D position estimation using an artificial neural network for a continuous scintillator PET detector," Phys Med Biol, vol. 58, no. 5, pp. 1375-90, Mar 7 2013, doi: 10.1088/0031-9155/58/5/1375.
[3] R. A. Valdés, M. Pereira, Daphne, and van, “Breast Cancer,” Springer eBooks, pp. 1–37, Jan. 2022, doi: https://doi.org/10.1007/978-3-031-05497-6_1.
Keywords: Monolithic Scintillator, Integrated Corner Reflectors, Axillary Lymph Node Scanner, Breast Cancer, LYSO
12:47 M-LT-02-02

Spatial Characterization of PET Detectors Based on BGO Pseudo‑slabs (#1981)

N. Cucarella1, L. Gil-Bosch1, F. Paganao1, F. Loignon-Houle1, M. Gil1, M. Freire1, D. Sanchez2, A. Gonzalez-Montoro1, A.J. Gonzalez1

1 Instituto de Instrumentación para Imagen Molecular (I3M), Centro Mixto CSIC-Universitat Politècnica de València,, DMIL, Valencia, Spain
2 Oncovision SA, Valencia, Spain

Abstract

Pixelated crystal arrays are the standard in commercial PET scanners. However, these detectors usually lack DOI information, essential to achieve a homogenous spatial resolution along the entire FOV. Yet, studies have been already performed with pixelated detectors with DOI capabilities but at the cost of using extra hardware material and/or crystal assembling. Inspired by semi‑monolithic crystal geometries, pseudo-slabs were proposed as an alternative. This geometry consists of an N×N pixel array optically isolated only in one direction, allowing light crosstalk in the orthogonal one. The N pixels sharing light in this direction form a pseudo-slab. The light sharing within a pseudo-slab is enough to enable DOI retrieval, while the physical separation between pixels confines the light more than in a pure slab geometry, yielding better time resolution.

In this work we evaluated the DOI resolution of a BGO block formed by 8×8 pixels, each measuring around 3×3×15 mm3. Two different readout configurations were employed and compared. First, all the 64 channels corresponding to the TOFPET2 ASIC channels were digitized. Then, a multiplexing circuit was utilized, reducing the signals to 8+8 by summing rows and columns. The DOI was obtained using neural networks trained on experimental data.

The results show a good DOI resolution for both readout configurations, achieving 3.8 ± 1.3 mm with the multiplexing circuit. These results show that pseudo-slabs are suitable for clinical PET scanners with DOI capabilities.
Acknowledgment

This work was partially funded by the AEI (Grant PID2023-151999OB-I00), co-funded by the ERC (Grant No. 101164363). Views and opinions expressed are, however, those of the author(s) only and do not necessarily reflect those of the European Union or the ERC. Neither the European Union nor the granting authorities can be held responsible for them.

N.C. is supported by PTA2022-022459-I. M.F. and F.L.-H. are supported by CIAPOS/2023/141 and CIAPOS/2023/133, respectively.

References
[1] M. Ito et al., Biom. Eng. Let., 2011. DOI: 10.1007/s13534-011-0019-6.
[2] N. Cucarella et al., TRPMS, 2025. DOI: 10.1109/TRPMS.2025.3546998.
[3] R. Bugalho et al., JINST, 2019. DOI: 10.1088/1748-0221/14/03/P03029.
[4] A.J. Gonzalez et al., arXiv, 2926. DOI: 10.48550/arXiv.2603.09898.
Keywords: BGO, DOI, PET detectors, pseudo-slab, semi monolithic
12:49 M-LT-02-03

SurfaceCam: A Close-Proximity Multi-Pinhole CZT Brain SPECT system with Enhanced Geometric Sensitivity and Spatial Resolution (#3121)

S. Mandot1, S. D. Metzler2, L.-J. Meng1, 3

1 University of Illinois at Urbana-Champaign, Department of Nuclear, Plasma and Radiological Engineering, Champaign, Illinois, United States of America
2 University of Pennsylvania at Philadelphia, Department of Radiology, Philadelphia, Pennsylvania, United States of America
3 University of Illinois at Urbana-Champaign, Beckman Institute for Advance Science and Technology, Champaign, Illinois, United States of America

Abstract

High-sensitivity, high-resolution brain-dedicated SPECT systems are critically important for advancing diagnosis and surgical planning in complex neurological disorders such as epilepsy, Parkinson's disease, and Alzheimer's disease. In this paper, we present SurfaceCam, a close-proximity multi-pinhole SPECT architecture with a radially adaptable rotating gantry, designed specifically for clinical brain imaging. The system comprises 15 detector-collimator panels distributed across three detector heads in a cylindrical arrangement, operating at a 1:8 minification geometry with a 28 cm bore. Each panel integrates 16 high-performance CZT sensors in a compact 2×8 configuration. The three detector heads translate radially to accommodate varying patient head sizes while maintaining close-proximity geometry, and the rotating gantry enables micro-rotation between acquisitions to facilitate angular super-sampling strategies. Pinholes are organized into groups of four with a twisted geometry, each subtending partial coverage of the 20 cm spherical FOV, while eliminating inter-pinhole projection multiplexing on the detector surface. Analytical comparison against a conventional 1:11 minification compound-eye reference design demonstrates approximately 1.4-fold higher geometric sensitivity and approximately 27% improvement in effective pinhole spatial resolution. The front-end electronics and data acquisition systems have been fully developed and validated, with Monte Carlo simulation studies and experimental phantom validation currently underway.

Acknowledgment

We would like to thank NIH/NIBIB (BRAIN Initiative) for supporting this work under the grant UG3 EB034715.

Keywords: Single Photon Emission Computed Tomography, Close-Proximity Geometry, Angular Super-Sampling, Multi-Pinhole Collimation, Cadmium Zinc Telluride
12:51 M-LT-02-04

In vivo Gold Nanoparticle Imaging in Mice for Tumor Detection Using SiPM-Based Photon-Counting CT (#1749)

S. Midorikawa1, M. Arimoto1, F. Lucyana1, Y. Furuta1, S. Kobayashi1, H. Kawashima1, K. Okumura1, A. Mizutani1, N. Ohno1, M. Oshima1, J. Kataoka2, R. Takahiro2, R. MInagawa2, K. Murakami3, Y. Maeno4, S. Terazawa4

1 Kanazawa University, Ishikawa, Japan
2 Waseda University, Tokyo, Japan
3 Ishikawa Prefectural University, Ishikawa, Japan
4 Proterial Ltd, Tokyo, Japan

Abstract

Photon-counting CT (PC-CT) is a powerful spectral imaging technique that identifies materials by counting individual X-ray photons. While direct-conversion systems are entering clinical practice, we are developing a cost-effective, large-area scintillator-based PC-CT system using YGAG:Ce coupled with silicon photomultipliers (SiPMs). PC-CT excels at identifying high-atomic-number elements like gold (Au) by utilizing its K-edge (80.7 keV). Gold nanoparticles (AuNPs) are promising for drug delivery, as their biodistribution can be controlled by particle size, enabling selective tumor accumulation via the enhanced permeability and retention (EPR) effect.

In this study, we evaluated the in vivo biodistribution of AuNPs in tumor-bearing mice using our SiPM-based PC-CT. Two sizes of AuNPs (1.9 nm and 15 nm) were intravenously administered, and their accumulation was quantitatively assessed using multi-energy CT concentration estimation. Results showed distinct size-dependent patterns: 1.9 nm AuNPs exhibited rapid renal clearance and bladder accumulation, while 15 nm AuNPs showed prolonged circulation and significant hepatic accumulation. Tumor accumulation was confirmed for both sizes, demonstrating the potential of our SiPM-based PC-CT for quantitative in vivo nanoparticle imaging.This presentation reports on the
experimental setup, the analysis methods, and the results obtained.

AcknowledgmentWe gratefully acknowledge financial support from JSPS KAKENHI Grant Numbers JP19H04483 and JP19K22924, JST ERATO Grant Number JPMJER2102, Japan, the Naito Foundation, Japan, the Uehara Memorial Foundation, Japan, the Casio Science Promotion Foundation, Japan, the Murata Science Foundation, Japan, and the JSPS Leading Initiative for Excellent Young Researchers program, Japan .
References
[1] H. Kiji, et al., “64-channel photon-counting computed tomography using
a new MPPC-CT system,” Nucl. Instrum. Methods Phys. Res. A, vol.
984, p. 164610, 2020, doi: 10.1016/j.nima.2020.164610.
[2] M. Arimoto, et al., “Development of 64-channel LSI with ultrafast
analog and digital signal processing dedicated for photon-counting
computed tomography with multi-pixel photon counter,” Nucl. In-
strum. Methods Phys. Res. A, vol. 1047, p. 167721, 2023, doi:
10.1016/j.nima.2022.167721.
[3]  Y. Furuta, et al., “X-ray imaging of anticancer drugs and a Platinum-
Iodine mixture encapsulated in liposomes using SiPM-based photon-
counting CT,” IEEE Trans. Nucl. Sci., vol. 73, no. 1, pp. 1–1, 2026, doi:
10.1109/TNS.2026.3662738
Keywords: photon counting CT, in-vivo, AuNP, MPPCs
12:53 M-LT-02-05

Quantitative Evaluation of Liver Disease through Multi-modality Approach Combining SiPM-based PC-CT and MRI—Approach to the Visualization of Inflammation through Water Fraction Estimation — (#1548)

Y. Furuta1, M. Arimoto1, F. Lucyana1, S. Midorikawa1, S. Kobayashi1, H. Kawashima1, K. Okumura1, N. Ohno1, T. Tomoda1, M. Oshima1, K. Murakami2, J. Kataoka3, R. Minagawa3, T. Ro3, S. Terazawa4, Y. Maeno4

1 Kanazawa University, Kanazawa, Japan
2 Ishikawa Prefectural University, Nonoichi, Japan
3 Waseda University, Shinzyuku, Japan
4 Proterial Ltd., Mishima, Japan

On behalf of the joint research group of Kanazawa University, Waseda University, Ishikawa Prefectural University, and Proterial Ltd.

Abstract

X-ray computed tomography (CT) is essential nondestructive internal imaging. Photon-counting computed tomography (PC-CT) acquires X-ray energy information to enable material identification and quantitative density evaluation. Combining PC-CT energy information with magnetic resonance imaging (MRI) data is expected to provide a novel diagnostic approach to visualize inflammation through tissue water fraction estimation. We developed a PC-CT system combining high-speed scintillators (YGAG:Ce) and silicon photomultipliers (SiPMs), specifically utilizing multi-pixel photon counters, MPPCs, which are optical semiconductor devices. This study evaluated healthy and fatty liver rat models (fed a high-fat diet for two or four weeks). Liver fat fractions, estimated from energy-dependent linear attenuation coefficients, showed good agreement with MRI and pathological results. Furthermore, we performed a detailed pathological evaluation focusing on inflammation induced changes in tissue water content. By integrating high reliability MRI fat fraction data with PC-CT data, we improved the estimation accuracy of water content and quantitatively visualized tissue ”water-richness” (edema). Increased water proportions in fatty liver groups suggested inflammation progression. These findings demonstrate that PC-CT is highly valuable for quantitative evaluation of liver disease within a multi-modality imaging approach.

Acknowledgment

We gratefully acknowledge financial support from JSPS KAKENHI Grant Numbers JP19H04483 and JP19K22924, JST ERATO Grant Number JPMJER2102, Japan, the Naito Foundation, Japan, the Uehara Memorial Foundation, Japan, the Casio Science Promotion Foundation, Japan, the Murata Science Foundation, Japan, and the JSPS Leading Initiative for Excellent Young Researchers program, Japan.

References
[1] H. Kiji et al. Nucl. Instr. Meth. A, 984:164610, 2020, DOI. 10.1016/j.nima.2020.164610.
[2] M. Arimoto et al. Nucl. Instr. Meth. A, 1047:167721, 2023, DOI. 10.1016/j.nima.2022.167721.
Keywords: photon-counting computed tomography (PC-CT), multi-pixel photon counters (MPPC), Multi-modality, magnetic resonance imaging (MRI)
12:55 M-LT-02-06

Position-, direction- and contrast-dependent measurement of MLEM-reconstructed PET resolution using a 3D-printed cube (#1967)

S. Alvarado Orozco1, M. Sevenois2, G. Schramm1

1 KU Leuven, Department of Imaging and Pathology, Leuven, Belgium
2 UZ Leuven, Division of Nuclear Medicine, Leuven, Belgium

Abstract

The reconstructed spatial resolution of a PET image is determined by both intrinsic scanner properties and the chosen reconstruction algorithm. Standard NEMA measurements use point sources reconstructed with FBP and  therefore mainly characterize the scanner.  They are thus suboptimal for predicting the resolution of images reconstructed with  MLEM and point-spread-function modeling, as is common in clinical practice.

We propose a 3D-printed resolution cube with an off-center inner compartment which, combined with a mid-acquisition phantom flip and a single reconstruction, produces sharp edges at finite (2:1) contrast along three orthogonal directions. Position-, direction- and contrast-dependent edge and line spread functions (ESFs, LSFs) are then extracted using smoothing-spline fits.  As a proof of concept, the method is applied to the NeuroExplorer brain PET system at the isocenter and 10 cm off-center, with and without depth-of-interaction (DOI) modeling.

The LSFs are non-Gaussian with pronounced negative side lobes.  With DOI, the resolution is approximately 2.5 mm FWHM in all directions at both positions; without DOI it degrades radially to about 4 mm at 10 cm off-center.  Notably, differences between finite and infinite contrast are small in most configurations.

The method provides a position-, direction- and contrast-dependent characterization  of PET resolution from clinically reconstructed images and complements NEMA point-source values for image quality comparison, harmonization and partial volume correction.

Keywords: PET, spatial resolution, edge spread function
12:57 M-LT-02-07

A Clinically Viable, Scalable Implementation of Monte Carlo Scatter Correction for PET/CT (#2119)

D. Zeltser1, E. Hait-Fraenkel1, X. Zhang2, M. Edelstein1, G. Elkin-Basudo1, Y. Mizrachi1, Y. Grobshtein1

1 GE HealthCare, Molecular Imaging, Haifa, Israel
2 GE HealthCare, Molecular Imaging, Waukesha, Wisconsin, United States of America

Abstract

Monte Carlo scatter correction is well established as a highly accurate approach for scatter estimation in PET/CT imaging. It demonstrates superior robustness compared with model based scatter estimation methods, particularly under extreme variations in activity distribution and in the presence of patient motion, conditions under which model based approaches often produce artifacts. Despite these advantages, Monte Carlo scatter correction is computationally intensive, and does not inherently account for out-of-field scatter. In this work, we address both limitations and propose an implementation suitable for integration into routine clinical workflows. A multi-GPU implementation, as well as parallel processing with efficient data streaming significantly reduce the processing time. To handle out-of-field scatter, we separately handle regions where information about out-of-field distribution is known, and can be used for a highly precise estimation. In regions where out-of-field information is inaccessible, the effect on scatter is approximated by the measured data, while ensuring a smooth transition in which the high-accuracy region governs the final estimate. Our approach is scalable and robust across system designs ranging from a standard axial field-of-view to a long axial field-of-view, each presenting distinct considerations for scatter estimation. We demonstrate the effectiveness of our method using phantom and clinical data.

Keywords: PET, scatter, Monte Carlo scatter
12:59 M-LT-02-08

Evaluation of Motion Correction Methods on the NeuroEXPLORER Using Multi-Tracer Human Data (#2130)

J. Zhang1, T. Zeng1, Y. Hu1, K. Fontaine1, T. Volpi1, T. Toyonaga1, R. E. Carson1

1 Yale University, New Haven, Connecticut, United States of America

Abstract

Accurate motion correction is critical for dynamic brain PET, particularly on the high-resolution NeuroEXPLORER (NX), where residual head motion can blur focal targets and bias time-activity curves (TACs). We evaluated multiple motion correction (MC) strategies using 17 multi-tracer human NX PET scans, including no motion correction (NMC), post-reconstruction registration (PRR), registration before attenuation correction (RBA), UMT-based frame-averaged correction (AVG), and UMT-based event-by-event correction (EBE). Performance was assessed using qualitative image comparison and ROI-based residual TAC variability (%SDres) from 20-70 min. All MC methods reduced %SDres compared with NMC. Across all tracers, subjects, and ROIs, EBE achieved the lowest residual variability (1.40 ± 1.27%), followed by PRR (1.79 ± 1.49%), RBA (1.94 ± 1.64%), AVG (2.01 ± 1.63%), and NMC (4.68 ± 4.26%). Poisson noise-floor analysis further showed that residual motion effects were tracer- and count-statistics-dependent. Overall, UMT-based EBE provided the most effective reduction of residual motion effects in dynamic NX brain PET.

AcknowledgmentResearch funding support: U01EB029811.
References
[1] T. Volpi, et al., "Exceptional brain PET images from the NeuroEXPLORER: scans with targeted radiopharmaceuticals and comparison to HRRT," EJNMMI, 2026.
[2] E. M. Revilla, et al., "Adaptive data-driven motion detection and optimized correction for brain PET," Neuroimage, 2022.
[3] Z. Cai, et al., "PET Head Motion Estimation Using Supervised Deep Learning with Attention," IEEE TMI, 2025.
[4] T. Zeng, et al., "Validation and application of markerless head motion tracking for a next-generation brain PET scanner," IEEE MIC, 2023.
[5] T. Zeng, et al., "Optimization of Brain PET Markerless Head Motion Correction in a Large Human Cohort," EJNMMI Physics, 2026.
[6] K. Fontaine, et al., "MOLAR-NX: building a PET reconstruction framework for exploring the novel features provided by the NeuroEXPLORER," IEEE MIC, 2024.
[7] J. Zhang, et al., "Cranberry: Combined Rigid and Nonrigid Event-by-Event Motion Correction at Yale," IEEE MIC, 2025.
[8] C. Sun, et al., "An objective evaluation method for head motion estimation in PET—motion corrected centroid-of-distribution," Neuroimage, 2022.
[9] J. Zhang, et al., "Evaluation of motion correction quality in brain, face, and neck for NeuroEXPLORER," SNMMI, 2024.
Keywords: NeuroEXPLORER, brain PET, motion correction, event-by-event, time-activity curve
13:01 M-LT-02-09

SSS Optimised for Monte Carlo Scatter Estimation for PET (#3060)

A. Reeves3, 2, S. ZamanPour1, S. Noe1, S. Arridge4, K. Thielemans2, 3

1 KU Leuven, Department of Imaging and Pathology, Leuven, Belgium
2 University College London, Institute of Nuclear Medicine and the Hawkes Institute, London, United Kingdom
3 University College London, Department of Medical Physics and Biomedical Engineering, London, United Kingdom
4 University College London, Department of computer science, London, United Kingdom

Abstract

Recent increases in computational power as well as new tools being more readily available have led to increased interest in Monte Carlo scatter estimation. However, Monte Carlo scatter simulation is still computationally expensive. This is exacerbated by the fact that model-based scatter estimation requires multiple iterations of scatter simulation and image reconstruction. We propose to combine the current clinical standard (analytic single scatter simulation and tail-fitting) with Monte Carlo scatter simulation.

AcknowledgmentThis work was supported by the i4Health Centre for Doctoral Training, UCL and GE Healthcare. Software used in this project is maintained by CCP SyneRBI (UKRI support: EPSRC grant EP/T026693/1 and STFC through the DRI programme).
Keywords: PET, Energy Window Selection, Monte Carlo Scatter Estimation
13:03 M-LT-02-10

Positron Range Confinement in Preclinical PET/MRI: A Multi-Isotope, Multi-Density Experimental Study (#1930)

F. Lopez-Berenguer1, A. Gonzalez-Montoro1, S. Berr2, M. Williams2, A. J. Gonzalez1

1 Instituto de Instrumentación para Imagen Molecular (i3M), Centro Mixto CSIC-Universitat Politècnica de València, Valencia, Spain
2 University of Virginia, Department of Radiology and Medical Imaging, Charlottesville, Virginia, United States of America

Abstract

The spatial resolution of PET imaging is fundamentally limited by the positron range, whose extent depends on both the emission energy of the radionuclide and the electron density of the surrounding medium. In integrated PET/MRI systems, the static magnetic field exerts a Lorentz force on the positrons, confining their trajectories in the transverse plane to the main magnetic field, potentially improving image resolution. Two complementary experiments were performed using a dedicated monolithic-crystal based PET insert operated at 0T and inside a 9.4T MRI: capillary line sources embedded in three tissue-equivalent phantoms of increasing density (EPF, PLA, and Rigid10k resin), and a microDerenzo phantom. Both tests were carried out with three positron emitters (¹⁸F, ⁸⁹Zr, and ⁶⁸Ga) eventually use in pre-clinical imaging. The results demonstrate that the magnetic field significantly enhances transverse spatial resolution for high-energy positron emitters in medium- and high-density materials, with FWTM reductions of ~60% for ⁶⁸Ga and ~18% for ⁸⁹Zr, while the axial direction remains largely unaffected. Rod resolvability improved at 9.4T from 1.5 mm to ~1.0 mm and from 1.2 mm to 0.9 mm for ⁶⁸Ga and  ⁸⁹Zr, respectively. . These findings highlight the relevance of both positron energy and tissue-equivalent density in determining the magnitude of magnetic-field-induced resolution improvements in preclinical PET/MRI.

Acknowledgment

This work was supported by the Generalitat Valenciana under Grant No. CIPROM/2021/003 and funds from the Univ, VA FUS Cancer Immunotherapy Center.

F.Lopez Berenguer is supported by the European Research Council (ERC) under the European Union’s Horizon Europe research and innovation programme (Grant agreement No. 101164363).
References
[1] Moses, WW 2011, 'Fundamental limits of spatial resolution in PET', Nuclear Instruments and Methods in Physics Research A, vol. 648, pp. S236–S240.
[2] Ku-Toval, D et al. 2024, 'Quantitative analysis of the effect of the magnetic field generated by a PET/MR scanner on positron range', Physics in Medicine and Biology, vol. 69, no. 18, p. 185006.
[3] Lopez-Berenguer, F et al. 2025, 'Positron Range Confinement in High-Field MRI, Enhancing PET Image Quality', IEEE NSS/MIC.
[4] Lopez-Berenguer, F et al. 2025, 'Evaluation of a PET insert for trimodal imaging: a step towards PET/MRI-guided focused ultrasound', IEEE Transactions on Radiation and Plasma Medical Sciences, in press.
Keywords: Positron Emission Tomography (PET), Magnetic Resonance Imaging (MRI), positron range, preclinical imaging.